Advertisement for orthosearch.org.uk
Results 1 - 2 of 2
Results per page:
Bone & Joint Research
Vol. 5, Issue 9 | Pages 419 - 426
1 Sep 2016
Leichtle CI Lorenz A Rothstock S Happel J Walter F Shiozawa T Leichtle UG

Objectives

Cement augmentation of pedicle screws could be used to improve screw stability, especially in osteoporotic vertebrae. However, little is known concerning the influence of different screw types and amount of cement applied. Therefore, the aim of this biomechanical in vitro study was to evaluate the effect of cement augmentation on the screw pull-out force in osteoporotic vertebrae, comparing different pedicle screws (solid and fenestrated) and cement volumes (0 mL, 1 mL or 3 mL).

Materials and Methods

A total of 54 osteoporotic human cadaver thoracic and lumbar vertebrae were instrumented with pedicle screws (uncemented, solid cemented or fenestrated cemented) and augmented with high-viscosity PMMA cement (0 mL, 1 mL or 3 mL). The insertion torque and bone mineral density were determined. Radiographs and CT scans were undertaken to evaluate cement distribution and cement leakage. Pull-out testing was performed with a material testing machine to measure failure load and stiffness. The paired t-test was used to compare the two screws within each vertebra.


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_9 | Pages 8 - 8
1 May 2016
Lorenz A Mueller A Lange B Herzog Y Schnauffer P Wuelker N Leichtle U
Full Access

Introduction

Persistent anterior knee pain, subluxation or dislocation of the patella as well as early aseptic loosening and increased polyethylene wear of the patella implant are common clinical problems after total knee arthroplasty (TKA) which are associated with the patellofemoral joint. In addition to patellar resurfacing, the design of the patellofemoral joint surfaces is attributed a large influence. While for patients without patella resurfacing, the native patella is sliding on the standardized femoral component and therefore the possibility of a reduced surface matching is high, patella resurfacing has been shown to decrease the joint contact area and yield to increased patellofemoral pressure. With regard to a further design optimization, the current study examined patellar biomechanics after TKA without and with resurfacing, comparing 5 differently designed patellofemoral joint surfaces of the femoral implant.

Methods

The femoral implant of the Genesis II prosthesis (Smith & Nephew) was scanned and an adaptable CAD-model was built using CATIA. Five different designs of the patellofemoral groove were created:

original

completely flat

laterally elevated (+2mm lateral, −1mm medial)

medially elevated (+2mm medial, −1mm lateral)

laterally & medially elevated (+3mm lateral+medial)

The tibiofemoral joint as well as patellofemoral groove path and radius remained unchanged. Rapid Prototyping was used to produce prototypes made of polyamide.

A dynamic muscle loaded knee squat was simulated on 10 fresh frozen knee specimens with an upright knee simulator. The patellofemoral pressure distribution was measured using a flexible, resistive force sensor (TEKSCAN) while tibiofemoral and patellofemoral kinematics were recorded with an ultrasonic motion tracking system (ZEBRIS). In addition, patellar stability was measured in different flexion angles on another 10 specimens using a robot (KUKA). Measurements were taken on the native knee as well as after TKA and after additional patellar resurfacing with alternating femoral implant.