Vol. 105-B, Issue SUPP_12 | Pages 11 - 11
23 Jun 2023
Lombardi AV Alexander JS Berend KR Houserman DJ Adams JB Crawford DA
Full Access

Previous studies have reported excellent results with tapered, titanium alloy, porous plasma-sprayed components in patients undergoing uncemented primary total hip arthroplasty (THA). The purpose of this study was to examine survival and clinical results at minimum 25-year follow-up.

We reviewed all patients who underwent primary THA at our center through 1995 with a specific femoral component, the Mallory-Head Porous (MHP; Zimmer Biomet, Warsaw, IN). This device, marketed in the U.S. until December 2021, was essentially unchanged since its 1984 introduction, except the porous coating was continued circumferentially along the lateral aspect in 1987, a hydroxyapatite-coated option was offered in 1988, and an offset option was added in 1999 after the study period. Three hundred thirty-two patients (396 THA) had a minimum of 25-year follow-up. Mean age at surgery was 47.6 years (range, 21–70 years).

Mean follow-up in non-failed patients was 28.7 years (range, 25 to 37 years). There were 31 femoral revisions (7.8%): 9 infection, 3 failure of ingrowth, 5 aseptic loosening, 8 osteolysis revised well-fixed, 2 periprosthetic fracture, 2 polyethylene wear with trochanteric avulsion, 1 component breakage, and 1 malalignment well-fixed. Kaplan-Meier survival with endpoint of stem revision for all causes was 94.8% (95% CI: ±0.9%) at 36.7 years, and survival with endpoint of aseptic loosening/failure of ingrowth was 98.7% (95% CI: ±0.5) at 36.7 years. Harris hip scores improved significantly from 43 preoperatively to 76 most recently.

This tapered, titanium, porous plasma spray-coated femoral component continues to demonstrate high long-term survival with a low rate of femoral component revision for any reason or aseptic loosening/failure of ingrowth.


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 91 - 97
1 Jul 2021
Crawford DA Lombardi AV Berend KR Huddleston JI Peters CL DeHaan A Zimmerman EK Duwelius PJ

Aims

The purpose of this study is to evaluate early outcomes with the use of a smartphone-based exercise and educational care management system after total hip arthroplasty (THA) and demonstrate decreased use of in-person physiotherapy (PT).

Methods

A multicentre, prospective randomized controlled trial was conducted to evaluate a smartphone-based care platform for primary THA. Patients randomized to the control group (198) received the institution’s standard of care. Those randomized to the treatment group (167) were provided with a smartwatch and smartphone application. PT use, THA complications, readmissions, emergency department/urgent care visits, and physician office visits were evaluated. Outcome scores include the Hip disability and Osteoarthritis Outcome Score (HOOS, JR), health-related quality-of-life EuroQol five-dimension five-level score (EQ-5D-5L), single leg stance (SLS) test, and the Timed Up and Go (TUG) test.


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 145 - 149
1 Jun 2021
Crawford DA Passias BJ Adams JB Berend KR Lombardi AV

Aims

A limited number of investigations with conflicting results have described perivascular lymphocytic infiltration (PVLI) in the setting of total knee arthroplasty (TKA). The purpose of this study was to determine if PVLI found in TKAs at the time of aseptic revision surgery was associated with worse clinical outcomes and survivorship.

Methods

A retrospective review was conducted on 617 patients who underwent aseptic TKA revision who had histological analysis for PVLI at the time of surgery. Clinical and radiological data were obtained pre- and postoperatively, six weeks postoperatively, and then every year thereafter.


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 3 - 12
1 Jun 2021
Crawford DA Duwelius PJ Sneller MA Morris MJ Hurst JM Berend KR Lombardi AV

Aims

The purpose is to determine the non-inferiority of a smartphone-based exercise educational care management system after primary knee arthroplasty compared with a traditional in-person physiotherapy rehabilitation model.

Methods

A multicentre prospective randomized controlled trial was conducted evaluating the use of a smartphone-based care management system for primary total knee arthroplasty (TKA) and partial knee arthroplasty (PKA). Patients in the control group (n = 244) received the respective institution’s standard of care with formal physiotherapy. The treatment group (n = 208) were provided a smartwatch and smartphone application. Early outcomes assessed included 90-day knee range of movement, EuroQoL five-dimension five-level score, Knee Injury and Osteoarthritis Outcome Score for Joint Replacement (KOOS JR) score, 30-day single leg stance (SLS) time, Time up and Go (TUG) time, and need for manipulation under anaesthesia (MUA).


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_10 | Pages 34 - 34
1 Oct 2020
Lombardi AV
Full Access

Background

Ultraporous metals have now been used in acetabular reconstruction for two decades with excellent survival. The purpose of this study is to evaluate a newer porous metal made from Ti6Al4V titanium alloy in complex primary and revision hip arthroplasty.

Methods

A retrospective review as performed on all total hip arthroplasty (THA) procedures in which a G7 Osseo-Ti (Zimmer Biomet, Warsaw, IN) acetabular component was used between 2015 and 2017. Patients with 2-year minimum follow-up or failure were included, yielding a cohort of 123 patients (126 hips). There were 50 male patients (41%; 51 hips) and 73 females (59%; 75 hips). Mean age was 65 years (range, 43–88) and mean BMI was 30.7 kg/m2 (range, 18–56). Indications for ultraporous metal components were in hips with compromised bone stock or severe acetabular deformity. Procedures were 35 complex primary THA and 91 revision THA that included 12 conversions and 24 reimplantations as part of 2-staged exchange for treatment of infection.


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_10 | Pages 56 - 56
1 Oct 2020
Lombardi AV Berend KR Huddleston J Crawford D Peters C VanAndel D Anderson M DeHaan A Southgate R Duwelius PJ
Full Access

Background

The purpose of this study is to evaluate the early outcomes with the use of a smartphone-based exercise and educational platform after primary total hip arthroplasty compared to a standard of care control group.

Methods

A multicenter prospective randomized control trial was conducted evaluating the use of the mymobility smartphone-based care platform for primary total hip arthroplasty (THA). Patients randomized to the control group (198 patients) received the respective institution's standard of care. Those randomized to mymobility treatment group (167 patients) were provided an Apple Watch and mymobility smartphone application. The application provides pre and postoperative educational content, video directed exercise programs as well as tracks the patient's activity. Patients in the treatment group were not initially prescribed physical therapy, but could be if their surgeon deemed it necessary. Early outcomes assessed included 90-day hip range of motion, HOOS JR scores, 30-day single leg stance (SLS) and time up and go (TUG) test.

We also evaluated PT utilization, THA complications associated with readmissions, ER visits not associated with readmissions, urgent care (non standard of care) visits, and physician office visits.

Outcome scores include HOOS-Jr, EQ-5D-5L, single stance (SLS), Timed up and go (TUG).

Satisfaction scores for the procedure and the mymobility study group were also recorded.


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_9 | Pages 30 - 30
1 Oct 2020
Lombardi AV Duwelius PJ Morris MJ Hurst JM Berend KR Crawford DA
Full Access

Background

The purpose of this study is to evaluate the early perioperative outcomes after primary knee arthroplasty with the use of a smartphone-based exercise and educational platform compared to a standard of care control group.

Methods

A multicenter prospective randomized control trial was conducted evaluating the use of the mymobility smartphone-based care platform for primary total knee arthroplasty (TKA) and unicondylar knee arthroplasty (UKA). Patients in the control group (224 patients) received the respective institution's standard of care typically with formal physical therapy. Those randomized to mymobility treatment group (192 patients) were provided an Apple Watch and mymobility smartphone application. The treatment group was not initially prescribed physical therapy, but could if their surgeon determined it necessary. Early outcomes assessed included 90-day knee range of motion, KOOS Jr scores, 30-day single leg stance (SLS) time, Time up and Go (TUG) time and need for manipulation under anesthesia (MUA). There was no significant difference in age, BMI or gender between groups.


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_10 | Pages 17 - 17
1 Oct 2020
Berend KR Morris MJ Lombardi AV Crawford DA
Full Access

Background

The impact of a patient's activity level following total hip arthroplasty (THA) remains controversial, with some concerned about increased polyethylene wear, aseptic loosening and revisions. The purpose of this study is to report on implant survivorship and outcomes of high activity patients compared to low activity patients after THA with current polyethylene.

Methods

A retrospective review identified 2002 patients (2532 hip) that underwent a primary THA with vitamin E infused highly crosslinked polyethylene liner and 2-year minimum follow-up or revision. Patients were divided in two groups based on their University of California Los Angeles (UCLA) activity level: Low activity (LA) (UCLA ≤5) and high activity (HA) (UCLA ≥ 6). Outcomes included Harris Hip Score, UCLA activity score, complications and reoperations. Multivariate nominal regression analysis was performed to evaluate the significance of postoperative activity level on survivorship while controlling for age, gender, preoperative pain, HHS and body mass index (BMI).


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_9 | Pages 50 - 50
1 Oct 2020
Berend KR Passias BJ Lombardi AV Crawford DA
Full Access

Background

A limited number of investigations with conflicting results have described perivascular lymphocytic infiltration (PVLI) in the setting of total knee arthroplasty. The purpose of this study was to determine if PVLI found in total knee replacements at the time of aseptic revision surgery was associated with worse clinical outcomes and survivorship.

Methods

A retrospective review was conducted on 617 patients that underwent aseptic total knee arthroplasty revision who had histologic analysis for PVLI at the time of surgery. Clinical and radiographic data was obtained pre and postoperatively, 6 weeks post operatively, and then every year thereafter.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_11 | Pages 19 - 19
1 Oct 2019
Berend KR Lombardi AV Crawford DA Hurst JM Morris MJ
Full Access

Introduction

A smartphone-based care platform allows a customizable educational and exercise interface with patients, allowing many to recover after surgery without the need for formal physical therapy (PT). Furthermore, advances in wearable technology to monitor physical activity (PA) provides patients and physicians quantifiable metrics of the patient's recovery. The purpose of this study is to determine the feasibility of a smartphone-based exercise educational platform after primary knee arthroplasty as well as identifying factors that may predict the need for formal physical therapy.

Methods

This study is part of a multi-institution, prospective study of patients after primary total knee arthroplasty (TKA) and partial knee arthroplasty (PKA) enrolled in a smartphone with smartwatch-based episode of care platform that recorded multimodal PA (steps, kcal, stairs). Postoperatively, all patients initially followed the smartphone-based exercise program. At the surgeon's discretion, patients were prescribed therapy if needed. The outcome of this study was the need for PT outside the app-based exercise program as well as time to return to preoperative step count. Variables assessed were preoperative weekly step counts (steps/day), weekly postoperative activity level (weekly step count compared to preoperative level), compliance with the exercise program (>75% completion) and patient demographic data including gender, age, BMI and narcotic use. One hundred eighty-eight patients were included in analysis: 45 PKA (24%) and 143 TKA (76%). Step count data was available on 135 patients and physical therapy data on 174.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_12 | Pages 14 - 14
1 Oct 2019
Lombardi AV Crawford DA Morris MJ Adams JB Berend KR
Full Access

Background

Thigh pain following a well-fixed total hip arthroplasty (THA) remains problematic and a source of patient dissatisfaction. The purpose of this study is to evaluate if the development of distal femoral cortical hypertrophy (DFCH) is associated with chronic postoperative thigh pain after THA with a short stem implant.

Methods

All patients who underwent an uncomplicated primary THA via a direct anterior approach with the short stem Taperloc Microplasty® (Zimmer Biomet, Warsaw, IN) implant between 2011 and 2015 were mailed a pain drawing questionnaire. Radiographs were reviewed at 1-year minimum to determine cortical thickness change from immediate post-op. Thigh pain was compared to DFCH as well as patient demographics and femoral stem size. 293 patients were included in the study


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_12 | Pages 37 - 37
1 Oct 2019
Berend KR Crawford DA Adams JB Morris MJ Lombardi AV
Full Access

Background

Recurrent instability and dislocation after total hip arthroplasty are difficult complications. In certain cases, a constrained acetabular device can address these issues. The purpose of this study is to report the midterm outcomes and survivorship of a single constrained liner device.

Methods

A retrospective review as performed on all procedures (except first stage exchange for infection) in which a Freedom® Constrained (Zimmer Biomet, Warsaw, IN) liner was used between December 2003 to November 2016. Patients with 2-year minimum follow-up or failure were included, yielding a cohort of 177 patients. Procedures were 130 revisions, 40 reimplantations following infection eradication, and 7 complex primaries. The constrained mechanism was implanted in 46 hips (26%) to treat active instability and 131 hips (74%) for increased risk of instability or intraoperative instability. Patients had on average 3.4 previous surgeries.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_11 | Pages 63 - 63
1 Oct 2019
Berend KR Crawford DA Adams JB Lombardi AV
Full Access

Introduction

Patients are often cautioned against a high level of activity following knee arthroplasty. The purpose of this study is to report on implant survivorship and outcomes of high activity patients compared to low activity patients after TKA.

Methods

We identified 1611 patients (2038 knees) who underwent TKA with 5-year minimum follow-up. Patients were divided in two groups: Low activity (LA) (UCLA ≤5) and high activity (HA) (UCLA ≥ 6). Pre-and postoperative ROM, Knee Society scores, complications and reoperations were evaluated.


The Bone & Joint Journal
Vol. 101-B, Issue 6_Supple_B | Pages 31 - 36
1 Jun 2019
Nam D Nunley RM Clohisy JC Lombardi AV Berend KR Barrack RL

Aims

Whether patient-reported pain differs among surgical approaches in total hip arthroplasty (THA) remains unclear. This study’s purposes were to determine differences in pain based on surgical approach (direct anterior (DA) vs posterolateral (PL)) and PL approach incision length.

Patients and Methods

This was a retrospective investigation from two centres and seven surgeons (three DA, three PL, one both) of primary THAs. PL patients were categorized for incision length (6 cm to 8 cm, 8 cm to 12 cm, 12 cm to 15 cm). All patients had cementless femoral and acetabular fixation, at least one year’s follow-up, and well-fixed components. Patients completed a pain-drawing questionnaire identifying the location and intensity of pain on an anatomical diagram. Power analysis indicated 800 patients in each cohort for adequate power to detect a 4% difference in pain (alpha = 0.05, beta = 0.80).


The Bone & Joint Journal
Vol. 101-B, Issue 1_Supple_A | Pages 25 - 31
1 Jan 2019
Greco NJ Manocchio AG Lombardi AV Gao SL Adams J Berend KR

Aims

Despite declining frequency of blood transfusion and electrolyte supplementation following total joint arthroplasty, postoperative blood analyses are still routinely ordered for these patients. This study aimed to determine the rate of blood transfusion and electrolyte restoration in arthroplasty patients treated with a perioperative blood conservation protocol and to identify risk factors that would predict the need for transfusion and electrolyte supplementation.

Patients and Methods

Patients undergoing primary total joint arthroplasty of the hip or knee between July 2016 and February 2017 at a single institution were included in the study. Standard preoperative and postoperative laboratory data were collected and reviewed retrospectively. A uniform blood conservation programme was implemented for all patients. Need for blood transfusion or potassium supplementation was determined through a coordinated decision by the care team. Rates of transfusion and supplementation were observed, and patient risk factors were noted.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_12 | Pages 35 - 35
1 Oct 2018
Lombardi AV Berend KR Adams JB Mallory NT
Full Access

Introduction

Modern implant systems offer a variety of options to address the posterior cruciate ligament (PCL) and afford stability in primary total knee arthroplasty (TKA). One system has three bearing options for cruciate retaining (CR) TKA: standard CR bearing (CR-S) with 3° posterior slope and no posterior lip; CR lipped bearing (CR-L) with no slope and small posterior lip, and deep-dished anterior stabilized bearing (CR-AS) with large anterior build-up to compensates for deficient or lax PCL. We previously reported on 1940 patients (2449 CR-TKA) at early follow-up comparing these three CR bearings and found CR-AS bearings had greater improvement in ROM and lower rate of manipulation despite their use in patients with greater preoperative deformity and lower ROM. The purpose of this study was to review a broader series by a single surgeon-designer with minimum 2-year follow-up to determine if there were differences between bearing types in terms of clinical outcomes and survival.

Methods

From 2003 to 2014, 3348 patients (4500 knees) underwent primary CR-TKA and were available for review with minimum 2-year follow-up. An algorithmic approach to constraint was employed, choosing the least amount necessary to afford stability. Bearings used were 3883 CR-S (86%), 256 CR-L (6%), and 361 CR-AS (8%). CR-S bearings were used in patients with less preoperative tibiofemoral deformity (p=0.0316), less flexion contracture (p<0.0001), and better ROM (p<0.0001), but more mediolateral instability (p<0.0001) than CR-AS bearings.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_13 | Pages 13 - 13
1 Oct 2018
Greco NJ Lombardi AV Morris MJ Berend KR
Full Access

Background

Despite growing interest in direct anterior approach total hip arthroplasty, perioperative femoral fracture and early aseptic loosening are increasingly recognized complications. Previous research has documented the role of surgeon experience in association with these femoral complications. The purpose of this study was to explore the relationship between femoral component design and early periprosthetic femoral complications.

Methods

This was an extension of previous work with an updated patient cohort of 5090 consecutive direct anterior primary total hip arthroplasties at a single institution with a single-taper, wedge femoral stem with 4 variants involving length and geometry: Group 1) full length, standard profile; Group 2) full length, reduced distal profile; Group 3) short length, standard profile; and Group 4) short length, reduced distal profile. Records were reviewed retrospectively for the incidence of early periprosthetic fracture or aseptic loosening and analyzed with regard to patient demographics and femoral stem type.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_12 | Pages 5 - 5
1 Oct 2018
Gerow D Greco NJ Berend ME Berend KR Lombardi AV
Full Access

Intro/Purpose

Lateral partial knee replacement is indicated as an alternative to total knee replacement for isolated end stage lateral compartment degenerative joint disease. The purpose of this study is to report the surgical technique and clinical results of a large series of lateral partial knee replacements from two institutions.

Methods

A retrospective review identified 205 lateral unicompartmental arthroplasty procedures performed in two practices over a six-year period (Aug 2011 – June 2017). Patients indicated for surgery met specific preoperative clinical and radiographic criteria confirming lateral compartment arthritic disease with a correctable deformity, intact full-thickness medial cartilage, competent anterior cruciate ligament, and minimal disease in the patellofemoral compartment. A single cemented fixed bearing tibial component design was used in all cases specifically designed for lateral compartment anatomy.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_13 | Pages 68 - 68
1 Oct 2018
Lombardi AV Berend KR Morris MJ Crawford DA Adams JB
Full Access

Total hip arthroplasty (THA) continues moving to the outpatient arena, and may be feasible for some conversion and revision scenarios. Controversy surrounds appropriate patient selection. The purpose of this study is to report complications associated with outpatient revision and conversion THA, and to determine if comorbidities are associated with complications or overnight stay.

From June 2013 through March 2018, 43 patients (44 hips) underwent conversion (n=9) or revision (n=35) THA at a free-standing ambulatory surgery center. Mean patient age was 58.4 years, and 52% of patients were male. Conversion procedures were failed fracture fixation with retained hardware and all involved both femoral and acetabular replacement. Revision procedures involved head only in one, head and liner in 20, cup and head in 8, stem only in one, stem and liner in 4, and full revision in one.

Forty-one (93%) were discharged same day without incident, none required transfer to acute facility, and 3 required overnight stay with 2 of these for convenience and only one for a medical reason, urinary retention. Three patients with early superficial infection, including 2 diagnosed by positive intraoperative cultures, were successfully treated with oral antibiotics. There were no major complications, readmissions, or subsequent surgeries within 90 days. One or more major comorbidities were present in 15 patients (34%) including 1 valvular disease, 7 arrhythmia, 2 thromboembolism history, 3 obstructive sleep apnea, 3 chronic obstructive pulmonary disease, 2 asthma, 4 frequent urination, and 1 renal disease. The single patient who stayed overnight for a medical reason had no major medical comorbidities.

Outpatient arthroplasty, including revision THA in some scenarios, is safe for many patients. Presence of medical comorbidities was not associated with risk of complications. The paradigm change of patient education, medical optimization, and a multimodal program to mitigate risk of blood loss and reduce need for narcotics facilitates performing arthroplasty safely in an outpatient setting.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_13 | Pages 11 - 11
1 Oct 2018
Nam D Salih R Riegler V Nunley RM Clohisy JC Lombardi AV Berend KR Barrack RL
Full Access

Introduction

Despite well-fixed implants, persistent pain following total hip arthroplasty (THA) remains a concern. Various surgical approaches have been advocated, yet whether patient-reported pain differs amongst techniques has not been investigated. This study's purposes were to determine differences in patient-reported pain based on surgical approach (direct anterior –DA versus posterolateral-PL) or PL approach incision length. Our hypothesis was that no differences in patient-reported pain would be present.

Methods

A retrospective, IRB-approved investigation from 2 centers was performed. 7 fellowship trained arthroplasty surgeons (3 DA, 3 PL, 1 both) enrolled patients undergoing primary THA for non-inflammatory arthritis. PL approach patients were categorized based on incision length (6–8cm, 8–12cm, 12–15cm). Exclusion criteria were a prior hip surgery, revision procedure, or limited postoperative mobility. All THAs were performed using a cementless titanium, proximally coated, tapered femoral stem and hemispherical acetabular component. All patients had a minimum of 1-year clinical follow-up with radiographically well-fixed components.

A pain-drawing questionnaire was administered in which patients identify the location and intensity of pain on an anatomic diagram. Independent Student's t-tests and Chi-square analyses were performed (p<0.05 = significant). Power analysis indicated 800 patients in each cohort would provide adequate power to detect a 4% difference in patient-reported pain (alpha = 0.05, beta = 0.80).