Advertisement for orthosearch.org.uk
Results 1 - 3 of 3
Results per page:
Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_24 | Pages 19 - 19
1 May 2013
Legg AJ Flowers MJ
Full Access

Aim

Kite manipulation and casting for congenital talipes equinovarus (CTEV) was noted to require a subsequent posteromedial release (PMR) in almost all of the children treated, with variable outcomes including overcorrection and stiffness. Introduction of the Ponseti serial manipulation and casting technique dramatically reduced the need for PMR. This study assesses the medium term outcomes in these two treatment groups.

Methods

We retrospectively identified patients treated for idiopathic CTEV between 1997 and 2007 under a single surgeon. Two cohorts with a minimum 4 years' follow-up were treated with Kite (1997–2001) and Ponseti (2002–2007) manipulation and casting. The entire Kite cohort (14 patients) and none of the Ponseti cohort (16 patients) required PMR. All patients were assessed by clinical and functional examination, questionnaire and medical notes review. There was a total of 40 feet (10 bilateral and 20 unilateral) with 20 feet in each cohort.


The Bone & Joint Journal
Vol. 95-B, Issue 3 | Pages 407 - 410
1 Mar 2013
Legg AJ Hamer AJ

We have recently shown that waste heat from forced-air warming blankets can increase the temperature and concentration of airborne particles over the surgical site. The mechanism for the increased concentration of particles and their site of origin remained unclear. We therefore attempted to visualise the airflow in theatre over a simulated total knee replacement using neutral-buoyancy helium bubbles. Particles were created using a Rocket PS23 smoke machine positioned below the operating table, a potential area of contamination. The same theatre set-up, warming devices and controls were used as in our previous study. This demonstrated that waste heat from the poorly insulated forced-air warming blanket increased the air temperature on the surgical side of the drape by > 5°C. This created convection currents that rose against the downward unidirectional airflow, causing turbulence over the patient. The convection currents increased the particle concentration 1000-fold (2 174 000 particles/m3 for forced-air warming vs 1000 particles/m3 for radiant warming and 2000 particles/m3 for the control) by drawing potentially contaminated particles from below the operating table into the surgical site.

Cite this article: Bone Joint J 2013;95-B:407–10.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 2 | Pages 254 - 256
1 Feb 2012
Legg AJ Cannon T Hamer AJ

Patient warming significantly decreases the risk of surgical site infection. Recently there have been concerns that forced air warming may interfere with unidirectional airflow, potentially posing an increased risk of infection. Our null hypothesis was that forced air and radiant warming devices do not increase the temperature and the number of particles over the surgical site when compared with no warming device. A forced air warming device was compared with a radiant warming device and no warming device as a control. The temperature and number of particles were measured over the surgical site. The theatre was prepared as for a routine lower-limb arthroplasty operation, and the same volunteer was used throughout the study.

Forced air warming resulted in a significant mean increase in the temperature (1.1°C vs 0.4°C, p < 0.0001) and number of particles (1038.2 vs 274.8, p = 0.0087) over the surgical site when compared with radiant warming, which raises concern as bacteria are known to require particles for transport.