For over a decade, modular titanium fluted tapered (TFT) stems have demonstrated excellent clinical success for femoral revision total hip arthroplasty (THA) surgery. The aim of this study was to report the short-term outcomes of a novel modern monoblock TFT stem used for revision and complex primary THA with a minimum of 2 years follow-up. We identified 126 patients who received a single monoblock TFT stem - 26 patients for complex THA (failed fracture fixation) and 100 patients for revision THA. The reasons for revision THA included 40 for previous prosthetic joint infection (PJI), 42 for aseptic loosening, 9 for trunnionosis, 9 for periprosthetic fractures. The Paprosky grading for femoral bone loss at the time of surgery and the measured subsidence of femoral stems at 3 months follow-up were determined. We evaluated the number and indications for re-operations. The mean time from surgery was 3.9 years (range 2.0 to 6.9 years). A paired t-test analysis showed significant improvement from pre-operative versus post-operative clinical outcome scores (p<0.001) for HHS (38.76 +/- 15.24vs. 83.42 +/- 15.38), WOMAC (45.6 ± 19.0 vs. 69.9 ± 21.3) and SF-12 Physical component (31.7 ± 8.1 vs. 37.8 ± 11.3) and SF-12 Mental component (48.2 ± 12.2 vs. 51.6 ± 12.5). The Paprosky grading for femoral bone loss was Grade 1 (3.9%), Grade 2 (35.7%), Grade 3A (47.6%), Grade 3B (11.1%) and Grade 4 (1.6%) cases. There were 18 re-operations (14.7%) with 13 for PJI (7 treated with implant retention, 6 treated with a two-staged revision), 4 for instability and one for acetabular aseptic loosening. There were no aseptic failures of the stem. This novel modern monoblock TFT stem provided reliable femoral fixation and has increasingly supplanted the use of modular TFT stems for complex primary and revision surgery in our institution.
The direct anterior (DA) approach has been associated with rapid patient recovery after total hip arthroplasty (THA) but may be associated with more frequent femoral complications including implant loosening. The objective of this study was to determine whether the addition of a collar to the femoral stem affects implant migration, patient activity, and patient function following primary THA using the DA approach. Patients were randomized to either a collared (n = 23) or collarless (n = 26) cementless femoral stem implanted using the DA approach. Canal fill ratio (CFR) was measured on the first postoperative radiographs. Patients underwent a supine radiostereometric analysis (RSA) exam postoperatively on the day of surgery and at two, four, six, 12, 26, and 52 weeks postoperatively. Patient-reported outcome measures (Western Ontario and McMaster Universities Osteoarthritis (WOMAC) Index, the 12-item Short Form Health Survey Mental and Physical Score, and University of California, Los Angeles (UCLA) Activity Score) were measured preoperatively and at each post-surgery clinic visit. Activity and function were also measured as the weekly average step count recorded by an activity tracker, and an instrumented timed up-and-go (TUG) test in clinic, respectively.Aims
Methods
Total hip arthroplasty (THA) is an effective surgery for the treatment of advanced osteoarthritis but increasing numbers of these procedures are having a significant impact on healthcare budgets. One route to mitigate the increasing costs is outpatient THA, discharging patients on the same day as their surgery. The purpose of this study was to determine the cost of outpatient THA compared to standard overnight stay in hospital. This was a prospective-randomized controlled trial for patients undergoing primary THA through a direct anterior approach. Participants were randomized to be discharged on the same day as surgery, as outpatients, or on day one post-surgery, as inpatients, using a Zelen consent model. Adverse events were assessed, and participants completed self-reported cost questionnaires at two-, six- and 12-weeks post-surgery, and the WOMAC preoperatively and at 12-weeks post-surgery. We performed a cost analysis from health care payer (HCP) and societal perspectives.Introduction
Methods
Trunnion wear has been reported as a cause for failure of modern total hip replacement implants. While there are several reports on the prevalence of trunnion corrosion with specific stem designs, little is known about the prevalence of this problem across other femoral stem designs. The purpose of this study is to review three commonly used uncemented femoral stems and to correlate any established risk factors with rates of revision. We reviewed all patients followed prospectively over a 15-year period (March 2000 to September 2015) who underwent total hip arthroplasty with one of three modern uncemented femoral stem designs. All of the stems were made of titanium (Ti6Al4V alloy) with a 12/14 taper design. We included only metal (cobalt chromium) on highly cross-linked polyethylene articulations. We evaluated age, gender, body mass index (BMI), femoral head size, head length, neck angle and offset as potential risk factors for all cause revision and revision excluding infection by univariate analysis. We performed a logistic regression analysis with odds ratios (OR) for each parameter for both all cause revision as well as with infection excluded. We used a p value of <0.05 as a level of significance.Introduction
Methods
The aim of this study was to determine whether there is a difference
in the rate of wear between acetabular components positioned within
and outside the ‘safe zones’ of anteversion and inclination angle. We reviewed 100 hips in 94 patients who had undergone primary
total hip arthroplasty (THA) at least ten years previously. Patients
all had the same type of acetabular component with a bearing couple
which consisted of a 28 mm cobalt-chromium head on a highly crosslinked
polyethylene (HXLPE) liner. A supine radiostereometric analysis
(RSA) examination was carried out which acquired anteroposterior
(AP) and lateral paired images. Acetabular component anteversion
and inclination angles were measured as well as total femoral head
penetration, which was divided by the length of implantation to
determine the rate of polyethylene wear.Aims
Patients and Methods
Tribocorrosion at the head–neck taper interface
– so-called ‘taperosis’ – may be a source of metal ions and particulate
debris in metal-on-polyethylene total hip arthroplasty (THA). We examined the effect of femoral head length on fretting and
corrosion in retrieved head–neck tapers Head length was observed to affect fretting (p = 0.03), with
28 mm + 8 mm femoral heads showing greater total fretting scores
than all other head lengths. The central zone of the femoral head
bore taper was subject to increased fretting damage (p = 0.01),
regardless of head length or stem offset. High-offset femoral stems
were associated with greater total fretting of the bore taper (p
= 0.04). Increased fretting damage is seen with longer head lengths and
high-offset femoral stems, and occurs within a central concentric
zone of the femoral head bore taper. Further investigation is required
to determine the effect of increased head size, and variations in
head–neck taper design. Cite this article:
Total hip replacement (THR) is a very common
procedure undertaken in up to 285 000 Americans each year. Patient
satisfaction with THR is very high, with improvements in general
health, quality of life, and function while at the same time very
cost effective. Although the majority of patients have a high degree
of satisfaction with their THR, 27% experience some discomfort,
and up to 6% experience severe chronic pain. Although it can be
difficult to diagnose the cause of the pain in these patients, this
clinical issue should be approached systematically and thoroughly.
A detailed history and clinical examination can often provide the
correct diagnosis and guide the appropriate selection of investigations, which
will then serve to confirm the clinical diagnosis made. Cite this article: