Advertisement for orthosearch.org.uk
Results 1 - 1 of 1
Results per page:
Applied filters
Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 90-B, Issue SUPP_II | Pages 306 - 306
1 Jul 2008
Davis E Kureshi S Olsen M Papini M Zdero R Waddell J Schemitsch E
Full Access

Introduction: Notching of the femoral neck during preparation of the femur during hip resurfacing has been associated with an increased risk of femoral neck fracture. We aimed to evaluate this with the use of a finite element model.

Methods: A three dimensional femoral model was used and molded with a femoral component constructed from the dimensions of a Birmingham Hip Resurfacing. Multiple constructs were made with the component inferiorly translated in order to cause a notch in the superior femoral neck. The component angulation was kept constant. Once constructed the model was imported into the Ansys finite element model software for analysis. Elements within the femoral model were assigned different material properties depending on cortical and cancellous bone distributions. Von Misses stresses were evaluated near the notches and compared in each of the cases.

Results: In the un-notched case the maximum Von Mises stress was only 40MPa. However, with the formation of a 1mm notch the stress rose to 144MPa and in the 4 mm notch the stress increased to 423MPa. These values demonstrated that a 1mm notch increased the maximum stress by 361% while a 4mm notch increased the maximum stress by 1061%.

Discussion: This study demonstrated that causing a notch in the superior femoral neck dramatically increases the stress within the femoral neck. This may result in the weakening of the femoral neck and potentially predispose it to subsequent femoral neck fracture. The data suggests that even a small notch of 1mm may be detrimental in weakening the femoral neck by dramatically increasing the stress in the superior neck. This study suggests that any femoral neck notching should be avoided during hip resurfacing.