Advertisement for orthosearch.org.uk
Results 1 - 1 of 1
Results per page:
Applied filters
Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_IV | Pages 393 - 394
1 Apr 2004
Tanaka C Shikata J Ikenaga M Takahashi M
Full Access

In revision THA, the solid acetabular reconstruction in the true acetabulum is often challenging. We are using the Kerboull acetabular reinforcement devices after packing hydroxyapatite granules for acetabular bone defects. We report our 3–7 year clinical and roentgenographical results.

Twenty-one acetabuli in 20 patients were reconstructed in the true acetabuli with Kerboull-type acetabular reinforcement devices and porous hydroxyapatite granules (Sumitomo Pharmaceutical Co. Ltd.). The mean age of the patients at operation was 68 years. The mean follow-up period was 5 years and 3 months (38–88 months). Acetabular bone deficiencies were evaluated according to AAOS classification (type II: 5 hips, type III: 16 hips). Porous hydroxyapatite granules (20–60 grams) were grafted to all the acetabuli. Autologous cortico-cancellous bone grafts or hydroxyapatite blocks were used to reconstruct the segmental defects in 6 hips. All the acetabuli were reinforced with Ker-boull-type reinforcement devices and Charnley-type cemented prostheses were implanted. Clinical and roent-genographical results were evaluated using Japanese Orthopaedic Association hip score and Hodgkinson classification. Average hip score was improved from 42 points to 75 points. No re-revision was done. No infection was noted. The roentgenograms showed neither migration nor loosening of the acetabular components.

Porous hydroxyapatite granules are one of the best bone substitutes because of their mechanical and biochemical properties. Oonishi reported very good results of his acetabular reconstruction using this material. The Kerboull-type acetabular devices are very effective to the reconstruction in the true acetabulum. Excellent results of these devices were reported by Kerboull. Ace-tabular reconstrution using both of them showed very good clinical and roentgenographical results during 3 to 7 post-operative years in our series.