Advertisement for orthosearch.org.uk
Results 1 - 1 of 1
Results per page:
Applied filters
Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_X | Pages 25 - 25
1 Apr 2012
Kamal T Hegazy A Elikety S Shah N Casha J
Full Access

We are presenting a prospective analysis and assessment for the results of the Coflex@ inter-spinous dynamic stabilization device with a 2 year follow up. The purpose of this study is to assess the efficacy and safety of the Coflex@ Posterior Dynamic Stabilization Implant.

A prospective analysis was performed on 121 patients (176 devices) treated with the Coflex@ Interspinous Implant. Indication for the treatment was spinal canal foraminal stenosis with nerve root irritation.

Pre- and postoperative disability and pain scores were measured using Oswestry disability score, functional assessment a radiological evaluation at pre-operative - three - six twelve and twenty-four month interval.

Data analysis revealed a high rate of patient satisfaction 92% of patients are satisfied and will have the surgery again 81% of patients had major improvement of their leg symptoms. 72% of patients reported improvement of their back pain symptoms this was more noted in the decompression group.

Two revision surgeries were carried out. One due to implant back-out and the second due to infection.

The Coflex@inter-spinous implant is a simple surgical treatment strategy with a low risk. Early results show a good improvement of both clinical and radiological parameters combined with patient satisfaction with better outcome achieved if combined with spinal decompression.