A-70-year old woman underwent uncomplicated total hip arthroplasty using a titanium modular stem with a 46mm CoCr femoral head, a titanium shell, and a metal linear (Wright Medical Technology). Eight years after implantation, she presented with a painful left hip. A pelvic radiograph revealed adequate positioning of both hip implants without any signs of wear of loosening. CT scanning confirmed the presence of a 5 × 5 cm soft tissue mass in the ilium above the cup component accompanied by the iliac fracture. The patient was diagnosed as having an adverse reaction to metal debris (ARMD) after a metal-on-metal THA and revision was performed. Perioperatively?tissue necrosis and partial destruction of the abductor mechanism were found in the absence of any macroscopic infection. Both the neck trunnion and bore of the head showed slight signs of corrosion. The modular neck was revised with a ceramic 28mm head and a new dual-mobility liner(Zimmer Biomet). The iliac fracture was fixed with a porous trabecular metal augment(Zimmer Biomet). The histopathology of tissue sample revealed extensively necrotic material with focal cellular areas of inflammatory cells containing macrophages and neutrophilas. Metalic debris was also scattered in the necrotic materials. After the revision, the patient was recovered without pain or dislocation, and iliac fracture was well fixed. Instability is a substantial problem in the revision of ARMD. Extensive necrosis with gross deficiency of the abductor mechanism is associated with postoperative dislocation. Revision of failed MoM THA a dual-mobility device an effective strategy.