Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 207 - 207
1 Mar 2013
Johns B Pelletier M Bertollo N Hancock N Walsh W Bruce W
Full Access

Introduction

Initial stability of the tibial component influences the success of uncemented total knee arthroplasty. In uncemented components, osseointegration provides long-term fixation which is particularly important for the tibial component. Osseointegration is facilitated by minimising bone-implant interface micromotion to within acceptable limits. To investigate initial stability, this study compares the micromotion and initial seating of two uncemented hydroxyapatite-coated tibial components, the Genesis II and Profix. This is the first stability comparison of two hydroxyapatite-coated tibial components.

Methods

Six components of each type were implanted into synthetic tibias by a single orthopaedic surgeon. Good coverage was achieved. No screws or articular inserts were used. Initial seating was measured using ImageJ software at five areas on each tibia. Tibias were transected and their proximal section implanted into a molten alloy parallel to horizontal. Dynamic mechanical testing was performed using a hydraulic 858-Bionix machine. Prostheses underwent unilateral axial point-loading of 700N cyclically applied four times. The load was applied to three locations approximating femoral loading points. The loading cycle was repeated six times at each point, allowing micromotion to be recorded at three contralateral locations. Micromotion was measured by optical lasers. After dynamic testing, two tibial components of each type were removed with claw pliers while measuring the force required on the 858-Bionix machine. Implant under-surfaces were photographed for wear.