header advert
Results 1 - 10 of 10
Results per page:
Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 16 - 16
23 Jun 2023
Laboudie P Hallé A Anract P Hamadouche M
Full Access

Periprosthetic femoral fracture (PFF) following primary total hip arthroplasty (THA) is of raising concern with the aging of the population. The aim of this retrospective study was to assess the incidence of early PFF associated with Charnley- Kerboull (CK) stems cemented line-to-line according to the “French paradox” principles through the anterior approach (AA) in patients older than 70 years old.

This monocentric study involved 25 surgeons with various level of experience. From a prospectively collected database, all CK stems (AmisK®, Medacta, Castel San Pietro, Switzerland) done consecutively from January 2018 to May 2022 through the AA in patients older than 70 years were included. The measured parameters included canal flare index (CFI), morphological cortical index (MCI), canal-calcar ratio (CCR), ilium-ischial ratio (IIR), ilium overhang, and anterior superior iliac spine (ASIS) to greater trochanter distance. Dorr classification was also recorded.

A total of 416 CK stems performed by 25 surgeons were included. All patients had a minimal 3-month follow-up. The mean age was 77.4 years, and the mean BMI was 25.3 kg/m2. According to the Dorr classification 240 hips were of Dorr type A, 144 Dorr type B, and 2 Dorr type C. Mean CFI, MCI, CCR, IIR and ASIS-GT were 3.8, 0.5, 2.8, 2.5, and 101mm, respectively. Amongst the 416 THAs, one (0.24%; 95% CI 0.20 – 0.28%) early PFF Vancouver type B2 fracture that occurred 24 days postoperatively was observed.

This study demonstrated that CK stems cemented according to the “French Paradox” were associated with an extremely low rate of early PFF in patients over 70 years. The observed results are better than those reported with uncemented or taper slip cemented stems. Longer follow-up is warranted to evaluate further rate of fracture that may occur during the bone remodeling process around the femoral component.


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_14 | Pages 14 - 14
1 Nov 2021
Nicoules S Zaoui A Hage SE Scemama C Langlois J Courpied J Hamadouche M
Full Access

The purpose of this study was to compare oxinium versus metal-on-polyethylene wear in two consecutive prospective randomized series of low friction total hip arthroplasty at a minimum 10-year follow-up.

A total of 100 patients with a median age of 60.9 years were randomized to receive either oxinium (50 hips) or metal (50 hips) femoral head. The polyethylene socket was EtO sterilized in the first 50 patients, whereas it was highly cross-linked and remelted (XLPE) in the following 50 patients. The primary criterion for evaluation was linear head penetration measurement using the Martell system by an investigator blinded to the material. Also, a survivorship analysis was performed using wear related loosening revised or not as the end point.

Complete data were available for analysis in 40 hips at a median follow-up of 12.9 years (11 to 14), and in 36 hips at a median follow-up of 12.3 years (10 to 13) in the EtO sterilized and XLPE series, respectively.

In the EtO sterilized series, the mean steady-state wear rate was 0.245 ± 0.080 mm/year in the oxinium group versus 0.186 ± 0.062 mm/year in the metal group (p = 0.009). In the XLPE series, the mean steady-state wear rate was 0.037±0.016 mm/year in the oxinium group versus 0.036±0.015 mm/year in the metal group (p = 0.94). The survival rate at 10 years was 100% in both XLPE series, whereas it was 82.9% (IC 95%, 65–100) and 70.5% (IC95%, 50.1–90.9) in the metal-EtO and oxinium-EtO series, respectively.

This RCT demonstrated that up to 14-year follow-up, wear was significantly reduced when using XLPE, irrespective of the femoral head material. Also, no osteolysis related complication was observed in the XLPE series. In the current study, oxinium femoral heads showed no advantage over metal heads and therefore their continued used should be questioned related to their cost.


The Bone & Joint Journal
Vol. 100-B, Issue 6 | Pages 725 - 732
1 Jun 2018
Gibon E Barut N Courpied J Hamadouche M

Aims

The purpose of this retrospective study was to evaluate the minimum five-year outcome of revision total hip arthroplasty (THA) using the Kerboull acetabular reinforcement device (KARD) in patients with Paprosky type III acetabular defects and destruction of the inferior margin of the acetabulum.

Patients and Methods

We identified 36 patients (37 hips) who underwent revision THA under these circumstances using the KARD, fresh frozen allograft femoral heads, and reconstruction of the inferior margin of the acetabulum. The Merle d’Aubigné system was used for clinical assessment. Serial anteroposterior pelvic radiographs were used to assess migration of the acetabular component.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_1 | Pages 27 - 27
1 Jan 2018
Zaoui A Langlois J El Hage S Scemama C Courpied J Hamadouche M
Full Access

The purpose of this study was to compare the effect of femoral head material (delta ceramic versus metal) on polyethylene wear in a consecutive prospective randomized series of low friction total hip arthroplasty.

A total of 110 patients with a mean age of 60.6 ± 9.3 (34–75) years were randomized (power of 90%, alpha of 5%) to receive either a metal (55 hips) or a delta ceramic (55 hips) femoral head. The polyethylene socket was moderately cross-linked (3 Mrads of gamma radiation in nitrogen) and annealed at 130°C in all hips. All other parameters were identical in both groups. The primary criterion for evaluation was linear head penetration measurement using the Martell system, performed by an investigator blinded to the material of the femoral head. Creep and steady state wear values were calculated.

At the minimum of 3-year follow-up, complete data were available for analysis in 38 hips at a median follow-up of 4.4 years (3.0 to 5.7), and in 42 hips at a median follow-up of 4.0 years (3.0 to 5.4) in the metal and delta ceramic groups, respectively. The mean creep, measured as the linear head penetration at one year follow-up, was 0.42 ± 1.0 mm in the metal group versus 0.30 ± 0.81 mm in the delta ceramic group (Mann and Whitney test, p = 0.56). The mean steady state penetration rate from one year onwards measured 0.17 ± 0.44 mm/year (median 0.072) in the metal group versus 0.074 ± 0.44 mm/year (median 0.072) in the delta ceramic group (Mann and Whitney test, p = 0.48). No case of delta ceramic femoral head fracture was recorded, and no hip had signs of periprosthetic osteolysis.

This study demonstrated that up to 5-year follow-up, delta ceramic femoral head did not significantly influence creep neither wear of a contemporary annealed polyethylene. Longer follow-up is necessary to further evaluate the potential clinical benefits of delta ceramic.


The Bone & Joint Journal
Vol. 97-B, Issue 11 | Pages 1458 - 1462
1 Nov 2015
Langlois J Atlan F Scemama C Courpied JP Hamadouche M

Most published randomised controlled trials which compare the rates of wear of conventional and cross-linked (XL) polyethylene (PE) in total hip arthroplasty (THA) have described their use with a cementless acetabular component.

We conducted a prospective randomised study to assess the rates of penetration of two distinct types of PE in otherwise identical cemented all-PE acetabular components.

A total of 100 consecutive patients for THA were randomised to receive an acetabular component which had been either highly XL then remelted or moderately XL then annealed.

After a minimum of eight years follow-up, 38 hips in the XL group and 30 hips in the annealed group had complete data (mean follow-up of 9.1 years (7.6 to 10.7) and 8.7 years (7.2 to 10.2), respectively). In the XL group, the steady state rate of penetration from one year onwards was -0.0002 mm/year (sd 0.108): in the annealed group it was 0.1382 mm/year (sd 0.129) (Mann–Whitney U test, p < 0.001). No complication specific to either material was recorded.

These results show that the yearly linear rate of femoral head penetration can be significantly reduced by using a highly XLPE cemented acetabular component.

Cite this article: Bone Joint J 2015;97-B:1458–62.


The Bone & Joint Journal
Vol. 97-B, Issue 2 | Pages 177 - 184
1 Feb 2015
Felden A Vaz G Kreps S Anract P Hamadouche M Biau DJ

Conventional cemented acetabular components are reported to have a high rate of failure when implanted into previously irradiated bone. We recommend the use of a cemented reconstruction with the addition of an acetabular reinforcement cross to improve fixation.

We reviewed a cohort of 45 patients (49 hips) who had undergone irradiation of the pelvis and a cemented total hip arthroplasty (THA) with an acetabular reinforcement cross. All hips had received a minimum dose of 30 Gray (Gy) to treat a primary nearby tumour or metastasis. The median dose of radiation was 50 Gy (Q1 to Q3: 45 to 60; mean: 49.57, 32 to 72).

The mean follow-up after THA was 51 months (17 to 137). The cumulative probability of revision of the acetabular component for a mechanical reason was 0% (0 to 0%) at 24 months, 2.9% (0.2 to 13.3%) at 60 months and 2.9% (0.2% to 13.3%) at 120 months, respectively. One hip was revised for mechanical failure and three for infection.

Cemented acetabular components with a reinforcement cross provide good medium-term fixation after pelvic irradiation. These patients are at a higher risk of developing infection of their THA.

Cite this article: Bone Joint J 2015;97-B:177–84.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 3 | Pages 342 - 348
1 Mar 2010
El Masri F Kerboull L Kerboull M Courpied JP Hamadouche M

We have evaluated the in vivo migration patterns of 164 primary consecutive Charnley-Kerboull total hip replacements which were undertaken in 155 patients. The femoral preparation included removal of diaphyseal cancellous bone to obtain primary rotational stability of the stem before line-to-line cementing. We used the Ein Bild Roentgen Analyse femoral component method to assess the subsidence of the femoral component.

At a mean of 17.3 years (15.1 to 18.3) 73 patients were still alive and had not been revised, eight had been revised, 66 had died and eight had been lost to follow-up. The mean subsidence of the entire series was 0.63 mm (0.0 to 1.94). When using a 1.5 mm threshold, only four stems were considered to have subsided. Our study showed that, in most cases, a highly polished double-tapered stem cemented line-to-line does not subside at least up to 18 years after implantation.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 3 | Pages 304 - 309
1 Mar 2009
Kerboull L Hamadouche M Kerboull M

We describe 129 consecutive revision total hip replacements using a Charnley-Kerboull femoral component of standard length with impaction allografting. The mean follow-up was 8.2 years (2 to 16). Additionally, extramedullary reinforcement was performed using struts of cortical allograft in 49 hips and cerclage wires in 30.

There was one intra-operative fracture of the femur but none later. Two femoral components subsided by 5 mm and 8 mm respectively, and were considered to be radiological failures. No further revision of a femoral component was required. The rate of survival of the femoral component at nine years, using radiological failure as the endpoint, was 98%. Our study showed that impaction grafting in association with a Charnley-Kerboull femoral component has a low rate of subsidence. Reconstruction of deficiencies of distal bone with struts of cortical allograft appeared to be an efficient way of preventing postoperative femoral fracture for up to 16 years.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 11 | Pages 1439 - 1445
1 Nov 2007
Triclot P Grosjean G El Masri F Courpied JP Hamadouche M

We carried out a prospective randomised study designed to compare the penetration rate of acetabular polyethylene inserts of identical design but different levels of cross-linking at a minimum of four years follow-up. A total of 102 patients (102 hips) were randomised to receive either highly cross-linked Durasul, or contemporary Sulene polyethylene inserts at total hip replacement. A single blinded observer used the Martell system to assess penetration of the femoral head. At a mean follow-up of 4.9 years (4.2 to 6.1) the mean femoral head penetration rate was 0.025 mm/year (sd 0.128) in the Durasul group compared with 0.106 mm/year (sd 0.109) in the Sulene group (Mann-Whitney test, p = 0.0027). The mean volumetric penetration rate was 29.24 mm3/year (sd 44.08) in the Durasul group compared with 53.32 mm3/year (sd 48.68) in the Sulene group. The yearly volumetric penetration rate was 55% lower in the Durasul group (Mann-Whitney test, p = 0.0058). Longer term results are needed to investigate whether less osteolysis will occur.


The Journal of Bone & Joint Surgery British Volume
Vol. 83-B, Issue 7 | Pages 979 - 987
1 Sep 2001
Hamadouche M Witvoet J Porcher R Meunier A Sedel L Nizard R

We have carried out a prospective, randomised study designed to compare the long-term stability of the stem of cementless femoral implants with differing surface configurations. A total of 50 hips (46 patients) was randomised into two groups, according to whether the medullary stem had been grit blasted (GB) or coated with hydroxyapatite (HA). Both femoral prostheses were of the same geometrical design. We used Ein Bild Roentgen Analyse femoral component analysis (EBRA-FCA) to assess the stability of the stem. The mean follow-up was for 8.66 years.

The mean migration of the stem was 1.26 mm in the HA group compared with 2.57 mm in the GB group (Mann-Whitney U test, p = 0.04). A mixed model ANOVA showed that the development of subsidence was statistically different in the two groups during the first 24 months. After this subsidence increased in both groups with no difference between them. Our results indicate that, with the same design of stem, HA coating enhanced the stability of the femoral stem when compared with GB stems.