Full Access

Introduction and Method: For 16 years, now, we have been using custom made femoral stems (titanium stem, HA coated) based on preoperative patient CT scans in young patients presenting with symptomatic osteoarthritis of the hip. The aim was to provide optimum initial fit-and-fill of the stem in the femoral medullary canal, conferring the best chance of secondary osteoin-tegration. This, with a goal of long term survivorship. The custom stem also enables dialling-in correction of the (often abnormal) femoral neck version (in young patients presenting with arthritis) to a more normalised 15 degrees of anteversion.

Results: We present the long term results (5–16 year, mean of 10 years), clinical, survivorship and radiological, of 312 primary total hip arthroplasties in 280 patients, all of whom were under the age of 50 years of age (mean age 40).

At 10 years we have a survivorship, if femoral aseptic loosening is used as an end point, of 97.6%. There was a deep infection rate of 1.2%, and a dislocation rate of 1.9%. There were no cases of thigh pain, and no intra-operative femoral neck/shaft fractures.

Discussion: The under 50’s with hip arthritis requiring total hip arthroplasty are a highly challenging group, they are young, active and tend to have distorted anatomy. Our results are superior to those previously published using either cemented or standard uncemented stems. Thus, justifying the increased initial financial outlay on the custom-made stem.