Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 14 - 14
23 Jun 2023
Hooper G Thompson D Frampton C Lash N Sharr J Fulkner D Gilchrist N
Full Access

Femoral stem design affects periprosthetic bone mineral density (BMD), which may impact long term survival of cementless implants in total hip arthroplasty (THA). The aim of this study was to examine proximal femoral BMD in three morphologically different uncemented femoral stems designs to investigate whether one particular design resulted in improved preservation of BMD.

119 patients were randomized to receive either a proximally coated dual taper wedge stem, a proximally coated anatomic stem or a fully coated collarless triple tapered stem. Dual energy x-ray absorptiometry scans (Lunar iDXA, GE Healthcare, Madison, WI) assessed BMD across the seven Gruen zones pre-operatively, and post-operatively at 6-weeks, 1-year, and 2-years comparing the unoperated contralateral femur as a control.

BMD increased in zones one (2.5%), two (17.1%), three (13.0%), five (10%) and six (17.9%) for all stems. Greater preservation of BMD was measured on the lateral cortex (zone 2) for both the dual taper wedge and anatomic stems (p = 0.019). The dual taper wedge stem also demonstrated preservation of BMD in the medial calcar (zone 7) whilst the anatomic and triple taper stem declined in this region, however this was not statistically significant (p = 0.059). BMD decreased on average by 2.1% in the mid-diaphysis region, distal to the stem tip (zone 4) for all implants. All stems performed equivalently at final follow-up in all patient reported outcome measures.

Implant design was shown to impact BMD changes. All stems preserved bone at the metaphyseal level in early follow-up, which theoretically indicates favorable implant fixation. There was no significant stress shielding observed, however longer follow-up is required to elucidate the impact of this finding on implant survivorship.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_2 | Pages 77 - 77
10 Feb 2023
Hooper G Thompson D Lash N Sharr J Faulkner D Frampton C Gilchrist N
Full Access

Femoral stem design affects periprosthetic bone mineral density (BMD), which may impact long term survival of cementless implants in total hip arthroplasty (THA). The aim of this study was to examine proximal femoral BMD in three morphologically different uncemented femoral stems designs to investigate whether one particular design resulted in improved preservation of BMDMethods: 119 patients were randomised to receive either a proximally coated dual taper wedge stem, a proximally coated anatomic stem or a fully coated collarless triple tapered stem. All surgeries were performed via the posterior approach with mobilization on the day of surgery. Dual energy x-ray absorptiometry scans (Lunar iDXA, GE Healthcare, Madison, WI) assessed BMD across the seven Gruen zones pre-operatively, and post-operatively at 6-weeks, 1-year, and 2-years and compared to the unoperated contralateral femur as a control. Patient reported outcome measures of pain, function and health were also included at these corresponding follow-ups.

BMD increased in zones one (2.5%), two (17.1%), three (13.0%), five (10%) and six (17.9%) for all stems. Greater preservation of BMD was measured on the lateral cortex (zone 2) for both the dual taper wedge and anatomic stems (p = 0.019). The dual taper wedge stem also demonstrated preservation of BMD in the medial calcar (zone 7) whilst the anatomic and triple taper stem declined in this region, however this was not statistically significant (p = 0.059). BMD decreased on average by 2.1% inthe mid-diaphysis region, distal to the stem tip (zone 4) for all implants. All stems performed equivalently at final follow-up in all patient reported outcome measures.

This study demonstrated maintenance of femoral BMD in three different cementless femoral stem designs, with all achieving excellent improvements in patient reported outcomes. There was no significant stress shielding observed, however longer follow-up is required to elucidate the impact of this finding on implant survivorship.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_11 | Pages 10 - 10
1 Aug 2018
Hooper G Gilchrist N Maxwell R Frampton C
Full Access

Stress shielding has been a well-recognised problem with uncemented femoral components resulting in proximal bone loss and dysfunction, but less attention has been paid to the preservation of acetabular bone stock. Uncemented acetabular components often demonstrate reduced bone density on plain radiographs in the mid-portion of the cup (zone 2), which may be due to the rigidity of the outer shell. This study compares the change in bone density around three different cups with varying moduli of elasticity at a minimum of 2 years. Our hypothesis was that less rigid cups would be associated with improved bone density and less stress shielding.

This prospective randomised controlled trial compared the bone mineral content (BMC) adjacent to three different cups with marked differences in stiffness. Cup A was an all titanium shell, cup B was a titanium coated all polyethylene implant and cup C was a tantalum backed shell. All articulations used a 32mm ceramic femoral head. Cup B used polyethylene modified by treatment with vitamin E whereas cups A and C used a liner made of irradiated cross linked polyethylene.

Five regions of interest (ROI) were established adjacent to the cup, regions 2, 3 and 4 where similar to the DeLee and Charnley regions 1, 2 and 3. Bone density was measured using IDXA preoperatively, postoperatively, 6 months, 1 and 2 years and compared for each ROI and implant.

Precision measurements showed significant reliability. All areas showed a reduction in BMC following insertion of the acetabular cup. Bone loss was less in ROI 1 and 4 in the area of rim fit for all cups and the maximal bone loss was seen in ROI 2 and 3 at the dome of the cup. The more elastic cup (Cup B) produced the least bone loss in this area (p<0.05). Cup C produced the largest bone loss at ROI 2 (40%) which continued increasing at 2 years.

Cup stiffness is related to bone loss adjacent to the acetabulum, presumably due to a similar process of stress shielding as seen in the femur. All cups produced similar changes at the periphery of the cup but the more elastic cup retained bone density beneath the cup which continued past 2 years. This improvement in bone quality is likely to be associated with better acetabular bone stock into the future and more reliable long term cup fixation.


The Bone & Joint Journal
Vol. 95-B, Issue 11 | Pages 1480 - 1483
1 Nov 2013
Hooper GJ Gilchrist N Maxwell R March R Heard A Frampton C

We studied the bone mineral density (BMD) and the bone mineral content (BMC) of the proximal tibia in patients with a well-functioning uncemented Oxford medial compartment arthroplasty using the Lunar iDXA bone densitometer. Our hypothesis was that there would be decreased BMD and BMC adjacent to the tibial base plate and increased BMD and BMC at the tip of the keel.

There were 79 consecutive patients (33 men, 46 women) with a mean age of 65 years (44 to 84) with a minimum two-year follow-up (mean 2.6 years (2.0 to 5.0)) after unilateral arthroplasty, who were scanned using a validated standard protocol where seven regions of interest (ROI) were examined and compared with the contralateral normal knee. All had well-functioning knees with a mean Oxford knee score of 43 (14 to 48) and mean Knee Society function score of 90 (20 to 100), showing a correlation with the increasing scores and higher BMC and BMD values in ROI 2 in the non-implanted knee relative to the implanted knee (p = 0.013 and p = 0.015, respectively).

The absolute and percentage changes in BMD and BMC were decreased in all ROIs in the implanted knee compared with the non-implanted knee, but this did not reach statistical significance. Bone loss was markedly less than reported losses with total knee replacement.

There was no significant association with side, although there was a tendency for the BMC to decrease with age in men. The BMC was less in the implanted side relative to the non-implanted side in men compared with women in ROI 2 (p = 0.027), ROI 3 (p = 0.049) and ROI 4 (p = 0.029).

The uncemented Oxford medial compartment arthroplasty appears to allow relative preservation of the BMC and BMD of the proximal tibia, suggesting that the implant acts more physiologically than total knee replacement. Peri-prosthetic bone loss is an important factor in assessing long-term implant stability and survival, and the results of this study are encouraging for the long-term outcome of this arthroplasty.

Cite this article: Bone Joint J 2013;95-B:1480–3.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XLI | Pages 42 - 42
1 Sep 2012
Hooper G Gilchrist N Frampton C Maxwell R Heard A Mcguire P
Full Access

Bone mineral density (BMD) and bone mineral content (BMC) have not been previously assessed in unicompartmental knee replacement (UKR). We studied the early bone changes beneath the uncemented Oxford medial UKR. Our hypothesis was that this implant should decrease the shear stresses across the bone-implant interface and result in improved BMD and BMC beneath the tibial component.

Using the Lunar iDXA and knee specific software we developed 7 regions of interest (ROI) in the proximal tibia and assessed 38 patients with an uncemented Oxford UKR at 2 years. We measured the replaced knee and contralateral unreplaced knee using the same ROI and compared the BMD and BMC. The initial precision study in 20 patients demonstrated high precision in all areas.

There were 12 males and 16 females with an average age of 65.8 years (46–84 years). ROI 1 and 2 were beneath the tibial tray and had significantly less BMC (p=0.023 and 0.001) and BMD (p=0.012 and 0.002). ROI 3 was the lateral tibial plateau and this area also had significantly less BMC (p=0.007) and BMD (p=0.0001). ROI 4 and 5 immediately below the tibial keel had no significant change. These changes were independent of gender and age.

These results were surprising in that the universal loss of BMC and BMD suggested that bone loading of the proximal tibia was not improved even after a UKR. The better BMD and BMC adjacent to the keel confirms other studies that show improved bone in-growth around keels and pegs in the uncemented tibial component. A prospective longitudinal study has been developed to compare BMD and BMC changes over time to see whether these changes are dynamic.