Large-head metal-on-metal (MoM) total hip replacements
(THR) have given rise to concern. Comparative studies of small-head
MoM THRs over a longer follow-up period are lacking. Our objective
was to compare the incidence of complications such as infection,
dislocation, revision, adverse local tissue reactions, mortality
and radiological and clinical outcomes in small-head (28 mm) MoM
and ceramic-on-polyethylene (CoP) THRs up to 12 years post-operatively. A prospective cohort study included 3341 THRs in 2714 patients.
The mean age was 69.1 years (range 24 to 98) and 1848 (55.3%) were
performed in women, with a mean follow-up of 115 months (18 to 201).
There were 883 MoM and 2458 CoP bearings. Crude incidence rates
(cases/1000 person-years) were: infection 1.3 In conclusion, we found similar results for small-head MoM and
CoP bearings up to ten years post-operatively, but after ten years
MoM THRs had a higher risk of all-cause revision. Furthermore, the
presence of an adverse response to metal debris seen in the small-head
MOM group at revision is a cause for concern. Cite this article:
A recent review of the literature on metal-on-metal total hip arthroplasties (THA) revealed the lack of comparative clinical studies with a sufficient sample size and the inclusion of patient-reported outcomes as well as patient activity levels. We conducted a prospective cohort study including all metal-on-metal and conventional polyethylene (PE)-ceramic THAs with an uncemented cup (Morscher press-fit cup), a 28mm head and operated upon via a lateral approach at our University hospital between 1/1999 and 12/2008. Only THAs for primary osteoarthritis were included. The study population is part of the Geneva Hip Arthroplasty Registry, a prospective cohort followed since 1996. The following outcomes were compared between the two groups (metal-on-metal=group 1 vs. PE-ceramic bearing=group 2): (1) Complication rates with respect to infection, dislocation and revision, (2) Radiographic outcomes (presence of linear or focal femoral osteolysis, loosening), and (3) Clinical outcomes (Harris Hip score increase, SF-12, activity and patient satisfaction evaluation, presence of groin pain). Patients operated between 1/1999 and 12/2004 were evaluated five years postoperatively by an independent assessor. Cox regression analysis was used to compare incidence rates while adjusting for differences in baseline characteristics.Introduction
Methods
We conducted a longitudinal study including patients with the same type of primary hybrid total hip replacement and evaluated patient activity and femoral osteolysis at either five or ten years post-operatively. Activity was measured using the University of California, Los Angeles scale. The primary outcome was the radiological assessment of femoral osteolysis. Secondary outcomes were revision of the femoral component for aseptic loosening and the patients’ quality of life. Of 503 hip replacements in 433 patients with a mean age of 67.7 years (30 to 91), 241 (48%) were seen at five and 262 (52%) at ten years post-operatively. Osteolytic lesions were identified in nine of 166 total hip replacements (5.4%) in patients with low activity, 21 of 279 (7.5%) with moderate activity, and 14 of 58 (24.1%) patients with high activity. The risk of osteolysis increased with participation in a greater number of sporting activities. In multivariate logistic regression adjusting for age, gender, body mass index and the inclination angle of the acetabular component, the adjusted odds ratio for osteolysis comparing high In conclusion, of patients engaged in high activity, 24% had developed femoral osteolysis five to ten years post-operatively.
None of the patients required cup revision for aseptic loosening. Complications included 12 dislocations and 3 deep infections resulting in 2 total revisions. In 3 patients the stem was revised for aseptic loosening at a mean of 63 months. At 10-years the survivorship was 98.6% (95% CI 96.7; 99.4) with endpoint revision for any cause. Mean total wear was 0.89 mm (±0.5). 32 cups (18.8%) with a cup inclination >
45° had a mean wear of 1.06 mm (±0.5), whereas 138 cups (81.2%) with inclination <
45° had a mean wear of 0.86 mm (±0.5), p=0.036. In 16 cases osteolytic defects around the stem were present. The outcome scores at 10 years were: HHS 85.9 (±14.1), WOMAC pain 70.7 (±24.7), WOMAC function 68.8 (±24.5), SF-12 physical score 40.3 (±9.2) and mental score 47.0 (±10.4). Ninety-four percent of the patients were satisfied or very satisfied.