Vol. 106-B, Issue SUPP_6 | Pages 55 - 55
2 May 2024
McCann C Ablett A Feng T Macaskill V Oliver W Keating J
Full Access

Subtrochanteric femoral fractures are a subset of hip fractures generally treated with cephalomedullary nail fixation\[1\]. Single lag screw devices are most commonly-used, but integrated dual screw constructs have become increasingly popular\[2,3\]. The aim of this study was to compare outcomes of fixation of subtrochanteric femoral fractures using a single lag screw (Gamma3 nail, GN) with a dual screw device (InterTAN nail, IN). The primary outcome was mechanical failure, defined as lag screw cut-out, back-out, nail breakage or peri-implant fracture.

Consecutive adult patients (18yrs) with subtrochanteric femoral fracture treated in a single centre were retrospectively identified using electronic records. Patients that underwent surgical fixation using either a long GN (2010–2017) or IN (2017–2022) were included. Medical records and radiographs were reviewed to identify complications of fixation. Cox regression analysis was used to determine the risk of mechanical failure and secondary outcomes by implant design. Multivariable regression models were used to identify predictors of mechanical failure.

The study included 622 patients, 354 in the GN group (median age 82yrs, 72% female) and 268 in the IN group (median age 82yrs, 69% female). The risk of any mechanical failure was increased two-fold in the GN group (HR 2.44 \[95%CI 1.13 to 5.26\]; _p=0.024_). Mechanical failure comprising screw cut-out (_p=0.032_), back-out (_p=0.032_) and nail breakage (_p=0.26_) was only observed in the GN group. Technical predictors of failure included varus >5° for cut-out (OR 19.98 \[2.06 to 193.88\]; _p=0.01_), TAD;25mm for back-out (8.96 \[1.36 to 58.86\]; p=0.022) and shortening 1cm for peri-implant fracture (7.81 \[2.92 to 20.91\]; _p=<0.001_).

Our results demonstrate that an intercalated screw construct is associated with a lower risk of mechanical failure compared with the a single lag screw device. Intercalated screw designs may reduce the risk of mechanical complications for patients with subtrochanteric femoral fractures.


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_4 | Pages 5 - 5
8 Feb 2024
Ablett AD McCann C Feng T Macaskill V Oliver WM Keating JF
Full Access

This study compares outcomes of fixation of subtrochanteric femoral fractures using a single lag screw (Gamma3 nail, GN) with a dual lag screw device (InterTAN nail, IN). The primary outcome was mechanical failure, defined as lag screw cut-out, back-out, nail breakage or peri-implant fracture. Technical factors associated with mechanical failure were also identified.

All adult patients (>18yrs) with a subtrochanteric femoral fracture treated in a single centre were retrospectively identified using electronic records. Included patients underwent surgical fixation using either a long GN (2010–2017) or IN (2017–2022). Cox regression analysis was used to determine the risk of mechanical failure and technical predictors of failure.

The study included 587 patients, 336 in the GN group (median age 82yrs, 73% female) and 251 in the IN group (median age 82yrs, 71% female). The IN group exhibited a higher prevalence of osteoporosis (p=0.002) and CKD□3 (p=0.007). There were no other baseline differences between groups. The risk of any mechanical failure was increased two-fold in the GN group (HR 2.51, p=0.020). Mechanical failure comprising screw cut-out (p=0.040), back-out (p=0.040) and nail breakage (p=0.51) was only observed in the GN group. The risk of peri-implant fracture was similar between the groups (HR 1.10, p=0.84). Technical predictors of mechanical included varus >5° for cut-out (HR 15.61, p=0.016), TAD>25mm for back-out (HR 9.41, p=0.020) and shortening >1cm for peri-implant fracture (HR 6.50, p=<0.001).

Dual lag screw designs may reduce the risk of mechanical complications for patients with subtrochanteric femoral fractures.


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_4 | Pages 7 - 7
8 Feb 2024
Martin DH Ng N Armstong B Brennan J Feng T Lekuse K White TO Mackenzie SP
Full Access

Myriad protocols exist for isolated Weber B lateral malleolus fractures with a congruent tibiotalar joint on initial radiographs. Stress and weight-bearing radiographs, all at various timepoints, may be employed to identify those injuries that develop significant talar shift but consensus is elusive. This study outlines a safe and reproducible protocol for such injuries, utilising a removable orthosis, immediate weight bearing and standard supine radiographs.

A retrospective analysis of a prospective trauma database was analysed to identify patients with an isolated Weber B ankle fracture with adequate presentation radiographs demonstrating a congruent mortise. Patient records and radiographs were evaluated a minimum of 5 years after initial presentation to determine ankle stability, complications, and the burden on outpatient services.

Between 2014 and 2016, 657 patients were referred to the specialist trauma clinic from the emergency department. Of the 657, 52 patients had inadequate ED radiographs to determine ankle congruity. At the two-week assessment, 11 of the 52 demonstrated talar shift and required intervention. Therefore 646 patients demonstrated ankle congruity at two weeks after weight bearing. No patient demonstrated talar shift at the six-week assessment. Average number of follow up appointments was 2.4 with 3.5 radiographs. Our new treatment protocol advocates discharge after a single orthopaedic assessment after two weeks of weight bearing.

This study supports immediate weight-bearing of Weber B ankle fractures with a congruent mortise in an orthosis. Follow up beyond two weeks is unnecessary and our protocol offers a safe means of significantly reducing the outpatient burden.