Advertisement for orthosearch.org.uk
Results 1 - 1 of 1
Results per page:
Applied filters
General Orthopaedics

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_28 | Pages 41 - 41
1 Aug 2013
Ecker T Steppacher S Haimerl M Murphy S
Full Access

Introduction

Correct postoperative leg length restoration is among the most important goals of hip arthroplasty. Therefore, we developed, validated and clinically applied a novel software algorithm based on surgical navigation, which allows the surgeon to restore a defined femur position without establishing a femoral coordinate system or the hip joint center and measure the leg length accurately and simply.

Material and Methods

This new leg length algorithm was used in 154 hips (145 patients) that underwent CT-based computer-assisted THA (VectorVision Build 274 prototype; BrainLAB AG, Helmstetten, Germany) with a tissue preserving superior capsulotomy. Intraoperatively, a pelvic and a femoral dynamic reference bases (DRB) were applied and the anterior pelvic plane (APP) was set as the pelvic coordinate system. Then, the hip joint was put in a neutral position and this position, and the relative position of the femoral DRB relative to the pelvic DRB, was captured and stored by the navigation system. After implantation of the prosthesis the same above described femoral position with the same amplitude of flexion/extension, abduction/adduction and rotation was restored. Now, any resulting difference was due to linear changes. Validation of this new algorithm was performed by comparing the navigated results to measurements from calibrated antero-posterior pre- and postoperative radiographs. The radiographic results were compared to the mean leg length change measured with the navigation system.