Advertisement for orthosearch.org.uk
Results 1 - 7 of 7
Results per page:
Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 388 - 388
1 Dec 2013
Latham J Cook R Bolland B Wakefield A Culliford D Tilley C
Full Access

Introduction

Metal on metal hip arthroplasty continues to be controversial. Emerging evidence suggests that there are multiple modes of failure, and that the results of revision surgery are influenced by host and implant factors.

Methods

This study compares a single surgeon series of hip resurfacings (Birmingham Hip Resurfacing {BHR}) and large diameter metal on metal total hip replacements (LDMOMTHR). Primary outcome measures included survival rates, failure secondary to histologically identified Adverse Reaction to Metal Debris (ARMD), and patient reported outcome measures (Oxford Hip Score {OHS}) following revision.

Between 1999 and 2005, 458 BHR and 175 LDMOMTHR were performed. At latest review 43 BHR's (9.4%) and 28 LDMOMTHR's (14%) have been revised.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 115 - 115
1 Dec 2013
Latham J Cook R Bolland B Yasen S
Full Access

Background

Large diameter chrome cobalt modular heads in hip replacement surgery have caused one of the biggest challenges to face orthopaedic surgeons, manufacturers and regulators in a generation. Increasing numbers of retrieved implants offer the opportunity to look in detail at the modes of failure of these devices.

Methods

We have examined the mechanical behaviour of the head taper/trunnion using a variety of techniques including:

mechanical testing in a custom-made jig

3D nano-precision form profilometry (RedLux)

optical 3D microscopy with focus variation technology

electro-chemical testing.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_34 | Pages 36 - 36
1 Dec 2013
Cook R Shearwood-Porter N Nicolae C Bolland B Latham J Wood R
Full Access

Introduction:

Cemented femoral components have been used in hip replacement surgery since its inception. For many patients this works well, but recent retrieval studies1–4 and more fundamental studies5, 6 have highlighted the issues of damage and material loss from the both matt and polished cemented stems.

Materials and methods:

This study will focus on a cohort of retrievals from the Southampton Orthopaedics Centre for Arthroplasty Retrieval Surgery (SOCARS). The cohort consisted of a number of hybrid modular total hip replacements with cemented femoral components, both from mixed and matched manufacturer stem and head combinations. Femoral stems were polished, collarless, tapered designs; head sizes ranged from 28–54 mm. For each femoral stem, samples of Palacos R + G cement (Heraeus Medical GmbH, Hanau, Germany) were retrieved from the proximal region of the cement mantle (Gruen zones 1 and 7), corresponding to both macroscopically damaged and undamaged surfaces of the stem. The areas of damage were determined using calibrated digital photography; damaged surfaces were then imaged in detail using an Alicona InfiniteFocus microscope (Alicona Imaging GmbH, Graz, Austria). The technique uses optical microscopy and focus variation technology to extract 3D morphology and depth information from the surface with a resolution of 10 nm. A series of measurements were made and two different analysis routes were used to provide volumetric material loss measurements from the stem surface. High-resolution microscopy and elemental analysis of the cement and stem surfaces was conducted via SEM and EDX to identify the mechanisms leading to material loss at the cement-stem interface.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 18 - 18
1 Mar 2013
Bolland B Maul C Cook R Roques A Tuke M Wood R Latham J
Full Access

The poor outcome of large head metal on metal total hip replacements (LHMOMTHR) in the absence of abnormal wear at the articulating surfaces has focussed attention on the trunnion / taper interface. The RedLux ultra-precision 3D form profiler provides a novel indirect optical method to detect small changes in the form and surface finish of the head taper as well as a quantitative assessment of wear volume. This study aimed to assess and compare qualitatively the tapers from well functioning small diameter, with poorly functioning LHMOMTHR's using the above technique.

Method

3 groups of retrieval tapers were analysed (Group 1: 28 mm CoCr heads from well functioning MOMTHRs (n=5); Group 2: Large diameter CoCr heads from LHMOMTHRs revised for failure secondary to adverse reaction to metal debris (n=5); Gp 3 (control): 28 mm heads from well functioning metal on Polyethylene (MOP) THRs; n=3). Clinical data on the retrievals was collated. The Redlux profiling of modular head tapers involves a non direct method whereby an imprint of the inside surface of a modular head is taken, and this is subsequently scanned by an optical non contact sensor using dedicated equipment [1]. The wear was also measured on the bearing surface [1]. RedLux profiling of the tapers produced a taper angle and 3D surface maps. The taper angles obtained with the Redlux method were compared to those obtained using CMM measurement on 3 parts. The Redlux profiling, including imprints, was also repeated 3 times to gauge potential errors.

Results

There was no difference in mean 12/14 taper angles between groups. There was no difference in volumetric and linear wear at the bearing surface between groups. Only the LHMOMs showed transfer of pattern from the stem to the internal head taper, with clear demarcation of the contact and damaged area between head taper and stem trunnion (see figure 1 – interpretation of head taper surface features demonstrated using Redlux optical imaging). 3D surface mapping demonstrated wear patterns compatible with motion or deformations between taper and trunnion in the LHMOM group. These appearances were not seen in tapers from small diameter MOM and MOP THRs (see Figure 2).


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_15 | Pages 19 - 19
1 Mar 2013
Bolland B Tilley S Tilley C Cook R Latham J
Full Access

Pseudotumour formation is being reported with increasing frequency in failing metal-on-metal hip resurfacings and replacements. This mode of failure complication has also been reported with metal-on-polyethylene bearing bearings when it is usually associated with evidence of surface corrosion and no apparent wear at the head–neck taper.

We present a case with evidence of taper wear and damage secondary to corrosion in an uncemented total hip replacement with a metal on polyethylene articulation (TMZF (Titanium, Molybdenum, Zirconium and Ferrous) Accolade® stem, Trident® HA coated acetabular shell, Low Friction Ion Treatment (LFIT™) Cobalt-Chrome anatomic head (40 mm), X3® polyethylene liner)

Case

A 69 year old woman had a THR in 2008. A year later she started to complain of lateral based hip pain. Clinical examination and initial imaging indicated trochanteric bursitis and heterotopic bone formation. The symptoms became worse over the next 3 years and the patient was listed for exploration and excision of heterotopic bone.

Surgical findings

Extensive pseudotumour was encountered deep to bursa and adherent to capsule. Abductors, external rotators and vastus lateralis were spared. There was minimal calcar osteolysis and marginal erosion in superior acetabulum. There was no obvious wear on the articulating surfaces of the femoral head and polyethylene inlay of the socket nor was there any evidence of neck-to-rim impingement or edge loading. There was visible blackening of both the taper and trunnion after femoral head removal.


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_1 | Pages 139 - 139
1 Jan 2013
Bolland B Maul C Cook R Roques A Wood R Latham J
Full Access

The poor outcome of large head metal on metal total hip replacements (LHMOMTHR) in the absence of abnormal wear at the articulating surfaces has focussed attention on the trunnion/taper interface. The RedLux ultra-precision 3D form profiler provides a novel indirect optical method to detect small changes in form and surface finish of the head taper as well as a quantitative assessment of wear volume. This study aimed to assess and compare qualitatively the tapers from small diameter with LHMOMTHR's. 3 groups of retrieval tapers were analysed (Group 1: 28mm CoCr heads from MOMTHRs (n=5); Group 2: Large diameter CoCr heads from LHMOMTHRs (n=5); Group 3: 28mm heads from metal on polyethylene (MOP) THRs; n=3). Clinical data on the retrievals was collated. Both bearing surfaces and head tapers were measured for wear using the Redlux profiling non contact measurement system. Measurements included taper angle and 3D surface maps. Taper angles obtained with the Redlux method were compared to those obtained using CMM measurement on 3 parts. The Redlux profiling, including imprints, was also repeated 3 times to gauge potential errors. There was no difference in mean 12/14 taper angles between groups. There was no difference in volumetric and linear wear at the bearing surface between groups. Only the LHMOMs showed transfer of pattern from the stem to the internal head taper, with clear demarcation of the contact and damaged area between head taper and stem trunnion. 3D surface mapping demonstrated wear patterns compatible with motion or deformations between taper and trunnion in the LHMOM group alone. Discussion: Differences in appearance of the taper surface between LHMOMTHRs and MOP or MOM small diameter devices highlight an area of concern and potential contributor to the mode of early failure. Further work is required to fully qualify the Redlux method capabilities.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_III | Pages 6 - 6
1 Feb 2012
Amin A Clayton R Patton J Gaston M Cook R Brenkel I
Full Access

Aim

To compare the results of total knee replacement in a consecutive series of morbidly obese patients (body mass index (BMI) > 40 kg/m2) with a matched group of non-obese (BMI< 30 kg/m2) patients.

Methods

41 consecutive total knee replacements performed in morbidly obese patients were matched pre-operatively with 41 total knee replacements performed in non-obese patients for age, sex, diagnosis, type of prosthesis, laterality, knee score and function score components of the Knee Society Score (KSS). All patients were prospectively followed up and the post-operative KSS, radiographs, complications (superficial wound infection, deep joint infection, deep venous thrombosis, peri-operative mortality) and five-year survivorship compared for the two groups. No patients were lost to follow-up (mean follow-up in morbidly obese: 38.5 (range 6-66) months; non-obese: 44 (range 6-67) months).