Advertisement for orthosearch.org.uk
Results 1 - 1 of 1
Results per page:
Applied filters
General Orthopaedics

Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_2 | Pages 8 - 8
1 Jan 2013
Jenkins P Bulkeley M Mackenzie S Simpson H
Full Access

Introduction

The Taylor Spatial Frame (TSF) is an hexapod external ring fixation system that can move with six degrees of freedom to correct complex limb deformities. The lengths of the struts between the rings are independently adjusted to correct the deformity. The struts form an acute and obtuse “ring-strut” angle with the ring with the sum of these angles totalling 180°. In the course of a correction schedule a strut may need to be exchanged for one of longer or shorter length. The manufacturer's instructions direct that a temporary seventh strut can be placed in any orientation to ensure stability during the exchange. We have noted several episodes of temporary frame instability during this procedure resulting in discomfort. The aim of this study was to investigate which temporary strut positions gave maximal stability.

Methods

A TSF frame was constructed in a neutral alignment with a neutral strut height of 130mm. Strut 1 (red) was identified for exchange. There were 169 theoretical placement options for the temporary strut that were sequentially tested. Fast-FX™ struts were used. Strut 1 was released and the shortening that occurred was recorded using the strut length gauge. Shortening of over 10mm was considered grossly unstable.