Advertisement for orthosearch.org.uk
Results 1 - 1 of 1
Results per page:
Applied filters
Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_III | Pages 452 - 452
1 Sep 2009
Loosli Y Baumgartner D Bigolin G Gasser B Heini P
Full Access

Posterior internal fixation systems undergo internal constraints resulting in high load bearing requirement for the pedicular screw/bone interface. Only few studies deal with the impact of the vertebral augmentation on the migration of pedicular screws. In this study, the impact of the pedicular screw augmentation has been investigated under physiological load for osteoporotic vertebras. The data have been proceeded to reduce the influence of vertebral geometry, which generally leads to results devoid of statistical meaning

In 8 osteoporotic vertebrae, two screws have been inserted in each vertebra: a non-augmented on one side and an augmented one on the contralateral side.

Compression tests have been performed (two consecutive 50 cycles load steps -100N and 200N-) to observe the displacement of the screw’s head. Two different setups have been employed: a free connection (FC) and a blocked connection (BC). A load step is successful if the migration between two consecutive cycles tends to zero. To reduce the impact of the vertebras’ geometry, the screws’ migration have been compared contra-laterally using the migration ratio (MR). MR of vertebrae is defined as the division of the augmented screw’s migration with the non-augmented screw’s migration.

All the augmented screws survived both test setups whereas the non-augmented failed the 200N FC load step. Significant differences are observable only for the highest successful load steps for each test setup: T-tests (P=0.039 and P=0.007 respectively) put into evidence that the results are statistically smaller than one. It is observable as well, that the BC induced fewer loads into the vertebrae: even non-augmented screw can withstand 200N load step.

As expected, augmentation of pedicular perforated screws increases their stability in osteoporotic vertebras undergoing large physiological load. This could be explained by the fact that the presence of PMMA increases the load transfer interface improving screw/PMMA complex bearing capacity. Smaller loads induce only small differences that are not significant.