There are no recent studies comparing cable with
wire for the fixation of osteotomies or fractures in total hip replacement
(THR). Our objective was to evaluate the five-year clinical and
radiological outcomes and complication rates of the two techniques.
We undertook a review including all primary and revision THRs performed
in one hospital between 1996 and 2005 using cable or wire fixation.
Clinical and radiological evaluation was performed five years post-operatively.
Cables were used in 51 THRs and wires in 126, and of these, 36 THRs
with cable (71%) and 101 with wire (80%) were evaluated at follow-up.
The five-year radiographs available for 33 cable and 91 wire THRs
revealed rates of breakage of fixation of 12 of 33 (36%) and 42
of 91 (46%), respectively. With cable there was a significantly
higher risk of metal debris (68% In conclusion, we found a higher incidence of complications and
a trend towards increased infection and foreign-body reaction with
the use of cables.
Large diameter metal on metal cups have been used in total hip arthroplasty advocating superior results with respect to dislocation rates, range of motion and long term survival. The Durom cup used as part of the Durom hip resurfacing system has been incriminated with poor short term results sometimes correlated to incorrect positioning of either the femoral or acetabular component. Our objective was to evaluate short term results of the Durom cup used in conjunction with standard stems. We prospectively followed all patients with a large diameter metal-on-metal articulation (Durom) and a standard stem operated upon between 9/2004 and 9/2008. Patients were seen at follow-up for a clinical (Harris hip score=HHS, UCLA scale and patient satisfaction), radiographic and questionnaire assessment.Introduction
Methods
The prosthetic impingement occurs if the range of motion of the hip exceeds implant mobility or in case of component malorientation. This retrieval study was designed to assess the frequency and the risk factors of this phenomenon. The frequency and the severity of the impingement were calculated from a continuous series of 311 cups retrievals collected between 1989 and 2004 by a single surgeon. The reason for retrieval was loosening (131 cases), infection (43 cases), instability (56 cases), osteolysis (28 cases), unexplained pain (48 cases) and prosthetic impingent (5 cases all with hard bearings). The notching at the cup rim was assessed twice by two examiners with optic magnification. The risk factors were analyzed from clinical charts by univariate and cox multihazard.Introduction
Material and Methods
Implantation of a large-diameter femoral head prosthesis with a metal-on-metal bearing surface reduces the risk of dislocation, increases the range of movement, minimises the risk of impingement and, in theory, results in little wear. Between February 2004 and March 2007 we implanted 100 consecutive total hip replacements with a metal-on-metal bearing and a large femoral head into 92 patients. There were 51 men and 41 women with a mean age of 50 years (18 to 70) at the time of surgery. Outcome was assessed using the Western Ontario McMaster University osteoarthritis index and the Harris hip score as well as the Devane activity score. These all improved significantly (p <
0.0001). At the last follow-up there were no cases of dislocation, no impingement, a good range of movement and no osteolysis, but seven revisions, two for infection and five for aseptic loosening. The probability of groin pain increased if the other acetabular component inclination exceeded 50° (p = 0.0007). At 4.8 years of follow-up, the projected survival of the Durom acetabular component, with revision for any reason, was 92.4% ( The design of the component made it difficult both to orientate and seat, which when combined with a poor porous coating, produced unpredictable fixation and a low survival at five years.