Orthopaedic Proceedings
Vol. 86-B, Issue SUPP_I | Pages 23 - 23
1 Jan 2004
Lecuire F Benareau P Rubini J Basso M
Full Access

Purpose: The double mobility of the G. Bousquet cup—head-polyethylene,polyethylene metal cup—considerably reduces the risk of total hip arthroplasty displacement.

Material and methods: We observed seven cases of “intra-prosthetic” displacement in patients with this type of cup. Progressive wear of the polyethylene retention ring allowed displacement of the femoral head. These displacements occurred in six patients (one bilateral case) aged 43–58 years, on the average ten years after the implantation (range 3.5 – 15 years).

All patients underwent revision.

Results: The polyethylene retention ring was worn in all cases allowing the femoral head to escape. Revision surgery was performed very early in six hips to change the modular base and the polyethylene ring and re-establish retaining capacity. A satisfactory result was achieved in all cases. One patient required revision five years later to change the prosthesis subsequent to haematogenous infection. Another patient underwent a late revision to change the modular base and the entire cup of a press fit implant. This hip was in excellent condition but revision was decided on because of four dislocations.

Discussion: Intra-prosthetic displacement of the polyethylene head of a double-mobility cup is exceptional in our experience. Wear of the retention ring is favoured by different elements: 1) direct phenomena: neck and head diameter causing early contact between the neck and the cup; 2) indirect phenomena: factors limiting the mobility of the polyethylene metal cup couple (fibrosis, repeated interventions, ossification).

In our experience, intra-prosthetic displacement occurs more readily in patients with favouring conditions (alcoholism, muscle deficiency, psychiatric disorder, obesity) observed in six of our seven cases.

Conclusion: Intra-prosthetic displacement of a double-mobility cup is a possible but exceptional complication requiring a technically simple revision. Some teams use this type of cup systematically and for us is an essential element for preventive treatment in high-risk patients and for curative revision of recurrent dislocation.