Advertisement for orthosearch.org.uk
Results 1 - 2 of 2
Results per page:
Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_11 | Pages 1 - 1
1 Jul 2014
Gao B Angibaud L
Full Access

Summary Statement

Femorotibial constraint is a key property of a total knee arthroplasty (TKA) prosthesis and should reflect the intended function of the device. With a validated simulation methodology, this study evaluated the constraint of two TKA prostheses designed for different intentions.

Introduction

TKA prostheses are semi-constrained artificial joints. Femorotibial constraint level is a major property of a prosthesis and should be designed to match the device's intended function. Cruciate Retaining (CR) prostheses are usually indicated for patients with a functioning posterior cruciate ligament (PCL). For patients without a fully functioning PCL, CR-Constrained (CRC) prostheses with additional built-in constraint may be indicated. A CRC prosthesis usually consists of a CR femoral component and a tibial insert which has a more conforming sagittal profile to offer an increased femorotibial constraint. This study evaluated the anterior-posterior (AP) constraint behavior of two lines of prostheses (CR and CRC) from a same TKA product family. Using a validated computer simulation approach, multiple sizes of each product line were evaluated.


Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_11 | Pages 2 - 2
1 Jul 2014
Gao B Angibaud L
Full Access

Summary Statement

The constraint behavior of total knee arthroplasty (TKA) prosthesis usually has to be physically tested. This study presents a computer simulation model using finite element analysis (FEA) and demonstrates its effectiveness in predicting the femorotibial constraint behavior of TKA implants.

Introduction

TKA prostheses are semi-constrained artificial joints. A well-functioning TKA prosthesis should be designed with a good balance between stability and mobility, meaning the femorotibial constraint of the artificial joint cannot be excessive or too lax. To assess the constraint behavior of a TKA prosthesis, physical testing is usually required, and an industrial test standard has been developed for this purpose. Benefiting from technological advancement, computer simulation has become increasingly useful in many industries, including medical device research and development. FEA has been extensively used in stress analysis and structural evaluation of various orthopaedic implants. This study presented an FEA-based simulation to evaluate the femorotibial constraint behavior of TKA prosthesis, and demonstrated the effectiveness of the method by validating it through physical testing.