Advertisement for orthosearch.org.uk
Results 1 - 1 of 1
Results per page:
Applied filters
Include Proceedings
Dates
Year From

Year To
Orthopaedic Proceedings
Vol. 91-B, Issue SUPP_I | Pages 27 - 27
1 Mar 2009
DAMBREVILLE A PHILIPPE M AMEIL M
Full Access

The topic of this study was to research the survival rate of ATLAS hip prosthesis (acetabular cup) performed by one operating surgeon only and with a minimum of 10 years follow up.

Material and method: The ATLAS cup was hemispheric, in titanium alloy TA6V4, with a thin layer (2.5mm). It had a large central orifice as well as a radial slot allowing a certain elasticity enabling a pressfit impaction. Between January 1989 and December 1995, 297 ATLAS were implanted in the Clinique Saint André of Reims, 171 ATLAS II non hydroxyapatite coated and 126 ATLAS III hydroxyapatite coated. There were 176 women (59,2%) and 121 men (40,8%). The average age was 66 (20 to 94).

Each patient was contacted by phone to find out if the prosthesis was still in place or whether a new operation had been performed. For the deceased patients, the family or the usual doctor were contacted by phone to answer the question with a maximum of details.

The non-parametric survival rates were performed using the actuarial method according to Kaplan-Meier. The results were given with a reliability rate of 95%. The PRISM program was used.

RESULTS: On the 1st of January 2006, out of the 297 patients present at the start, 38 patients (12,8%) could not be contacted, 70 patients (23,5%) had died, 35 were reviewed and 120 answered the phone inquiry.

32 patients had undergone a revision: 23 due to the cup: there were 15 cases of wear of the polyethylene, 4 osteolysis, 3 cases of recurring dislocation and a secondary tilt of the cup, 5 cases of acetabular and femoral revision for 4 femoral loosening (change of cup by principal) and 4 cases of revision of the femoral components only.

The survival rate of the global series of 297 ATLAS (coated and non coated with hydroxyapatite) taking into consideration only the revisions due to the acetabular cup (wear of polyethylene, wrong position) was evaluated at 90% after 10 years and 85,5% after 15 years.

The results were better for the ATLAS III coated in hydroxyapatite: 92,3% after 10 years and 88,4% after 14 years, which confirms the advantages of this surface treatment. In this series, the revision rate for wear of the polyethylene was less important (3%)

With a maximum of 17 years follow up no mobilisation of the insert in the cup had been observed and no metallosis.

Conclusion: The survival rate of the global series of 297 ATLAS (coated and non coated with hydroxyapatite), taking into consideration only the revisions due to the cup (wear of the polyethylene, wrong position) was of 90% after 10 years and 85,5% after 15 years.

The rate was higher for the ATLAS III coated with hydroxyapatite: 92,3% after 10 years and 88,4 % after 14 years, which confirmed the advantages of this surface treatment.