Aims. This study reports the ten-year wear rates, incidence of osteolysis, clinical outcomes, and complications of a multicentre randomized controlled trial comparing oxidized
We investigated the wear characteristics and clinical performance of four different total hip joint articulations in 114 patients. Wear and migration was measured by roentgenstereophotogrammetric analysis at five years or at the last follow-up. The mean annual wear was 0.11 mm for a stainless steel/Enduron articulation, 0.34 mm for stainless steel/Hylamer cup, 0.17 mm for
Aims. The aim of this study was to evaluate fretting and corrosion in retrieved oxidized
Oxidized
The purpose of this study was to precisely measure the 10-year polyethylene wear rate of primary total hips using Radiostereometric analysis (RSA) comparing Oxidized
Oxidised
This study reports the ten-year polyethylene liner wear rates, incidence of osteolysis, clinical outcomes and complications of a three-arm, multicentre randomised controlled trial comparing Cobalt-Chrome (CoCr) and Oxidised
This study reports the ten-year outcomes of a three-arm, multicentre randomised controlled trial comparing Cobalt-Chrome (CoCr) and Oxidised
We investigated the revision rates of primary total hip replacement (THR) reported in the National Joint Registry (NJR) by types of bearing surfaces used. We analysed THR procedures across all orthopaedic units in England and Wales. Our analyses estimated all-cause and cause-specific revision rates. We identified primary THRs with heads and monobloc cups or modular acetabular component THRs with head and shell/liner combinations. We used flexible parametric survival models to estimate adjusted hazard ratios (HR). A total of 1,026,481 primary THRs performed between 2003–2019 are included in the primary analysis (Monobloc: n=378,979 and Modular: n=647,502) with 20,869 (2%) of these primary THRs subsequently undergoing a revision episode (Monobloc: n=7,381 and Modular: n=13,488). Compared to implants with a cobalt chrome head and highly crosslinked polyethylene (HCLPE) cup, the all-cause risk of revision for monobloc acetabular implant was higher for patients with cobalt chrome or stainless steel head and non-HCLPE cup. The risk of revision was lower for patients with a delta ceramic head and HCLPE cup implant, at any post-operative period. Compared to patients with a cobalt chrome head and HCLPE liner primary THR, the all-cause risk of revision for modular acetabular implant varied non-constantly. THRs with a delta ceramic or oxidised
Hip bearing surfaces materials are typically broadly reported in national registry (metal-on-polyethylene, ceramic-on-ceramic etc). We investigated the revision rates of primary total hip replacement (THR) reported in the National Joint Registry (NJR) by detailed types of bearing surfaces used. We analysed THR procedures across all orthopaedic units in England and Wales. Our analyses estimated all-cause and cause-specific revision rates. We identified primary THRs with heads and monobloc cups or modular acetabular component THRs with detailed head and shell/liner bearing material combinations. We used flexible parametric survival models to estimate adjusted hazard ratios (HR). A total of 1,026,481 primary THRs performed between 2003–2019 were included in the primary analysis (Monobloc cups: n=378,979 and Modular cups: n=647,502) with 20,869 (2%) of these primary THRs subsequently undergoing a revision episode (Monobloc: n=7,381 and Modular: n=13,488). Compared to implants with a cobalt chrome head and highly crosslinked polyethylene (HCLPE) cup, the overall risk of revision for monobloc acetabular implant was higher for patients with cobalt chrome or stainless steel head and non-HCLPE cup. The risk of revision was lower for patients with a delta ceramic head and HCLPE cup implant, at any post-operative period. Compared to patients with a cobalt chrome head and HCLPE liner primary THR, the overall risk of revision for modular acetabular implant varied non-constantly. THRs with a delta ceramic or oxidised
We report the five-year outcome of a randomised
controlled trial which used radiostereometric analysis (RSA) to assess
the influence of surface oxidised
Head-taper corrosion is a cause of failure in total hip arthroplasty (THA). Recent reports have described an increasing number of V40 taper failures with adverse local tissue reaction (ALTR). However, the real incidence of V40 taper damage and its cause remain unknown. The aim of this study was to evaluate the long-term incidence of ALTR in a consecutive series of THAs using a V40 taper and identify potentially related factors. Between January 2006 and June 2007, a total of 121 patients underwent THA using either an uncemented (Accolade I, made of Ti12Mo6Zr2Fe; Stryker, USA) or a cemented (ABG II, made of cobalt-chrome-molybdenum (CoCrMo); Stryker) femoral component, both with a V40 taper (Stryker). Uncemented acetabular components (Trident; Stryker) with crosslinked polyethylene liners and CoCr femoral heads of 36 mm diameter were used in all patients. At a mean folllow-up of 10.8 years (SD 1.1), 94 patients (79%) were eligible for follow-up (six patients had already undergone a revision, 15 had died, and six were lost to follow-up). A total of 85 THAs in 80 patients (mean age 61 years (24 to 75); 47 (56%) were female) underwent clinical and radiological evaluation, including the measurement of whole blood levels of cobalt and chrome. Metal artifact reduction sequence MRI scans of the hip were performed in 71 patients.Aims
Methods
Highly cross-linked polyethylene (HXLPE) has greatly improved the durability of total hip arthroplasty (THA) in young patients because of its improved wear characteristics. Few studies have followed this population into the second decade, and therefore the purpose of this investigation was to evaluate the clinical outcome for THA patients 50 years of age and younger at a minimum of 15 years postoperatively. The second purpose was to evaluate the radiological findings secondary to wear or mechanical failure of the implant. Between October 1999 and December 2005, 105 THAs were performed in 95 patients (53 female, 42 male) aged 50 years and younger (mean 42 years (20 to 50)). There were 87 patients (96 hips) that were followed for a minimum of 15 years (mean 17.3 years (15 to 21)) for analysis. Posterior approach was used with cementless fixation with a median head size of 28 mm. HXLPE was the acetabular bearing for all hips. Radiographs were evaluated for polyethylene wear, radiolucent lines, and osteolysis.Aims
Methods
Dual mobility implants in total hip arthroplasty are designed to increase the functional head size, thus decreasing the potential for dislocation. Modular dual mobility (MDM) implants incorporate a metal liner (e.g. cobalt-chromium alloy) in a metal shell (e.g. titanium alloy), raising concern for mechanically assisted crevice corrosion at the modular liner-shell connection. We sought to examine fretting and corrosion on MDM liners, to analyze the corrosion products, and to examine histologically the periprosthetic tissues. A total of 60 retrieved liners were subjectively scored for fretting and corrosion. The corrosion products from the three most severely corroded implants were removed from the implant surface, imaged using scanning electron microscopy, and analyzed using Fourier-transform infrared spectroscopy.Aims
Methods
Early evidence has emerged suggesting that ceramic-on-ceramic
articulations induce a different tissue reaction to ceramic-on-polyethylene
and metal-on-metal bearings. Therefore, the aim of this study was
to investigate the tissue reaction and cellular response to ceramic
total hip arthroplasty (THA) materials We investigated tissue collected at revision surgery from nine
ceramic-on-ceramic articulations. we compared our findings with
tissue obtained from five metal-on-metal THA revisions, four ceramic-on-polyethylene
THAs, and four primary osteoarthritis synovial membranes. The latter
were analyzed to assess the amount of tissue fibrosis that might
have been present at the time of implantation to enable evaluation,
in relation to implantation time, of any subsequent response in
the tissues.Aims
Patients and Methods
To present a surgically relevant update of trunnionosis. Systematic review performed April 2017.Aims
Materials and Methods
We reviewed the literature on the currently available
choices of bearing surface in total hip replacement (THR). We present
a detailed description of the properties of articulating surfaces
review the understanding of the advantages and disadvantages of
existing bearing couples. Recent technological developments in the
field of polyethylene and ceramics have altered the risk of fracture
and the rate of wear, although the use of metal-on-metal bearings has
largely fallen out of favour, owing to concerns about reactions
to metal debris. As expected, all bearing surface combinations have
advantages and disadvantages. A patient-based approach is recommended,
balancing the risks of different options against an individual’s
functional demands. Cite this article:
The widely used and well-proven Palacos R (a.k.a. Refobacin Palacos
R) bone cement is no longer commercially available and was superseded
by Refobacin bone cement R and Palacos R + G in 2005. However, the
performance of these newly introduced bone cements have not been
tested in a phased evidence-based manner, including roentgen stereophotogrammetric
analysis (RSA). In this blinded, randomised, clinical RSA study, the migration
of the Stanmore femoral component was compared between Refobacin
bone cement R and Palacos R + G in 62 consecutive total hip arthroplasties.
The primary outcome measure was femoral component migration measured
using RSA and secondary outcomes were Harris hip score (HHS), Hip
disability and Osteoarthritis Outcome Score (HOOS), EuroQol 5D (EQ-5D)
and Short Form 36 (SF-36).Aims
Patients and Methods
The aim of this study was to obtain detailed long-term data on the cement-bone interface in patients with cemented stems, implanted using the constrained fixation technique. A total of eight stems were removed together with adjacent bone during Our results confirm that a complete cement mantle is not essential for the survival of Müller straight stems into the mid term, and support our hypothesis that no benefit to long-term survival can be expected from modern cementing techniques.
Following the recall of modular neck hip stems
in July 2012, research into femoral modularity will intensify over
the next few years. This review aims to provide surgeons with an
up-to-date summary of the clinically relevant evidence. The development
of femoral modularity, and a classification system, is described.
The theoretical rationale for modularity is summarised and the clinical
outcomes are explored. The review also examines the clinically relevant problems
reported following the use of femoral stems with a modular neck. Joint replacement registries in the United Kingdom and Australia
have provided data on the failure rates of modular devices but cannot
identify the mechanism of failure. This information is needed to
determine whether modular neck femoral stems will be used in the
future, and how we should monitor patients who already have them implanted. Cite this article: