Bone & Joint Open
Vol. 5, Issue 7 | Pages 592 - 600
18 Jul 2024
Faschingbauer M Hambrecht J Schwer J Martin JR Reichel H Seitz A

Aims. Patient dissatisfaction is not uncommon following primary total knee arthroplasty. One proposed method to alleviate this is by improving knee kinematics. Therefore, we aimed to answer the following research question: are there significant differences in knee kinematics based on the design of the tibial insert (cruciate-retaining (CR), ultra-congruent (UC), or medial congruent (MC))?. Methods. Overall, 15 cadaveric knee joints were examined with a CR implant with three different tibial inserts (CR, UC, and MC) using an established knee joint simulator. The effects on coronal alignment, medial and lateral femoral roll back, femorotibial rotation, bony rotations (femur, tibia, and patella), and patellofemoral length ratios were determined. Results. No statistically significant differences were found regarding coronal alignment (p = 0.087 to p = 0.832). The medial congruent insert demonstrated restricted femoral roll back (mean medial 37.57 mm; lateral 36.34 mm), while the CR insert demonstrated the greatest roll back (medial 42.21 mm; lateral 37.88 mm; p < 0.001, respectively). Femorotibial rotation was greatest with the CR insert with 2.45° (SD 4.75°), then the UC insert with 1.31° (SD 4.15°; p < 0.001), and lowest with the medial congruent insert with 0.8° (SD 4.24°; p < 0.001). The most pronounced patella shift, but lowest patellar rotation, was noted with the CR insert. Conclusion. The MC insert demonstrated the highest level of constraint of these inserts. Femoral roll back, femorotibial rotation, and single bony rotations were lowest with the MC insert. The patella showed less shifting with the MC insert, but there was significantly increased rotation. While the medial congruent insert was found to have highest constraint, it remains uncertain if this implant recreates native knee kinematics or if this will result in improved patient satisfaction. Cite this article: Bone Jt Open 2024;5(7):592–600


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_IX | Pages 95 - 95
1 Mar 2012
Wilding CP Maruthainar K Malikian R Stammers J Blunn GW
Full Access

Purpose

To determine the effect that Titanium Nitride (TiN) coatings have on wear rates of ultra high molecular weight polyethylene (UHMWPE)

Background

Ceramic coatings have been shown to confer advantageous countersurface scratch resistance in knee arthroplasty. This may reduce UHMWPE wear rates and revision rates. Dermal hypersensitivity is a common problem with metals; TiN, a ceramic surface, has been used to prevent it. There is little data in the literature regarding the effect of TiN on UHMWPE.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_13 | Pages 58 - 58
7 Aug 2023
Saghir R Watson K Martin A Cohen A Newman J Rajput V
Full Access

Abstract. Introduction. Knee arthroscopy can be used for ligamentous repair, reconstruction and to reduce burden of infection. Understanding and feeling confident with knee arthroscopy is therefore a highly important skillset for the orthopaedic surgeon. However, with limited training or experience, furthered by reduced practical education due to COVID-19, this skill can be under-developed amongst trainee surgeons. Methods. At a single institution, ten junior doctors (FY1 to CT2), were recruited as a part of a five, two-hour session, training programme utilising the Simbionix® ARTHRO Mentor knee arthroscopy simulator, supplemented alongside educational guidance with a consultant orthopaedic knee surgeon. All students had minimal to no levels of prior arthroscopic experience. Exercises completed included maintaining steadiness, image centering and orientation, probe triangulation, arthroscopic knee examination, removal of loose bodies, and meniscectomy. Pre and post-experience questionnaires and quantitative repeat analysis on simulation exercises were undertaken to identify levels of improvement. Results. Comparing pre and post-experience questionnaires significant improvements in levels of confidence were noted in the following domains: naming arthroscopic instruments, port positioning and insertion, recognising normal anatomy arthroscopically, holding and using arthroscopic instruments and assisting in a live theatre setting (p<0.05). Significant improvements were noted in time taken to complete, distance covered in metres and roughness of instruments used on the simulated exercises on repeat performance (p<0.05). Conclusion. With only five sessions under senior guidance, using a simulator such as the ARTHRO Mentor, significant improvements in both levels of confidence and skill can be developed even among individuals with no prior experience


The Bone & Joint Journal
Vol. 103-B, Issue 6 Supple A | Pages 94 - 101
1 Jun 2021
Roy ME Whiteside LA Ly KK Gauvain MJ

Aims. The aims of this study were to evaluate wear on the surface of cobalt-chromium (CoCr) femoral components used in total knee arthroplasty (TKA) and compare the wear of these components with that of ceramic femoral components. Methods. Optical profilometry was used to evaluate surface roughness and to examine the features created by the wear process in a knee wear simulator. We developed a method of measuring surface changes on five CoCr femoral components and quantifying the loss of material from the articular surface during the wear process. We also examined the articular surface of three ceramic femoral components from a previous test for evidence of surface damage, and compared it with that of CoCr components. Results. We found that the surface roughness of CoCr components rapidly increased during the first 1,000 wear cycles, then reached a steady state, but material loss from the surface continued at a rate of 1,778,000 μm. 3. per million cycles as carbides were removed from its matrix. These carbides formed third-body wear particles, leading to the formation of new scratches even as older scratches were worn away. In contrast, no scratching, loss of material, or other surface damage, when evaluated with one nanometer resolution, was found on the surface of the ceramic components after a 15 M wear cycle test. Conclusion. This study showed wear and loss of CoCr material from scratching and microabrasive wear in TKA. The material loss from the surface continued in a linear relationship with increasing cycles. We also found the absence of scratching and roughening of ceramic femoral components in simulated wear, suggesting an advantage in wear rate and avoiding metal sensitivity. This may have implications in the management of persistent pain after TKA. Cite this article: Bone Joint J 2021;103-B(6 Supple A):94–101


The Bone & Joint Journal
Vol. 102-B, Issue 11 | Pages 1527 - 1534
3 Nov 2020
Orita K Minoda Y Sugama R Ohta Y Ueyama H Takemura S Nakamura H

Aims. Vitamin E-infused highly cross-linked polyethylene (E1) has recently been introduced in total knee arthroplasty (TKA). An in vitro wear simulator study showed that E1 reduced polyethylene wear. However there is no published information regarding in vivo wear. Previous reports suggest that newly introduced materials which reduce in vitro polyethylene wear do not necessarily reduce in vivo polyethylene wear. To assist in the evaluation of the newly introduced material before widespread use, we established an in vivo polyethylene wear particle analysis for TKA. The aim of this study was to compare in vivo polyethylene wear particle generation between E1 and conventional polyethylene (ArCom) in TKA. Methods. A total of 34 knees undergoing TKA (17 each with ArCom or E1) were investigated. Except for the polyethylene insert material, the prostheses used for both groups were identical. Synovial fluid was obtained at a mean of 3.4 years (SD 1.3) postoperatively. The in vivo polyethylene wear particles were isolated from the synovial fluid using a previously validated method and examined by scanning electron microscopy. Results. The total number of polyethylene wear particles obtained from the knees with E1 (mean 6.9, SD 4.0 × 10. 7. counts/knee) was greater than that obtained from those with ArCom (mean 2.2, SD 2.6 × 10. 7. counts/knee) (p = 0.001). The particle size (equivalent circle of diameter) from the knees with E1 was smaller (mean 0.5 μm, SD 0.1) than that of knees with ArCom (mean 1.5, SD 0.3 μm) (p = 0.001). The aspect ratio of particles from the knees with E1 (mean 1.3, SD 0.1) was smaller than that with ArCom (mean 1.4, SD 0.1) (p < 0.001 ). Conclusion. This is the first report of in vivo wear particle analysis of E1. E1 polyethylene did not reduce the number of in vivo polyethylene wear particles compared with ArCom in early clinical stage. Further careful follow-up of newly introduced E1 for TKA should be carried out. Cite this article: Bone Joint J 2020;102-B(11):1527–1534


Introduction. Total knee arthroplasty (TKA) femoral components are known to wear and roughen with clinical use, and reaction to metal is a well-documented complication of TKA. Ceramic materials are resistant to wear and corrosion, but the surface wear of ceramic femoral components in TKA has not been reported. This study measured the changes in roughness and mass lost from cobalt-chromium (CoCr) femoral components tested in a knee simulator, and compared them to those observed in ceramic components. Methods. Six cast CoCr femoral components were tested in a knee simulator bearing against UHMWPE, with high-kinematics/high-load waveforms. Roughness and scratches were measured via optical profilometry. Scratch volume was equated to mass lost from abrasive wear, while nano-particulate wear and corrosion were estimated from the change in the distance of the surface to the depth of the scratches after increasing numbers of cycles. Three magnesia-stabilized zirconia femoral components that had been wear tested to 15 million cycles were compared with the CoCr components. Results. After 500,000 cycles, the CoCr femoral components were visibly scratched and measurably roughened (average roughness Sa = 38.4nm, vs. 22.2nm for never-implanted components; p < 0.0001). The ceramic components were not scratched or roughened (Sa = 18.5nm). The CoCr components had an average of 7.5μg lost due to scratches (0.0018mm. 3. per million cycles). The material lost due to nanoparticulate wear and corrosion was estimated at 1.5μg. Discussion. This study demonstrates rapid loss of CoCr material from wear in TKA, and reports absence of scratching and roughening of ceramic TKA femoral components in simulated wear. This observation suggests a mechanism for the lower polyethylene wear rate against ceramic femoral components, and suggests an advantage in avoiding metal sensitivity


Bone & Joint Research
Vol. 12, Issue 4 | Pages 285 - 293
17 Apr 2023
Chevalier A Vermue H Pringels L Herregodts S Duquesne K Victor J Loccufier M

Aims

The goal was to evaluate tibiofemoral knee joint kinematics during stair descent, by simulating the full stair descent motion in vitro. The knee joint kinematics were evaluated for two types of knee implants: bi-cruciate retaining and bi-cruciate stabilized. It was hypothesized that the bi-cruciate retaining implant better approximates native kinematics.

Methods

The in vitro study included 20 specimens which were tested during a full stair descent with physiological muscle forces in a dynamic knee rig. Laxity envelopes were measured by applying external loading conditions in varus/valgus and internal/external direction.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 3 | Pages 344 - 350
1 Mar 2009
Luyckx T Didden K Vandenneucker H Labey L Innocenti B Bellemans J

The purpose of this study was to test the hypothesis that patella alta leads to a less favourable situation in terms of patellofemoral contact force, contact area and contact pressure than the normal patellar position, and thereby gives rise to anterior knee pain. A dynamic knee simulator system based on the Oxford rig and allowing six degrees of freedom was adapted in order to simulate and record the dynamic loads during a knee squat from 30° to 120° flexion under physiological conditions. Five different configurations were studied, with variable predetermined patellar heights. The patellofemoral contact force increased with increasing knee flexion until contact occurred between the quadriceps tendon and the femoral trochlea, inducing load sharing. Patella alta caused a delay of this contact until deeper flexion. As a consequence, the maximal patellofemoral contact force and contact pressure increased significantly with increasing patellar height (p < 0.01). Patella alta was associated with the highest maximal patellofemoral contact force and contact pressure. When averaged across all flexion angles, a normal patellar position was associated with the lowest contact pressures. Our results indicate that there is a biomechanical reason for anterior knee pain in patients with patella alta


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_11 | Pages 48 - 48
1 Oct 2019
Jacobs JJ Knowlton CB Lundberg HJ Wimmer MA
Full Access

Introduction. Studies of retrieved total knee replacement (TKR) components demonstrate that in vivo wear on the articular surface of polyethylene liners exhibits a much higher variability than their in vitro counterparts tested on simulators. 1. Only one study has attempted to validate a patient-specific model of wear with a clinically retrieved component. 2. The purpose of this study is to investigate the relationship between observed TKR contact conditions during gait and measured volume loss on retrieved tibial components. Methods. Eleven retrieved ultra-high molecular weight polyethylene (UHMWPE) cruciate-retaining tibial liner components from ten separate patients (implantation time = 8.6±5.6 years) had matching gait trials of normal level walking for each knee. Volume loss on retrieved components was calculated using a coordinate measuring machine and autonomous reconstruction. 3. Motion analysis of normal level walking gait had been conducted between 1986 and 2005 for various previous studies and stored in a consented Human Mechanics Repository, ranging from pre-operative to long-term post-operative testing. Contact location between the femoral component and the tibial component on the medial and lateral plateaus were calculated throughout stance. 4. A previously validated and fine-tuned parametric numerical model was used to calculate TKR contact forces for each gait trial. 5. Vertical contact forces and contact paths on the medial and lateral plateaus were input as normal force and sliding distance to a simplified Archard equation for wear with material wear constant = 2.42 × 10. −7. mm. 3. /Nm. 2,6. to compute average wear per gait cycle. Wear rates were calculated using linear regression, and Pearson correlation examined correlations between modeled and measured wear. Results. Secondary motions at the knee from gait testing showed distinct grouping between trials of each patient (Fig. 1). Three components demonstrated severe polyethylene delamination and were excluded from wear rate analyses. All calculated wear rates for measured and modeled volume loss, shown in Fig. 2, showed excellent agreement and were not significantly different (Table 1). Measured wear rates were comparable to a previous study of a large population of retrieved Miller-Galante II components. 7. As seen in Fig 2b, medial wear volumes for six of eight mild wearing components were closely tracked by their modeled counterparts. Volumes were significantly correlated between measured and modeled wear for the total part and on the medial side, but not for the lateral side (Table 1). Conclusion. Because the Archard equation produces wear volumes that are linearly related to time in situ, deviations from linear predictions arise from patient-specific variations in contact forces and tibiofemoral pathways during normal walking gait. As suggested by the results of the current study, these variations in gait between patients result in meaningful differences to the wear of the UHMWPE component. Despite many assumptions, this study demonstrates the feasibility of a patient-specific model of wear using a rare population of gait-matched retrievals. This study suggests that gait analysis may play an important role in individualized medicine in orthopedics. For figures, tables, or references, please contact authors directly


The Bone & Joint Journal
Vol. 103-B, Issue 11 | Pages 1695 - 1701
1 Nov 2021
Currier JH Currier BH Abdel MP Berry DJ Titus AJ Van Citters DW

Aims

Wear of the polyethylene (PE) tibial insert of total knee arthroplasty (TKA) increases the risk of revision surgery with a significant cost burden on the healthcare system. This study quantifies wear performance of tibial inserts in a large and diverse series of retrieved TKAs to evaluate the effect of factors related to the patient, knee design, and bearing material on tibial insert wear performance.

Methods

An institutional review board-approved retrieval archive was surveyed for modular PE tibial inserts over a range of in vivo duration (mean 58 months (0 to 290)). Five knee designs, totalling 1,585 devices, were studied. Insert wear was estimated from measured thickness change using a previously published method. Linear regression statistical analyses were used to test association of 12 patient and implant design variables with calculated wear rate.


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_9 | Pages 1 - 1
1 Oct 2020
Springer B Haddad FS
Full Access

The COVID-19 pandemic has led to unprecedented times worldwide. From lockdowns to masks now being part of our everyday routine, to the halting of elective surgeries, the virus has touched everyone and every part of our personal and professional lives. Perhaps, now more than ever, our ability to adapt, change and persevere is critical to our survival. This year's closed meeting of The Knee Society demonstrated exactly those characteristics. When it became evident that an in-person meeting would not be feasible, The Knee Society leadership, under the direction of President John Callaghan, MD and Program Chair Craig Della Valle, MD created a unique and engaging meeting held on September 10–12, 2020. Special recognition should be given to Olga Foley and Cynthia Garcia at The Knee Society for their flexibility and creativeness in putting together a world-class flawless virtual program. The Bone & Joint Journal is very pleased to partner with The Knee Society to once again publish the proceedings of the closed meeting of the Knee Society. The Knee Society is a United States based society of highly selected members who have shown leadership in education and research in knee surgery. It invites up to 15% international members; this includes some of the key opinion leaders in knee surgery from outside the USA. Each year, the top research papers from The Knee Society meeting will be published and made available to the wider orthopaedic community in The Bone & Joint Journal. The first such proceedings were published in BJJ in 2019. International dissemination should help to fulfil the mission and vision of the Knee Society of advancing the care of patients with knee disorders through leadership, education and research. The quality of dissemination that The Bone & Joint Journal provides should enhance the profile of this work and allow a larger body of surgeons, associated healthcare professionals and patients to benefit from the expertise of the members of The Knee Society. The meeting is one of the highlights of the annual academic calendar for knee surgeons. With nearly every member in attendance virtually throughout the 3 days, the top research papers from the membership were presented and discussed in a virtual format that allowed for lively interaction and discussion. There are 75 abstracts presented. More selective proceedings with full papers will be available after a robust peer review process in 2021, both online and in The Bone & Joint Journal. The meeting commenced with the first group of scientific papers focused on Periprosthetic Joint Infection. Dr Berry and colleagues from the Mayo Clinic further help to clarify the issue of serology and aspirate results to diagnose TKA PJI in the acute postoperative setting. 177 TKA's had an aspiration within 12 weeks and 22 were proven to have PJI. Their results demonstrated that acute PJI after TKA should be suspected within 6 weeks if CRP is ≥81 mg/L, synovial WBCs are ≥8500 cells/μL, and/or synovial neutrophils≥86%. Between 6– 12 weeks, concerning thresholds include a CRP ≥ 32 mg/L, synovial WBC ≥7450, and synovial neutrophils ≥ 84%. While historically the results of a DAIR procedure for PJI have been variable, Tom Fehring's study showed promise with the local delivery of vancomycin through the Intraosseous route improved early results. New member Simon Young contrasted the efficacy of the DAIR procedure when comparing early infections to late acute hematogenous PJI. DAIR failed in 63% of late hematogenous PJIs (implant age>1 year) compared to 36% of early (<1year) PJIs. Dr Masri demonstrated in a small group of patients that those with well-functioning articulating spacers can retain their spacers for over 12 months with no difference in infection from those that had a formal two stage exchange. The mental toll of PJI was demonstrated in a longitudinal study by Doug Dennis, where patient being treated with 2 stage exchange had 4x higher rates of depression compared to patient undergoing aseptic revision. The second session focused on both postoperative issues with regards to anticoagulation and manipulation. Steven Haas demonstrated high complication rates with utilization of anticoagulation for treatment of postoperative pulmonary embolism with modern therapeutic anticoagulation (warfarin, enoxaparin, Xa inhibitors) with the Xa inhibitors demonstrating lower complication rates. Two papers focused on the topic of manipulation. Mark Pagnano presented data on timing of manipulation under anesthesia up to even past 12 months. While gains were modest, a subset of patients did achieve substantial gains in ROM > 20degrees even after 3 months post op. Dr Westrich's study demonstrated no difference in MUA outcomes with either IV sedation or neuraxial anesthesia although the length of stay was shorter in the IV sedation group. Several studies in Session II focused on kinematics and femoral component position. Dr Li's in vivo kinematic study during weightbearing flexion and gait demonstrated that several knees rotated with a lateral pivot motion and not all knees can be described with a single motion character. Dr Mayman and his group utilized a computational knee model to demonstrate that additional distal femoral resection results in increasing levels of mid -flexion instability and cautioned against the use of additional bony resection as the first line for flexion contractures. Using computer navigation, Dr Huddleston's study nicely outlined the variability in femoral component rotation to achieve a rectangular flexion gap utilizing a gap balanced method. The third session opened the meeting on Friday morning. The focus was on unicompartmental knee arthroplasty and the increasing utilization of robotic assisted total knee arthroplasty. David Murray showed using registry data that for patient with higher comorbidities (ASA >3), UKA was safer and more cost effective than TKA while Dr Della Valle's group demonstrated overall lower average healthcare costs in UKA patients compared to TKA in the first 10 years after surgery. Dr Geller assessed UKA survivorship among 3 international registries. While survivorship varied by nation and designs, certain designs consistently had better overall performance. Dr Nunley and his group showed robotic navigation UKA significantly reduced outliers in alignment and overhang compared to manual UKA. Dr Catani's data demonstrated that full thickness cartilage loss should still be considered a requirement for UKA success even with robotic assistance. Despite a high dislocation rate of 4%, Mr Dodd demonstrated high survivorship for lateral UKA despite historical contraindications. The growing evidence for robotics TKA was demonstrated in two studies. Professor Haddad showed less soft tissue injury, reduced bone trauma and improved accuracy or rTKA compared to manual TKA while Dr Gustke single surgeon study showed his rTKA had improved forgotten joint scores and less ligament releasing required for balancing. Despite these finding, Dr Lee's study demonstrated that a robotic TKA could not guarantee excellent pain relief and other factors such a patient expectations and psychological factors play a role. Our fourth session was devoted to machine learning and smart tools and modeling. Dr Meneghini used machine learning algorithms to identify optimal alignment outcomes that correlated with patient outcomes. Several parameters such as native tibial slope, femoral sagittal position and coronal limb alignment correlated with outcomes. Along the same lines, Bozic and coauthors demonstrated that using AI algorithms incorporated with PROM's improved levels of shared decision making and patient satisfaction. Dr Lombardi demonstrated that a mobile patient engagement platform that provided smart phone-based exercise and education was comparable to traditional methods. Dr Mahfouz demonstrated the accuracy of using ultrasound to produce 3D models of the bone compared to conventional CT based strategies and Dr Mahoney showed the valued of a preop 3D model in reproducing more normal knee kinematics. The last two talks of the session focused on some of the positives of the COVID-19 pandemic, namely the embracing of telemedicine by patients and surgeons as demonstrated by Dr Slover and the increasing and far reaching educational opportunities made available to residents and fellows during the pandemic. Session five focused on risk stratification and optimization prior to TKA. Dr O'Connor demonstrated that that the implementation of an optimization program preoperatively reduced length of stay and ED visits, and Charles Nelson's study showed that risk stratification tool can lower complication rates in obese patients undergoing TKA comparable to those that are nonobese. Dr Markel's study demonstrated that those who have preoperative depression and anxiety are at higher risk of complications and readmissions after surgery and these issues should be addressed preoperatively. Interestingly, a study by Dr Callaghan demonstrated that care improvement pathways have not lowered the gap in complications for morbidly obese patients undergoing TKA, Dr Barsoum argued that the overall complication rates were low and this patient cohort had significant gains in PROMS after TKA that would not be experienced if arbitrary cutoff for limited surgery were established. The final session on Friday, Session six, had several well done and interesting studies. There continues to be mounting evidence that liposomal bupivacaine has little effect on managing post-operative pain to warrant its increased use. Bill Macaulay and colleagues showed no change in pain scores, opioid consumption and functional scores when liposomal bupivacaine was discontinued at a large academic medical center. Dr Bugbee importantly demonstrated that a supervised ambulation program reduced falls in the early postoperative period. Several paper on healthcare economics were presented. Rich Iorio showed that stratifying complexity of total joint cases between hospitals with a system can be efficient and cost savings while Dr Jiranek demonstrated in his study that complex TKAs can be identified preoperatively and are associated with prolonged operative time and cost of care and consideration should be given in future reimbursement models to a complexity modifier. Dr Springer, in their evaluation of Medicare bundled payment models, demonstrated that providers and hospitals in historical bundled models that became efficient were penalized in the new model, forcing many groups to drop out and return to a fee for service model. Ron Delanois important work showed that social determinants can have a major negative impact on outcomes following TKA. Our final day on Saturday opened with Session seven, and several interesting paper on metal ions/debris in TKA. Dr Whitesides simulator study showed the absence of scratches and material loss in a ceramic TKA compared with Co-Cr TKA and suggested an advantage to this material in patients with metal sensitivity. Conversely, in a histological study of failed TKA, perivascular lymphocytic infiltration was not associated with worse clinical outcomes or differences in revision in a series of 617 aseptic revisions, 19% of which had PVLI found on histology. The Mayo group and Dr Trousdale however, noted that serum metal ion levels can be helpful in identifying implant failure in a group of revision TKAs, especially those with metallic junctions. Dr Dalury demonstrated nicely that use of maximally conforming inserts did not have a negative effect on implant loosening in a series of 76 revision TKA's at an average follow up of 7 years, while Kevin Garvin and his group showed no difference in end of stem pain between cemented and cementless stems in revision TKA. The final two studies in the session by Bolognesi and Peters respectively showed that metaphyseal cones continue to demonstrate excelled survivorship in rTKA setting despite extensive bone loss. Session eight was highlighted by a large series of revision reported by new member Dr Schwarzkopf, who showed that revision TKA done by high volume surgeons demonstrated better outcomes and lower revision rates compared to surgeon who did less than 18 rTKA's per year. Dr Maniar importantly showed that preoperatively, patients with high activity level and low pain and indicated by a high preop forgotten joint score did poorly following TKA while David Ayers nicely demonstrated that KOOS scores that assess specific postoperative outcomes can predict patient dissatisfaction after TKA. The final paper in this session by Max Courtney showed that the majority of surgical cancellations are due to medical issues, yet a minority of these undergo any intervention specifically for that condition, but they resulted in a delay of 5 months. The first two studies of Session nine focused on polyethylene thickness. Dr Backstein demonstrated no difference in KSS scores, change in ROM and aseptic revision rates based on polyethylene thickness in a series of 195 TKA's. An interesting lab study by Dr Tim Wright showed a surprising consistency in liner thickness choice among varying levels of surgeon experience that did not correlate with applied forces or gap stability estimates. Two studies looked specifically at the issue of tibial loosening and implant design. Nam and colleagues were not able to demonstrate concerning findings for increasing tibial loosening in a tibial baseplate with a shortened tibial keel at short term follow up, while Lachiewicz demonstrated a 19% revision or revision pending rate in 223 cemented fixed bearing ATTUNE TKA at a mean of 30 months. Our final session of the meeting, began with encouraging news, that despite only currently capturing about 40% of TJA's done in the US, the American Joint Replacement Registry data is representative of data in other representative US databases. An interesting study presented by Robert Barrack looked at bone remodeling in the proximal tibia after cemented and cementless TKA of two different designs. No significant difference was noted among the groups with the exception of the cemented thicker cobalt chrome tray which demonstrated significantly more bone mineral density loss. Along the same lines, a study out of Dr Bostrom's lab demonstrated treatment of a murine tibial model with iPTH prevents fibrous tissue formation and enhances bone formation in cementless implants. New Member Jamie Howard showed no difference in implant migration and kinematics of a single radius cementless design using either a measured resection or gap balancing technique and Dr Cushner show no difference in blood loss with cemented or cementless TKA with the use of TKA. The final two studies looked at staging and bilateral TKA's. Peter Sharkey showed that simultaneous TKA's were associated with higher complication compared to staged TKA and that staged TKA with less than a 90-day interval was not associated with higher risk. However, Mark Figgie showed that patients undergoing simultaneous TKA compared to staged TKA, missed 17 fewer days of work. In spite of the virtual nature of the meeting, there were some outstanding scientific interactions and the material presented will continue to generate debate and to guide the direction of knee arthroplasty as we move forwards


The Bone & Joint Journal
Vol. 102-B, Issue 6 Supple A | Pages 129 - 137
1 Jun 2020
Knowlton CB Lundberg HJ Wimmer MA Jacobs JJ

Aims

A retrospective longitudinal study was conducted to compare directly volumetric wear of retrieved polyethylene inserts to predicted volumetric wear modelled from individual gait mechanics of total knee arthroplasty (TKA) patients.

Methods

In total, 11 retrieved polyethylene tibial inserts were matched with gait analysis testing performed on those patients. Volumetric wear on the articular surfaces was measured using a laser coordinate measure machine and autonomous reconstruction. Knee kinematics and kinetics from individual gait trials drove computational models to calculate medial and lateral tibiofemoral contact paths and forces. Sliding distance along the contact path, normal forces and implantation time were used as inputs to Archard’s equation of wear to predict volumetric wear from gait mechanics. Measured and modelled wear were compared for each component.


The Bone & Joint Journal
Vol. 102-B, Issue 6 Supple A | Pages 151 - 157
1 Jun 2020
Gil D Atici AE Connolly RL Hugard S Shuvaev S Wannomae KK Oral E Muratoglu OK

Aims

We propose a state-of-the-art temporary spacer, consisting of a cobalt-chrome (CoCr) femoral component and a gentamicin-eluting ultra-high molecular weight polyethylene (UHMWPE) tibial insert, which can provide therapeutic delivery of gentamicin, while retaining excellent mechanical properties. The proposed implant is designed to replace conventional spacers made from bone cement.

Methods

Gentamicin-loaded UHMWPE was prepared using phase-separated compression moulding, and its drug elution kinetics, antibacterial, mechanical, and wear properties were compared with those of conventional gentamicin-loaded bone cement.


Bone & Joint Research
Vol. 8, Issue 11 | Pages 563 - 569
1 Nov 2019
Koh Y Lee J Lee H Kim H Kang K

Objectives

Unicompartmental knee arthroplasty (UKA) is an alternative to total knee arthroplasty with isolated medial or lateral compartment osteoarthritis. However, polyethylene wear can significantly reduce the lifespan of UKA. Different bearing designs and materials for UKA have been developed to change the rate of polyethylene wear. Therefore, the objective of this study is to investigate the effect of insert conformity and material on the predicted wear in mobile-bearing UKA using a previously developed computational wear method.

Methods

Two different designs were tested with the same femoral component under identical kinematic input: anatomy mimetic design (AMD) and conforming design inserts with different conformity levels. The insert materials were standard or crosslinked ultra-high-molecular-weight polyethylene (UHMWPE). We evaluated the contact pressure, contact area, wear rate, wear depth, and volumetric wear under gait cycle loading conditions.


Bone & Joint Research
Vol. 8, Issue 2 | Pages 65 - 72
1 Feb 2019
Cowie RM Aiken SS Cooper JJ Jennings LM

Objectives

Bone void fillers are increasingly being used for dead space management in arthroplasty revision surgery. The aim of this study was to investigate the influence of calcium sulphate bone void filler (CS-BVF) on the damage and wear of total knee arthroplasty using experimental wear simulation.

Methods

A total of 18 fixed-bearing U2 total knee arthroplasty system implants (United Orthopedic Corp., Hsinchu, Taiwan) were used. Implants challenged with CS-BVF were compared with new implants (negative controls) and those intentionally scratched with a diamond stylus (positive controls) representative of severe surface damage (n = 6 for each experimental group). Three million cycles (MC) of experimental simulation were carried out to simulate a walking gait cycle. Wear of the ultra-high-molecular-weight polyethylene (UHMWPE) tibial inserts was measured gravimetrically, and damage to articulating surfaces was assessed using profilometry.


The Bone & Joint Journal
Vol. 101-B, Issue 11 | Pages 1348 - 1355
1 Nov 2019
Gascoyne T Parashin S Teeter M Bohm E Laende E Dunbar M Turgeon T

Aims

A retrospective study was conducted to measure short-term in vivo linear and volumetric wear of polyethylene (PE) inserts in 101 total knee arthroplasty (TKA) patients using model-based radiostereometric analysis (MBRSA).

Patients and Methods

Nonweightbearing supine RSA exams were performed postoperatively and at six, 12, and 24 months. Weightbearing standing RSA exams were performed on select patients at 12 and 24 months. Wear was measured both linearly (joint space) and volumetrically (digital model overlap) at each available follow-up. Precision of both methods was assessed by comparing double RSA exams. Patient age, sex, body mass index, and Oxford Knee Scores were analyzed for any association with PE wear.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXIX | Pages 74 - 74
1 Jul 2012
Al-Ali S Alvand A Gill HS Beard DJ Jackson W Price AJ Rees JL
Full Access

Rheumatology and Musculoskeletal Sciences, NIHR Biomedical Research Unit, University of Oxford and the Nuffield Orthopaedic Hospital, Oxford. Purpose. The aim of this study was to use motion analysis to objectively study the learning curve of surgical trainees performing arthroscopic meniscal repair on a training model in a skills laboratory. Background. With improving technology and an appreciation of its likely chondroprotective effects, meniscal repair surgery is becoming more common. It remains a difficult procedure and is not routinely learnt during surgical training. Methodology. 19 orthopaedic surgical trainees watched an instructional video of a meniscal repair method (Smith & Nephew Fast-Fix) and then performed 12 meniscal repair episodes on a ‘sawbones’ knee simulator with a standardised lateral meniscal tear. The 12 repair episodes were performed during over a 3 week period. A validated motion analysis system was used to record: distance travelled by each hand; number of hand movements; and time taken to complete the task. Results. Time taken, number of hand movements and total path travelled all showed improvements over the twelve episodes. Time taken improved by 34%, Total path travelled by 21%, and Hand movements improved by 27% for the camera hand (right) and 19% for the instrument hand (left). There was evidence of plateau on the learning curve over the 12 episodes, with larger improvements in measured outcomes over the initial episodes compared to the last episodes. Conclusion. This study objectively demonstrates a learning curve for surgeons performing arthroscopic meniscal repair in a skills laboratory. It indicates the benefits to surgical trainees of practicing such arthroscopic techniques in a skills centre prior to progressing to the operating theatre


Objectives

Unicompartmental knee arthroplasty (UKA) is an alternative to total knee arthroplasty for patients who require treatment of single-compartment osteoarthritis, especially for young patients. To satisfy this requirement, new patient-specific prosthetic designs have been introduced. The patient-specific UKA is designed on the basis of data from preoperative medical images. In general, knee implant design with increased conformity has been developed to provide lower contact stress and reduced wear on the tibial insert compared with flat knee designs. The different tibiofemoral conformity may provide designers the opportunity to address both wear and kinematic design goals simultaneously. The aim of this study was to evaluate wear prediction with respect to tibiofemoral conformity design in patient-specific UKA under gait loading conditions by using a previously validated computational wear method.

Methods

Three designs with different conformities were developed with the same femoral component: a flat design normally used in fixed-bearing UKA, a tibia plateau anatomy mimetic (AM) design, and an increased conforming design. We investigated the kinematics, contact stress, contact area, wear rate, and volumetric wear of the three different tibial insert designs.


The Bone & Joint Journal
Vol. 101-B, Issue 7_Supple_C | Pages 33 - 39
1 Jul 2019
Lachiewicz PF O’Dell JA

Aims

There is insufficient evidence to recommend the use of alternative polyethylene bearings in modular, fixed-bearing total knee arthroplasty (TKA). The purpose of this study was to compare standard polyethylene (SP) and highly crosslinked polyethylene (XLP) tibial liners in posterior-stabilized TKA, with osteolysis as the primary outcome and clinical results and the rate of re-operation as the secondary outcomes.

Patients and Methods

This is a single-surgeon, prospective randomized study involving one design of modular posterior-stabilized TKA. An analysis of 122 TKAs with an SP compression moulded liner and 123 with an XLP liner was performed, with a mean follow-up of six years (2 to 11). Patients were evaluated clinically using the Knee Society score, Lower Extremity Activity Score (LEAS), and the presence of an effusion, and standard radiographs were assessed for radiolucent lines and osteolytic lesions.


Bone & Joint Research
Vol. 6, Issue 1 | Pages 43 - 51
1 Jan 2017
Nakamura S Tian Y Tanaka Y Kuriyama S Ito H Furu M Matsuda S

Objectives

Little biomechanical information is available about kinematically aligned (KA) total knee arthroplasty (TKA). The purpose of this study was to simulate the kinematics and kinetics after KA TKA and mechanically aligned (MA) TKA with four different limb alignments.

Materials and Methods

Bone models were constructed from one volunteer (normal) and three patients with three different knee deformities (slight, moderate and severe varus). A dynamic musculoskeletal modelling system was used to analyse the kinematics and the tibiofemoral contact force. The contact stress on the tibial insert, and the stress to the resection surface and medial tibial cortex were examined by using finite element analysis.