Introduction. The Precice nail is the latest intramedullary lengthening nail with excellent early outcomes. Implant complications have led to modification of the nail design. The aim of this study was to perform a
Oxidized zirconium (OxZr) is used as a ceramic surface for femoral components in total knee arthroplasty (TKA). The aim of this study was to investigate its performance by examining retrieved femoral components and their corresponding PE inserts in matched comparison with conventional chrome/cobalt/molybdenum alloy (CrCoMo). 11 retrieved posterior stabilized (PS) TKA with an OxZr femoral component were included. From a cohort of 56 retrieved TKA with CrCoMo femoral components, pairs were matched according to duration of implantation, patient age, reason for revision, and BMI. The retrieved tibial polyethylene (PE) inserts were analyzed for wear using the Hood classification. Femoral components were optically viewed at 8–32x magnification and screened for scratching, pitting, delamination, and striation. Profilometry was performed to measure surface roughness of the OxZr components using a non-contact white light profiler.Introduction
Methods
Introduction. The prosthetic impingement occurs if the range of motion of the hip exceeds implant mobility or in case of component malorientation. This
Aims. The management of completely displaced fractures of the distal radius in children remains controversial. This study evaluates the outcomes of surgical and non-surgical management of ‘off-ended’ fractures in children with at least two years of potential growth remaining. Methods. A total of 34 boys and 22 girls aged 0 to ten years with a closed, completely displaced metaphyseal distal radial fracture presented between 1 November 2015 and 1 January 2020. After 2018, children aged ten or under were offered treatment in a straight plaster or manipulation under anaesthesia with Kirschner (K-)wire stabilization. Case notes and radiographs were reviewed to evaluate outcomes. In all, 16 underwent treatment in a straight cast and 40 had manipulation under anaesthesia, including 37 stabilized with K-wires. Results. Of the children treated in a straight cast, all were discharged with good range of mo (ROM). Five children were discharged at six to 12 weeks with no functional limitations at six-month follow-up. A total of 11 children were discharged between 12 and 50 weeks with a normal ROM and radiological evidence of remodelling. One child had a subsequent diaphyseal fracture proximal to the original injury four years after the initial fracture. Re-displacement with angulation greater than 10° occurred for 17 children who had manipulation under anaesthesia. Four had a visible cosmetic deformity at discharge and nine had restriction of movement, with four requiring physiotherapy. One child developed over- granulation at the pin site and one wire became buried, resulting in a difficult
Background. Pseudotumours have been associated with metal-on-metal (MoM) hip replacements. We define it as a solid mass which may have cystic components that is neither neoplastic nor infectious in aetiology. The cause of a pseudotumour is not fully understood but could be due to excessive wear, metal hypersensitivity or due to an as-yet unknown cause. Aim. We present the
Background. Skeletal stem cells (SSCs) have been used for the treatment of osteonecrosis of the femoral head to prevent subsequent collapse. In isolation SSCs do not provide structural support but an innovative case series in Southampton, UK, has used SSCs in combination with impaction bone grafting (IBG) to improve both the biological and mechanical environment and to regenerate new bone at the necrotic site. Aims. Analysis of retrieved tissue-engineered bone as part of ongoing follow-up of this translational case series. Methods. With Proof-of-Concept established in vitro and in vivo, the use of a living bone composite of SSCs and allograft has been translated to four patients (five hips) for treatment of osteonecrosis of their femoral heads. Parallel in vitro culture of the implanted cell-graft construct was performed. Patient follow-up was by serial clinical and radiological examination. In one patient collapse occurred in both hips due to more advanced disease than was originally appreciated. This necessitated bilateral hip arthroplasty, but allowed
Background. An increased incidence of periprosthetic osteolysis, resulting in loss of biologic fixation, has been recently reported in contemporary THAs with low-carbide metal-on-metal compared to metal-on-polyethylene couple bearings. A hypersensitivity reaction due to Co and Cr debris is reported as a potential cause for failure of THAs with high-carbide bearings, but there are no evidence-based data for this reaction in low-carbide metal-on-metal bearings. Questions/purposes. We investigated whether there were differences in immunologic hypersensitivity reactions in
Complex fractures of the femur and tibia with associated severe soft tissue injury are often devastating for the individual. The aim of this study was to describe the two-year patient-reported outcomes of patients in a civilian population who sustained a complex fracture of the femur or tibia with a Mangled Extremity Severity Score (MESS) of ≥ 7, whereby the score ranges from 2 (lowest severity) to 11 (highest severity). Patients aged ≥ 16 years with a fractured femur or tibia and a MESS of ≥ 7 were extracted from the Victorian Orthopaedic Trauma Outcomes Registry (January 2007 to December 2018). Cases were grouped into surgical amputation or limb salvage. Descriptive analysis were used to examine return to work rates, three-level EuroQol five-dimension questionnaire (EQ-5D-3L), and Glasgow Outcome Scale-Extended (GOS-E) outcomes at 12 and 24 months post-injury.Aims
Methods
INTRODUCTION. Autologous bone grafts are considered gold standard in the repair of bone defects. However they are limited in supply and are associated with donor site morbidity. This has led to the development of synthetic bone graft substitute (BGS) materials, many of which have been reported as being osteoinductive. The structure of the BGS is important and bone formation has been observed in scaffolds with a macroporous morphology. Smaller pores termed ‘strut porosity’ may also be important for osteoinduction. The aim of this study was to compare the osteoinductive ability of one silicate-substituted calcium phosphate (SiCaP) with differing strut porosities in an ectopic ovine model. Our hypothesis was that SiCaP with greater strut porosity would be more osteoinductive. METHODS. The osteoinduction of SiCaP BGS with two different strut porosities (AF and AF++) was investigated. The materials had an identical chemical composition and morphological structure but differing strut porosity (AF=22.5%, AF++=47%). Implants were inserted into the paraspinal muscles in skeletally mature sheep. Procedures were carried out in compliance with UK Home Office regulations. There were 12 implants in each group. Implants remained in vivo for 8 and 12 weeks and on
INTRODUCTION. Analysis of retrieved ceramic components have shown areas of localized ‘stripe wear’, which have been attributed to joint laxity and/or impingement resulting in subluxation of the head, causing wear on the edge of the cup. Studies have been conducted into the effects of mild subluxation, however few in vitro tests have looked at severe subluxation. The aim of this study was to develop a more clinically relevant subluxation protocol. MATERIALS & METHODS. Seven (Subluxation n=4; standard test n=3) of 36mm Biolox Forte (R3, Smith & Nephew) ceramic devices were tested for 0.5m cycles (mc). Two of the subluxed joints were further tested to 1 Mc. The devices were subjected to subluxation under standard testing conditions. The flex/ext was 30° and 15° respectively, with internal/external rotation of ±10°. The force was Paul type stance phase loading with a maximum load of 3 kN, and a standard ISO swing phase load of 0.3 kN at 1 Hz. The test was conducted on a ProSim hip joint wear simulator (SimSol, UK). The simulator is equipped with a novel mechanism to achieve translation of the head, to achieve subluxation. During the ISO swing phase load of 0.3kN, a controlled lateral force required for the translation of the head is applied by a cam mechanism, head retraction then occurs during heel strike. The lubricant used was new born calf serum diluted with de-ionised water to achieve average protein concentration of 20 g/l, with 0.2 wt % concentration NaN3, and changed every 250k cycles. Measurements have been taken at 0.5 & 1 mc stages. RESULTS. Linear wear measurements conducted on the subluxed joints resulted in stripe wear similar to that reported in vivo. Average length, width and depth dimensions were 25.34±1.96 mm, 8±1.60 mm and 16.95±3.87 μm (± 95% CL) respectively. Linear wear at 0.5 Mc for standard joints, were undistinguishable from the original profile. Gravimetrically, weight loss was undetectable for joints tested under standard conditions. The volume loss of the joints under subluxation was 1.9± 0.7 mm3 at 0.5 mc. Two joints tested to 1mc generated an average volume loss of 3.1±2.3 mm3. The stripe wear length, width and depth at 1 Mc were 25.30±3.33mm, 8±3.92mm and 35±17.07 μm respectiveley. DISCUSSION. The current study presents test results of a hip joint simulator with a novel subluxation mechanism to simulate severe and clinically relevant hip joint. Past techniques have had to reduce the swing phase load to achieve stripe wear patches of varying size and depth. The subluxed joints produced significantly higher volumetric wear than the standard joints. Dimensional measurements in terms of length, width and depth of wear patches of subluxed joints generated similar results to that which have been observed following