Objectives. Favourable results for collarless
We retrospectively reviewed 175 patients (191 hips) who had undergone primary cemented total hip replacement between November 1992 and November 1995 using a collarless polished double-tapered femoral component after a minimum of ten years (mean 11.08; 10 to 12.8). All stems were implanted using contemporary cementing techniques with a distal cement restrictor, pressurised lavage, retrograde cementing with a gun and proximal pressurisation. Clinical outcome was assessed using the Harris Hip score. Radiological analysis was performed on calibrated plain radiographs taken in two planes. Complete radiological data on 110 patients (120 hips) and clinical follow-up on all the surviving 111 patients (122 hips) was available. The fate of all the hips was known. At final follow-up, the mean Harris Hip score was 86 (47 to 100), and 87 of 116 patients (75%) had good or excellent scores. Survival with revision of the stem for aseptic loosening as the endpoint was 100%; and survival with revision of the stem for any reason was 95.9% (95% confidence interval 87.8 to 96.8) at ten years. All the stems subsided vertically at the stem-cement interface in a predictable pattern, at an overall mean rate of 0.18 mm per year (0.02 to 2.16), but with a mean rate of 0.80 mm (0.02 to 2.5) during the first year. The mean total subsidence was 1.95 mm (0.21 to 24). Only three stems loosened at the cement-bone interface. There was excellent preservation of proximal femoral bone stock. There was a high incidence of Brooker III and IV heterotopic ossification affecting 25 patients (22%). The collarless
The Unified Classification System (UCS), or Vancouver system, is a validated and widely used classification system to guide the management of periprosthetic femoral fractures. It suggests that well-fixed stems (type B1) can be treated with fixation but that loose stems (types B2 and B3) should be revised. Determining whether a stem is loose can be difficult and some authors have questioned how to apply this classification system to
We prospectively followed 191 consecutive collarless
polished tapered (CPT) femoral stems, implanted in 175 patients
who had a mean age at operation of 64.5 years (21 to 85). At a mean
follow-up of 15.9 years (14 to 17.5), 86 patients (95 hips) were
still alive. The fate of all original stems is known. The 16-year
survivorship with re-operation for any reason was 80.7% (95% confidence
interval 72 to 89.4). There was no loss to follow-up, with clinical
data available on all 95 hips and radiological assessment performed
on 90 hips (95%). At latest follow-up, the mean Harris hip score
was 78 (28 to 100) and the mean Oxford hip score was 36 (15 to 48).
Stems subsided within the cement mantle, with a mean subsidence
of 2.1 mm (0.4 to 19.2). Among the original cohort, only one stem
(0.5%) has been revised due to aseptic loosening. In total seven
stems were revised for any cause, of which four revisions were required
for infection following revision of the acetabular component. A
total of 21 patients (11%) required some sort of revision procedure;
all except three of these resulted from failure of the acetabular
component. Cemented acetabular components had a significantly lower
revision burden (three hips, 2.7%) than Harris Galante uncemented
components (17 hips, 21.8%) (p <
0.001). The CPT stem continues to provide excellent radiological and
clinical outcomes at 15 years following implantation. Its results
are consistent with other
Aims. The aims of this study were to evaluate the incidence of reoperation (all cause and specifically for periprosthetic femoral fracture (PFF)) and mortality, and associated risk factors, following a hemiarthroplasty incorporating a cemented collarless
Femoral impaction bone allografting has been developed as a means of restoring bone stock in revision total hip replacement. We report the results of 75 consecutive patients (75 hips) with a mean age of 68 years (35 to 87) who underwent impaction grafting using the Exeter collarless,
The aim of this study was to radiologically evaluate the quality of cement mantle and alignment achieved with a polished tapered cemented femoral stem inserted through the anterior approach and compared with the posterior approach. A comparative retrospective study of 115 consecutive hybrid total hip arthroplasties or cemented hemiarthroplasties in 110 patients, performed through anterior (n = 58) or posterior approach (n = 57) using a collarless polished taper-slip femoral stem, was conducted. Cement mantle quality and thickness were assessed in both planes. Radiological outcomes were compared between groups.Aims
Methods
Cemented femoral stems with force closed fixation
designs have shown good clinical results despite high early subsidence.
A new triple-tapered stem in this category (C-stem AMT) was introduced
in 2005. This study compares this new stem with an established stem
of similar design (Exeter) in terms of migration (as measured using radiostereometric
analysis), peri-prosthetic bone remodelling (measured using dual
energy x-ray densitometry, DXA), Oxford Hip Score, and plain radiographs. A total of 70 patients (70 hips) with a mean age of 66 years
(53 to 78) were followed for two years. Owing to missing data of
miscellaneous reasons, the final analysis represents data from 51
(RSA) and 65 (DXA) patients. Both stems showed a typical pattern
of migration: Subsidence and retroversion that primarily occurred
during the first three months. C-stem AMT subsided less during the
first three months (p = 0.01), before stabilising at a subsidence rate
similar to the Exeter stem from years one to two. The rate of migration
into retroversion was slightly higher for C-stem AMT during the
second year (p = 0.03). Whilst there were slight differences in
movement patterns between the stems, the C-stem AMT exhibits good
early clinical outcomes and displays a pattern of migration and
bone remodelling that predicts good clinical performance. Cite this article:
The advantages of high viscosity Simplex AF cement (PMMA) compared to low viscosity Simplex P are the low porosity, the high fatigue strength, the lower polymerization time, and the lower maximum polymerization temperature. A prospective, randomized, double-blind clinical study was conducted to assess the in vivo effects of viscosity of bone cement on the micromotion of a
Studies have reported stem subsidence without loosening in cemented
Aims. The aim of this large registry-based study was to compare mid-term survival rates of cemented femoral stems of different designs used in hemiarthroplasty for a fracture of the femoral neck. Patients and Methods. From the Norwegian Hip Fracture Register (NHFR), 20 532 primary cemented bipolar hemiarthroplasties, which were undertaken in patients aged > 70 years with a femoral neck fracture between 2005 and 2016, were included. Polished tapered stems (n = 12 065) (Exeter and CPT), straight stems (n = 5545) (Charnley, Charnley Modular, and Spectron EF), and anatomical stems (n = 2922) (Lubinus SP2) were included. The survival of the implant with any reoperation as the endpoint was calculated using the Kaplan–Meier method and hazard ratios (HRs), and the different indications for reoperation were calculated using Cox regression analysis. Results. The one-year survival was 96.0% (95% confidence interval (CI) 95.6 to 96.4) for the Exeter stem, 97.0% (95% CI 96.4 to 97.6) for the Lubinus SP2 stem, 97.6% (95% CI 97.0 to 98.2) for the Charnley stem, 98.1% (95% CI 97.3 to 98.9) for the Spectron EF stem, and 96.4% (95% CI 95.6 to 97.2) for the Charnley Modular stem, respectively. The hazard ratio for reoperation after one year was lower for Lubinus SP2 (HR 0.77, 95% CI 0.60 to 0.97), Charnley (HR 0.64, 95% CI 0.48 to 0.86), and Spectron EF stems (HR 0.44, 95% CI 0.29 to 0.67) compared with the Exeter stem. Reoperation for periprosthetic fracture occurred almost exclusively after the use of
Aims. Periprosthetic hip fractures (PPFs) after total hip arthroplasty are difficult to treat. Therefore, it is important to identify modifiable risk factors such as stem selection to reduce the occurrence of PPFs. This study aimed to clarify differences in fracture torque, surface strain, and fracture type analysis between three different types of cemented stems. Methods. We conducted biomechanical testing of bone analogues using six cemented stems of three different types: collarless
Aims. Debate continues about whether it is better to use a cemented or uncemented hemiarthroplasty to treat a displaced intracapsular fracture of the hip. The aim of this study was to attempt to resolve this issue for contemporary prostheses. Methods. A total of 400 patients with a displaced intracapsular fracture of the hip were randomized to receive either a cemented
Dual mobility (DM) is most often used by surgeons to reduce instability in high risk patients. NJR data on DM has not demonstrated a reduction in all cause revision and has reported an increase in revision for peri-prosthetic fracture (PPF). The aim of our study was:. Report outcome of DM used in high-risk patients including non-revision re-operations (dislocation & PPF). Comparison with conventional bearing THA (cTHA) with local, national and NJR benchmarking data. Retrospective cohort assessment of falls risk for patients receiving DM. Prospective F/U of a DM implant since 2016 and enrolled into Beyond Compliance (BC). Primary outcome measure all-cause revision with secondary outcome including any re-operation and Oxford Hip Score (OHS). All patients were risk stratified and considered high risk for instability. Complications were identified via hospital records, clinical coding linkage, NJR and BC. Benchmarking data for comparison was obtained from same data sources we also considered all B type PPF that occurred with cemented
Aim: Small amounts of subsidence may be beneficial in stabilising a stem &
appears to be a featue of
We report the outcome at a minimum of 10 years follow-up for 80
Introduction. The success of total hip replacement in the young has consistently been worse both radiologically and clinically when compared to the standard hip replacement population. Methods. We describe the clinical and radiological outcome of 58 consecutive
Cement-in-cement revision of the femoral component represents a widely practiced technique for a variety of indications. In this study we compare the clinical and radiological outcomes of two
The long term survival of well-cemented,
The success of total hip replacement in the young has consistently been worse both radiologically and clinically when compared to the standard hip replacement population. Methods: We describe the clinical and radiological outcome of 58 consecutive