Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_6 | Pages 12 - 12
2 May 2024
Selim A Al-Hadithy N Diab N Ahmed A Kader KA Hegazy M Abdelazeem H Barakat A
Full Access

Lag screw cut-out is a serious complication of dynamic hip screw fixation in trochanteric hip fractures. Lag screw position is recognised as a crucial factor influencing the occurrence of lag screw cut-out. We propose a modification of the Tip Apex Distance (TAD) and hypothesize that it could enhance the reliability of predicting lag screw cut-out in these injuries. A retrospective study of hip fracture cases was conducted from January 2018 to July 2022. A total of 109 patients were eligible for the final analysis. The modified TAD was measured in millimetres, based on the sum of the traditional TAD in the lateral view and the net value of two distances in the anteroposterior (AP) view. The first distance is from the lag screw tip to the opposite point on the femoral head along the lag screw axis, while the second distance is from that point to the femoral head apex. The first distance is a positive value, whereas the second distance is positive if the lag screw is superior and negative if it is inferior. Receiver operating characteristic (ROC) curve analysis was used to assess the reliability of various parameters for evaluating the lag screw position within the femoral head. Factors such as reduction quality, fracture pattern according to the AO/OTA classification, TAD, Calcar-Referenced TAD, Axis Blade Angle, Parker’s ratio in the AP view, Cleveland Zone 1, and modified TAD were statistically associated with lag screw cut-out. Among the tested parameters, the novel parameter exhibited 90.1% sensitivity and 90.9% specificity for predicting lag screw cut-out at a cut-off value of 25 mm, with a p-value < 0.001. The modified TAD demonstrated the highest reliability in predicting lag screw cut-out. A value of 25 mm may potentially reduce the risk of lag screw cut-out in trochanteric hip fractures


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_13 | Pages 52 - 52
1 Oct 2018
Parry J Langford J Koval K Haidukewych G
Full Access

Introduction. The vast majority of intertrochanteric fractures treated with cephalomedullary nails (CMN) will heal. Occasionally even though bony union occurs excessive lag screw sliding can cause persistent pain and soft tissue irritation and return to surgery for hardware removal. The purpose of this study was to evaluate if fracture stability, lag screw tip-apex distance (TAD), and quality of reduction have any impact excessive lag screw sliding and potential cutout. Methods. As part of our level one trauma center's institutional hip fracture registry, a retrospective analysis identified 199 intertrochanteric fractures fixed with CMN between 2009 and 2015 with follow up to union or a minimum of three months. The mean follow-up was 22 months (3 to 94 months). Mean patient age was 75 years (50 to 97 years) and 72% were women. Postoperative radiographs were used to measure the TAD, quality of reduction, neck-shaft angle (NSA), and lateral lag screw prominence. Follow-up radiographs were reviewed to assess fracture union, translation, and progression of lateral lag screw prominence. Complications and reoperations were recorded. Results. The average lag screw sliding was 5±5 mm. Excessive lag screw sliding (defined as > 10 mm; one standard deviation above the mean) was present in 12% of patients. Lag screw sliding was more common in unstable fracture patterns (21% vs. 5%, p<0.01) and patients with calcar fracture gapping > 4 mm (26% vs. 4%, p<0.01). Lag screw sliding was not associated with age (p=0.9), sex (p=0.4), TAD (p=0.3), implant (p=0.8), distal interlocking screws (p=0.3), or NSA (p=0.2). There were seven (3%) patients with prominent lag screws that required removal. These patients experience more lag screw sliding than those that did not require removal (9 mm vs. 5 mm, p<0.01). The average TAD was 17±5 mm. 15 (7%) of patients had TAD of 25 mm or more. There were 2 cutouts (1%). The average TAD was larger in the cutout group (26 vs. 17 mm, p<0.01). Conclusion. In this series, the incidence of cutout was low and associated with a larger tip-apex distance. Excessive lag screw sliding was associated with unstable fracture patterns, calcar fracture gapping, and more reoperations for symptomatic hardware. Careful attention to calcar fracture reduction may minimize excessive lag screw sliding


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_6 | Pages 55 - 55
2 May 2024
McCann C Ablett A Feng T Macaskill V Oliver W Keating J
Full Access

Subtrochanteric femoral fractures are a subset of hip fractures generally treated with cephalomedullary nail fixation\[1\]. Single lag screw devices are most commonly-used, but integrated dual screw constructs have become increasingly popular\[2,3\]. The aim of this study was to compare outcomes of fixation of subtrochanteric femoral fractures using a single lag screw (Gamma3 nail, GN) with a dual screw device (InterTAN nail, IN). The primary outcome was mechanical failure, defined as lag screw cut-out, back-out, nail breakage or peri-implant fracture. Consecutive adult patients (18yrs) with subtrochanteric femoral fracture treated in a single centre were retrospectively identified using electronic records. Patients that underwent surgical fixation using either a long GN (2010–2017) or IN (2017–2022) were included. Medical records and radiographs were reviewed to identify complications of fixation. Cox regression analysis was used to determine the risk of mechanical failure and secondary outcomes by implant design. Multivariable regression models were used to identify predictors of mechanical failure. The study included 622 patients, 354 in the GN group (median age 82yrs, 72% female) and 268 in the IN group (median age 82yrs, 69% female). The risk of any mechanical failure was increased two-fold in the GN group (HR 2.44 \[95%CI 1.13 to 5.26\]; _p=0.024_). Mechanical failure comprising screw cut-out (_p=0.032_), back-out (_p=0.032_) and nail breakage (_p=0.26_) was only observed in the GN group. Technical predictors of failure included varus >5° for cut-out (OR 19.98 \[2.06 to 193.88\]; _p=0.01_), TAD;25mm for back-out (8.96 \[1.36 to 58.86\]; p=0.022) and shortening 1cm for peri-implant fracture (7.81 \[2.92 to 20.91\]; _p=<0.001_). Our results demonstrate that an intercalated screw construct is associated with a lower risk of mechanical failure compared with the a single lag screw device. Intercalated screw designs may reduce the risk of mechanical complications for patients with subtrochanteric femoral fractures


Bone & Joint Open
Vol. 4, Issue 5 | Pages 329 - 337
8 May 2023
Khan AQ Chowdhry M Sherwani MKA McPherson EJ

Aims. Total hip arthroplasty (THA) is considered the preferred treatment for displaced proximal femoral neck fractures. However, in many countries this option is economically unviable. To improve outcomes in financially disadvantaged populations, we studied the technique of concomitant valgus hip osteotomy and operative fixation (VOOF). This prospective serial study compares two treatment groups: VOOF versus operative fixation alone with cannulated compression screws (CCSs). Methods. In the first series, 98 hip fixation procedures were performed using CCS. After fluoroscopic reduction of the fracture, three CCSs were placed. In the second series, 105 VOOF procedures were performed using a closing wedge intertrochanteric osteotomy with a compression lag screw and lateral femoral plate. The alignment goal was to create a modified Pauwel’s fracture angle of 30°. After fluoroscopic reduction of fracture, lag screw was placed to achieve the calculated correction angle, followed by inter-trochanteric osteotomy and placement of barrel plate. Patients were followed for a minimum of two years. Results. Mean follow-up was 4.6 years (4.1 to 5.0) in the CCS group and 5.5 years (5.25 to 5.75) in the VOOF group. The mean Harris Hip Score at two-year follow-up was 83.85 in the CCS group versus 88.00 in the VOOF group (p < 0.001). At the latest follow-up, all-cause failure rate was 29.1% in the CCS group and 11.7% in the VOOF group (p = 0.003). The total cost of the VOOF technique was 7.2% of a THA, and total cost of the CCS technique was 6.3% of a THA. Conclusion. The VOOF technique decreased all-cause failure rate compared to CCS. The total cost of VOOF was 13.5% greater than CCS, but 92.8% less than a THA. Increased cost of VOOF was considered acceptable to all patients in this series. VOOF technique provides a reasonable alternative to THA in patients who cannot afford a THA procedure. Cite this article: Bone Jt Open 2023;4(5):329–337


Bone & Joint Open
Vol. 1, Issue 9 | Pages 594 - 604
24 Sep 2020
James HK Pattison GTR Griffin J Fisher JD Griffin DR

Aims. To develop a core outcome set of measurements from postoperative radiographs that can be used to assess technical skill in performing dynamic hip screw (DHS) and hemiarthroplasty, and to validate these against Van der Vleuten’s criteria for effective assessment. Methods. A Delphi exercise was undertaken at a regional major trauma centre to identify candidate measurement items. The feasibility of taking these measurements was tested by two of the authors (HKJ, GTRP). Validity and reliability were examined using the radiographs of operations performed by orthopaedic resident participants (n = 28) of a multicentre randomized controlled educational trial (ISRCTN20431944). Trainees were divided into novice and intermediate groups, defined as having performed < ten or ≥ ten cases each for DHS and hemiarthroplasty at baseline. The procedure-based assessment (PBA) global rating score was assumed as the gold standard assessment for the purposes of concurrent validity. Intra- and inter-rater reliability testing were performed on a random subset of 25 cases. Results. In total, 327 DHS and 248 hemiarthroplasty procedures were performed by 28 postgraduate year (PGY) 3 to 5 orthopaedic trainees during the 2014 to 2015 surgical training year at nine NHS hospitals in the West Midlands, UK. Overall, 109 PBAs were completed for DHS and 80 for hemiarthroplasty. Expert consensus identified four ‘final product analysis’ (FPA) radiological parameters of technical success for DHS: tip-apex distance (TAD); lag screw position in the femoral head; flushness of the plate against the lateral femoral cortex; and eight-cortex hold of the plate screws. Three parameters were identified for hemiarthroplasty: leg length discrepancy; femoral stem alignment; and femoral offset. Face validity, content validity, and feasibility were excellent. For all measurements, performance was better in the intermediate compared with the novice group, and this was statistically significant for TAD (p < 0.001) and femoral stem alignment (p = 0.023). Concurrent validity was poor when measured against global PBA score. This may be explained by the fact that they are measuring difference facets of competence. Intra-and inter-rater reliability were excellent for TAD, moderate for lag screw position (DHS), and moderate for leg length discrepancy (hemiarthroplasty). Use of a large multicentre dataset suggests good generalizability of the results to other settings. Assessment using FPA was time- and cost-effective compared with PBA. Conclusion. Final product analysis using post-implantation radiographs to measure technical skill in hip fracture surgery is feasible, valid, reliable, and cost-effective. It can complement traditional workplace-based assessment for measuring performance in the real-world operating room . It may have particular utility in competency-based training frameworks and for assessing skill transfer from the simulated to live operating theatre. Cite this article: Bone Joint Open 2020;1-9:594–604


The Bone & Joint Journal
Vol. 96-B, Issue 8 | Pages 1029 - 1034
1 Aug 2014
Kashigar A Vincent A Gunton MJ Backstein D Safir O Kuzyk PRT

The purpose of this study was to identify factors that predict implant cut-out after cephalomedullary nailing of intertrochanteric and subtrochanteric hip fractures, and to test the significance of calcar referenced tip-apex distance (CalTAD) as a predictor for cut-out. We retrospectively reviewed 170 consecutive fractures that had undergone cephalomedullary nailing. Of these, 77 met the inclusion criteria of a non-pathological fracture with a minimum of 80 days radiological follow-up (mean 408 days; 81 days to 4.9 years). The overall cut-out rate was 13% (10/77). The significant parameters in the univariate analysis were tip-apex distance (TAD) (p <  0.001), CalTAD (p = 0.001), cervical angle difference (p = 0.004), and lag screw placement in the anteroposterior (AP) view (Parker’s ratio index) (p = 0.003). Non-significant parameters were age (p = 0.325), gender (p = 1.000), fracture side (p = 0.507), fracture type (AO classification) (p = 0.381), Singh Osteoporosis Index (p = 0.575), lag screw placement in the lateral view (p = 0.123), and reduction quality (modified Baumgaertner’s method) (p = 0.575). In the multivariate analysis, CalTAD was the only significant measurement (p = 0.001). CalTAD had almost perfect inter-observer reliability (interclass correlation coefficient (ICC) 0.901). Our data provide the first reported clinical evidence that CalTAD is a predictor of cut-out. The finding of CalTAD as the only significant parameter in the multivariate analysis, along with the univariate significance of Parker’s ratio index in the AP view, suggest that inferior placement of the lag screw is preferable to reduce the rate of cut-out. Cite this article: Bone Joint J 2014; 96-B:1029–34


Bone & Joint Research
Vol. 6, Issue 4 | Pages 204 - 207
1 Apr 2017
Fernandez MA Aquilina A Achten J Parsons N Costa ML Griffin XL

Objectives. The Sliding Hip Screw (SHS) is commonly used to treat trochanteric hip fractures. Fixation failure is a devastating complication requiring complex revision surgery. One mode of fixation failure is lag screw cut-out which is greatest in unstable fracture patterns and when the tip-apex distance of the lag screw is > 25 mm. The X-Bolt Dynamic Hip Plating System (X-Bolt Orthopaedics, Dublin, Ireland) is a new device which aims to reduce this risk of cut-out. However, some surgeons have reported difficulty minimising the tip-apex distance with subsequent concerns that this may lead to an increased risk of cut-out. Patients and Methods. We measured the tip-apex distance from the intra-operative radiographs of 93 unstable trochanteric hip fractures enrolled in a randomised controlled trial (Warwick Hip Trauma Evaluation, WHiTE One trial). Participants were treated with either the sliding hip screw or the X-Bolt dynamic hip plating system. We also recorded the incidence of cut-out in both groups, at a median follow-up time of 17 months. Results. There was a significantly increased tip-apex distance with the use of the X-Bolt (mean difference 3.7mm (95% confidence interval 1.58 to 5.73); SHS mean 17.1 mm, X-Bolt mean 20.8; p = 0.001. However, this was not associated with an increased incidence of cut-out at a median follow-up time of 17 months, with three cut-outs (6%) in the SHS group and 0 (0%) in the X-Bolt group. Conclusion. The X-Bolt is a safe implant with no increased risk for cut-out. Concerns about minimising the tip-apex distance may be justified but do not appear to be clinically important. Cite this article: M. A. Fernandez, A. Aquilina, J. Achten, N. Parsons, M. L. Costa, X. L. Griffin. The tip-apex distance in the X-Bolt dynamic plating system. Bone Joint Res 2017;6:–207. DOI: 10.1302/2046-3758.64.BJR-2015-0016.R2


The Bone & Joint Journal
Vol. 96-B, Issue 9 | Pages 1198 - 1201
1 Sep 2014
Magu NK Singla R Rohilla R Gogna P Mukhopadhyay R Singh A

We report the outcome of 39 patients who underwent a modified Pauwels’ intertrochanteric osteotomy for nonunion of a femoral neck fracture following failed osteosynthesis. There were 31 men and eight women with a mean age of 47.2 years (34 to 59). By Pauwels’ classification, there were 11 Type II fractures and 28 Type III fractures. The mean follow-up was 7.9 years (2 to 19). In the 11 patients whose initial treatment had been osteotomy, union was achieved in nine (81.8%). In 28 patients whose initial treatment had been with a lag screw or a dynamic hip screw, union was achieved in 27 (96.4%). Limb lengths were equalised in 14 of 16 patients (87.5%) with pre-operative shortening. The mean neck-shaft angle improved significantly from 100.5° (80° to 120°) to 131.6° (120° to 155°) (p = 0.004). The mean modified Harris hip score was 85.6 points (70 to 97) and the mean modified Merle d’Aubigné score was 14.3 (11 to 18). Good to excellent functional outcomes were achieved in 32 patients (88.8%). A modified Pauwels’ intertrochanteric osteotomy is a reliable method of treating ununited fractures of the femoral neck following failed osteosynthesis: coxa vara and shortening can also simultaneously be addressed. Cite this article: Bone Joint J 2014;96-B:1198–1201


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_11 | Pages 43 - 43
1 Jun 2016
Mehta N Reddy G Goldsmith T Ramakrishnan M
Full Access

Background. Sub-trochanteric fractures are challenging to treat due to various anatomical and biomechanical factors. High tensile forces contribute to the challenge of fracture reduction. Intramedullary nailing has become the treatment of choice. If anatomical reduction is not achieved, any mal-alignment will predispose to implant failure. Open reduction with cerclage wires can add to construct stability and improve the quality of reduction. There is no consensus or classification to guide surgeons on when to perform open reduction, which is often performed intra-operatively when closed reduction fails often with no planning. This can lead to intraoperative delays as theatre staff would not have prepared the correct equipment necessary for open reduction. Objectives. The purpose of this study was to assess outcomes of closed and open reduction of traumatic sub-trochanteric fractures treated with intramedullary nailing and to propose a new classification system to dictate management. Methods. After a review of current classification systems, a 3-tier classification was proposed (Type 1, 2 and 3). Type 1 indicated a transverse fracture, Type 2 was a spiral fracture with an intact posterior and medial wall and a Type 3 fracture were fractures with no posterior and/or medial walls. Over a two-year period (2013–2015), patients with sub trochanteric fractures were classified into Type 1, 2 or 3 injuries based on radiographic appearances by two senior clinicians. Patients with Type 3 injuries were divided into two groups based on whether they were treated with open or closed reduction. A clinical and radiographic review was performed. The primary outcome measure was the incidence of implant failure, whereas secondary outcome measures were related to fracture reduction. Statistical analysis was performed using GraphPad Prism Version 6 (GraphPad Software Inc. California, USA). Fisher's exact test was used for independent categorical data and Mann–U Whitney for continuous nonparametric data. Statistical significance was set at p<0.05. Results. 75 patients had intramedullary nailing for subtrochanteric fractures over the study period with a mean age of 82.6 years. There were 48 patients who had a Type 3 fracture pattern with a deficient medial and/or posterior wall. Reduction was achieved open with cerclage wires in 42% of patients (n=20 and closed in 58% (n=28). Overall there were a total of 18 (37.5%) major complications. In patients treated with closed reduction, 9 patients suffered mechanical complications (6 distal locking screw failures, 3 lag screw cut outs). There was a significantly increased risk of implant failure in patients treated with closed reduction compared to open reduction (p=0.006). No cases with cerclage wire had implant failure. Open reduction with cerclage wires improved the quality of reduction (p=0.0001) compared with closed reduction. There was no significant increase in operating time in patients treated with cerclage wires (p=0.4334). Conclusions. Open reduction with cerclage wires should be considered in patients with Type 3 sub-trochanteric fractures as it has shown to significantly reduce the risk of implant failure and improve the quality of reduction with no significant increase in operating time


Bone & Joint Open
Vol. 5, Issue 2 | Pages 123 - 131
12 Feb 2024
Chen B Duckworth AD Farrow L Xu YJ Clement ND

Aims

This study aimed to determine whether lateral femoral wall thickness (LWT) < 20.5 mm was associated with increased revision risk of intertrochanteric fracture (ITF) of the hip following sliding hip screw (SHS) fixation when the medial calcar was intact. Additionally, the study assessed the association between LWT and patient mortality.

Methods

This retrospective study included ITF patients aged 50 years and over treated with SHS fixation between 2019 and 2021 at a major trauma centre. Demographic information, fracture type, delirium status, American Society of Anesthesiologists grade, and length of stay were collected. LWT and tip apex distance were measured. Revision surgery and mortality were recorded at a mean follow-up of 19.5 months (1.6 to 48). Cox regression was performed to evaluate independent risk factors associated with revision surgery and mortality.


Aims

Surgical treatment of hip fracture is challenging; the bone is porotic and fixation failure can be catastrophic. Novel implants are available which may yield superior clinical outcomes. This study compared the clinical effectiveness of the novel X-Bolt Hip System (XHS) with the sliding hip screw (SHS) for the treatment of fragility hip fractures.

Methods

We conducted a multicentre, superiority, randomized controlled trial. Patients aged 60 years and older with a trochanteric hip fracture were recruited in ten acute UK NHS hospitals. Participants were randomly allocated to fixation of their fracture with XHS or SHS. A total of 1,128 participants were randomized with 564 participants allocated to each group. Participants and outcome assessors were blind to treatment allocation. The primary outcome was the EuroQol five-dimension five-level health status (EQ-5D-5L) utility at four months. The minimum clinically important difference in utility was pre-specified at 0.075. Secondary outcomes were EQ-5D-5L utility at 12 months, mortality, residential status, mobility, revision surgery, and radiological measures.


Bone & Joint Research
Vol. 8, Issue 7 | Pages 313 - 322
1 Jul 2019
Law GW Wong YR Yew AK Choh ACT Koh JSB Howe TS

Objectives

The paradoxical migration of the femoral neck element (FNE) superomedially against gravity, with respect to the intramedullary component of the cephalomedullary device, is a poorly understood phenomenon increasingly seen in the management of pertrochanteric hip fractures with the intramedullary nail. The aim of this study was to investigate the role of bidirectional loading on the medial migration phenomenon, based on unique wear patterns seen on scanning electron microscopy of retrieved implants suggestive of FNE toggling.

Methods

A total of 18 synthetic femurs (Sawbones, Vashon Island, Washington) with comminuted pertrochanteric fractures were divided into three groups (n = 6 per group). Fracture fixation was performed using the Proximal Femoral Nail Antirotation (PFNA) implant (Synthes, Oberdorf, Switzerland; n = 6). Group 1 was subjected to unidirectional compression loading (600 N), with an elastomer (70A durometer) replacing loose fracture fragments to simulate surrounding soft-tissue tensioning. Group 2 was subjected to bidirectional loading (600 N compression loading, 120 N tensile loading), also with the elastomer replacing loose fracture fragments. Group 3 was subjected to bidirectional loading (600 N compression loading, 120 N tensile loading) without the elastomer. All constructs were tested at 2 Hz for 5000 cycles or until cut-out occurred. The medial migration distance (MMD) was recorded at the end of the testing cycles.


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 122 - 128
1 Jul 2021
Tibbo ME Limberg AK Gausden EB Huang P Perry KI Yuan BJ Berry DJ Abdel MP

Aims

The prevalence of ipsilateral total hip arthroplasty (THA) and total knee arthroplasty (TKA) is rising in concert with life expectancy, putting more patients at risk for interprosthetic femur fractures (IPFFs). Our study aimed to assess treatment methodologies, implant survivorship, and IPFF clinical outcomes.

Methods

A total of 76 patients treated for an IPFF from February 1985 to April 2018 were reviewed. Prior to fracture, at the hip/knee sites respectively, 46 femora had primary/primary, 21 had revision/primary, three had primary/revision, and six had revision/revision components. Mean age and BMI were 74 years (33 to 99) and 30 kg/m2 (21 to 46), respectively. Mean follow-up after fracture treatment was seven years (2 to 24).


The Bone & Joint Journal
Vol. 100-B, Issue 4 | Pages 443 - 449
1 Apr 2018
Kalsbeek JH van Walsum ADP Vroemen JPAM Janzing HMJ Winkelhorst JT Bertelink BP Roerdink WH

Aims

The objective of this study was to investigate bone healing after internal fixation of displaced femoral neck fractures (FNFs) with the Dynamic Locking Blade Plate (DLBP) in a young patient population treated by various orthopaedic (trauma) surgeons.

Patients and Methods

We present a multicentre prospective case series with a follow-up of one year. All patients aged ≤ 60 years with a displaced FNF treated with the DLBP between 1st August 2010 and December 2014 were included. Patients with pathological fractures, concomitant fractures of the lower limb, symptomatic arthritis, local infection or inflammation, inadequate local tissue coverage, or any mental or neuromuscular disorder were excluded. Primary outcome measure was failure in fracture healing due to nonunion, avascular necrosis, or implant failure requiring revision surgery.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 9 | Pages 1215 - 1221
1 Sep 2010
Sakai T Ohzono K Nishii T Miki H Takao M Sugano N

The long-term results of grafting with hydroxyapatite granules for acetabular deficiency in revision total hip replacement are not well known. We have evaluated the results of revision using a modular cup with hydroxyapatite grafting for Paprosky type 2 and 3 acetabular defects at a minimum of ten years’ follow-up. We retrospectively reviewed 49 acetabular revisions at a mean of 135 months (120 to 178). There was one type 2B, ten 2C, 28 3A and ten 3B hips. With loosening as the endpoint, the survival rate was 74.2% (95% confidence interval 58.3 to 90.1). Radiologically, four of the type 3A hips (14%) and six of the type 3B hips (60%) showed aseptic loosening with collapse of the hydroxyapatite layer, whereas no loosening occurred in type 2 hips. There was consolidation of the hydroxyapatite layer in 33 hips (66%). Loosening was detected in nine of 29 hips (31%) without cement and in one of 20 hips (5%) with cement (p = 0.03, Fisher’s exact probability test). The linear wear and annual wear rate did not correlate with loosening.

These results suggest that the long-term results of hydroxyapatite grafting with cement for type 2 and 3A hips are encouraging.


The Bone & Joint Journal
Vol. 95-B, Issue 11_Supple_A | Pages 7 - 10
1 Nov 2013
Petrie J Sassoon A Haidukewych GJ

Most hip fractures treated with modern internal fixation techniques will heal. However, failures occasionally occur and require revision procedures. Salvage strategies employed during revision are based on whether the fixation failure occurs in the femoral neck, or in the intertrochanteric region. Patient age and remaining bone stock also influence decision making. For fractures in young patients, efforts are generally focused on preserving the native femoral head via osteotomies and repeat internal fixation. For failures in older patients, some kind of hip replacement is usually selected. Disuse osteopenia, deformity, bone loss, and stress-risers from previous internal fixation devices all pose technical challenges to successful reconstruction. Attention to detail is important in order to minimise complications. In the majority of cases, good outcomes have been reported for the various salvage strategies.

Cite this article: Bone Joint J 2013;95-B, Supple A:7–10.


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 9 | Pages 1203 - 1209
1 Sep 2005
Mattsson P Alberts A Dahlberg G Sohlman M Hyldahl HC Larsson S

We undertook a multicentre, prospective study of a series of 112 unstable trochanteric fractures in order to evaluate if internal fixation with a sliding screw device combined with augmentation using a calcium phosphate degradable cement (Norian SRS) could improve the clinical, functional and radiological outcome when compared with fractures treated with a sliding screw device alone. Pain, activities of daily living, health status (SF-36), the strength of the hip abductor muscles and radiological outcome were analysed.

Six weeks after surgery, the patients in the augmented group had significantly lower global and functional pain scores (p < 0.003), less pain after walking 50 feet (p < 0.01), and a better return to the activities of daily living (p < 0.05). At follow-up at six weeks and six months, those in the augmented group showed a significant improvement compared with the control group in the SF-36 score. No other significant differences were found between the groups. We conclude that augmentation with calcium phosphate cement in unstable trochanteric fractures provides a modest reduction in pain and a slight improvement in the quality of life during the course of healing when compared with conventional fixation with a sliding screw device alone.


The Bone & Joint Journal
Vol. 95-B, Issue 11_Supple_A | Pages 11 - 16
1 Nov 2013
Sierra RJ Mabry TM Sems SA Berry DJ

Total hip replacement (THR) after acetabular fracture presents unique challenges to the orthopaedic surgeon. The majority of patients can be treated with a standard THR, resulting in a very reasonable outcome. Technical challenges however include infection, residual pelvic deformity, acetabular bone loss with ununited fractures, osteonecrosis of bone fragments, retained metalwork, heterotopic ossification, dealing with the sciatic nerve, and the difficulties of obtaining long-term acetabular component fixation. Indications for an acute THR include young patients with both femoral head and acetabular involvement with severe comminution that cannot be reconstructed, and the elderly, with severe bony comminution. The outcomes of THR for established post-traumatic arthritis include excellent pain relief and functional improvements. The use of modern implants and alternative bearing surfaces should improve outcomes further.

Cite this article: Bone Joint J 2013;95-B, Supple A:11–16.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 6 | Pages 715 - 723
1 Jun 2008
McMinn DJW Daniel J Ziaee H Pradhan C

The dysplasia cup, which was devised as an adjunct to the Birmingham Hip Resurfacing system, has a hydroxyapatite-coated porous surface and two supplementary neutralisation screws to provide stable primary fixation, permit early weight-bearing, and allow incorporation of morcellised autograft without the need for structural bone grafting.

A total of 110 consecutive dysplasia resurfacing arthroplasties in 103 patients (55 men and 48 women) performed between 1997 and 2000 was reviewed with a minimum follow-up of six years. The mean age at operation was 47.2 years (21 to 62) and 104 hips (94%) were Crowe grade II or III.

During the mean follow-up of 7.8 years (6 to 9.6), three hips (2.7%) were converted to a total hip replacement at a mean of 3.9 years (2 months to 8.1 years), giving a cumulative survival of 95.2% at nine years (95% confidence interval 89 to 100). The revisions were due to a fracture of the femoral neck, a collapse of the femoral head and a deep infection. There was no aseptic loosening or osteolysis of the acetabular component associated with either of the revisions performed for failure of the femoral component. No patient is awaiting a revision.

The median Oxford hip score in 98 patients with surviving hips at the final review was 13 and the 10th and the 90th percentiles were 12 and 23, respectively.


The Journal of Bone & Joint Surgery British Volume
Vol. 94-B, Issue 11_Supple_A | Pages 22 - 25
1 Nov 2012
Haidukewych GJ

Many tumors metastasise to bone, therefore, pathologic fracture and impending pathologic fractures are common reasons for orthopedic consultation. Having effective treatment strategies is important to avoid complications, and relieve pain and preserve function. Thorough pre-operative evaluation is recommended for medical optimization and to ensure that the lesion is in fact a metastasis and not a primary bone malignancy. For impending fractures, various scoring systems have been proposed to determine the risk of fracture, and therefore the need for prophylactic stabilisation. Lower score lesions can often be treated with radiation, while more problematic lesions may require internal fixation. Intramedullary fixation is generally preferred due to favorable biomechanics. Arthroplasty may be required for lesions with massive bony destruction where internal fixation attempts are likely to fail. Radiation may also be useful postoperatively to minimise construct failure due to tumor progression.