Advertisement for orthosearch.org.uk
Results 1 - 20 of 856
Results per page:

Aims. The aim of this study was to evaluate medium-term outcomes and complications of the S-ROM NOILES Rotating Hinge Knee System (DePuy, USA) in revision total knee arthroplasty (rTKA) at a tertiary unit. Methods. A retrospective consecutive study of all patients who underwent a rTKA using this implant from January 2005 to December 2018. Outcome measures included reoperations, revision for any cause, complications, and survivorship. Patients and implant survivorship data were identified through both local hospital electronic databases and linked data from the National Joint Registry/NHS Personal Demographic Service. Kaplan-Meier survival analysis was used at ten years. Results. A total of 89 consecutive patients (89 knees) were included with 47 females (52.8%) and a median age of 74 years (interquartile range 66 to 79). The main indications were aseptic loosening with instability (39.4%; n = 35) and infection (37.1%; n = 33) with the majority of patients managed through two-stage approach. The mean follow-up was 7.4 years (2 to 16). The overall rate of reoperation, for any cause, was 10.1% (n = 9) with a rate of implant revision of 6.7% (n = 6). Only two cases required surgery for patellofemoral complications. Kaplan-Meier implant-survivorship analysis was 93.3% at ten years, using revision for any cause as an endpoint. Conclusion. This implant achieved high ten-year survivorship with a low complication rate, particularly patellofemoral complications. These can be avoided by ensuring central patella tracking and appropriate tension of the patellofemoral joint in this posterior hinge design. Cite this article: Bone Jt Open 2022;3(3):205–210


The Bone & Joint Journal
Vol. 105-B, Issue 3 | Pages 277 - 283
1 Mar 2023
Gausden EB Puri S Chiu Y Figgie MP Sculco TP Westrich G Sculco PK Chalmers BP

Aims. The purpose of this study was to assess mid-term survivorship following primary total knee arthroplasty (TKA) with Optetrak Logic components and identify the most common revision indications at a single institution. Methods. We identified a retrospective cohort of 7,941 Optetrak primary TKAs performed from January 2010 to December 2018. We reviewed the intraoperative findings of 369 TKAs that required revision TKA from January 2010 to December 2021 and the details of the revision implants used. Kaplan-Meier analysis was used to determine survivorship. Cox regression analysis was used to examine the impact of patient variables and year of implantation on survival time. Results. The estimated survivorship free of all-cause revision was 98% (95% confidence interval (CI) 97% to 98%), 95% (95% CI 95% to 96%), and 86% (95% CI 83% to 88%) at two, five, and ten years, respectively. In 209/369 revisions there was a consistent constellation of findings with varying severity that included polyethylene wear and associated synovitis, osteolysis, and component loosening. This failure mode, which we refer to as aseptic mechanical failure, was the most common revision indication. The mean time from primary TKA to revision for aseptic mechanical failure was five years (5 months to 11 years). Conclusion. In this series of nearly 8,000 primary TKAs performed with a specific implant, we identified a lower-than-expected mid-term survivorship and a high number of revisions with a unique presentation. This study, along with the recent recall of the implant, confirms the need for frequent monitoring of patients with Optetrak TKAs given the incidence of polyethylene failure, osteolysis, and component loosening. Cite this article: Bone Joint J 2023;105-B(3):277–283


The Bone & Joint Journal
Vol. 105-B, Issue 8 | Pages 880 - 887
1 Aug 2023
Onodera T Momma D Matsuoka M Kondo E Suzuki K Inoue M Higano M Iwasaki N

Aims. Implantation of ultra-purified alginate (UPAL) gel is safe and effective in animal osteochondral defect models. This study aimed to examine the applicability of UPAL gel implantation to acellular therapy in humans with cartilage injury. Methods. A total of 12 patients (12 knees) with symptomatic, post-traumatic, full-thickness cartilage lesions (1.0 to 4.0 cm. 2. ) were included in this study. UPAL gel was implanted into chondral defects after performing bone marrow stimulation technique, and assessed for up to three years postoperatively. The primary outcomes were the feasibility and safety of the procedure. The secondary outcomes were self-assessed clinical scores, arthroscopic scores, tissue biopsies, and MRI-based estimations. Results. No obvious adverse events related to UPAL gel implantation were observed. Self-assessed clinical scores, including pain, symptoms, activities of daily living, sports activity, and quality of life, were improved significantly at three years after surgery. Defect filling was confirmed using second-look arthroscopy at 72 weeks. Significantly improved MRI scores were observed from 12 to 144 weeks postoperatively. Histological examination of biopsy specimens obtained at 72 weeks after implantation revealed an extracellular matrix rich in glycosaminoglycan and type II collagen in the reparative tissue. Histological assessment yielded a mean overall International Cartilage Regeneration & Joint Preservation Society II score of 69.1 points (SD 10.4; 50 to 80). Conclusion. This study provides evidence supporting the safety of acellular UPAL gel implantation in facilitating cartilage repair. Despite being a single-arm study, it demonstrated the efficacy of UPAL gel implantation, suggesting it is an easy-to-use, one-step method of cartilage tissue repair circumventing the need to harvest donor cells. Cite this article: Bone Joint J 2023;105-B(8):880–887


The Bone & Joint Journal
Vol. 106-B, Issue 5 | Pages 468 - 474
1 May 2024
d'Amato M Flevas DA Salari P Bornes TD Brenneis M Boettner F Sculco PK Baldini A

Aims. Obtaining solid implant fixation is crucial in revision total knee arthroplasty (rTKA) to avoid aseptic loosening, a major reason for re-revision. This study aims to validate a novel grading system that quantifies implant fixation across three anatomical zones (epiphysis, metaphysis, diaphysis). Methods. Based on pre-, intra-, and postoperative assessments, the novel grading system allocates a quantitative score (0, 0.5, or 1 point) for the quality of fixation achieved in each anatomical zone. The criteria used by the algorithm to assign the score include the bone quality, the size of the bone defect, and the type of fixation used. A consecutive cohort of 245 patients undergoing rTKA from 2012 to 2018 were evaluated using the current novel scoring system and followed prospectively. In addition, 100 first-time revision cases were assessed radiologically from the original cohort and graded by three observers to evaluate the intra- and inter-rater reliability of the novel radiological grading system. Results. At a mean follow-up of 90 months (64 to 130), only two out of 245 cases failed due to aseptic loosening. Intraoperative grading yielded mean scores of 1.87 (95% confidence interval (CI) 1.82 to 1.92) for the femur and 1.96 (95% CI 1.92 to 2.0) for the tibia. Only 3.7% of femoral and 1.7% of tibial reconstructions fell below the 1.5-point threshold, which included the two cases of aseptic loosening. Interobserver reliability for postoperative radiological grading was 0.97 for the femur and 0.85 for the tibia. Conclusion. A minimum score of 1.5 points for each skeletal segment appears to be a reasonable cut-off to define sufficient fixation in rTKA. There were no revisions for aseptic loosening at mid-term follow-up when this fixation threshold was achieved or exceeded. When assessing first-time revisions, this novel grading system has shown excellent intra- and interobserver reliability. Cite this article: Bone Joint J 2024;106-B(5):468–474


Bone & Joint Research
Vol. 11, Issue 2 | Pages 91 - 101
1 Feb 2022
Munford MJ Stoddart JC Liddle AD Cobb JP Jeffers JRT

Aims. Unicompartmental and total knee arthroplasty (UKA and TKA) are successful treatments for osteoarthritis, but the solid metal implants disrupt the natural distribution of stress and strain which can lead to bone loss over time. This generates problems if the implant needs to be revised. This study investigates whether titanium lattice UKA and TKA implants can maintain natural load transfer in the proximal tibia. Methods. In a cadaveric model, UKA and TKA procedures were performed on eight fresh-frozen knee specimens, using conventional (solid) and titanium lattice tibial implants. Stress at the bone-implant interfaces were measured and compared to the native knee. Results. Titanium lattice implants were able to restore the mechanical environment of the native tibia for both UKA and TKA designs. Maximum stress at the bone-implant interface ranged from 1.2 MPa to 3.3 MPa compared with 1.3 MPa to 2.7 MPa for the native tibia. The conventional solid UKA and TKA implants reduced the maximum stress in the bone by a factor of 10 and caused > 70% of bone surface area to be underloaded compared to the native tibia. Conclusion. Titanium lattice implants maintained the natural mechanical loading in the proximal tibia after UKA and TKA, but conventional solid implants did not. This is an exciting first step towards implants that maintain bone health, but such implants also have to meet fatigue and micromotion criteria to be clinically viable. Cite this article: Bone Joint Res 2022;11(2):91–101


Bone & Joint Open
Vol. 5, Issue 4 | Pages 277 - 285
8 Apr 2024
Khetan V Baxter I Hampton M Spencer A Anderson A

Aims. The mean age of patients undergoing total knee arthroplasty (TKA) has reduced with time. Younger patients have increased expectations following TKA. Aseptic loosening of the tibial component is the most common cause of failure of TKA in the UK. Interest in cementless TKA has re-emerged due to its encouraging results in the younger patient population. We review a large series of tantalum trabecular metal cementless implants in patients who are at the highest risk of revision surgery. Methods. A total of 454 consecutive patients who underwent cementless TKA between August 2004 and December 2021 were reviewed. The mean follow-up was ten years. Plain radiographs were analyzed for radiolucent lines. Patients who underwent revision TKA were recorded, and the cause for revision was determined. Data from the National Joint Registry for England, Wales, Northern Island, the Isle of Man and the States of Guernsey (NJR) were compared with our series. Results. No patients in our series had evidence of radiolucent lines on their latest radiological assessment. Only eight patients out of 454 required revision arthroplasty, and none of these revisions were indicated for aseptic loosening of the tibial baseplate. When compared to data from the NJR annual report, Kaplan-Meier estimates from our series (2.94 (95% confidence interval (CI) 1.24 to 5.87)) show a significant reduction in cumulative estimates of revision compared to all cemented (4.82 (95% CI 4.69 to 4.96)) or cementless TKA (5.65 (95% CI 5.23 to 6.10)). Our data (2.94 (95% CI 1.24 to 5.87)) also show lower cumulative revision rates compared to the most popular implant (PFC Sigma Cemented Knee implant fixation, 4.03 (95% CI 3.75 to 4.33)). The prosthesis time revision rate (PTIR) estimates for our series (2.07 (95% CI 0.95 to 3.83)) were lower than those of cemented cases (4.53 (95% CI 4.49 to 4.57)) from NJR. Conclusion. The NexGen trabecular (tantalum) cementless implant has lower revision rates in our series compared to all cemented implants and other types of cementless implants, and its use in younger patients should be encouraged. Cite this article: Bone Jt Open 2024;5(4):277–285


Bone & Joint Open
Vol. 5, Issue 10 | Pages 911 - 919
21 Oct 2024
Clement N MacDonald DJ Hamilton DF Gaston P

Aims. The aims were to assess whether joint-specific outcome after total knee arthroplasty (TKA) was influenced by implant design over a 12-year follow-up period, and whether patient-related factors were associated with loss to follow-up and mortality risk. Methods. Long-term follow-up of a randomized controlled trial was undertaken. A total of 212 patients were allocated a Triathlon or a Kinemax TKA. Patients were assessed preoperatively, and one, three, eight, and 12 years postoperatively using the Oxford Knee Score (OKS). Reasons for patient lost to follow-up, mortality, and revision were recorded. Results. A total of 94 patients completed 12-year functional follow-up (62 females, mean age 66 years (43 to 82) at index surgery). There was a clinically significantly greater improvement in the OKS at one year (mean difference (MD) 3.0 (95% CI 0.4 to 5.7); p = 0.027) and three years (MD 4.7 (95% CI 1.9 to 7.5); p = 0.001) for the Triathlon group, but no differences were observed at eight (p = 0.331) or 12 years’ (p = 0.181) follow-up. When assessing the OKS in the patients surviving to 12 years, the Triathlon group had a clinically significantly greater improvement in the OKS (marginal mean 3.8 (95% CI 0.2 to 7.4); p = 0.040). Loss to functional follow-up (53%, n = 109/204) was independently associated with older age (p = 0.001). Patient mortality was the major reason (56.4%, n = 62/110) for loss to follow-up. Older age (p < 0.001) and worse preoperative OKS (p = 0.043) were independently associated with increased mortality risk. An age at time of surgery of ≥ 72 years was 75% sensitive and 74% specific for predicting mortality with an area under the curve of 78.1% (95% CI 70.9 to 85.3; p < 0.001). Conclusion. The Triathlon TKA was associated with clinically meaningful greater improvement in knee-specific outcome when compared to the Kinemax TKA. Loss to follow-up at 12 years was a limitation, and studies planning longer-term functional assessment could limit their cohort to patients aged under 72 years. Cite this article: Bone Jt Open 2024;5(10):911–919


Bone & Joint Research
Vol. 11, Issue 4 | Pages 229 - 238
11 Apr 2022
Jaeger S Eissler M Schwarze M Schonhoff M Kretzer JP Bitsch RG

Aims. One of the main causes of tibial revision surgery for total knee arthroplasty is aseptic loosening. Therefore, stable fixation between the tibial component and the cement, and between the tibial component and the bone, is essential. A factor that could influence the implant stability is the implant design, with its different variations. In an existing implant system, the tibial component was modified by adding cement pockets. The aim of this experimental in vitro study was to investigate whether additional cement pockets on the underside of the tibial component could improve implant stability. The relative motion between implant and bone, the maximum pull-out force, the tibial cement mantle, and a possible path from the bone marrow to the metal-cement interface were determined. Methods. A tibial component with (group S: Attune S+) and without (group A: Attune) additional cement pockets was implanted in 15 fresh-frozen human leg pairs. The relative motion was determined under dynamic loading (extension-flexion 20° to 50°, load-level 1,200 to 2,100 N) with subsequent determination of the maximum pull-out force. In addition, the cement mantle was analyzed radiologically for possible defects, the tibia base cement adhesion, and preoperative bone mineral density (BMD). Results. The BMD showed no statistically significant difference between both groups. Group A showed for all load levels significantly higher maximum relative motion compared to group S for 20° and 50° flexion. Group S improved the maximum failure load significantly compared to group A without additional cement pockets. Group S showed a significantly increased cement adhesion compared to group A. The cement penetration and cement mantle defect analysis showed no significant differences between both groups. Conclusion. From a biomechanical point of view, the additional cement pockets of the component have improved the fixation performance of the implant. Cite this article: Bone Joint Res 2022;11(4):229–238


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_7 | Pages 25 - 25
1 Jul 2022
Phillips J Tucker K
Full Access

Abstract. Introduction. There are a wide variety of implant brands and types of knee replacement available to surgeons. With time, the options available within many implant brand portfolios has grown, with alternative tibial or femoral components, tibial insert materials or shapes and patella resurfacings. Aim. To investigate the effect of the expansion of implant brand portfolios, and to establish the potential numbers of compatible implant construct combinations. Methods. Hypothetical implant brand portfolios were proposed, and the number of compatible implant construct combinations was calculated. Results. A simple knee portfolio with cemented cruciate-retaining (CR) and posterior-stabilised (PS) components, with and without a patella has 4 combinations. If there are two options available for each, the numbers double for each option, resulting in 32 combinations. The effect of adding a third option multiplies the number by 1.3. Introducing compatible uncemented options, with the effect of hybrids multiplies the number by 4. An implant portfolio with two femoral components (both in CR and PS), with two insert options and a patella, all in cemented and uncemented versions leads to 192 possible compatible implant construct combinations. There are implant brands available with many more than two options available for use. Conclusion. This study demonstrates that the addition of multiple variants within a knee brand portfolio leads to a large number (many hundreds) of compatible implant construct combinations. Revision rates of implant combinations are not currently reviewed at this level of granularity, leading to the risk of implant camouflage


The Bone & Joint Journal
Vol. 104-B, Issue 4 | Pages 444 - 451
1 Apr 2022
Laende EK Mills Flemming J Astephen Wilson JL Cantoni E Dunbar MJ

Aims. Thresholds of acceptable early migration of the components in total knee arthroplasty (TKA) have traditionally ignored the effects of patient and implant factors that may influence migration. The aim of this study was to determine which of these factors are associated with overall longitudinal migration of well-fixed tibial components following TKA. Methods. Radiostereometric analysis (RSA) data over a two-year period were available for 419 successful primary TKAs (267 cemented and 152 uncemented in 257 female and 162 male patients). Longitudinal analysis of data using marginal models was performed to examine the associations of patient factors (age, sex, BMI, smoking status) and implant factors (cemented or uncemented, the size of the implant) with maximum total point motion (MTPM) migration. Analyses were also performed on subgroups based on sex and fixation. Results. In the overall group, only fixation was significantly associated with migration (p < 0.001). For uncemented tibial components in males, smoking was significantly associated with lower migration (p = 0.030) and BMI approached significance (p = 0.061). For females with uncemented components, smoking (p = 0.081) and age (p = 0.063) approached significance and were both associated with increased migration. The small number of self-reported smokers in this study warrants cautious interpretation and further investigation. For cemented components in females, larger sizes of tibial component were significantly associated with increased migration (p = 0.004). No factors were significant for cemented components in males. Conclusion. The migration of uncemented tibial components was more sensitive to patient factors than cemented implants. These differences were not consistent by sex, suggesting that it may be of value to evaluate female and male patients separately following TKA. Cite this article: Bone Joint J 2022;104-B(4):444–451


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_13 | Pages 42 - 42
7 Aug 2023
Al-Jabri T Brivio A Martin J Barrett D Maffulli N
Full Access

Abstract. Background. Infections are rare and poorly studied complications of unicompartmental knee athroplasty (UKA) surgery. They are significantly less common compared to infections after total knee arthroplasties (TKAs). Optimal management of periprosthetic joint infections (PJIs) after a UKA is not clearly defined in the literature. We present the results of a multicentre retrospective series of UKA PJIs treated with Debridement, Antibiotics and Implant Retention (DAIR). Methodology. Patients presenting between January 2016 and December 2019 with early UKA infection were identified at three specialist centres using the Musculoskeletal Infection Society (MSIS) criteria. All patients underwent a standardized treatment protocol consisting of the DAIR procedure and antibiotic therapy comprising two weeks of intravenous (IV) antibiotics followed by six weeks of oral therapy. The main outcome measure was overall survivorship free from reoperation for infection. Results. A total of 3225 UKAs (2793 (86.2%) medial and 432 (13.8%) lateral UKAs) were performed between January 2016 and December 2019. Nineteen patients had early infections necessitating DAIR. The mean follow-up period was 32.5 months. DAIR showed an overall survivorship free from septic reoperation of 84.2%, with an overall survivorship free from all-cause reoperation of 78.95%. The most common bacteria were Coagulase-negative Staphylococci, Staphylococcus aureus and Group B Streptococci. Three patients required a second DAIR procedure but remained free from re-infection at follow-up obviating the need for more demanding, staged revision surgery. Conclusions. In infected UKAs, the DAIR procedure produces a high rate of success, with a high survivorship of the implant


The Bone & Joint Journal
Vol. 103-B, Issue 10 | Pages 1555 - 1560
4 Oct 2021
Phillips JRA Tucker K

Aims. Knee arthroplasty surgery is a highly effective treatment for arthritis and disorders of the knee. There are a wide variety of implant brands and types of knee arthroplasty available to surgeons. As a result of a number of highly publicized failures, arthroplasty surgery is highly regulated in the UK and many other countries through national registries, introduced to monitor implant performance, surgeons, and hospitals. With time, the options available within many brand portfolios have grown, with alternative tibial or femoral components, tibial insert materials, or shapes and patella resurfacings. In this study we have investigated the effect of the expansion of implant brand portfolios and where there may be a lack of transparency around a brand name. We also aimed to establish the potential numbers of compatible implant construct combinations. Methods. Hypothetical implant brand portfolios were proposed, and the number of compatible implant construct combinations was calculated. Results. A simple knee portfolio with cemented cruciate-retaining (CR) and posterior-stabilized (PS) components, with and without a patella, has four combinations. If there are two options available for each component, the numbers double for each option, resulting in 32 combinations. The effect of adding a third option multiplies the number by 1.3. Introducing compatible uncemented options, with the effect of hybrids, multiplies the number by 4. An implant portfolio with two femoral components (both in CR and PS), with two insert options and a patella, all in cemented and uncemented versions leads to 192 possible compatible implant construct combinations. There are implant brands available to surgeons with many more than two options. Conclusion. This study demonstrates that the addition of multiple variants within a knee brand portfolio leads to a large number (many hundreds) of compatible implant construct combinations. Revision rates of implant combinations are not currently reviewed at this level of granularity, leading to the risk of camouflage of true outcomes. Cite this article: Bone Joint J 2021;103-B(10):1555–1560


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_13 | Pages 15 - 15
7 Aug 2023
Deo S Jonas S Jhaj J
Full Access

Abstract. INTRODUCTION. The most frequent mode of aseptic failure of primary total knee replacements is tibial baseplate loosening. This is influenced by stresses across the implant-bone interface which can be increased in obese patients leading to potentially higher rates of early failure. The evidence is mixed as to the true effect of elevated BMI (body mass index) on revision rates. We present the experience of early tibial failures in our department and how our implant choices have evolved. METHODOLOGY. We retrospectively reviewed our unit's arthroplasty database and identified all patients who had sustained mechanical tibial failure. Data were collected on patient demographics, operative details of primary and revision operations, components used, alignment pre and post operatively and indication. Complications and further surgery performed were recorded. RESULTS. 12 patients were identified who had been revised for mechanical tibial failure. All were female, mean age 60 years (46–76). Mean BMI was 11/12 had significant comorbidities and 6/12 had risk factors for reduced bone density (inflammatory arthritis, diabetes, hypothyroid). CONCLUSION. In our consecutive series of 12 patients who presented with tibial mechanical failure, all were female and had a high BMI with the majority over 40. We recommend that a stemmed tibia construct should be used in female patients with an elevated BMI, particularly if it is over 40


The Bone & Joint Journal
Vol. 102-B, Issue 6 Supple A | Pages 85 - 90
1 Jun 2020
Blevins JL Rao V Chiu Y Lyman S Westrich GH

Aims. The purpose of this investigation was to determine the relationship between height, weight, and sex with implant size in total knee arthroplasty (TKA) using a multivariate linear regression model and a Bayesian model. Methods. A retrospective review of an institutional registry was performed of primary TKAs performed between January 2005 and December 2016. Patient demographics including patient age, sex, height, weight, and body mass index (BMI) were obtained from registry and medical record review. In total, 8,100 primary TKAs were included. The mean age was 67.3 years (SD 9.5) with a mean BMI of 30.4 kg/m. 2. (SD 6.3). The TKAs were randomly split into a training cohort (n = 4,022) and a testing cohort (n = 4,078). A multivariate linear regression model was created on the training cohort and then applied to the testing cohort . A Bayesian model was created based on the frequencies of implant sizes in the training cohort. The model was then applied to the testing cohort to determine the accuracy of the model at 1%, 5%, and 10% tolerance of inaccuracy. Results. Height had a relatively strong correlation with implant size (femoral component anteroposterior (AP) Pearson correlation coefficient (ρ) = 0.73, p < 0.001; tibial component mediolateral (ML) ρ = 0.77, p < 0.001). Weight had a moderately strong correlation with implant size, (femoral component AP ρ = 0.46, p < 0.001; tibial ML ρ = 0.48, p < 0.001). There was a significant linear correlation with height, weight, and sex with implant size (femoral component R. 2. = 0.607, p < 0.001; tibial R. 2. = 0.695, p < 0.001). The Bayesian model showed high accuracy in predicting the range of required implant sizes (94.4% for the femur and 96.6% for the tibia) accepting a 5% risk of inaccuracy. Conclusion. Implant size was correlated with basic demographic variables including height, weight, and sex. The linear regression and Bayesian models accurately predicted required implant sizes across multiple manufacturers based on height, weight, and sex alone. These types of predictive models may help improve operating room and implant supply chain efficiency. Level of Evidence: Level IV. Cite this article: Bone Joint J 2020;102-B(6 Supple A):85–90


The Bone & Joint Journal
Vol. 101-B, Issue 7_Supple_C | Pages 61 - 63
1 Jul 2019
Lawrie CM Schwabe M Pierce A Nunley RM Barrack RL

Aims. The aim of this study was to compare the actual cost of a cemented and cementless total knee arthroplasty (TKA) procedure. Materials and Methods. The cost of operative time, implants, cement, and cementing accessories were included in the overall cost of the TKA procedure. Operative time was determined from a previously published study comparing cemented and cementless implants of the same design. The cost of operative time, implants, cement, and cementing accessories was determined from market and institutional data. Results. Mean operative time for cemented TKA was 11.6 minutes longer for cemented TKA than cementless TKA (93.7 minutes (. sd. 16.7) vs 82.1 minutes (. sd. 16.6); p = 0.001). Using a conservative published standard of $36 per minute for operating theatre time cost, the total time cost was $418 higher for cementing TKA. The cost of cement and accessories ranged from $170 to $625. Overall, the calculated cost of cemented TKA is $588 to $1043, depending on technique. The general increased charge for cementless TKA implants over cemented TKA implants was $366. Conclusion. The overall procedural cost of implanting a cementless TKA is less than implanting a cemented TKA. Cost alone should not be a barrier to using cementless TKA. Cite this article: Bone Joint J 2019;101-B(7 Supple C):61–63


The Bone & Joint Journal
Vol. 103-B, Issue 5 | Pages 840 - 845
1 May 2021
Rossi SMP Perticarini L Clocchiatti S Ghiara M Benazzo F

Aims. In the last decade, interest in partial knee arthroplasties and bicruciate retaining total knee arthroplasties has increased. In addition, patient-related outcomes and functional results such as range of movement and ambulation may be more promising with less invasive procedures such as bicompartmental arthroplasty (BCA). The purpose of this study is to evaluate clinical and radiological outcomes after a third-generation patellofemoral arthroplasty (PFA) combined with a medial or lateral unicompartmental knee arthroplasty (UKA) at mid- to long-term follow-up. Methods. A total of 57 procedures were performed. In 45 cases, a PFA was associated with a medial UKA and, in 12, with a lateral UKA. Patients were followed with validated patient-reported outcome measures (Oxford Knee Score (OKS), EuroQol five-dimension questionnaire (EQ-5D), EuroQoL Visual Analogue Scale (EQ-VAS)), the Knee Society Score (KSS), the Forgotten Joint Score (FJS), and radiological analysis. Results. The mean follow-up was nine years (6 to 13). All scores significantly improved from preoperatively to final follow-up (mean and SD): OKS from 23.2 (8.1) to 42.5 (3.5), EQ-5D from 0.44 (0.25) to 0.815 (0.1), EQ-VAS from 46.7 (24.9) to 89.1 (9.8), KSS (Knee) from 51.4 (8.5) to 94.4 (4.2), and KSS (Function) from 48.7 (5.5) to 88.8 (5.2). The mean FJS at final follow-up was 79.2 (4.2). All failures involved the medial UKA + PFA group. Overall, survival rate was 91.5% for all the combined implants at ten years with 95% confidence intervals and 22 knees at risk. Conclusion. Excellent clinical and radiological outcomes were achieved after a third-generation PFA combined with a medial or lateral UKA. BCA with unlinked partial knee prostheses showed a good survival rate at mid- to long-term follow-up. Cite this article: Bone Joint J 2021;103-B(5):840–845


The Bone & Joint Journal
Vol. 105-B, Issue 12 | Pages 1279 - 1285
1 Dec 2023
Baker JF Nadar AC Jouflas AC Smith LS Sachdeva S Yakkanti MR Malkani AL

Aims

The use of cementless total knee arthroplasty (TKA) components has increased during the past decade. The initial design of cementless metal-backed patellar components had shown high failure rates due to many factors. The aim of this study was to evaluate the clinical results of a second-generation cementless, metal-backed patellar component of a modern design.

Methods

This was a retrospective review of 707 primary TKAs in 590 patients from a single institution, using a cementless, metal-backed patellar component with a mean follow-up of 6.9 years (2 to 12). A total of 409 TKAs were performed in 338 females and 298 TKAs in 252 males. The mean age of the patients was 63 years (34 to 87) and their mean BMI was 34.3 kg/m2 (18.8 to 64.5). The patients were chosen to undergo a cementless procedure based on age and preoperative radiological and intraoperative bone quality. Outcome was assessed using the Knee Society knee and function scores and range of motion (ROM), complications, and revisions.


Abstract. Introduction. Medial fix bearing unicompartmental knee replacement (UKR) designs are consider safe and effective implants with many registries data and big cohort series showing excellent survivorship and clinical outcome comparable to that reported for the most expensive and surgically challenging medial UKR mobile bearing designs. However, whether all polyethylene tibial components (all-poly) provided comparable results to metal-backed modular components during medial fix bearing UKR remains unclear. There have been previous suggestions that all-poly tibia UKR implants might show unacceptable higher rates of early failure due to tibial component early loosening especially in high body max index (BMI) patients. This study aims to find out the short and long-term survival rate of all-poly tibia UKR and its relationship with implant thickness and patient demographics including sex, age, ASA and BMI. Material and Methods. we present the results of a series of 388 medial fixed bearing all-polly tibia UKR done in our institution by a single surgeon between 2007–2019. Results. We found out excellent implant survival with this all-poly tibia UKR design with 5 years survival rate: 96.42%, 7 years survival rate: 95.33%, and 10 years survival rate: 91.87%. Only 1.28% had early revision within 2 years. Conclusion. Fixed bearing medial all-poly tibia UKR shows excellent survivor rate at 2, 5, 7 and 10 years follow up and the survival rate is not related with sex, age, BMI, ASA grade or implant thickness. Contrary to the popular belief, we found out that only 1.71% of all implants was revised due to implant loosening


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_9 | Pages 70 - 70
1 Oct 2020
Staats K Sosa BR Kuyl E Niu Y Suhardi VJ Turajane K Windhager R Greenblatt MB Ivashkiv L Bostrom MP Yang X
Full Access

Introduction. Initial post-operative implant instability leads to impaired osseointegration, one of the most common reasons for aseptic loosening and revision surgery. In this study, we developed a novel murine model of implant instability and demonstrated the anabolic effect of immediate and delayed intermittent Parathyroid Hormone (iPTH) treatment in the setting of instability-induced osseointegration failure. Methods. 3D-printed titanium implants were inserted in an oversized drill-hole in the tibia of C57Bl/6 mice (n=54). After implantation, the mice were randomly divided in 3 treatment groups (control: PBS-vehicle; iPTH; delayed iPTH). Radiographic analysis was performed to confirm signs of implant loosening. Peri-implant tissue formation was assessed through histology. Osseointegration was assessed through µCT and biomechanical pullout testing. Results. Immediate iPTH treatment reduced radiolucencies and induced a distinct pedestal sign distal to the implant stem (white arrow Fig 1A). The PBS treated mice had fibrous tissue implant encapsulation, whereas new mineralized tissue and no fibrous tissue was observed with immediate iPTH treatment (Fig 1E). Delayed iPTH treatment was also able to exhibit significant peri-implant bone mineralization, osteoblasts, angiogenesis, and a reduction of fibrous tissue (Fig 2A-B). iPTH treatment increased the force required to pull out the implant significantly from 8.41 ± 8.15N in the PBS group to 21.49 ± 10.45N and 23.68 ± 8.99N, in the immediate and delayed iPTH treatment groups, respectively (Fig 2D). PBS vehicle resulted in a bone volume/trabecular volume (BV/TV) of 0.23 ± 0.03, while immediate and delayed iPTH treatment increased BV/TV significantly to 0.46 ± 0.07 and 0.34 ± 0.10, respectively (Fig 2E). Conclusion. Immediate iPTH treatment prevents peri-implant fibrous tissue formation and enhances peri-implant bone formation in our murine model of mechanical instability. Delayed iPTH treatment was able to resolve the peri-implant fibrous tissue and stimulate bone formation. This study potentially addresses a leading cause of aseptic loosening by demonstrating that initial implant instability can be rescued by iPTH even with delayed start of treatment. For any figures or tables, please contact authors directly


The Bone & Joint Journal
Vol. 101-B, Issue 7_Supple_C | Pages 48 - 54
1 Jul 2019
Kahlenberg CA Lyman S Joseph AD Chiu Y Padgett DE

Aims. The outcomes of total knee arthroplasty (TKA) depend on many factors. The impact of implant design on patient-reported outcomes is unknown. Our goal was to evaluate the patient-reported outcomes and satisfaction after primary TKA in patients with osteoarthritis undergoing primary TKA using five different brands of posterior-stabilized implant. Patients and Methods. Using our institutional registry, we identified 4135 patients who underwent TKA using one of the five most common brands of implant. These included Biomet Vanguard (Zimmer Biomet, Warsaw, Indiana) in 211 patients, DePuy/Johnson & Johnson Sigma (DePuy Synthes, Raynham, Massachusetts) in 222, Exactech Optetrak Logic (Exactech, Gainesville, Florida) in 1508, Smith & Nephew Genesis II (Smith & Nephew, London, United Kingdom) in 1415, and Zimmer NexGen (Zimmer Biomet) in 779 patients. Patients were evaluated preoperatively using the Knee Injury and Osteoarthritis Outcome Score (KOOS), Lower Extremity Activity Scale (LEAS), and 12-Item Short-Form Health Survey questionnaire (SF-12). Demographics including age, body mass index, Charlson Comorbidity Index, American Society of Anethesiologists status, sex, and smoking status were collected. Postoperatively, two-year KOOS, LEAS, SF-12, and satisfaction scores were compared between groups. Results. Outcomes were available for 4069 patients (98%) at two years postoperatively. In multiple regression analysis, which separately compared each implant group with the aggregate of all others, there were no clinically significant differences in the change of KOOS score from baseline to two-year follow-up between any of the groups. More than 80% of patients in each group were satisfied at this time in all domains. In a multivariate regression model, patients in the NexGen group were the most likely to be satisfied (odds ratio (OR) 1.63; p = 0.006) and Optetrak Logic patients were the least likely to be satisfied (OR 0.60; p < 0.001). Conclusion. TKA provides improvement in function and satisfaction regardless of the type of implant. We could not demonstrate superiority of one design above others across these groups of implants, and any price premium for one above the other systems may not be justified. Healthcare administrators may find these similarities in outcomes helpful when negotiating purchasing contracts. Cite this article: Bone Joint J 2019;101-B(7 Supple C):48–54