Although femoral neck fractures remain a concern in short term failures for hip resurfacing, acetabular component position and fixation are increasingly being recognized as causes of mid term failures for hip resurfacing. This study aimed to evaluate the migration pattern of a cobalt chrome, beaded acetabular component for a metal on metal hip resurfacing. 130 patients underwent metal on metal hip resurfacing: 66 hips in 60 patients had sufficient films to be included in this analysis. Forty-eight patients were male and 12 were female, with a mean age of 50 (range, 32–66). Ninety-five percent of the procedures were performed for osteoarthritis. In all cases acetabular migration was measured both vertically and horizontally, on serial radiographs using the computer-assisted Ein Bild Röntgen Analyse method. A minimum of three comparable radiographs is necessary for calculating the migration curves. We scored medial migration as negative horizontal movement.Introduction
Methods
Over the last decade stemless shoulder arthroplasty has become increasingly popular. However, stability of metaphyseal loading humeral components remains a concern. This study aimed to assess the stability of the Affinis stemless humeral component using Radiostereometric analysis (RSA). Patients underwent total shoulder arthroplasty via a standardised technique with a press-fit stemless humeral component and a cemented pegged glenoid. Tantalum beads were inserted into the humerus at the time of operation. RSA of the relaxed shoulder was completed at weeks 1, 6, 13, 26, 52 and 104 post-operatively. Stressed RSA with 12 newtons of abduction force was completed from week 13 onwards. ABRSA 5.0 software (Downing Imaging Limited, Aberdeen) was used to calculate humeral
Introduction. Current artificial discs include 1 or 2 bearing surfaces, providing 3 or 5 degrees of freedom. The ESP® is a one-piece e implant made of silicon and polycarbonate -urethane securely fixed to titanium endplates. It allows limited rotation and translation with elastic return. This cushion without fixed rotation center achieves 6 degrees of freedom including shock absorption. This objective of this study was to evaluate the safety and efficacy of the concept in a prospective nonrandomized trial. Material and methods. Prior to clinical implantations, the device was endurance tested at least 40 millions cycles. The polymer core weight and properties, the geometrical characteristics and cohesion of the implants remained stable. A prospective trial was initiated in 2004 for L3L4, L4L5 and L5S1 levels. Total disc replacements have been performed in 153 lumbar levels through extra-peritoneal mini-invasive anterior approach with a minimum 2 years follow-up. Results. There was no clinical or radiological device related complication, except 2 early revisions for post-traumatic
There is a lack of long-term data for minimally invasive acromioclavicular (AC) joint repair. Furthermore, it is not clear if good early clinical results can be maintained over time. The purpose of this study was to report long-term results of minimally invasive AC joint reconstruction (MINAR) and compare it to corresponding short-term data. We assessed patients with a follow-up of at least five years after minimally invasive flip-button repair for high-grade AC joint dislocation. The clinical outcome was evaluated using the Constant score and a questionnaire. Ultrasound determined the coracoclavicular (CC) distance. Results of the current follow-up were compared to the short-term results of the same cohort.Aims
Methods
The spiral blade modification of the Dynamic
Hip Screw (DHS) was designed for superior biomechanical fixation
in the osteoporotic femoral head. Our objective was to compare clinical
outcomes and in particular the incidence of loss of fixation. In a series of 197 consecutive patients over the age of 50 years
treated with DHS-blades (blades) and 242 patients treated with conventional
DHS (screw) for AO/OTA 31.A1 or A2 intertrochanteric fractures were
identified from a prospectively compiled database in a level 1 trauma
centre. Using propensity score matching, two groups comprising 177
matched patients were compiled and radiological and clinical outcomes
compared. In each group there were 66 males and 111 females. Mean
age was 83.6 (54 to 100) for the conventional DHS group and 83.8
(52 to 101) for the blade group. Loss of fixation occurred in two blades and 13 DHSs. None of
the blades had observable migration while nine DHSs had gross migration
within the femoral head before the fracture healed. There were two
versus four implant cut-outs respectively and one side plate pull-out
in the DHS group. There was no significant difference in mortality
and eventual walking ability between the groups. Multiple logistic
regression suggested that poor reduction (odds ratio (OR) 11.49,
95% confidence intervals (CI) 1.45 to 90.9, p = 0.021) and fixation
by DHS (OR 15.85, 95%CI 2.50 to 100.3, p = 0.003) were independent
predictors of loss of fixation. The spiral blade design may decrease the risk of implant migration
in the femoral head but does not reduce the incidence of cut-out
and reoperation. Reduction of the fracture is of paramount importance
since poor reduction was an independent predictor for loss of fixation
regardless of the implant being used. Cite this article:
The aim of this study was to analyse the functional outcome after a displaced intracapsular fracture of the femoral neck in active patients aged over 70 years without osteoarthritis or rheumatoid arthritis of the hip, randomised to receive either a hemiarthroplasty or a total hip replacement (THR). We studied 252 patients of whom 47 (19%) were men, with a mean age of 81.1 years (70.2 to 95.6). They were randomly allocated to be treated with either a cemented hemiarthroplasty (137 patients) or cemented THR (115 patients). At one- and five-year follow-up no differences were observed in the modified Harris hip score, revision rate of the prosthesis, local and general complications, or mortality. The intra-operative blood loss was lower in the hemiarthroplasty group (7% >
500 ml) than in the THR group (26% >
500 ml) and the duration of surgery was longer in the THR group (28% >
1.5 hours Because of a higher intra-operative blood loss (p <
0.001), an increased duration of the operation (p <
0.001) and a higher number of early and late dislocations (p = 0.002), we do not recommend THR as the treatment of choice in patients aged ≥ 70 years with a fracture of the femoral neck in the absence of advanced radiological osteoarthritis or rheumatoid arthritis of the hip.