Introduction. Large diameter metal-on-metal hip arthroplasty (LDMMTHA) provides benefits of reduced dislocation rates and low wear. The use of modular systems allows better restoration of hip biomechanics. There have been reports of modular LDMMTHAs with tapered sleeves generating excessively high metal ions, due to possible mismatch between the titanium stem and the
While Oxidized Zirconium (OxZr) femoral heads matched with highly cross-linked polyethylene (XLPE) have demonstrated the lowest rate of revision compared to other bearing couples in the Australian National Joint Registry, it has been postulated that these results may, in part, be due to the fact that a single company offers this bearing option with a limited combination of femoral and acetabular prostheses. The purpose of this study was to assess clinical and radiographic outcomes in a matched cohort of total hip replacements (THR) utilizing an identical cementless femoral stem and acetabular component with either an Oxidized Zirconium (OxZr) or
Aim. To provide proof of concept in an in vivo animal model for the prevention of prosthetic joint infection prevention using electric fields along with conventional antibiotic prophylaxis. Corresponding Author: Marti Bernaus. Method. First, we standardized the animal model to simulate implant contamination during the surgical procedure. We then implanted
Total Knee Arthroplasty (TKA) patients may present with effusion, pain, stiffness and functional impairment. A positive metal hypersensitivity (positive LTT) may be an indication for a revision surgery with a custom-made implant devoid of any hypersensitivity-related metal or an implant with the least possible ion content of the metal hypersensitivity, if no custom-made is available. The purpose of the current study is to assess the prevalence of metal hypersensitivity in subjects requiring a primary TKA and assess their early functional outcomes. We are recruiting 660 subjects admitted for TKA. Subjects are randomly assigned to 2 groups: oxidized zirconium implant group or
We report a case of fatal heart failure caused by cobalt intoxication after revision THR in the patient who successfully underwent re-revision THR. 53-year old male presented to emergency room in our hospital with progressive shortness of breath. Symptom was started about 6 months ago so he visited local hospital. He worked up for worsening dyspnea. Simple chest radiograph and enhanced heart MRI study were performed and they showed bilateral pericardial and pleural effusion. There was no evidence of ischemic change. Transthoracic echocardiogram showed the evidence of heart failure, left ventricular ejection fraction(EF) was 40%. He was admitted at local hospital and started on vasopressors but urine output was decreased and follow-up echocardiogram showed a 25% of EF. Patient recommended heart transplantation and transferred our hospital emergency room. He underwent sequential bilateral total hip arthroplasties using CoP bearing surfaces. At 12 years postoperatively, he presented to the other hospital with acute onset of left hip pain. He was diagnosed ceramic head fracture on his left hip. Head and liner change revision surgery was performed using
A multitude of different bearing combinations exist to recreate the artificial hip joint. To date, there is no particular ‘gold-standard’ total hip arthroplasty (THA) couple since none is faultless. Strategies to improve performance are aimed either at modifying the shape and design of components or their material properties. Wear particle generation is now well recognised as a cause of aseptic loosening which consistently features amongst the most common indication for revision THA and thus minimising wear lies at the cornerstone of developing bearing couples. However, history has shown the use of supposed newer and improved materials have not been without occasional catastrophic failure. Hard-on-hard bearings are theoretically more resistant to wear but component fracture and squeaking has been witnessed with ceramic-on-ceramic articulations whilst metal-on-metal articulations have been plagued by reports of pseudotumor and ALVAL formation. This has all led to resurgence in the hard-on-soft couple. More recently, corrosion at taper junctions has been identified as a significant factor in hip arthroplasty failure. Femoral head materials, surface changes or coatings may therefore have an increasing role to play. In 2005, a multi-center, prospective, assessor and patient-blinded, randomised control trial was initiated. This was designed as a three armed study with either
INTRODUCTION. Recently there have been case reports of component fractures and elevated metal ion levels potentially resulting from the use of
Uncemented metal-on-polyethylene total hip arthroplasties (THAs) have had a modular
Background. Periprosthetic joint infection (PJI) is a devastating complication and interest exists in finding lower cost alternatives to current management strategies. Current strategies include a two-stage revision with placement of an antibiotic spacer and delayed placement of a new arthroplasty implant. This study aimed to show that biofilm residue can be reliably eradicated on infected implants, safely allowing re-implantation in a spacer. Methods. Strains of Staphylococcus aureus MRSA252 or Staphylococcus epidermidis RP62A were grown on
The purpose of this investigation is to assess the rate of wear the effect once the “bedding in period”/ poly creep had been eliminated. Creep is the visco-elastic deformation that polyethylene exhibits in the first 6–12 weeks. We also assessed the wear pattern of four different bearing couples in total hip arthroplasty (THA):
Introduction. Metal ion and particle release, particularly cobalt, has become an important subject in total hip arthroplasty, as it has shown to induce metal hypersensitivity, adverse local tissue reactions and systemic ion related diseases. The purpose of the following study was compare the ion release barrier function of a zirconium nitride (ZrN) multilayer coated hip stem for cemented use, designed for patients with metal ion hypersensitivity, against its uncoated version in a test configuration simulating the worst case scenario of a severely debonded hip stem. The ZrN multilayer coating is applied on a CoCrMo hip stem and consists of a thin adhesive chromium layer, five alternating intermediate layers out of chromium nitride (CrN) and chromium carbonitride (CrCN) and a final zirconium nitride (ZrN) shielding layer [1]. Methods. Hip stems with a ZrN multilayer coating (CoreHip AS, Aesculap AG, Germany) were tested in comparison with a
Periacetabular osteolysis in association with well-fixed cementless components was first recognised as a serious clinical problem in the early 1990s. By the mid-1990s, revision surgery for pelvic osteolysis secondary to polyethylene wear was the most common revision hip procedure performed. As a result, new bearing surfaces were introduced in hopes of reducing wear volume and thus reducing pelvic osteolysis. These included highly crosslinked polyethylene, ceramic-on-ceramic and metal-on-metal bearing surfaces. Metal-on-metal has for the most part been eliminated in conventional hip replacement because of the concerns centered around adverse local tissue reactions. Both highly crosslinked polyethylene and ceramic-on-ceramic bearings have been successful in limiting wear and all but eliminating clinically significant osteolysis. Multiple reports on highly crosslinked polyethylene have documented wear rates below the lysis threshold. No reports of revision for wear have been reported despite twenty years of in-vivo use. Of import to the surgeons, all manufacturers commonly used in North America have performed well. In addition, highly crosslinked polyethylene has been relatively insensitive to head size allowing the use of 36mm femoral heads routinely. Similar reports are noted with ceramic-on-ceramic bearings. However, highly crosslinked has dominated the North American market because it is a relatively forgiving bearing surface and comes at a lower cost. Currently, there is a trend towards the use of ceramic femoral heads – not because of wear concerns, but concerns related to taper corrosion and large
There has been much discussion and controversy in the media recently regarding metal toxicity following large head metal on metal (MoM) total hip replacement (THR). Patients have been reported as having hugely elevated levels of metal ions with, at times, devastating systemic, neurolgical and/or orthopaedic sequelae. However, no direct correlation between metal ion level and severity of metallosis has yet been defined. Normative levels of metal ions in well functioning, non
Aims. Metal-on-metal total hip replacements (MoM THRs) are frequently revised. However, there is a paucity of data on clinical outcomes following revision surgery in this cohort. We report on outcomes from the largest consecutive series of revisions from MoM THRs and consider pre-revision factors which were prognostic for functional outcome. Materials and Methods. A single-centre consecutive series of revisions from MoM THRs performed during 2006–2015 was identified through a prospectively maintained, purpose-built joint registry. The cohort was subsequently divided by the presence or absence of symptoms prior to revision. The primary outcome was functional outcome (Oxford Hip Score (OHS)). Secondary outcomes were complication data, pre- and post-revision serum metal ions and modified Oxford classification of pre-revision magnetic resonance imaging (MRI). In addition, the study data along with demographic data was interrogated for prognostic factors informing on post-revision functional outcome. Results. 180 revisions in 163 patients were identified at a median follow-up of 5.48 (2–11.7) years. There were 152 (84.4%) in the symptomatic subgroup and 28 (15.6%) in the asymptomatic group. Overall median OHS improved from 29 to 37 with revision (P<0.001). Symptomatic patients experienced greater functional benefit (DOHS 6.5 vs. 1.4, p=0.012) compared to asymptomatic patients, though they continued to report inferior outcomes (OHS 36.5 vs 43, p=0.004). The functional outcome of asymptomatic patients was unaffected by revision surgery (pre-revision OHS 41, post-revision OHS 43, p=0.4). Linear regression analysis confirmed use of a
There are pros and cons of all bearing surface options for our young patients. I pick the bearing surface for my young patients trying to maximise durability and minimise risks. For the ultra-young, ≤30 years of age patient, I use ceramic-on-ceramic. The pros of this are the best wear couple available and a favorable track record (with well designed implants). The risks can be minimised: fracture risk now decreased, runaway wear minimised with good surgical technique, impingement problems minimised with good technique and well designed implants, as well as squeaking is minimised with good design (majority of reported squeakers are of one designed socket). I don't use metal-on-metal because I am not willing to subject young patients to potentially 50+ years of high metal ion exposure. I also don't use HCLPE. This would be okay from a biologic standpoint but I still have concerns about long-term wear durability. So the marked superior wear characteristics of ceramic-ceramic win in my view. For my middle age patients, 30–60, I use HCLPE I don't use ceramic-ceramic because at some point between 30 and 60 years of age the improved wear properties are outweighed by their potential risks (fracture, impingement, squeak). HCLPE at short F/U (<15 years) appears to be durable, reliable with good wear properties so it is a reasonable choice. Using a ceramic head versus CoCr provides minimal improvement in wear properties in the lab but no marked advantage in vivo. Concerns persist about
In recent years, cementless stems have dominated the North American market. There are several categories of cementless stems, but in the past 20 years, the two most popular designs in the United States have been the extensively coated cylindrical
Background/Purpose. Cross-linked polyethylene (XLPE) has shown reduced wear rates as compared to conventional polyethylene, but the long-term effect of this on the incidence of osteolysis remains unclear. In addition, the measurement of osteolysis on plain radiographs can underestimate the incidence and extent of osteolysis. Therefore, we evaluated the wear rate, incidence and volume of osteolysis at a minimum follow-up of ten years using three-dimensional computed tomography (3-D CT), a more accurate and sensitive method for detecting and measuring the size of osteolysis than plain radiographs. Materials and Methods. Between 2000 and 2004, 233 primary THAs were performed using 28-mm
Introduction. Corrosion of the femoral head-trunnion junction in modular hip components has become a concern as the corrosion products may lead to adverse local tissue reactions. A simple way to avoid trunnion corrosion is to manufacture the femoral head with a non-metallic material, such as ceramics that are widely. An alternative solution may lie in advanced polymers like polyaryletherketones (PAEKs). These thermoplastics have high mechanical strength necessary for use as femoral heads in hip arthroplasty, but they must be tested to ensure that they do not adversely affect the wear of the ultrahigh molecular weight polyethylene (UHMWPE) liner counterface. Pin-on-disc (POD) wear testing has been extensively used to evaluate the wear properties of UHMWPE prior to more extensive and costly analysis with joint simulators. We hypothesized that the wear of crosslinked UHMWPE would not be adversely affected in POD tests when articulated against an advanced thermoplastic counterface. Methods. 0.1 wt.% VitE blended UHMWPE stock was e-beam irradiated to 100, 125, 140, 160, and 175 kGy and machined into cylindrical pins for testing. An additional group of 100 kGy e-beam irradiated and melted UHMWPE (with no vitamin E) was also machined and tested. Three different counterface materials were tested: (1)
Introduction. Manifestation of high interface stresses coupled with micromotion at the interface can render the taper lock joint in a modular hip replacement prosthesis at risk for failure. Bending can lead to crevice formation between the trunnion and the head and can potentially expose the interface to the biological fluids, generating interface corrosion. Additionally, development of high stresses can cause the material to yield, ultimately leading to irreversible damage to the implant. The objective of this study is to elucidate the mechanical response of taper junction in different material combination assemblies, under the maximum loads applied during everyday activities. Methods. Computer simulations were executed using a verified FE model. A stable hexahedral mesh (33648 elements) was generated for the trunnion (taper size: 12/14mm) and a tetrahedral mesh (51182 elements) for the head (CoCr, size: 32mm). An assembly load of 4000N was applied along the trunnion axis followed by the application of a load of 230–4300N at 25° and 10° angle to the trunnion axis in the frontal and sagittal planes. A linear static solution was set up using Siemens NX Nastran. Two material combinations were tested - cobalt-chrome head with a titanium alloy trunnion and cobalt chrome head with a
Concerns with long-term fixation of cemented all-poly sockets have led to the near-universal acceptance of cementless fixation of the acetabulum. The sockets most often utilised today are bone-ingrowth porous metal, made of titanium,