Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_12 | Pages 43 - 43
1 Oct 2019
Zochowski K Cheung J Argentieri EC Lin BQ Kaushik SS Burge AJ Koff MF Potter HG
Full Access

Introduction. MAVRIC-SL (MVSL), a multi-spectral magnetic resonance imaging (MRI) sequence [1] used for visualization of tissues near hip arthroplasty, can be accelerated by a short calibration scan [2], permitting an isotropic acquisition in a clinically feasible scan time. Decreasing the repetition time (TR) can further reduce scan time. This study evaluated the feasibility of acquiring isotropic MAVRIC-SL images (MVISO) and reduced TR isotropic MAVRIC-SL images (TRMVISO) of hip arthroplasties to elucidate the added clinical benefit of such acquisitions. Methods. Following IRB approval, MVSL, MVISO and TRMVISO images were acquired for 93 hip arthroplasties (84 subjects). Images were reviewed by 2 radiologists to evaluate clinical and image quality features. Comparison of features between scanning sequences was assessed by mixed-effects ordinal logistic regression and odds ratios (OR). Inter/intra-rater agreement was determined using Gwet's agreement coefficient (AC). Results. A calibration scan permitted acquisition of isotropic image sets in under 8 minutes. MVISO and TRMVISO had decreased blurring and improved visualization of the synovium and periprosthetic bone as compared to MVSL (p<0.001, Fig. 1). MVISO and TRMVISO improved image quality features over MVSL, OR range: 1 to 968. MVISO acquisitions also had better image quality, visualization of synovium and bone, lesion conspicuity, and decreased blurring than TRMVISO (p<0.032). Substantial to perfect (0.61 ≥ AC ≥ 1.0) agreement was found for 86% of factors evaluated by the 2 readers. Conclusions. The isotropic acquisitions displayed clinically relevant features better than MVSL, with notable improvement of blurring and overall image quality. The TRMVISO altered the image contrast and contributed to poorer visualization of synovial reactions as compared to MVISO. The study demonstrated that isotropic MAVRIC scans are clinically feasible, and that MVISO scans added clinical value through improved visualization by reducing blurring. For any tables or figures, please contact the authors directly


Bone & Joint Open
Vol. 5, Issue 1 | Pages 28 - 36
18 Jan 2024
Selmene MA Moreau PE Zaraa M Upex P Jouffroy P Riouallon G

Aims

Post-traumatic periprosthetic acetabular fractures are rare but serious. Few studies carried out on small cohorts have reported them in the literature. The aim of this work is to describe the specific characteristics of post-traumatic periprosthetic acetabular fractures, and the outcome of their surgical treatment in terms of function and complications.

Methods

Patients with this type of fracture were identified retrospectively over a period of six years (January 2016 to December 2021). The following data were collected: demographic characteristics, date of insertion of the prosthesis, details of the intervention, date of the trauma, characteristics of the fracture, and type of treatment. Functional results were assessed with the Harris Hip Score (HHS). Data concerning complications of treatment were collected.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 12 | Pages 1584 - 1590
1 Dec 2006
Hook S Moulder E Yates PJ Burston BJ Whitley E Bannister GC

We reviewed 142 consecutive primary total hip replacements implanted into 123 patients between 1988 and 1993 using the Exeter Universal femoral stem. A total of 74 patients (88 hips) had survived for ten years or more and were reviewed at a mean of 12.7 years (10 to 17). There was no loss to follow-up.

The rate of revision of the femoral component for aseptic loosening and osteolysis was 1.1% (1 stem), that for revision for any cause was 2.2% (2 stems), and for re-operation for any cause was 21.6% (19 hips). Re-operation was because of failure of the acetabular component in all but two hips.

All but one femoral component subsided within the cement mantle to a mean of 1.52 mm (0 to 8.3) at the final follow-up. One further stem had subsided excessively (8 mm) and had lucent lines at the cement-stem and cement-bone interfaces. This was classified as a radiological failure and is awaiting revision. One stem was revised for deep infection and one for excessive peri-articular osteolysis. Defects of the cement mantle (Barrack grade C and D) were found in 28% of stems (25 hips), associated with increased subsidence (p = 0.01), but were not associated with endosteal lysis or failure.

Peri-articular osteolysis was significantly related to the degree of polyethylene wear (p < 0.001), which was in turn associated with a younger age (p = 0.01) and male gender (p < 0.001).

The use of the Exeter metal-backed acetabular component was a notable failure with 12 of 32 hips (37.5%) revised for loosening. The Harris-Galante components failed with excessive wear, osteolysis and dislocation with 15% revised (5 of 33 hips). Only one of 23 hips with a cemented Elite component (4%) was revised for loosening and osteolysis.

Our findings show that the Exeter Universal stem implanted outside the originating centre has excellent medium-term results.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 1 | Pages 32 - 33
1 Jan 2007
Konangamparambath S Wilkinson JM Cleveland T Stockley I

Bleeding is a major complication of revision total hip replacement. We report a case where the inflated balloon of a urinary catheter was used to temporarily control intrapelvic bleeding from the superior gluteal artery, while definitive measures for endovascular embolisation were made.


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 1 | Pages 16 - 22
1 Jan 2008
Yates PJ Burston BJ Whitley E Bannister GC

We retrospectively reviewed 175 patients (191 hips) who had undergone primary cemented total hip replacement between November 1992 and November 1995 using a collarless polished double-tapered femoral component after a minimum of ten years (mean 11.08; 10 to 12.8). All stems were implanted using contemporary cementing techniques with a distal cement restrictor, pressurised lavage, retrograde cementing with a gun and proximal pressurisation. Clinical outcome was assessed using the Harris Hip score. Radiological analysis was performed on calibrated plain radiographs taken in two planes. Complete radiological data on 110 patients (120 hips) and clinical follow-up on all the surviving 111 patients (122 hips) was available. The fate of all the hips was known.

At final follow-up, the mean Harris Hip score was 86 (47 to 100), and 87 of 116 patients (75%) had good or excellent scores. Survival with revision of the stem for aseptic loosening as the endpoint was 100%; and survival with revision of the stem for any reason was 95.9% (95% confidence interval 87.8 to 96.8) at ten years. All the stems subsided vertically at the stem-cement interface in a predictable pattern, at an overall mean rate of 0.18 mm per year (0.02 to 2.16), but with a mean rate of 0.80 mm (0.02 to 2.5) during the first year. The mean total subsidence was 1.95 mm (0.21 to 24). Only three stems loosened at the cement-bone interface. There was excellent preservation of proximal femoral bone stock. There was a high incidence of Brooker III and IV heterotopic ossification affecting 25 patients (22%).

The collarless polished tapered stem has an excellent clinical and radiological outcome at a minimum of ten years’ follow-up. The pattern and magnitude of subsidence of the stem within the cement mantle occurred in a predictable pattern, consistent with the design philosophy.