Aims. Head-taper corrosion is a cause of failure in total hip arthroplasty (THA). Recent reports have described an increasing number of V40 taper failures with
Background.
Adverse reaction to wear and corrosion debris
is a cause for concern in total hip arthroplasty (THA). Modular junctions
are a potential source of such wear products and are associated
with secondary pseudotumour formation. We present a consecutive series of 17 patients treated at our
unit for this complication following metal-on-highly cross-linked
polyethylene (MoP) THA. We emphasise the risk of misdiagnosis as
infection, and present the aggregate laboratory results and pathological
findings in this series. The clinical presentation was pain, swelling or instability.
Solid, cystic and mixed soft-tissue lesions were noted on imaging
and confirmed intra-operatively. Corrosion at the head–neck junction
was noted in all cases. No bacteria were isolated on multiple pre-
and intra-operative samples yet the mean erythrocyte sedimentation
rate was 49 (9 to 100) and C-reactive protein 32 (0.6 to 106) and
stromal polymorphonuclear cell counts were noted in nine cases. Adverse soft–tissue reactions can occur in MoP THA owing to corrosion
products released from the head–neck junction. The diagnosis should
be carefully considered when investigating pain after THA. This
may avoid the misdiagnosis of periprosthetic infection with an unidentified
organism and mitigate the unnecessary management of these cases
with complete single- or two-stage exchange. Cite this article:
Fretting and corrosion at the modular head/neck junction, known
as trunnionosis, in total hip arthroplasty (THA) is a cause of adverse
reaction to metal debris (ARMD). We describe the outcome of revision
of metal-on-polyethylene (MoP) THA for ARMD due to trunnionosis
with emphasis on the risk of major complications. A total of 36 patients with a MoP THA who underwent revision
for ARMD due to trunnionosis were identified. Three were excluded
as their revision had been to another metal head. The remaining
33 were revised to a ceramic head with a titanium sleeve. We describe
the presentation, revision findings, and risk of complications in
these patients.Aims
Patients and Methods
Aims. The purpose of this study is to examine six types of bearing surfaces implanted at a single institution over three decades to determine whether the reasons for revision vary among the groups and how long it takes to identify differences in survival. Methods. We considered six cohorts that included a total of 1,707 primary hips done between 1982 and 2010. These included 223 conventional polyethylene sterilized with γ irradiation in air (CPE-GA), 114 conventional polyethylene sterilized with gas plasma (CPE-GP), 116 crosslinked polyethylene (XLPE), 1,083 metal-on-metal (MOM), 90 ceramic-on-ceramic (COC), and 81 surface arthroplasties (SAs). With the exception of the COC, all other groups used cobalt-chromium (CoCr) femoral heads. The mean follow-up was 10 (0.008 to 35) years. Descriptive statistics with revisions per 100 component years (re/100 yr) and survival analysis with revision for any reason as the endpoint were used to compare bearing surfaces. Results. XLPE liners demonstrated a lower cumulative incidence of revision at 15 years compared to the CPE-GA and CPE-GP groups owing to the absence of wear-related revisions (4% for XLPE vs 18%, p = 0.02, and 15%, p = 0.003, respectively). Revisions for
Aims. Advances in surgical technique and implant design may influence the incidence and mechanism of failure resulting in revision total hip arthroplasty (rTHA). The purpose of the current study was to characterize aetiologies requiring rTHA, and to determine whether temporal changes existed in these aetiologies over a ten-year period. Methods. All rTHAs performed at a single institution from 2009 to 2019 were identified. Demographic information and mode of implant failure was obtained for all patients. Data for rTHA were stratified into two time periods to assess for temporal changes: 2009 to 2013, and 2014 to 2019. Operative reports, radiological imaging, and current procedural terminology (CPT) codes were cross-checked to ensure the accurate classification of revision aetiology for each patient. Results. In all, 2,924 patients with a mean age of 64.6 years (17 to 96) were identified. There were 1,563 (53.5%) female patients, and the majority of patients were Caucasian (n = 2,362, 80.8%). The three most frequent rTHA aetiologies were infection (27.2%), aseptic loosening (25.2%), and wear (15.2%). The frequency of rTHA for
The purpose of this randomized controlled trial was to evaluate serum metal levels in patients undergoing total hip arthroplasty (THA) with a conventional compared to a modular dual-mobility bearing. Patients undergoing primary THA for osteoarthritis were randomized to receive either a modular dual-mobility or conventional polyethylene bearing. All patients received the same titanium acetabular and femoral component and a ceramic femoral head. Serum metal levels were drawn pre-operatively then annually for a minimum of two years postoperatively. An a priori power analysis determined that 40 patients (20 per cohort) were needed to identify a clinically relevant difference in serum cobalt of 0.35 ng/ml (ppb) at 90% power. Forty-six patients were randomized to a modular dual-mobility (n=25) or conventional bearing (n=21) with 40 at a minimum follow-up of two years. No differences in serum cobalt (mean 0.14 ppb [range, 0.075–0.29] vs. 0.20 ppb [range, 0.075–0.57], p=0.39) or chromium levels (mean 0.14 ppb [range, 0.05–0.50] vs. 0.12 ppb [range, 0.05–0.35], p=0.65) were identified between the modular dual-mobility and conventional cohorts, respectively. There was no statistically significant difference in serum Co or Cr at two years postoperatively in subjects implanted with a ceramic head and this particular dual mobility bearing in comparison to a ceramic head and a conventional acetabular component. While modest expected elevations in serum Co and Cr were observed in the dual mobility group, in no case did the Co level exceed the laboratory reference range nor the threshold of one part per billion that has been associated with
Aims.
Cup-cage constructs are one of several methods commonly used to treat severe acetabular bone loss during contemporary revision total hip arthroplasty. The purpose of this study was to provide a long-term results of the technique with emphasis on implant survivorship, radiographic results, and clinical outcomes for both full and half cup-cage reconstructions. We identified 57 patients treated with a cup-cage reconstruction for major acetabular bone loss between 2002–2012. All patients had Paprosky Type 2B through 3B bone loss, with 60% having an associated pelvic discontinuity. Thirty-one patients received a full cup-cage construct, and 26 a half cup-cage. Mean age at reconstruction was 66 years, 75% were female, and the mean BMI was 27 kg/m. 2. Mean follow-up was 10 years. The 10-year cumulative incidences of any revision were 14% and 12% for the full and half cup-cage construct groups, respectively. Of the 9 revisions, 3 were for dislocation, 2 for aseptic loosening and construct failure (both were pelvic discontinuities), 1 for
Aims. Hip resurfacing remains a potentially valuable surgical procedure for appropriately-selected patients with optimised implant choices. However, concern regarding high early failure rates continues to undermine confidence in use. A large contributor to failure is
Aims. Modular dual-mobility constructs reduce the risk of dislocation after revision total hip arthroplasty (THA). However, questions about metal ions from the cobalt-chromium (CoCr) liner persist, and are particularly germane to patients being revised for
Aims. We present a case series of ten metal-on-polyethylene total hip
arthroplasties (MoP THAs) with delayed dislocation associated with
unrecognised
Introduction. Modular neck (MN) components in total hip arthroplasty (THA) offer intraoperative flexibility, but
Dislocation remains among the most common complications
of, and reasons for, revision of both primary and revision total
hip replacements (THR). Hence, there is great interest in maximising
stability to prevent this complication. Head size has been recognised
to have a strong influence on the risk of dislocation post-operatively.
As femoral head size increases, stability is augmented, secondary
to an increase in impingement-free range of movement. Larger head
sizes also greatly increase the ‘jump distance’ required for the
head to dislocate in an appropriately positioned cup. Level-one
studies support the use of larger diameter heads as they decrease
the risk of dislocation following primary and revision THR. Highly cross-linked
polyethylene has allowed us to increase femoral head size, without
a marked increase in wear. However, the thin polyethylene liners
necessary to accommodate larger heads may increase the risk of liner
fracture and larger heads have also been implicated in causing soft-tissue
impingement resulting in groin pain. Larger diameter heads also
impart larger forces on the femoral trunnion, which may contribute
to corrosion, metal release, and
Introduction. MDM implants can enhance stability in total hip replacement (THR), but complications include malseated liners and corrosion between the cobalt-chrome liner and titanium acetabular shell increased systemic metal ion levels. The liner-shell junction has the potential for fretting corrosion, and the corrosion could be exacerbated in malseated liners. We determined the potential for fretting corrosion in malseated versus well-seated liners using a mechanical electrochemical corrosion chamber. Methods. Four pristine MDM liners and shells were tested. Two liners were well-seated into their shells; two were canted at 6°. The liner-shell couples were assembled with a 2kN force after wetting the surfaces to promote a crevice environment conducive to corrosion. Couples were fixed in an electrochemical chamber at 40° inclination/20° anteversion to the load axis. The chamber was filled with phosphate buffered saline and setup as a three-electrode configuration: the shell as the working, a saturated calomel electrode as the reference, and a carbon rod as the counter electrode. A potentiostat held the system at −50mV throughout testing. After equilibration, couples underwent cyclic loading of increasing magnitudes from 100 to 3400N at 3 Hz. Fretting current was measured throughout, and the onset load for fretting was determined from the increase in average current. Results. Well-seated liners showed lower fretting current values at all peak compressive loads greater than 800 N (p<0.05). Canted liners demonstrated a fretting onset load of 2400 N, and fretting currents at greater than 2400 N were larger than those at lower peak compressive loads (p<0.05). Conclusion. The clinical consequences of MDM liner malseating remain unknown, but our results demonstrate earlier fretting current onset at lower peak loads when compared to well-seated liners. The onset loads were consistent with physiologic loads for daily activities. Our findings are significant given the potential for metallosis and
Introduction.
Introduction. Modular dual-mobility constructs reduce the risk of dislocation after total hip arthroplasty (THA). However, questions about metal ions from the cobalt-chrome (CoCr) liner persist, and are particularly germane to patients being revised for
Surface replacement arthroplasty (SRA) has been proposed as a viable option for the treatment of osteoarthritis in young, active patients. Positive results of the Birmingham Hip Resurfacing (BHR) in select patient groups have been described in international series and registry data. We report 5–10 year U.S. follow-up for the BHR at our high volume institution. 314 patients (361 hips) between 2006–2011 underwent BHR at our institution and agreed to participate in research. Demographic features, modified Harris Hip Score, UCLA Activity Score, and satisfaction were recorded for patients with minimum 5-year follow-up (90%). Radiographs were evaluated for implant position and “at risk” signs. Complications, reoperations, and revisions were investigated. Mean modified Harris Hip and UCLA scores significantly improved postoperatively to scores of 89.96 and 7.90 (p < 0.001), respectively. Kaplan-Meier estimated survival for all-cause revision was 96.7% [95% CI 94.7 – 98.7%] at 5 years and 91.5% [95% CI 85.3 – 97.6%] at 10 years. Estimated survival for aseptic revision in males less than 60 years old with a primary diagnosis of osteoarthritis was 99.5% [95% CI 98.7 – 100%] at 5 years and 98.8% [95% CI 97.0 – 100%] at 10 years. 14 patients required revision, including 5 revisions for
Background. Hip resurfacing arthoplasty (HRA) showed promising early and mid-terms results at the beginning of the new millennium.
Recently, corrosion at the head-neck junction in metal-on-polyethylene bearing total hip arthroplasty (THA) has been recognized as a cause of