Ensuring correct rotation of the femoral component
is a challenging aspect of patellofemoral replacement surgery. Rotation
equal to the epicondylar axis or marginally more external rotation
is acceptable. Internal rotation is associated with poor outcomes.
This paper comprises two studies evaluating the use of the medial
malleolus as a landmark to guide rotation. We used 100 lower-leg anteroposterior radiographs to evaluate
the reliability of the medial malleolus as a landmark. Assessment
was made of the angle between the tibial shaft and a line from the
intramedullary rod entry site to the medial malleolus. The femoral
cut was made in ten cadaver knees using the inferior tip of the
medial malleolus as a landmark for rotation. Rotation of the cut
relative to the anatomical epicondylar axis was assessed using CT.
The study of radiographs found the position of the medial malleolus
relative to the tibial axis is consistent. Using the inferior tip
of the medial malleolus in the cadaver study produced a mean external
rotation of 1.6° (0.1° to 3.7°) from the anatomical epicondylar
axis. Using the inferior tip of the medial malleolus to guide the
femoral cutting jig avoids internal rotation and introduces an acceptable
amount of external rotation of the femoral component.
Mobile-bearing TKRs allow some axial rotation and may provide a more natural patellar movement. The aim was to compare patellar kinematics among the normal knee, fixed-bearing and mobile-bearing TKR. Optical computer navigation (Brainlab) was used to track the position of the femur, tibia and patella in 9 whole lower extremities (5 fresh cadavers) in the natural knee, in the same knee with the trial components of a posterior stabilised fixed-bearing TKR (FB) (Sigma PFC, Depuy) and a posterior stabilised mobile-bearing TKR (MB) (Sigma RP Stabilised). The patellae were not resurfaced. Values: mean+/−one standard deviation. Statistical analysis: two tailed paired Student's T-test.Introduction
Methods
The aim of our study was to assess lateral tracking of the patella with differing designs of Total Knee Arthroplasty (TKA) and compare to that of the native patella. A modified caliper was used to measure the width and position of the patella relative to the femur at different degrees of knee flexion. The relationship of the patella midpoint to that of the femur was subsequently assessed. Group 1 consisted of 25 native knees. Group 2 consisted of 25 patients with antero-posterior stabilised knee implant with a spherical medial condyle and a deep lateralised patellar groove, and Group 3 consisted of 25 patients with a conventional cam-and-post design with a midline patellar groove. The mean follow-up was 28 months.Aim
Method
Inadvertent soft tissue damage caused by the oscillating saw during total knee arthroplasty (TKA) occurs when the sawblade passes beyond the bony boundaries into the soft tissue. The primary objective of this study is to assess the risk of inadvertent soft tissue damage during jig-based TKA by evaluating the excursion of the oscillating saw past the bony boundaries. The second objective is the investigation of the relation between this excursion and the surgeon’s experience level. A conventional jig-based TKA procedure with medial parapatellar approach was performed on 12 cadaveric knees by three experienced surgeons and three residents. During the proximal tibial resection, the motion of the oscillating saw with respect to the tibia was recorded. The distance of the outer point of this cutting portion to the edge of the bone was defined as the excursion of the oscillating saw. The excursion of the sawblade was evaluated in six zones containing the following structures: medial collateral ligament (MCL), posteromedial corner (PMC), iliotibial band (ITB), lateral collateral ligament (LCL), popliteus tendon (PopT), and neurovascular bundle (NVB).Aims
Methods
Trochlear dysplasia is a developmental condition characterised by an abnormally flat or dome-shaped trochlea and is an important contributory factor to patellofemoral instability and recurrent dislocation. We studied prospectively a series of 54 consecutive patients (59 knees) with patellofemoral instability secondary to trochlear dysplasia, who were treated by a trochleoplasty by a single surgeon between June 2002 and June 2007. Pre- and post-operative scores were assessed by the patients and a satisfaction questionnaire was completed. Of the 54 patients (59 knees) in the series, 39 (44 knees) were female and 15 (15 knees) were male. Their mean age at surgery was 21 years and 6 months (14 years 4 months to 33 years 11 months). In 40 patients (42 knees) the mean follow-up was for 24 months (12 to 58). One patient was unable to attend for follow-up. An analysis of the results of those patients followed up for at least 12 months showed a statistically significant improvement in outcome (p <
0.001 for all scores). Overall, 50 patients (92.6%) were satisfied with the outcome of their procedure. The early results of trochleoplasty are encouraging in this challenging group of patients.