Advertisement for orthosearch.org.uk
Results 1 - 20 of 118
Results per page:
The Bone & Joint Journal
Vol. 102-B, Issue 6 Supple A | Pages 43 - 48
1 Jun 2020
D’Lima DP Huang P Suryanarayan P Rosen A D’Lima DD

Aims. The extensive variation in axial rotation of tibial components can lead to coronal plane malalignment. We analyzed the change in coronal alignment induced by tray malrotation. Methods. We constructed a computer model of knee arthroplasty and used a virtual cutting guide to cut the tibia at 90° to the coronal plane. The virtual guide was rotated axially (15° medial to 15° lateral) and with posterior slopes (0° to 7°). To assess the effect of axial malrotation, we measured the coronal plane alignment of a tibial tray that was axially rotated (25° internal to 15° external), as viewed on a standard anteroposterior (AP) radiograph. Results. Axial rotation of the cutting guide induced a varus-valgus malalignment up to 1.8° (for 15° of axial rotation combined with 7° of posterior slope). Axial malrotation of tibial tray induced a substantially higher risk of coronal plane malalignment ranging from 1.9° valgus with 15° external rotation, to over 3° varus with 25° of internal rotation. Coronal alignment of the tibial cut changed by 0.07° per degree of axial rotation and 0.22° per degree of posterior slope (linear regression, R. 2. > 0.99). Conclusion. While the effect of axial malalignment has been studied, the impact on coronal alignment is not known. Our results indicate that the direction of the cutting guide and malalignment in axial rotation alter coronal plane alignment and can increase the incidence of outliers. Cite this article: Bone Joint J 2020;102-B(6 Supple A):43–48


The Bone & Joint Journal
Vol. 103-B, Issue 12 | Pages 1791 - 1801
1 Dec 2021
Bhalekar RM Nargol ME Shyam N Nargol AVF Wells SR Collier R Pabbruwe M Joyce TJ Langton DJ

Aims. The aim of this study was to investigate whether wear and backside deformation of polyethylene (PE) tibial inserts may influence the cement cover of tibial trays of explanted total knee arthroplasties (TKAs). Methods. At our retrieval centre, we measured changes in the wear and deformation of PE inserts using coordinate measuring machines and light microscopy. The amount of cement cover on the backside of tibial trays was quantified as a percentage of the total surface. The study involved data from the explanted fixed-bearing components of four widely used contemporary designs of TKA (Attune, NexGen, Press Fit Condylar (PFC), and Triathlon), revised for any indication, and we compared them with components that used previous generations of PE. Regression modelling was used to identify variables related to the amount of cement cover on the retrieved trays. Results. A total of 114 explanted fixed-bearing TKAs were examined. This included 76 used with contemporary PE inserts which were compared with 15 used with older generation PEs. The Attune and NexGen (central locking) trays were found to have significantly less cement cover than Triathlon and PFC trays (peripheral locking group) (p = 0.001). The median planicity values of the PE inserts used with central locking trays were significantly greater than of those with peripheral locking inserts (205 vs 85 microns; p < 0.001). Attune and NexGen inserts had a characteristic pattern of backside deformation, with the outer edges of the PE deviating inferiorly, leaving the PE margins as the primary areas of articulation. Conclusion. Explanted TKAs with central locking mechanisms were significantly more likely to debond from the cement mantle. The PE inserts of these designs showed characteristic patterns of deformation, which appeared to relate to the manufacturing process and may be exacerbated in vivo. This pattern of deformation was associated with PE wear occurring at the outer edges of the articulation, potentially increasing the frictional torque generated at this interface. Cite this article: Bone Joint J 2021;103-B(12):1791–1801


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_13 | Pages 19 - 19
7 Aug 2023
Langton D Bhalekar R Wells S Nargol M Waller S Wildberg L Tilley S Nargol A
Full Access

Abstract. Introduction. At our national explant retrieval unit, we identified an unusual pattern of backside-deformation on polyethylene (PE) inserts of contemporary total-knee-replacements (TKRs). The PE backside's margins were inferiorly deformed in TKRs with central-locking trays. We reported that this backside-deformation appeared to be linked to tray debonding. Moreover, recent studies have shown high-rate of tray debonding in PS NexGen TKRs. Therefore, we hypothesised that backside deformation on PS inserts may be more than on CR inserts. Methodology. We used peer-reviewed techniques to analyse changes in the bearing (wear rate) and backside surfaces (deformation) of PE inserts using coordinate measuring machines [N=61 NexGen (CR-39 and PS-22) TKRs with non-augmented-trays]. Multiple regression was used to determine which variable had the greatest influence on backside-deformation. The amount of cement cover on trays was quantified as a %of the total surface using Image-J software. Results. There was no statistically significant difference (p=0.238) in median (IQR) wear rate of the CR PEs 18 (12–28) mm. 3. /year and PS PEs 14 (8–20) mm. 3. /year. The PE backside-deformation median (IQR) of PS [297(242–333) µm] was significantly higher (p=0.011), when compared with CR [241(161–259) µm]. Multiple regression modelling showed that duration in-vivo (p=0.037), central-clearance between insert and tray (p<0.001) and constraint (p=0.003) were significantly associated with PE backside-deformation. 27(69%) of CR and 20(91%) PS exhibited ≤10% of cement cover on tray. Conclusion. This explant study showed backside-deformation on PS inserts was more than on CR inserts. Therefore, indicating a high-rate of tibial tray debonding in PS compared to CR NexGen TKRs


The Journal of Bone & Joint Surgery British Volume
Vol. 78-B, Issue 2 | Pages 220 - 225
1 Mar 1996
Abernethy PJ Robinson CM Fowler RM

We reviewed 1567 elective knee replacements performed between 1980 and 1990, using either the Total Condylar prosthesis with an all-plastic tibial component, or the Kinematic prosthesis which has a metal tibial tray. The ten-year probability of survival was 92.1% for the Total Condylar design and 87.9% for the Kinematic. The difference was mainly due to 16 revisions required in the Kinematic series for fracture of the metal base-plate. This was the most common cause of aseptic failure in this group. These fractures were strongly associated with a preoperative varus deformity (hazard ratio (HR) 8.8) and there was a slightly increased risk in males (HR 1.9) and in osteoarthritic knees (HR 1.8). In the nine fractures which occurred within four years of primary implantation (group 1), failure to correct adequately a preoperative varus deformity and the use of a bone graft to correct such a deformity were both strongly associated with fracture (HR 13.9 and 15.8, respectively). In eight fractures which occurred more than five years after primary replacement (group 2) we could detect no significant risk factors. Early complications occurred in two patients after the 16 revision procedures for tray fracture. One had a deep infection and the other refracture of the tray


The Bone & Joint Journal
Vol. 95-B, Issue 3 | Pages 367 - 370
1 Mar 2013
Bini SA Chen Y Khatod M Paxton EW

We evaluated the impact of pre-coating the tibial component with polymethylmethacrylate (PMMA) on implant survival in a cohort of 16 548 primary NexGen total knee replacements (TKRs) in 14 113 patients. In 13 835 TKRs a pre-coated tray was used while in 2713 TKRs the non-pre-coated version of the same tray was used. All the TKRs were performed between 2001 and 2009 and were cemented. TKRs implanted with a pre-coated tibial component had a lower cumulative survival than those with a non-pre-coated tibial component (p = 0.01). After adjusting for diagnosis, age, gender, body mass index, American Society of Anesthesiologists grade, femoral coupling design, surgeon volume and hospital volume, pre-coating was an independent risk factor for all-cause aseptic revision (hazard ratio 2.75, p = 0.006). Revision for aseptic loosening was uncommon for both pre-coated and non-pre-coated trays (rates of 0.12% and 0%, respectively). Pre-coating with PMMA does not appear to be protective of revision for this tibial tray design at short-term follow-up. Cite this article: Bone Joint J 2013;95-B:367–70


Bone & Joint Research
Vol. 10, Issue 8 | Pages 467 - 473
2 Aug 2021
Rodríguez-Collell JR Mifsut D Ruiz-Sauri A Rodríguez-Pino L González-Soler EM Valverde-Navarro AA

Aims. The main objective of this study is to analyze the penetration of bone cement in four different full cementation techniques of the tibial tray. Methods. In order to determine the best tibial tray cementation technique, we applied cement to 40 cryopreserved donor tibiae by four different techniques: 1) double-layer cementation of the tibial component and tibial bone with bone restrictor; 2) metallic cementation of the tibial component without bone restrictor; 3) bone cementation of the tibia with bone restrictor; and 4) superficial bone cementation of the tibia and metallic keel cementation of the tibial component without bone restrictor. We performed CT exams of all 40 subjects, and measured cement layer thickness at both levels of the resected surface of the epiphysis and the endomedular metaphyseal level. Results. At the epiphyseal level, Technique 2 gave the greatest depth compared to the other investigated techniques. At the endomedular metaphyseal level, Technique 1 showed greater cement penetration than the other techniques. Conclusion. The best metaphyseal cementation technique of the tibial component is bone cementation with cement restrictor. Additionally, if full tibial component cementation is to be done, the cement volume used should be about 40 g of cement, and not the usual 20 g. Cite this article: Bone Joint Res 2021;10(8):467–473


The Bone & Joint Journal
Vol. 101-B, Issue 1_Supple_A | Pages 53 - 58
1 Jan 2019
Billi F Kavanaugh A Schmalzried H Schmalzried TP

Aims. Loosening of the tibial component after total knee arthroplasty (TKA) is a common indication for revision. Increasing the strength of the initial tibial implant/cement interface is desirable. There is little information about the surgical techniques that lead to the highest strength. We investigated the effects of eight variables on the strength of the initial tibial baseplate/cement interface. Materials and Methods. A total of 48 tibial trays were cemented into acrylic holders using cement from two manufacturers, at three different times (early, normal, and late) using two techniques: cementing the tibial plateau or the plateau and the keel; and involving two conditions of contamination with marrow fat (at the metal/cement and cement/cement interfaces). Push-out tests were performed with load continuously recorded. Results. Compared with normal conditions, early cementing increased the mean strength of the interface when using the two cements, Simplex and Palacos, by 48% and 72%, respectively. Late cementing reduced the strength by 47% and 73%, respectively. Cementing the keel increased the mean strength by 153% and 147%, respectively, for the two cements. Contamination of the metal/cement interface with fat reduced the mean strength by 99% and 94% for the two cements but adding cement to the underside of the tibial tray prior to insertion resulted in the mean strength being lowered by only 65% and 43%, respectively. Conclusion. In order to maximize the strength of the tibial tray/cement interface, cement should be applied to the component soon after mixing, contamination of the interface should be avoided, and the keel and the plateau should be cemented


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_13 | Pages 16 - 16
7 Aug 2023
Arthur L Amin A Rahman A Tu SJ Mellon S Murray D
Full Access

Abstract. Introduction. Anecdotal reports suggest some cementless Oxford Unicompartmental Knee Replacements (OUKRs) have painful early subsidence of the tibial component with valgus rotation and/or posterior tilting. The incidence of subsidence and its association with pain is poorly understood. This radiographic study aimed to evaluate the incidence of tibial subsidence and five-year patient reported outcome measures (PROMs) of a cementless OUKR cohort. Methodology. Cementless OUKRs from a high-volume centre with acceptable post-operative and five-year radiographs were included. Subsidence was determined by measuring the angle of the tibial tray/tibial axis angle and distance between the tibial tray axis and fibula head on anteroposterior and lateral radiographs using a custom MATLAB program. Analysis of 5-year PROMs assessed the relationship between subsidence and pain. Radiographs indicating tibial subsidence were validated by two observers. Results. Radiographs of 94 cementless OUKRs were analysed of which five tibial components had subsided (incidence=5%) with an average of 1.92° varus rotation and 2.97° posterior tilt. Subsidence appeared to occur within the first post-operative year with all tibial components fixed securely at 5 years. Two subsiders had moderate pain (ICOAP=20.5,15.9, AKSS Pain=20,45) and lower Oxford Knee Scores (OKS=26,31) compared to the cohort means (ICOAP=3.91, AKSS Pain=45.9, OKS = 43.6). These two subsiders had BMIs greater than 30. Conclusion. This study found 5% of cementless OUKRs in the cohort underwent tibial subsidence. Subsidence with moderate pain occurred in 2% of cases, and pain may be associated with obesity. A larger study is needed to study subsidence in greater detail


The Bone & Joint Journal
Vol. 103-B, Issue 11 | Pages 1695 - 1701
1 Nov 2021
Currier JH Currier BH Abdel MP Berry DJ Titus AJ Van Citters DW

Aims. Wear of the polyethylene (PE) tibial insert of total knee arthroplasty (TKA) increases the risk of revision surgery with a significant cost burden on the healthcare system. This study quantifies wear performance of tibial inserts in a large and diverse series of retrieved TKAs to evaluate the effect of factors related to the patient, knee design, and bearing material on tibial insert wear performance. Methods. An institutional review board-approved retrieval archive was surveyed for modular PE tibial inserts over a range of in vivo duration (mean 58 months (0 to 290)). Five knee designs, totalling 1,585 devices, were studied. Insert wear was estimated from measured thickness change using a previously published method. Linear regression statistical analyses were used to test association of 12 patient and implant design variables with calculated wear rate. Results. Five patient-specific variables and seven implant-specific variables were evaluated for significant association with lower insert wear rate. Six were significant when controlling for other factors: greater patient age, female sex, shorter duration in vivo, polished tray, highly cross-linked PE (HXLPE), and constrained knee design. Conclusion. This study confirmed that knee wear rate increased with duration in vivo. Older patients and females had significantly lower wear rates. Polished modular tibial tray surfaces, HXLPE, and constrained TKA designs were device design factors associated with significantly reduced wear rate. Cite this article: Bone Joint J 2021;103-B(11):1695–1701


Bone & Joint Open
Vol. 2, Issue 9 | Pages 737 - 744
1 Sep 2021
Øhrn F Lian ØB Tsukanaka M Röhrl SM

Aims. Medial pivot (MP) total knee arthroplasties (TKAs) were designed to mimic native knee kinematics with their deep medial congruent fitting of the tibia to the femur almost like a ball-on-socket, and a flat lateral part. GMK Sphere is a novel MP implant. Our primary aim was to study the migration pattern of the tibial tray of this TKA. Methods. A total of 31 patients were recruited to this single-group radiostereometric analysis (RSA) study and received a medial pivot GMK Sphere TKA. The distributions of male patients versus female patients and right versus left knees were 21:10 and 17:14, respectively. Mean BMI was 29 kg/m. 2. (95% confidence interval (CI) 27 to 30) and mean age at surgery was 63 years (95% CI 61 to 66). Maximum total point motions (MTPMs), medial, proximal, and anterior translations and transversal, internal, and varus rotations were calculated at three, 12, and 24 months. Patient-reported outcome measure data were also retrieved. Results. MTPMs at three, 12, and 24 months were 1.0 mm (95% CI 0.8 to 1.2), 1.3 mm (95% CI 0.9 to 1.7), and 1.4 mm (0.8 to 2.0), respectively. The Forgotten Joint Score was 79 (95% CI 39 to 95) and Knee Injury and Osteoarthritis Outcome Score obtained at two years was 94 (95% CI 81 to 100), 86 (95% CI 75 to 93), 94 (95% CI 88 to 100), 69 (95% CI 48 to 88), and 81 (95% CI59 to 100) for Pain, Symptoms, Activities of Daily Living, Sport & Recreation, and Quality of Life, respectively. Conclusion. In conclusion, we found that the mean increase in MTPM was lower than 0.2 mm between 12 and 24 months and thus apparently stable. Yet the GMK Sphere had higher migration at one and two years than anticipated. Based on current RSA data, we therefore cannot conclude on the long-term performance of the implant, pending further assessment. Cite this article: Bone Jt Open 2021;2(9):737–744


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 5 | Pages 656 - 663
1 May 2005
Toms AD McClelland D Chua L de Waal Malefijt M Verdonschot N Jones RS Kuiper J

Clinical experience of impaction bone grafting for revision knee arthroplasty is limited, with initial stability of the tibial tray emerging as a major concern. The length of the stem and its diameter have been altered to improve stability. Our aim was to investigate the effect of the type of stem, support of the rim and graft impaction on early stability of the tray. We developed a system for impaction grafting of trays which we used with morsellised bone in artificial tibiae. Trays with short, long thick or long thin stems were implanted, with or without support of the rim. They were cyclically loaded while measuring relative movement. Long-stemmed trays migrated 4.5 times less than short-stemmed trays, regardless of diameter. Those with support migrated 2.8 times less than those without. The migration of short-stemmed trays correlated inversely with the density of the impacted groups. That of impaction-grafted tibial trays was in the range reported for uncemented primary trays. Movements of short-stemmed trays without cortical support were largest and sensitive to the degree of compaction of the graft. If support of the rim was sufficient or a long stem was used, impacted morsellised bone graft achieved adequate initial stability


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_9 | Pages 69 - 69
1 Oct 2020
Lawrie CM Barrack RL Nunley RM
Full Access

Introduction. Bone mineral density (BMD) is correlated with component migration and aseptic loosening after total knee arthroplasty (TKA). Older implant designs have demonstrated BMD loss up to 23% in the first 6 months after TKA, and continued to BMD decline at an average of 5% per year for as long as 2 years after TKA. The impact of component design and fixation method on BMD loss after TKA in modern implant designs has not been fully elucidated. The purpose of this study is to determine the effect of tibial tray thickness and fixation method (cemented versus cementless) on BMD loss patterns of the proximal tibia in two different modern TKA implant systems. Methods. A prospective, nonrandomized, single center study of patients undergoing primary TKA by one of two surgeons was performed with four study cohorts: cemented DePuy Attune, cementless DePuy Attune, cemented Stryker Triathlon, cementless Stryker Triathlon. Target sample size was 80, with 20 per cohort based on adhoc power analysis. Exclusion criteria included: age over 75, BMI >40, inflammatory arthritis, previous knee surgery involving the femur, tibia or tibial bone, and diagnosis of osteopenia/osteoporosis. Implant fixation type was based on surgeon intraoperative assessment of patient bone quality. Demographic data was collected preoperatively. Dual Energy X-ray Absorptiometry (DEXA) Bone Density Monitoring was performed at 6 weeks and one year postoperatively. Bone mineral density was calculated from the DEXA scans for 4 zones for the tibia relative to the keel or central peg: anterior, posterior, medial and lateral. Results were reported as BMD at 1 year postoperatively as a percentage of BMD at 6 weeks postoperatively. Results. 81 knees with complete DEXA studies at 6 weeks and 1 year postoperatively were available for analysis. The mean (SD) age at the time of surgery was 65.4 years (6.1), with 39 men and 42 women and a mean (SD) BMI of 31.2 kg/m. 2. (4.7). The patients who received cementless implants were significantly younger with a mean (SD) age of 62.1 (6.0) than those who received cemented implants with a mean age of 69.1 (3.7) (p < 0.05). There was no difference in age between implant type (p > 0.05). There was no difference in BMI or gender distribution between implant types or fixation type (p > 0.05). There was no significant difference in mean BMD loss between 6 weeks and 1 year postoperatively between the cemented and cementless Triathlon in any zone, between the cemented Triathlon and Attune in any zone, the cementless Triathlon and Attune in any zone, and the cemented and cementless attune in the lateral, anterior and posterior proximal tibia (p > 0.05). There was significantly more BMD loss seen in the medial tibia in the cemented (BMD at 1 year 88% of BMD at 6 weeks) than cementless (BMD at 1 year 103% of BMD at 6 weeks) Attune (p = 0.043). Conclusion. BMD loss at 1 year postoperatively versus 6 weeks postoperatively after TKA with two modern implant designs is not significantly affected by tibial tray thickness. BMD loss was unaffected by fixation type for the Stryker Triathlon, but was significant less for the cementless DePuy Attune than the cemented version


The Bone & Joint Journal
Vol. 95-B, Issue 4 | Pages 472 - 477
1 Apr 2013
Liebs T Kloos S Herzberg W Rüther W Hassenpflug J

We investigated whether an asymmetric extension gap seen on routine post-operative radiographs after primary total knee replacement (TKR) is associated with pain at three, six, 12 and 24 months’ follow-up. On radiographs of 277 patients after primary TKR we measured the distance between the tibial tray and the femoral condyle on both the medial and lateral sides. A difference was defined as an asymmetric extension gap. We considered three groups (no asymmetric gap, medial-opening and lateral-opening gap) and calculated the associations with the Western Ontario and McMaster Universities osteoarthritis index pain scores over time. Those with an asymmetric extension gap of ≥ 1.5 mm had a significant association with pain scores at three months’ follow-up; patients with a medial-opening extension gap reported more pain and patients with a lateral-opening extension gap reported less pain (p = 0.036). This effect was still significant at six months (p = 0.044), but had lost significance by 12 months (p = 0.924). When adjusting for multiple cofounders the improvement in pain was more pronounced in patients with a lateral-opening extension gap than in those with a medial-opening extension gap at three (p = 0.037) and six months’ (p = 0.027) follow-up. Cite this article: Bone Joint J 2013;95-B:472–7


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 12 | Pages 1585 - 1593
1 Dec 2008
Henricson A Linder L Nilsson KG

We compared the performance of uncemented trabecular metal tibial components in total knee replacement with that of cemented tibial components in patients younger than 60 years over two years using radiostereophotogrammetric analysis (RSA). A total of 22 consecutive patients (mean age 53 years, 33 to 59, 26 knees) received an uncemented NexGen trabecular metal cruciate-retaining monobloc tibial component and 19 (mean 53 years, 44 to 59, 21 knees) a cemented NexGen Option cruciate-retaining modular tibial component. All the trabecular metal components migrated during the initial three months and then stabilised. The exception was external rotation, which did not stabilise until 12 months. Unlike conventional metal-backed implants which displayed a tilting migration comprising subsidence and lift-off from the tibial tray, most of the trabecular metal components showed subsidence only, probably due to the elasticity of the implant. This pattern of subsidence is regarded as being beneficial for uncemented fixation


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 8 | Pages 1016 - 1021
1 Aug 2006
Delport HP Banks SA De Schepper J Bellemans J

Mobile-bearing posterior-stabilised knee replacements have been developed as an alternative to the standard fixed- and mobile-bearing designs. However, little is known about the in vivo kinematics of this new group of implants. We investigated 31 patients who had undergone a total knee replacement with a similar prosthetic design but with three different options: fixed-bearing posterior cruciate ligament-retaining, fixed-bearing posterior-stabilised and mobile-bearing posterior-stabilised. To do this we used a three-dimensional to two-dimensional model registration technique. Both the fixed- and mobile-bearing posterior-stabilised configurations used the same femoral component. We found that fixed-bearing posterior stabilised and mobile-bearing posterior-stabilised knee replacements demonstrated similar kinematic patterns, with consistent femoral roll-back during flexion. Mobile-bearing posterior-stabilised knee replacements demonstrated greater and more natural internal rotation of the tibia during flexion than fixed-bearing posterior-stabilised designs. Such rotation occurred at the interface between the insert and tibial tray for mobile-bearing posterior-stabilised designs. However, for fixed-bearing posterior-stabilised designs, rotation occurred at the proximal surface of the bearing. Posterior cruciate ligament-retaining knee replacements demonstrated paradoxical sliding forward of the femur. We conclude that mobile-bearing posterior-stabilised knee replacements reproduce internal rotation of the tibia more closely during flexion than fixed-bearing posterior-stabilised designs. Furthermore, mobile-bearing posterior-stabilised knee replacements demonstrate a unidirectional movement which occurs at the upper and lower sides of the mobile insert. The femur moves in an anteroposterior direction on the upper surface of the insert, whereas the movement at the lower surface is pure rotation. Such unidirectional movement may lead to less wear when compared with the multidirectional movement seen in fixed-bearing posterior-stabilised knee replacements, and should be associated with more evenly applied cam-post stresses


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 7 | Pages 896 - 902
1 Jul 2009
Gulati A Chau R Pandit HG Gray H Price AJ Dodd CAF Murray DW

Narrow, well-defined radiolucent lines commonly observed at the bone-implant interface of unicompartmental knee replacement tibial components have been referred to as physiological radiolucencies. These should be distinguished from pathological radiolucencies, which are poorly defined, wide and progressive, and associated with loosening and infection. We studied the incidence and clinical significance of tibial radiolucent lines in 161 Oxford unicondylar knee replacements five years after surgery. All the radiographs were aligned with fluoroscopic control to obtain views parallel to the tibial tray to reveal the tibial bone-implant interface. We found that 49 knees (30%) had complete, 52 (32%) had partial and 60 (37%) had no radiolucent lines. There was no relationship between the incidence of radiolucent lines and patient factors such as gender, body mass index and activity, or operative factors including the status of the anterior cruciate ligament and residual varus deformity. Nor was any statistical relationship established between the presence of radiolucent lines and clinical outcome, particularly pain, assessed by the Oxford Knee score and the American Knee Society score. We conclude that radiolucent lines are common after Oxford unicompartmental knee replacement but that their aetiology remains unclear. Radiolucent lines were not a source of adverse symptoms or pain. Therefore, when attempting to identify a source of postoperative pain after Oxford unicompartmental knee replacement the presence of a physiological radiolucency should be ignored


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_12 | Pages 22 - 22
1 Oct 2018
Saffi M Young SW Spangehl MJ Clarke HD
Full Access

Introduction. Tibial component malrotation is associated with pain, stiffness and altered patellofemoral kinematics in total knee arthroplasty (TKA). However, accurately measuring tibial component rotation following TKA is difficult. Proposed protocols utilizing computed tomography (CT) are not well validated and can be time consuming. This study aimed to; 1) Validate and compare the reproducibility of the Berger (2D-CT) and Mayo (3D-CT) protocols; 2) Validate a simple, and potentially rapid screening measurement using an anatomical distance on 2D axial CT- the Centre of Tibial Tray to Tibial Tubercle (CTTT) distance. Methods. Rotational alignment of 70 TKA patients were evaluated by 3 independent observers using the Berger, and Mayo protocols, which have been previously described, and a new CTTT protocol (Figure 1). The inter and intra-rater interclass correlation coefficients (ICC's), mean difference between measurements and the mean measurement times were calculated. Linear regression analysis was performed to give a coefficient of determination (R2). Results. The intra-rater reliability for all 3 protocols was rated as “very good” (Mayo 0.96, Berger 0.85 and CTTT 0.85). The inter-rater reliability for the Mayo and the Berger method was rated as “very good” (0.87 and 0.83 respectively), the CTTT was rated as “good” (0.79). The Mayo method had a lower mean difference in intra-rater measurements than the Berger method (1.42° vs 2.60° p= <0.01). Comparing the CTTT to the Mayo method produced an R2 value of 0.73 indicating strong correlation. As a screening tool, 92% of patients with CTTT ≤ 6mm had < 9° of tibial component internal rotation (IR), and 93% of patients with a CTTT ≥ 10mm had ≥ 9° IR. The Mayo method takes 3 minutes, 29 seconds; Berger method: 2 minutes, 5 seconds; CTTT method: 39 seconds to perform. Conclusion. 3D CT is the gold standard for formally determining tibial component rotational alignment. The CTTT has the strongest correlation to the Mayo method, and is the least time consuming. The CTTT method can be used as a reliable, simple and rapid screening tool in daily clinical practice to assess tibial component rotational alignment following TKA, prior to formal measurement. For any figures or tables, please contact authors directly


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_12 | Pages 58 - 58
1 Oct 2018
Dalury D Chapman DM
Full Access

Introduction. The causes of revision total knee replacement are varied. There is a subset of these revision cases (poly wear and some cases of instability for example) where the ability to retain the metal femoral and tibial components and replacing just the polyethylene is an appealing option. We report on a series of RTKR where only the poly insert was replaced and the patients were followed for a minimum of ten years. Materials and Methods. Our study group consisted of 64 consecutive non infected RTKR patients who underwent a revision of the polyethylene alone between 1998 and 2006. All patients had been treated originally with the same cemented, CR, patella resurfaced primary TKR. Reasons for revision were 51 poly wear and/or osteolysis, 7 instability, and 6 other. The average time from the primary to the revision surgery was 9.1 years (range 2.2 to 16.1 years). All patients had an isolated poly liner change. No femurs or tibial trays were revised. Average age of the cohort at revision was 72.2 (range 48 to 88). Average BMI was 31.9 (range 23.6 to 43.9). There were 36 female patients. Of the group, 42 were ASA 1 or 2 and 22 were ASA 3. Pre-op alignment averaged 6.0 (range 2 varus to 8 valgus). Patients were followed for a minimum of 10 years (range 10 to 19 years). Results. At final follow up 13 patients had died, none were lost to follow up leaving 51 patients available for review. There were 7 reoperations in 6 patients (7 knees). Reasons for reoperation were: 4 aseptic tibial loosening at an average of 4.1 years (range 1.2 to 8.3) following first poly revision, two for polywear at an average 9 years post first poly revision, and one for deep infection (5.6 years post first poly revision). Knee society scores at 6 weeks post revision was 93.5 (range 38 to 100) and at final follow up for the non-revised knees was 91.6 (range 36 to 100). Conclusion. In appropriate settings, where the femoral and tibial components are satisfactorily aligned and well fixed and the soft tissues can be balanced, a poly change alone can provide a durable solution for the revision total knee patient. The ability to retain the metal components has advantages in terms of patient morbidity and cost


The Bone & Joint Journal
Vol. 105-B, Issue 6 | Pages 610 - 621
1 Jun 2023
Prodromidis AD Chloros GD Thivaios GC Sutton PM Pandit H Giannoudis PV Charalambous CP

Aims

Loosening of components after total knee arthroplasty (TKA) can be associated with the development of radiolucent lines (RLLs). The aim of this study was to assess the rate of formation of RLLs in the cemented original design of the ATTUNE TKA and their relationship to loosening.

Methods

A systematic search was undertaken using the Cochrane methodology in three online databases: MEDLINE, Embase, and CINAHL. Studies were screened against predetermined criteria, and data were extracted. Available National Joint Registries in the Network of Orthopaedic Registries of Europe were also screened. A random effects model meta-analysis was undertaken.


The Bone & Joint Journal
Vol. 106-B, Issue 9 | Pages 916 - 923
1 Sep 2024
Fricka KB Wilson EJ Strait AV Ho H Hopper, Jr RH Hamilton WG Sershon RA

Aims

The optimal bearing surface design for medial unicompartmental knee arthroplasty (UKA) remains controversial. The aim of this study was to compare outcomes of fixed-bearing (FB) and mobile-bearing (MB) UKAs from a single high-volume institution.

Methods

Prospectively collected data were reviewed for all primary cemented medial UKAs performed by seven surgeons from January 2006 to December 2022. A total of 2,999 UKAs were identified, including 2,315 FB and 684 MB cases. The primary outcome measure was implant survival. Secondary outcomes included 90-day and cumulative complications, reoperations, component revisions, conversion arthroplasties, range of motion, and patient-reported outcome measures. Overall mean age at surgery was 65.7 years (32.9 to 94.3), 53.1% (1,593/2,999) of UKAs were implanted in female patients, and demographics between groups were similar (p > 0.05). The mean follow-up for all UKAs was 3.7 years (0.0 to 15.6).