Purpose: To analyse preliminary outcomes of a new synthetic ligament used for reconstructing coraco-clavicular ligament in acromio-clavicular joint disruption. Methods: Patients who underwent acromio-clavicular joint stabilisation using Nottingham
The LockDown device (previously called Surgilig)
is a braided polyester mesh which is mostly used to reconstruct the
dislocated acromioclavicular joint. More than 11 000 have been implanted
worldwide. Little is known about the tissue reaction to the device
nor to its wear products when implanted in an extra-articular site
in humans. This is of importance as an adverse immunological reaction
could result in osteolysis or damage to the local tissues, thereby affecting
the longevity of the implant. We analysed the histology of five LockDown implants retrieved
from five patients over the last seven years by one of the senior
authors. Routine analysis was carried out in all five cases and
immunohistochemistry in one. The LockDown device acts as a scaffold for connective tissue
which forms an investing fibrous pseudoligament. The immunological
response at the histological level seems favourable with a limited
histiocytic and giant cell response to micron-sized wear particles.
The connective tissue envelope around the implant is less organised
than a native ligament. Cite this article:
Background. Nonsurgical treatment of Acromioclavicular joint dislocations is well established. Most patients treated conservatively do well, however, some of them develop persistent symptoms. We have used two different surgical reconstruction techniques for Chronic ACJ dislocation stabilization. The study evaluates the effectiveness of a braided polyester prosthetic ligament and modified Weaver-Dunn reconstruction methods. Methods. 55 patients (mean age 42) with Chronic Acromioclavicular joint dislocation were included in this study. They were treated either by a modified Weaver-Dunn method or a braided polyester prosthetic ligament. Patients were assessed using Oxford shoulder score preoperatively and a minimum of 12 months postoperatively. Results. 31 patients (mean age 40, M=24, F=7) were treated by Modified WD method and 24 patients (mean age 44, M=18, F=6) by
Chronic instability of the acromioclavicular joint is relatively common and normally occurs following a fall onto the point of the shoulder. Reconstruction of the joint (Weaver-Dunn procedure) is often required in service personnel, and numerous methods of fixation have been used, including vicryl tape, PDS loops and the use of a hook plate. Many of these operative methods require a second operation to remove the plates and/or screws, and are associated with a failure rate of up to 30%. The ‘Surgilig’ was designed as a method of revision for failed Weaver-Dunn procedures. However this study evaluates its use in the primary operation. We prospectively followed up the Modified Weaver Dunn procedures using
Osteolysis has been reported following ACJ reconstruction with a synthetic graft. We present the first study into its prevalence and pattern, and its effect on patient outcome. Patients who underwent treatment of an unstable ACJ injury using the
In the UK we have many surgeon inventors – surgeons who innovate and create new ways of doing things, who invent operations, who design new instruments to facilitate surgery or design new implants for using in patients. However truly successful surgeon inventors are a rare breed and they need to develop additional knowledge and skills during their career in order to push forward their devices and innovations. This article reviews my own experiences as a surgeon inventor and the highs and lows over the whole of my surgical career.