Aims. This prospective study reports longitudinal, within-patient, patient-reported outcome measures (PROMs) over a 15-year period following cemented
Total knee arthroplasty (TKA) with a highly congruent condylar-stabilized (CS) articulation may be advantageous due to increased stability versus cruciate-retaining (CR) designs, while mitigating the limitations of a posterior-stabilized construct. The aim was to assess ten-year implant survival and functional outcomes of a cemented single-radius TKA with a CS insert, performed without posterior cruciate ligament sacrifice. This retrospective cohort study included consecutive patients undergoing TKA at a specialist centre in the UK between November 2010 and December 2012. Data were collected using a bespoke electronic database and cross-referenced with national arthroplasty audit data, with variables including: preoperative characteristics, intraoperative factors, complications, and mortality status. Patient-reported outcome measures (PROMs) were collected by a specialist research team at ten years post-surgery. There were 536 TKAs, of which 308/536 (57.5%) were in female patients. The mean age was 69.0 years (95% CI 45.0 to 88.0), the mean BMI was 32.2 kg/m2 (95% CI 18.9 to 50.2), and 387/536 (72.2%) survived to ten years. There were four revisions (0.7%): two deep infections (requiring debridement and implant retention), one aseptic loosening, and one haemosiderosis.Aims
Methods
The objective of this study was to compare the early migration
characteristics and functional outcome of the Triathlon cemented
knee prosthesis with its predecessor, the Duracon cemented knee
prosthesis (both Stryker). A total 60 patients were prospectively randomised and tibial
component migration was measured by radiostereometric analysis (RSA)
at three months, one year and two years; clinical outcome was measured
by the American Knee Society score and the Knee Osteoarthritis and
Injury Outcome Score.Objectives
Methods
Aims. The aims of this study were to investigate the ability to kneel after total knee arthroplasty (TKA) without patellar resurfacing, and its effect on patient-reported outcome measures (PROMs). Secondary aims included identifying which kneeling positions were most important to patients, and the influence of radiological parameters on the ability to kneel before and after TKA. Methods. This prospective longitudinal study involved 209 patients who underwent
The aim of this study was to compare the migration of the femoral component, five years postoperatively, between patients with a highly cross-linked polyethylene (HXLPE) insert and those with a conventional polyethylene (PE) insert in an uncemented Triathlon fixed insert cruciate-retaining total knee arthroplasty (TKA). Secondary aims included clinical outcomes and patient-reported outcome measures (PROMs). We have previously reported the migration and outcome of the tibial components in these patients. A double-blinded randomized controlled trial was conducted including 96 TKAs. The migration of the femoral component was measured with radiostereometry (RSA) at three and six months and one, two, and five years postoperatively. PROMs were collected preoperatively and at all periods of follow-up.Aims
Methods
The primary objective of this study was to compare the five-year tibial component migration and wear between highly crosslinked polyethylene (HXLPE) inserts and conventional polyethylene (PE) inserts of the uncemented Triathlon fixed insert cruciate-retaining total knee arthroplasty (TKA). Secondary objectives included clinical outcomes and patient-reported outcome measures (PROMs). A double-blinded, randomized study was conducted including 96 TKAs. Tibial component migration and insert wear were measured with radiostereometric analysis (RSA) at three, six, 12, 24, and 60 months postoperatively. PROMS were collected preoperatively and at all follow-up timepoints.Aims
Methods
The aims were to assess whether joint-specific outcome after total knee arthroplasty (TKA) was influenced by implant design over a 12-year follow-up period, and whether patient-related factors were associated with loss to follow-up and mortality risk. Long-term follow-up of a randomized controlled trial was undertaken. A total of 212 patients were allocated a Triathlon or a Kinemax TKA. Patients were assessed preoperatively, and one, three, eight, and 12 years postoperatively using the Oxford Knee Score (OKS). Reasons for patient lost to follow-up, mortality, and revision were recorded.Aims
Methods
Mid-level constraint designs for total knee arthroplasty (TKA) are intended to reduce coronal plane laxity. Our aims were to compare kinematics and ligament forces of the Zimmer Biomet Persona posterior-stabilized (PS) and mid-level designs in the coronal, sagittal, and axial planes under loads simulating clinical exams of the knee in a cadaver model. We performed TKA on eight cadaveric knees and loaded them using a robotic manipulator. We tested both PS and mid-level designs under loads simulating clinical exams via applied varus and valgus moments, internal-external (IE) rotation moments, and anteroposterior forces at 0°, 30°, and 90° of flexion. We measured the resulting tibiofemoral angulations and translations. We also quantified the forces carried by the medial and lateral collateral ligaments (MCL/LCL) via serial sectioning of these structures and use of the principle of superposition.Aims
Methods
The COVID-19 pandemic has led to unprecedented times worldwide. From lockdowns to masks now being part of our everyday routine, to the halting of elective surgeries, the virus has touched everyone and every part of our personal and professional lives. Perhaps, now more than ever, our ability to adapt, change and persevere is critical to our survival. This year's closed meeting of The Knee Society demonstrated exactly those characteristics. When it became evident that an in-person meeting would not be feasible, The Knee Society leadership, under the direction of President John Callaghan, MD and Program Chair Craig Della Valle, MD created a unique and engaging meeting held on September 10–12, 2020. Special recognition should be given to Olga Foley and Cynthia Garcia at The Knee Society for their flexibility and creativeness in putting together a world-class flawless virtual program. The Bone & Joint Journal is very pleased to partner with The Knee Society to once again publish the proceedings of the closed meeting of the Knee Society. The Knee Society is a United States based society of highly selected members who have shown leadership in education and research in knee surgery. It invites up to 15% international members; this includes some of the key opinion leaders in knee surgery from outside the USA. Each year, the top research papers from The Knee Society meeting will be published and made available to the wider orthopaedic community in The Bone & Joint Journal. The first such proceedings were published in BJJ in 2019. International dissemination should help to fulfil the mission and vision of the Knee Society of advancing the care of patients with knee disorders through leadership, education and research. The quality of dissemination that The Bone & Joint Journal provides should enhance the profile of this work and allow a larger body of surgeons, associated healthcare professionals and patients to benefit from the expertise of the members of The Knee Society. The meeting is one of the highlights of the annual academic calendar for knee surgeons. With nearly every member in attendance virtually throughout the 3 days, the top research papers from the membership were presented and discussed in a virtual format that allowed for lively interaction and discussion. There are 75 abstracts presented. More selective proceedings with full papers will be available after a robust peer review process in 2021, both online and in The Bone & Joint Journal. The meeting commenced with the first group of scientific papers focused on Periprosthetic Joint Infection. Dr Berry and colleagues from the Mayo Clinic further help to clarify the issue of serology and aspirate results to diagnose TKA PJI in the acute postoperative setting. 177 TKA's had an aspiration within 12 weeks and 22 were proven to have PJI. Their results demonstrated that acute PJI after TKA should be suspected within 6 weeks if CRP is ≥81 mg/L, synovial WBCs are ≥8500 cells/μL, and/or synovial neutrophils≥86%. Between 6– 12 weeks, concerning thresholds include a CRP ≥ 32 mg/L, synovial WBC ≥7450, and synovial neutrophils ≥ 84%. While historically the results of a DAIR procedure for PJI have been variable, Tom Fehring's study showed promise with the local delivery of vancomycin through the Intraosseous route improved early results. New member Simon Young contrasted the efficacy of the DAIR procedure when comparing early infections to late acute hematogenous PJI. DAIR failed in 63% of late hematogenous PJIs (implant age>1 year) compared to 36% of early (<1year) PJIs. Dr Masri demonstrated in a small group of patients that those with well-functioning articulating spacers can retain their spacers for over 12 months with no difference in infection from those that had a formal two stage exchange. The mental toll of PJI was demonstrated in a longitudinal study by Doug Dennis, where patient being treated with 2 stage exchange had 4x higher rates of depression compared to patient undergoing aseptic revision. The second session focused on both postoperative issues with regards to anticoagulation and manipulation. Steven Haas demonstrated high complication rates with utilization of anticoagulation for treatment of postoperative pulmonary embolism with modern therapeutic anticoagulation (warfarin, enoxaparin, Xa inhibitors) with the Xa inhibitors demonstrating lower complication rates. Two papers focused on the topic of manipulation. Mark Pagnano presented data on timing of manipulation under anesthesia up to even past 12 months. While gains were modest, a subset of patients did achieve substantial gains in ROM > 20degrees even after 3 months post op. Dr Westrich's study demonstrated no difference in MUA outcomes with either IV sedation or neuraxial anesthesia although the length of stay was shorter in the IV sedation group. Several studies in Session II focused on kinematics and femoral component position. Dr Li's in vivo kinematic study during weightbearing flexion and gait demonstrated that several knees rotated with a lateral pivot motion and not all knees can be described with a single motion character. Dr Mayman and his group utilized a computational knee model to demonstrate that additional distal femoral resection results in increasing levels of mid -flexion instability and cautioned against the use of additional bony resection as the first line for flexion contractures. Using computer navigation, Dr Huddleston's study nicely outlined the variability in femoral component rotation to achieve a rectangular flexion gap utilizing a gap balanced method. The third session opened the meeting on Friday morning. The focus was on unicompartmental knee arthroplasty and the increasing utilization of robotic assisted total knee arthroplasty. David Murray showed using registry data that for patient with higher comorbidities (ASA >3), UKA was safer and more cost effective than TKA while Dr Della Valle's group demonstrated overall lower average healthcare costs in UKA patients compared to TKA in the first 10 years after surgery. Dr Geller assessed UKA survivorship among 3 international registries. While survivorship varied by nation and designs, certain designs consistently had better overall performance. Dr Nunley and his group showed robotic navigation UKA significantly reduced outliers in alignment and overhang compared to manual UKA. Dr Catani's data demonstrated that full thickness cartilage loss should still be considered a requirement for UKA success even with robotic assistance. Despite a high dislocation rate of 4%, Mr Dodd demonstrated high survivorship for lateral UKA despite historical contraindications. The growing evidence for robotics TKA was demonstrated in two studies. Professor Haddad showed less soft tissue injury, reduced bone trauma and improved accuracy or rTKA compared to manual TKA while Dr Gustke single surgeon study showed his rTKA had improved forgotten joint scores and less ligament releasing required for balancing. Despite these finding, Dr Lee's study demonstrated that a robotic TKA could not guarantee excellent pain relief and other factors such a patient expectations and psychological factors play a role. Our fourth session was devoted to machine learning and smart tools and modeling. Dr Meneghini used machine learning algorithms to identify optimal alignment outcomes that correlated with patient outcomes. Several parameters such as native tibial slope, femoral sagittal position and coronal limb alignment correlated with outcomes. Along the same lines, Bozic and coauthors demonstrated that using AI algorithms incorporated with PROM's improved levels of shared decision making and patient satisfaction. Dr Lombardi demonstrated that a mobile patient engagement platform that provided smart phone-based exercise and education was comparable to traditional methods. Dr Mahfouz demonstrated the accuracy of using ultrasound to produce 3D models of the bone compared to conventional CT based strategies and Dr Mahoney showed the valued of a preop 3D model in reproducing more normal knee kinematics. The last two talks of the session focused on some of the positives of the COVID-19 pandemic, namely the embracing of telemedicine by patients and surgeons as demonstrated by Dr Slover and the increasing and far reaching educational opportunities made available to residents and fellows during the pandemic. Session five focused on risk stratification and optimization prior to TKA. Dr O'Connor demonstrated that that the implementation of an optimization program preoperatively reduced length of stay and ED visits, and Charles Nelson's study showed that risk stratification tool can lower complication rates in obese patients undergoing TKA comparable to those that are nonobese. Dr Markel's study demonstrated that those who have preoperative depression and anxiety are at higher risk of complications and readmissions after surgery and these issues should be addressed preoperatively. Interestingly, a study by Dr Callaghan demonstrated that care improvement pathways have not lowered the gap in complications for morbidly obese patients undergoing TKA, Dr Barsoum argued that the overall complication rates were low and this patient cohort had significant gains in PROMS after TKA that would not be experienced if arbitrary cutoff for limited surgery were established. The final session on Friday, Session six, had several well done and interesting studies. There continues to be mounting evidence that liposomal bupivacaine has little effect on managing post-operative pain to warrant its increased use. Bill Macaulay and colleagues showed no change in pain scores, opioid consumption and functional scores when liposomal bupivacaine was discontinued at a large academic medical center. Dr Bugbee importantly demonstrated that a supervised ambulation program reduced falls in the early postoperative period. Several paper on healthcare economics were presented. Rich Iorio showed that stratifying complexity of total joint cases between hospitals with a system can be efficient and cost savings while Dr Jiranek demonstrated in his study that complex TKAs can be identified preoperatively and are associated with prolonged operative time and cost of care and consideration should be given in future reimbursement models to a complexity modifier. Dr Springer, in their evaluation of Medicare bundled payment models, demonstrated that providers and hospitals in historical bundled models that became efficient were penalized in the new model, forcing many groups to drop out and return to a fee for service model. Ron Delanois important work showed that social determinants can have a major negative impact on outcomes following TKA. Our final day on Saturday opened with Session seven, and several interesting paper on metal ions/debris in TKA. Dr Whitesides simulator study showed the absence of scratches and material loss in a ceramic TKA compared with Co-Cr TKA and suggested an advantage to this material in patients with metal sensitivity. Conversely, in a histological study of failed TKA, perivascular lymphocytic infiltration was not associated with worse clinical outcomes or differences in revision in a series of 617 aseptic revisions, 19% of which had PVLI found on histology. The Mayo group and Dr Trousdale however, noted that serum metal ion levels can be helpful in identifying implant failure in a group of revision TKAs, especially those with metallic junctions. Dr Dalury demonstrated nicely that use of maximally conforming inserts did not have a negative effect on implant loosening in a series of 76 revision TKA's at an average follow up of 7 years, while Kevin Garvin and his group showed no difference in end of stem pain between cemented and cementless stems in revision TKA. The final two studies in the session by Bolognesi and Peters respectively showed that metaphyseal cones continue to demonstrate excelled survivorship in rTKA setting despite extensive bone loss. Session eight was highlighted by a large series of revision reported by new member Dr Schwarzkopf, who showed that revision TKA done by high volume surgeons demonstrated better outcomes and lower revision rates compared to surgeon who did less than 18 rTKA's per year. Dr Maniar importantly showed that preoperatively, patients with high activity level and low pain and indicated by a high preop forgotten joint score did poorly following TKA while David Ayers nicely demonstrated that KOOS scores that assess specific postoperative outcomes can predict patient dissatisfaction after TKA. The final paper in this session by Max Courtney showed that the majority of surgical cancellations are due to medical issues, yet a minority of these undergo any intervention specifically for that condition, but they resulted in a delay of 5 months. The first two studies of Session nine focused on polyethylene thickness. Dr Backstein demonstrated no difference in KSS scores, change in ROM and aseptic revision rates based on polyethylene thickness in a series of 195 TKA's. An interesting lab study by Dr Tim Wright showed a surprising consistency in liner thickness choice among varying levels of surgeon experience that did not correlate with applied forces or gap stability estimates. Two studies looked specifically at the issue of tibial loosening and implant design. Nam and colleagues were not able to demonstrate concerning findings for increasing tibial loosening in a tibial baseplate with a shortened tibial keel at short term follow up, while Lachiewicz demonstrated a 19% revision or revision pending rate in 223 cemented fixed bearing ATTUNE TKA at a mean of 30 months. Our final session of the meeting, began with encouraging news, that despite only currently capturing about 40% of TJA's done in the US, the American Joint Replacement Registry data is representative of data in other representative US databases. An interesting study presented by Robert Barrack looked at bone remodeling in the proximal tibia after cemented and cementless TKA of two different designs. No significant difference was noted among the groups with the exception of the cemented thicker cobalt chrome tray which demonstrated significantly more bone mineral density loss. Along the same lines, a study out of Dr Bostrom's lab demonstrated treatment of a murine tibial model with iPTH prevents fibrous tissue formation and enhances bone formation in cementless implants. New Member Jamie Howard showed no difference in implant migration and kinematics of a
Abnormal sagittal kinematics after total knee replacement (TKR) can adversely affect functional outcome. Two important determinants of knee kinematics are component geometry and the presence or absence of a posterior-stabilising mechanism (cam-post). We investigated the influence of these variables by comparing the kinematics of a TKR with a polyradial femur with a
The results of kinematic total knee arthroplasty (KTKA) have been reported in terms of limb and component alignment parameters but not in terms of gap laxities and differentials. In kinematic alignment (KA), balance should reflect the asymmetrical balance of the normal knee, not the classic rectangular flexion and extension gaps sought with gap-balanced mechanical axis total knee arthroplasty (MATKA). This paper aims to address the following questions: 1) what factors determine coronal joint congruence as measured on standing radiographs?; 2) is flexion gap asymmetry produced with KA?; 3) does lateral flexion gap laxity affect outcomes?; 4) is lateral flexion gap laxity associated with lateral extension gap laxity?; and 5) can consistent ligament balance be produced without releases? A total of 192 KTKAs completed by a single surgeon using a computer-assisted technique were followed for a mean of 3.5 years (2 to 5). There were 116 male patients (60%) and 76 female patients (40%) with a mean age of 65 years (48 to 88). Outcome measures included intraoperative gap laxity measurements and component positions, as well as joint angles from postoperative three-foot standing radiographs. Patient-reported outcome measures (PROMs) were analyzed in terms of alignment and balance: EuroQol (EQ)-5D visual analogue scale (VAS), Knee Injury and Osteoarthritis Outcome Score (KOOS), KOOS Joint Replacement (JR), and Oxford Knee Score (OKS).Aims
Patients and Methods
Posterior condylar offset (PCO) and posterior tibial slope (PTS) are critical factors in total knee arthroplasty (TKA). A computational simulation was performed to evaluate the biomechanical effect of PCO and PTS on cruciate retaining TKA. We generated a subject-specific computational model followed by the development of ± 1 mm, ± 2 mm and ± 3 mm PCO models in the posterior direction, and -3°, 0°, 3° and 6° PTS models with each of the PCO models. Using a validated finite element (FE) model, we investigated the influence of the changes in PCO and PTS on the contact stress in the patellar button and the forces on the posterior cruciate ligament (PCL), patellar tendon and quadriceps muscles under the deep knee-bend loading conditions.Objectives
Methods
Little biomechanical information is available about kinematically aligned (KA) total knee arthroplasty (TKA). The purpose of this study was to simulate the kinematics and kinetics after KA TKA and mechanically aligned (MA) TKA with four different limb alignments. Bone models were constructed from one volunteer (normal) and three patients with three different knee deformities (slight, moderate and severe varus). A dynamic musculoskeletal modelling system was used to analyse the kinematics and the tibiofemoral contact force. The contact stress on the tibial insert, and the stress to the resection surface and medial tibial cortex were examined by using finite element analysis.Objectives
Materials and Methods
The aim of this study was to investigate differences in pain,
range of movement function and satisfaction at three months and
one year after total knee arthroplasty (TKA) in patients with an
oblique pattern of kinematic graph of the knee and those with a
varus pattern. A total of 91 patients who underwent TKA were included in this
retrospective study. Patients (59 women and 32 men with mean age
of 68.7 years; 38.6 to 88.4) were grouped according to kinematic
graphs which were generated during navigated TKA and the outcomes
between the groups were compared.Aims
Patients and Methods
The Advance Medial-Pivot total knee arthroplasty (TKA) was designed
to reflect contemporary data regarding the kinematics of the knee.
We wished to examine the long-term results obtained with this prosthesis
by extending a previous evaluation. We retrospectively evaluated prospectively collected data from
225 consecutive patients (41 men and 184 women; mean age at surgery
71 years, 52 to 84) who underwent 284 TKAs with a mean follow-up
of 13.4 years (11 to 15). Implant failure, complication rate, clinical (both
subjective and objective) and radiological outcome were assessed.
Pre- and post-operative clinical and radiographic data were available
at regular intervals for all patients. A total of ten patients (4.4%;
ten TKAs) were lost to follow-up.Aims
Patients and Methods
Total knee arthroplasty (TKA) is an established
and successful procedure. However, the design of prostheses continues
to be modified in an attempt to optimise the functional outcome
of the patient. The aim of this study was to determine if patient outcome after
TKA was influenced by the design of the prosthesis used. A total of 212 patients (mean age 69; 43 to 92; 131 female (62%),
81 male (32%)) were enrolled in a single centre double-blind trial
and randomised to receive either a Kinemax (group 1) or a Triathlon
(group 2) TKA. Patients were assessed pre-operatively, at six weeks, six months,
one year and three years after surgery. The outcome assessments
used were the Oxford Knee Score; range of movement; pain numerical
rating scales; lower limb power output; timed functional assessment
battery and a satisfaction survey. Data were assessed incorporating
change over all assessment time points, using repeated measures
analysis of variance longitudinal mixed models. Implant group 2
showed a significantly greater range of movement (p = 0.009), greater
lower limb power output (p = 0.026) and reduced report of ‘worst
daily pain’ (p = 0.003) over the three years of follow-up. Differences
in Oxford Knee Score (p = 0.09), report of ‘average daily pain’
(p = 0.57) and timed functional performance tasks (p = 0.23) did
not reach statistical significance. Satisfaction with outcome was
significantly better in group 2 (p = 0.001). These results suggest that patient outcome after TKA can be influenced
by the prosthesis used. Cite this article:
We performed a prospective, randomised trial of 44 patients to compare the functional outcomes of a posterior-cruciate-ligament-retaining and posterior-cruciate-ligament-substituting total knee arthroplasty, and to gain a better understanding of the At follow-up at five years, no statistically significant differences were found in the clinical outcome measurements for either design. The prevalence of radiolucent lines and the survivorship were the same. In a subgroup of 15 knees, additional image-intensifier analysis in the horizontal and sagittal planes was performed during step-up and lunge activity. Our analysis revealed striking differences. Lunge activity showed a mean posterior displacement of both medial and lateral tibiofemoral contact areas (roll-back) which was greater and more consistent in the cruciate-substituting than in the cruciate-retaining group (medial p <
0.0001, lateral p = 0.011). The amount of posterior displacement could predict the maximum flexion which could be achieved (p = 0.018). Forward displacement of the tibiofemoral contact area in flexion during stair activity was seen more in the cruciate-retaining than in the cruciate-substituting group. This was attributed mainly to insufficiency of the posterior cruciate ligament and partially to that of the anterior cruciate ligament. We concluded that, despite similar clinical outcomes, there are significant kinematic differences between cruciate-retaining and cruciate-substituting arthroplasties.
We investigated the changes in surface roughness of retrieved femoral components in 18 men and four women at revision knee surgery. The mean age at revision was 68.4 years and the mean period of implantation was for 55.6 months. Eighteen implants were retrieved for aseptic loosening and four for infection. The surface changes in the articulating areas were inspected visually and the roughness (Ra) analysed with a profilometer. Parallel scratching and burnishing were the two main forms of damage. The mean Ra measurements in the articulating areas showed no statistically significant difference when compared with those in a control area on either side of the patellar groove at the apex of the femoral flange. This suggests that it is not essential to revise a well-fixed and correctly aligned femoral component where the polished surface has become burnished or bears fine parallel scratches, if the revision is conducted solely for failure of the tibial component.