Thermal sensors have been used in bracing research as self-reported diaries are inaccurate. Little is known about new low-profile sensors, optimal location within a brace, locational thermal micro-climate and effect of brace lining. Our objective is to Determine an optimal temperature threshold for sensor-measured and true wear time agreement. Identify optimal sensor location. Assess all factors to determine the best sensor option for the Bracing AdoleScent Idiopathic Scoliosis (BASIS) multicentre RCT. Seven Orthotimer and five iButton (DS1925L) sensors were synchronised to record temperature at five-minute intervals. Three healthy participants donned a rigid spinal brace, embedded with both sensors across four anatomical locations (abdomen/axilla/lateral-gluteal/sacral). Universal-coordinated-time wear protocols were performed in/out-doors. Intraclass correlation coefficient (ICC) assessed sensor-measured and true wear time agreement at thresholds 15–36oC. Optimal thresholds, determined by largest ICC estimate: Orthotimer: Abdomen=26oC, axilla=27oC, lateral-gluteal=24.5oC, sacral=22.5oC. iButton: Abdomen=26oC, axilla=27oC, lateral-gluteal=23.5oC, sacral=23.5oC. Warm-up time and error at optimal thresholds increased for moulded sensors covered with 6mm lining. Location: anterior abdominal wall. Excellent reliability and higher optimal thresholds, less likely to be exceeded by ambient temperature; not a pressure area. Sensor: iButton, longer battery life and larger memory than Orthotimer; allows recording at 10 min intervals for life of brace. Orthotimer only able to record every 30 mins, increasing error between true and measured wear time; Orthotimer needs 6-monthly data download. Threshold: 26oC is optimal threshold to balance warm-up and cool-down times for accurately measuring wear time. Sensor should not be covered by lining foam as this significantly prolongs warm-up time.