Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_8 | Pages 110 - 110
11 Apr 2023
Lee K Lin J Lynch J Smith P
Full Access

Variations in pelvic anatomy are a major risk factor for misplaced percutaneous sacroiliac screws used to treat unstable posterior pelvic ring injuries. A better understanding of pelvic morphology improves preoperative planning and therefore minimises the risk of malpositioned screws, neurological or vascular injuries, failed fixation or malreduction. Hence a classification system which identifies the clinically important anatomical variations of the sacrum would improve communication among pelvic surgeons and inform treatment strategy. 300 Pelvic CT scans from skeletally mature trauma patients that did not have pre-existing posterior pelvic pathology were identified. Axial and coronal transosseous corridor widths at both S1 and S2 were recorded. Additionally, the S1 lateral mass angle were also calculated. Pelvises were classified based upon the sacroiliac joint (SIJ) height using the midpoint of the anterior cortex of L5 as a reference point. Four distinct types could be identified:. Type-A – SIJ height is above the midpoint of the anterior cortex of the L5 vertebra. Type-B – SIJ height is between the midpoint and the lowest point of the anterior cortex of the L5 vertebra. Type-C – SIJ height is below the lowest point of the anterior cortex of the L5 vertebra. Type-D – a subgroup for those with a lumbosacral transitional vertebra, in particular a sacralised L5. Differences in transosseous corridor widths and lateral mass angles between classification types were assessed using two-way ANOVAs. Type-B was the most common pelvic type followed by Type-A, Type-C, and Type-D. Significant differences in the axial and coronal corridors was observed for all pelvic types at each level. Lateral mass angles increased from Types-A to C, but were smaller in Type-D. This classification system offers a guide to surgeons navigating variable pelvic anatomy and understanding how it is associated with the differences in transosseous sacral corridors. It can assist surgeons’ preoperative planning of screw position, choice of fixation or the need for technological assistance


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_2 | Pages 107 - 107
2 Jan 2024
Pastor T Zderic I Berk T Souleiman F Vögelin E Beeres F Gueorguiev B Pastor T
Full Access

Recently, a new generation of superior clavicle plates was developed featuring the variable-angle locking technology for enhanced screw positioning and optimized plate-to-bone fit design. On the other hand, mini-fragment plates used in dual plating mode have demonstrated promising clinical results. However, these two bone-implant constructs have not been investigated biomechanically in a human cadaveric model. Therefore, the aim of the current study was to compare the biomechanical competence of single superior plating using the new generation plate versus dual plating with low-profile mini-fragment plates. Sixteen paired human cadaveric clavicles were assigned pairwise to two groups for instrumentation with either a 2.7 mm Variable Angle Locking Compression Plate placed superiorly (Group 1), or with one 2.5 mm anterior plate combined with one 2.0 mm superior matrix mandible plate (Group 2). An unstable clavicle shaft fracture AO/OTA15.2C was simulated by means of a 5 mm osteotomy gap. All specimens were cyclically tested to failure under craniocaudal cantilever bending, superimposed with bidirectional torsion around the shaft axis and monitored via motion tracking. Initial stiffness was significantly higher in Group 2 (9.28±4.40 N/mm) compared to Group 1 (3.68±1.08 N/mm), p=0.003. The amplitudes of interfragmentary motions in terms of craniocaudal and shear displacement, fracture gap opening and torsion were significantly bigger over the course of 12500 cycles in Group 1 compared to Group 2; p≤0.038. Cycles to 2 mm shear displacement were significantly lower in Group 1 (22792±4346) compared to Group 2 (27437±1877), p=0.047. From a biomechanical perspective, low-profile 2.5/2.0 dual plates demonstrated significantly higher initial stiffness, less interfragmentary movements, and higher resistance to failure compared to 2.7 single superior variable-angle locking plates and can therefore be considered as a useful alternative for diaphyseal clavicle fracture fixation especially in unstable fracture configurations


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_13 | Pages 119 - 119
1 Nov 2021
Facchini A Troiano E Saviori M Meglio MD Ghezzi R Mondanelli N Giannotti S
Full Access

Introduction and Objective. The aim of this study was to evaluate whether CT-based pre-operative planning, integrated with intra-operative navigation could improve glenoid baseplate fixation and positioning by increasing screw length, reducing number of screws required to obtain fixation and increasing the use of augmented baseplate to gain the desired positioning. Reverse total shoulder arthroplasty (RSA) successfully restores shoulder function in different conditions. Glenoid baseplate fixation and positioning seem to be the most important factors influencing RSA survival. When scapular anatomy is distorted (primitive or secondary), optimal baseplate positioning and secure screw purchase can be challenging. Materials and Methods. Twenty patients who underwent navigated RSA (oct 2018 and feb 2019) were compared retrospectively with twenty patients operated on with a conventional technique. All the procedures were performed by the same surgeon, using the same implant in cases of eccentric osteoarthritis or complete cuff tear. Exclusion criteria were: other diagnosis as proximal humeral fractures, post-traumatic OA previously treated operatively with hardware retention, revision shoulder arthroplasty. Results. The NAV procedure required mean 11 (range 7–16) minutes more to performed than the conventional procedure. Mean screw length was significantly longer in the navigation group (35.5+4.4 mm vs 29.9+3.6 mm; p . .001). Significant higher rate of optimal fixation using 2 screws only (17 vs 3 cases, p . .019) and higher rate of augmented baseplate usage (13 vs 4 cases, p . .009) was also present in the navigation group. Signficant difference there is all in function outcomes, DASH score is 15.7 vs 29.4 and constant scale 78.1 vs 69.8. Conclusions. The glenoid component positioning in RSA is crucial to prevent failure, loosening and biomechanical mismatch, coverage by the baseplate of the glenoid surface, version, inclination and offset are all essential for implant survival. This study showed how useful 3D CT-based planning helps in identifying the best position of the metaglena and the usefulness of receiving directly in the operation theater real-time feedback on the change in position. This study shows promising results, suggesting that improved baseplate and screw positioning and fixation is possible when computer-assisted implantation is used in RSA comparing to a conventional procedure


Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_11 | Pages 354 - 354
1 Jul 2014
Eraly K Stoffelen D Van Geel N Demol J Debeer P
Full Access

Summary Statement. In this study, excellent positioning of custom-made glenoid components was achieved using patient-specific guides. Achieving the preoperatively planned orientation of the component improved significantly and more screws were located inside the scapular bone compared to implantations without such guide. Introduction. Today's techniques for total or reverse shoulder arthroplasty are limited when dealing with severe glenoid defects. The available procedures, for instance the use of bone allografts in combination with available standard implants, are technically difficult and tend to give uncertain outcomes (Hill et al. 2001; Elhassan et al. 2008; Sears et al. 2012). A durable fixation between bone and implant with optimal fit and implant positioning needs to be achieved. Custom-made defect-filling glenoid components are a new treatment option for severe glenoid defects. Despite that the patient-specific implants are uniquely designed to fit the patient's bone, it can be difficult to achieve the preoperatively planned position of the component, resulting in less optimal screw fixation. We hypothesised that the use of a patient-specific guide would improve implant and screw positioning. The aim of this study was to evaluate the added value of a newly developed patient-specific guide for implant and screw positioning, by comparing glenoid implantations with and without such guide. Patients & Methods. Large glenoid defects, representative for the defects encountered in clinical practice, were created in ten cadaveric shoulders. A CT scan of each cadaver was taken to evaluate the defects and to generate three-dimensional models of the scapular bones. Based on these models, custom glenoid components were designed. Furthermore, a newly developed custom guide was designed for five randomly selected shoulders. New CT scans were taken after implantation to generate 3D models of the bone and the implanted component and screws. This enabled to compare the experimentally achieved and preoperatively planned reconstruction. The location and orientation of the glenoid component and screw positioning were determined and differences with the optimal preoperative planning were calculated. Results. An excellent component positioning (difference in location: 1.4±0, 7mm; difference in orientation: 2, 5±1, 2°) was achieved when using the guide compared to implantations without guidance (respectively 1, 7±0, 5mm; 5, 1±2, 3°). The guide improved component orientation significantly (P<0.1). After using the guide, all screws were positioned inside the scapular bone whereas 25% of the screws placed without guidance were positioned outside the scapular bone. Discussion/Conclusion. In this study, excellent positioning of custom-made glenoid components was achieved using patient-specific guides. Achieving the preoperatively planned orientation of the component improved significantly and more screws were located inside the scapular bone compared to implantations without such guide


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XI | Pages 7 - 7
1 Apr 2012
Mullen M Pillai A Fogg Q Kumar CS
Full Access

The extended lateral approach offers a safe surgical approach in the fixation of calcaneal fractures. Lateral plating of the calcaneum could put structures on the medial side at risk. The aim was to identify structures at risk on the medial side of the calcaneum from wires, drills or screws passed from lateral to medial. Ten embalmed cadaveric feet were dissected. A standard extended lateral approach was performed. The DePuy perimeter plate was first applied and 2mm K-wires were drilled through each of the holes. The medial side was now examined to determine the structures at risk through each hole. The process was repeated with the Stryker plate. The calcaneum was divided into 6 zones, by two vertical lines, from the margins of the posterior facet and a transverse line along the axis of the bone through the highest point of the peroneal tubercle. The DePuy and the Stryker plates have 12 screw positions, 5 of which are common. With both systems, screw positions in zone 1 risk injury to the medial plantar nerve and zone 3 the lateral plantar nerve. A screw through zone 2 compromises the medial plantar in both. Screws through zone 4 risk the lateral plantar nerve with the DePuy plate. Screws through zone 5 of the DePuy plate risk the medial calcaneal nerve. Zone 5 of the Stryker plate and Zone 6 of both are safe. There is significant risk to medial structures from laterally placed wires, drills or screws. Subtalar screws have the highest risk and have to be carefully measured and placed. The Stryker plating system is relatively safer than the DePuy perimeter plate with three safe zones out of six


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_4 | Pages 19 - 19
1 Mar 2021
Mischler D Schader JF Windolf M Varga P
Full Access

To date, the fixation of proximal humeral fractures with angular stable locking plates is still insufficient with mechanical failure rates of 18% to 35%. The PHILOS plate (DePuy Synthes, Switzerland) is one of the most used implants. However, this plate has not been demonstrated to be optimal; the closely symmetric plate design and the largely heterogeneous bone mineral density (BMD) distribution of the humeral head suggest that the primary implant stability may be improved by optimizing the screw orientations. Finite element (FE) analysis allows testing of various implant configurations repeatedly to find the optimal design. The aim of this study was to evaluate whether computational optimization of the orientation of the PHILOS plate locking screws using a validated FE methodology can improve the predicted primary implant stability. The FE models of nineteen low-density (humeral head BMD range: 73.5 – 139.5 mg/cm3) left proximal humeri of 10 male and 9 female elderly donors (mean ± SD age: 83 ± 8.8 years) were created from high-resolution peripheral computer tomography images (XtremeCT, Scanco Medical, Switzerland), using a previously developed and validated computational osteosynthesis framework. To simulate an unstable mal-reduced 3-part fracture (AO/OTA 11-B3.2), the samples were virtually osteotomized and fixed with the PHILOS plate, using six proximal screws (rows A, B and E) according to the surgical guide. Three physiological loading modes with forces taken from musculoskeletal models (AnyBody, AnyBody Technology A/S, Denmark) were applied. The FE analyses were performed with Abaqus/Standard (Simulia, USA). The average principal compressive strain was evaluated in cylindrical bone regions around the screw tips; since this parameter was shown to be correlated with the experimental number of cycles to screw cut-out failure (R2 = 0.90). In a parametric analysis, the orientation of each of the six proximal screws was varied by steps of 5 in a 5×5 grid, while keeping the screw head positions constant. Unfeasible configurations were discarded. 5280 simulations were performed by repeating the procedure for each sample and loading case. The best screw configuration was defined as the one achieving the largest overall reduction in peri-screw bone strain in comparison with the PHILOS plate. With the final optimized configuration, the angle of each screw could be improved, exhibiting significantly smaller average bone strain around the screw tips (range of reduction: 0.4% – 38.3%, mean ± SD: 18.49% ± 9.56%). The used simulation approach may help to improve the fixation of complex proximal humerus fractures, especially for the target populations of patients at high risk of failure


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_3 | Pages 80 - 80
1 Apr 2018
Sugand K van Duren B Wescott R Carrington R Hart A
Full Access

Background. Hip fractures cause significant morbidity and mortality, affecting 70,000 people in the UK each year. The dynamic hip screw (DHS) is used for the osteosynthesis of extracapsular neck of femur fractures, a procedure that requires complex psychomotor skills to achieve optimal lag screw positioning. The tip-apex distance (TAD) is a measure of the position of the lag screw from the apex of the femoral head, and is the most comprehensive predictor of cut-out (failure of the DHS construct). To develop these skills, trainees need exposure to the procedure, however with the European Working Time Directive, this is becoming harder to achieve. Simulation can be used as an adjunct to theatre learning, however it is limited. FluoroSim is a digital fluoroscopy simulator that can be used in conjunction with workshop bones to simulate the first step of the DHS procedure (guide-wire insertion) using image guidance. This study assessed the construct validity of FluoroSim. The null hypothesis stated that there would be no difference in the objective metrics recorded from FluoroSim between users with different exposure to the DHS procedure. Methods. This multicentre study recruited twenty-six orthopaedic doctors. They were categorised into three groups based on the number of DHS procedures they had completed as the primary surgeon (novice <10, intermediate 10≤x<40 and experienced ≥40). Twenty-six participants completed a single DHS guide-wire attempt into a workshop bone using FluoroSim. The TAD, procedural time, number of radiographs, number of guide-wire retires and cut-out rate (COR) were recorded for each attempt. Results. A significant construct effect was seen for TAD and COR between novice and other users (p < 0.05). The intermediate and experienced users were not significantly different for these metrics. For all other metrics, experienced users had the highest score, contrary to expectation. Conclusion. FluoroSim was able to separate novice users from other cohorts for the two clinically significant outcome metrics. We can therefore partially reject the null hypothesis as construct validity was present for TAD and COR. We have demonstrated that FluoroSim has the potential to be a useful adjunct when learning the psychomotor skills needed for the DHS procedure away from theatre


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXVIII | Pages 45 - 45
1 Jun 2012
Russell D Behbahani M Alakandy L
Full Access

Pedicle screw fixation is an effective and reliable method for achieving stabilization in lumbar degenerative disease. The procedure carries a risk of violating the spinal and neural canal which can lead to nerve injury. This audit examines the accuracy of screw placement using intra-operative image guidance. Retrospective audit of patients undergoing lumbar pedicle screw fixation using image guidance systems over an 18-month period. Case records were reviewed to identify complications related to screw placement and post-operative CT scans reviewed to study the accuracy of screw position. Of the 98 pedicle screws placed in 25 patients, pedicle violation occurred in 4 screw placements (4.1%). Medial or inferior breach of the pedicle cortex was seen in 2 screws (2%). Nerve root injury as a consequence of this violation was seen in one patient resulting in irreversible partial nerve root dysfunction. Mean set up time for the guidance system was 42 minutes. The mean operative time was 192 minutes. Violation of either the medial or inferior pedicle cortex during placement of fixation screws is a rare, but potentially serious complication bearing lasting consequences. Image-guided placement can be helpful and possibly improve accuracy; particularly in patients with distorted spinal anatomy


Bone & Joint Research
Vol. 6, Issue 8 | Pages 481 - 488
1 Aug 2017
Caruso G Bonomo M Valpiani G Salvatori G Gildone A Lorusso V Massari L

Objectives

Intramedullary fixation is considered the most stable treatment for pertrochanteric fractures of the proximal femur and cut-out is one of the most frequent mechanical complications. In order to determine the role of clinical variables and radiological parameters in predicting the risk of this complication, we analysed the data pertaining to a group of patients recruited over the course of six years.

Methods

A total of 571 patients were included in this study, which analysed the incidence of cut-out in relation to several clinical variables: age; gender; the AO Foundation and Orthopaedic Trauma Association classification system (AO/OTA); type of nail; cervical-diaphyseal angle; surgical wait times; anti-osteoporotic medication; complete post-operative weight bearing; and radiological parameters (namely the lag-screw position with respect to the femoral head, the Cleveland system, the tip-apex distance (TAD), and the calcar-referenced tip-apex distance (CalTAD)).


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 9 | Pages 1257 - 1263
1 Sep 2006
Richter M Droste P Goesling T Zech S Krettek C

Different calcaneal plates with locked screws were compared in an experimental model of a calcaneal fracture. Four plate models were tested, three with uniaxially-locked screws (Synthes, Newdeal, Darco), and one with polyaxially-locked screws (90° ± 15°) (Rimbus). Synthetic calcanei were osteotomised to create a fracture model and then fixed with the plates and screws. Seven specimens for each plate model were subjected to cyclic loading (preload 20 N, 1000 cycles at 800 N, 0.75 mm/s), and load to failure (0.75 mm/s).

During cyclic loading, the plate with polyaxially-locked screws (Rimbus) showed significantly lower displacement in the primary loading direction than the plates with uniaxially-locked screws (mean values of maximum displacement during cyclic loading: Rimbus, 3.13 mm (sd 0.68); Synthes, 3.46 mm (sd 1.25); Darco, 4.48 mm (sd 3.17); Newdeal, 5.02 mm (sd 3.79); one-way analysis of variance, p < 0.001).

The increased stability of a plate with polyaxially-locked screws demonstrated during cyclic loading compared with plates with uniaxially-locked screws may be beneficial for clinical use.