The Gamma nail is frequently used in unstable
peri-trochanteric hip fractures. We hypothesised that mechanical failure
of the Gamma nail was associated with inadequate proximal three-point
fixation. We identified a consecutive series of 299 Gamma nails
implanted in 299 patients over a five-year period, 223 of whom fulfilled
our inclusion criteria for investigation. The series included 61
men and 162 women with a mean age of 81 years (20 to 101). Their fractures
were classified according to the Modified Evans’ classification
and the quality of fracture reduction was graded. The technical
adequacy of three points of proximal fixation was recorded from
intra-operative fluoroscopic images, and technical inadequacy for
each point was defined. All patients were followed to final follow-up
and mechanical failures were identified. A multivariate statistical
analysis was performed, adjusting for confounders. A total of 16
failures (7.2%) were identified. The position of the lag screw relative
to the lateral cortex was the most important point of proximal fixation,
and when inadequate the failure rate was 25.8% (eight of 31: odds
ratio 7.5 (95% confidence interval 2.5 to 22.7), p <
0.001). Mechanical failure of the Gamma nail in peri-trochanteric femoral
fractures is rare (<
1%) when three-point proximal fixation is
achieved. However, when proximal fixation is inadequate, failure
rates increase. The strongest predictor of failure is positioning
the lateral end of the lag screw short of the lateral cortex. Adherence
to simple technical points minimises the risk of fixation failure
in this vulnerable patient group. Cite this article:
As the incidence of total hip arthroplasty (THA) rises, an increasing prevalence of peri-prosthetic femur fractures has been reported. This is likely due to the growing population with arthroplasties, increasing patient survival and a more active life-style following arthroplasty. It is the 3rd most common reason for THA reoperation (9.5%) and 5th most common reason for revision (5% with fracture risk after primary THA reported at 0.4%-1.1% and after revision at 2.1%-4%). High quality radiographs are usually sufficient to classify the fracture and plan treatment. Important issues in treatment include stem fixation status and fracture location relative to the stem. Additional comorbidities will also influence treatment choices, of which the most critical is the presence of infection and the quality of bone stock. The most commonly studied, and reported classification system is the Vancouver. Type A are
Patients with above knee amputation (AKA) often experience poor socket fit exacerbated by minor weight changes, sweating, and skin problems. A transcutaneous, press-fit distal femoral intra-medullary device was designed in 1999, such that the distal external aspect serves as a hard point for AKA prosthesis attachment. The implant is placed in a retrograde fashion, followed 6-8 weeks later by stomatisation and connected via direct extention to an AKA prosthesis. Thirty-seven patients with 39 limbs (30 males, 7 females) underwent two-stage prosthesis implantation with the Endo-Exo Femurprosthesis (EEFP) in Germany between 1999 and January 2008. Their indications for surgery were persistent AKA prosthesis socket difficulties with absence of major comorbid physical or mental illnesses. The patients were followed clinically and radiographically for a minimum of 2 years. Assessments included patient satisfaction ratings, functional surveys, pain scores, and oxygen consumption. Fifty-four percent of patients needed at least one revision (20/37); 80% were minor (16/20) and 20% were major (4/20). Most of the minor revisions were due to soft tissue stomal irritation and occurred prior to a design change in the prosthesis. 2 patients sustained traumatic,
In a retrospective study, we reviewed 45