The aim of this study was to present the first retrieval analysis findings of PRECICE STRYDE intermedullary nails removed from patients, providing useful information in the post-market surveillance of these recently introduced devices. We collected ten nails removed from six patients, together with patient clinical data and plain radiograph imaging. We performed macro- and microscopic analysis of all surfaces and graded the presence of corrosion using validated semiquantitative scoring methods. We determined the elemental composition of surface debris using energy dispersive x-ray spectroscopy (EDS) and used metrology analysis to characterize the surface adjacent to the extendable junctions.Aims
Methods
Introduction. Motorised intramedullary lengthening nails are considered more expensive than external fixators for limb lengthening. This research aims to compare the cost of femoral lengthening in children using the
Aims. Patients undergoing femoral lengthening by external fixation
tolerate treatment less well when compared to tibial lengthening.
Lengthening of the femur with an intramedullary device may have
advantages. Patients and Methods. We reviewed all cases of simple femoral lengthening performed
at our unit from 2009 to 2014. Cases of nonunions, concurrent deformities,
congenital limb deficiencies and lengthening with an unstable hip
were excluded, leaving 33 cases (in 22 patients; 11 patients had
bilateral procedures) for review. Healing index, implant tolerance
and complications were compared. Results. In 20 cases (15 patients) the
Background:. External fixators are not as well tolerated around the femur when compared to the tibia. Lengthening with an intramedullary device is therefore attractive. Method:. We reviewed all cases of femoral lengthening performed at our unit from 2007 to 2014. Cases of non-unions, concurrent deformities, congenital limb deficiencies and lengthening with an unstable hip were excluded. This left 33 cases for review. Healing index, implant tolerance and complications were compared. Results and Discussion:. In 20 cases the
Background:. Leg length discrepancy related to absence of the femoral head with proximal migration of the femur presents a treatment dilemma. Late sequelae of neonatal hip sepsis and chronic hip dislocation due to developmental dysplasia are the most common causes. Traditional teaching dictates that the hip is stabilised prior to limb lengthening. Reconstructive options alter the shape of the proximal femur which complicates future surgery. Methods:. We retrospectively reviewed 3 cases of femoral lengthening with an ‘unstable’ hip without prior stabilization. The aetiology was neonatal hip sepsis in 2 cases and chronic hip dislocation due to developmental dysplasia in 1 case. Lengthening was performed with the use of a retrograde
Distraction osteogenesis with intramedullary lengthening devices has undergone rapid development in the past decade with implant enhancement. In this first single-centre matched-pair analysis we focus on the comparison of treatment with the PRECICE and STRYDE intramedullary lengthening devices and aim to clarify any clinical and radiological differences. A single-centre 2:1 matched-pair retrospective analysis of 42 patients treated with the STRYDE and 82 patients treated with the PRECICE nail between May 2013 and November 2020 was conducted. Clinical and lengthening parameters were compared while focusing radiological assessment on osseous alterations related to the nail’s telescopic junction and locking bolts at four different stages.Aims
Methods
Limb-lengthening nails have largely replaced external fixation in limb-lengthening and reconstructive surgery. However, the adverse events and high prevalence of radiological changes recently noted with the STRYDE lengthening nail have raised concerns about the use of internal lengthening nails. The aim of this study was to compare the prevalence of radiological bone abnormalities between STRYDE, PRECICE, and FITBONE nails prior to nail removal. This was a retrospective case series from three centres. Patients were included if they had either of the three limb-lengthening nails (STYDE, PRECICE, or FITBONE) removed. Standard orthogonal radiographs immediately prior to nail removal were examined for bone abnormalities at the junction of the telescoping nail parts.Aims
Methods
The Precice nail is the latest intramedullary lengthening nail with excellent early outcomes. Implant complications have led to modification of the nail design. The aim of this study was to perform a retrieval study of Precice nails following lower-limb lengthening and to assess macroscopical and microscopical changes to the implants and evaluate differences following design modification, with the aim of identifying potential surgical, implant, and patient risk factors. A total of 15 nails were retrieved from 13 patients following lower-limb lengthening. Macroscopical and microscopical surface damage to the nails were identified. Further analysis included radiology and micro-CT prior to sectioning. The internal mechanism was then analyzed with scanning electron microscopy and energy dispersive x-ray spectroscopy to identify corrosion.Objectives
Methods
The Precice intramedullary limb-lengthening system has demonstrated significant benefits over external fixation lengthening methods, leading to a paradigm shift in limb lengthening. This study compares outcomes following antegrade and retrograde femoral lengthening in both adolescent and adult patients. A retrospective review of prospectively collected data was undertaken of a consecutive series of 107 femoral lengthening operations in 92 patients. In total, 73 antegrade nails and 34 retrograde nails were inserted. Outcome was assessed by the regenerate healing index (HI), hip and knee range of movement (ROM), and the presence of any complications.Aims
Patients and Methods