Conventional growing rods are the most commonly
used distraction-based devices in the treatment of progressive early-onset
scoliosis. This technique requires repeated lengthenings with the
patient anaesthetised in the operating theatre. We describe the
outcomes and complications of using a non-invasive magnetically
controlled growing rod (MCGR) in children with early-onset scoliosis.
Lengthening is performed on an outpatient basis using an external remote
control with the patient awake. Between November 2009 and March 2011, 34 children with a mean
age of eight years (5 to 12) underwent treatment. The mean length
of follow-up was 15 months (12 to 18). In total, 22 children were
treated with dual rod constructs and 12 with a single rod. The mean
number of distractions per patient was 4.8 (3 to 6). The mean pre-operative
Cobb angle was 69° (46° to 108°); this was corrected to a mean 47°
(28° to 91°) post-operatively. The mean Cobb angle at final review
was 41° (27° to 86°). The mean pre-operative distance from T1 to
S1 was 304 mm (243 to 380) and increased to 335 mm (253 to 400)
in the immediate post-operative period. At final review the mean distance
from T1 to S1 had increased to 348 mm (260 to 420). Two patients developed a superficial wound infection and a further
two patients in the single rod group developed a loss of distraction.
In the dual rod group, one patient had pull-out of a hook and one
developed prominent metalwork. Two patients had a rod breakage;
one patient in the single rod group and one patient in the dual
rod group. Our early results show that the MCGR is safe and effective
in the treatment of progressive early-onset scoliosis with the avoidance
of repeated surgical lengthenings. Cite this article:
We present the results of a new
Aims. The aim of this study was to describe, analyze, and compare the survival, functional outcome, and complications of minimally invasive (MI) and
Non-invasive expandable prostheses for limb salvage tumour surgery were first used in 2002. These implants allow ongoing lengthening of the operated limb to maintain limb-length equality and function while avoiding unnecessary repeat surgeries and the phenomenon of anniversary operations. A large series of skeletally immature patients have been treated with these implants at the two leading orthopaedic oncology centres in England (Royal National Orthopaedic Hospital, Stanmore, and Royal Orthopaedic Hospital, Birmingham). An up to date review of these patients has been made, documenting the relevant diagnoses, sites of tumour and types of implant used. 87 patients were assessed, with an age range of 5 to 17 years and follow up range of up to 88 months. Primary diagnosis was osteosarcoma, followed by Ewing's sarcoma. We implanted distal femoral, proximal femoral, total femoral and proximal tibial prostheses. All implants involving the knee joint used a rotating hinge knee. 6 implants reached maximum length and were revised. 8 implants had issues with lengthening but only 4 of these were identified as being due to failure of the lengthening mechanism and were revised successfully. Deep infection was limited to 5% of patients. Overall satisfaction was high with the patients avoiding operative lengthening and tolerating the
Non-invasive expandable prostheses for limb salvage tumour surgery were first used in 2002. These implants allow ongoing lengthening of the operated limb to maintain limb-length equality and function while avoiding unnecessary repeat surgeries and the phenomenon of anniversary operations. A large series of skeletally immature patients have been treated with these implants at the two leading orthopaedic oncology centres in England (Royal National Orthopaedic Hospital, Stanmore, and Royal Orthopaedic Hospital, Birmingham). An up to date review of these patients has been made, documenting the relevant diagnoses, sites of tumour and types of implant used. 74 patients were assessed, with an age range of 7 – 16 years and follow up range of 4 – 88 months. We identified five problems with lengthening. One was due to soft tissue restriction which resolved following excision of the hindering tissue. Another was due to autoclaving of the prosthesis prior to insertion and this patient, along with two others, all had successful further surgery to replace the gearbox. Another six patients required mechanism revision when the prosthesis had reached its maximal length. Complications included one fracture of the prosthesis that was revised successfully and six cases of metalwork infection (two of which were present prior to insertion of the implant and three of which were treated successfully with silver-coated implants). There were no cases of aseptic loosening. Overall satisfaction was high with the patients avoiding operative lengthening and tolerating the
We reviewed our initial seven-year experience
with a non-invasive extendible prosthesis in 34 children with primary bone
tumours. The distal femur was replaced in 25 cases, total femur
in five, proximal femur in one and proximal tibia in three. The
mean follow-up was 44 months (15 to 86) and 27 patients (79%) remain
alive. The prostheses were lengthened by an electromagnetic induction
mechanism in an outpatient setting and a mean extension of 32 mm
(4 to 80) was achieved without anaesthesia. There were lengthening
complications in two children: failed lengthening in one and the
formation of scar tissue in the other. Deep infection developed
in six patients (18%) and local recurrence in three. A total of
11 patients required further surgery to the leg. Amputation was
necessary in five patients (20%) and a two-stage revision in another.
There were no cases of loosening, but two patients had implant breakage
and required revision. The mean Musculoskeletal Tumor Society functional
score was 85% (60% to 100%) at last known follow-up. These early
results demonstrate that the non-invasive extendible prosthesis
allows successful lengthening without surgical intervention, but
the high incidence of infection is a cause for concern.