Advertisement for orthosearch.org.uk
Results 1 - 20 of 78
Results per page:
The Bone & Joint Journal
Vol. 102-B, Issue 5 | Pages 573 - 579
1 May 2020
Krueger DR Guenther K Deml MC Perka C

Aims. We evaluated a large database with mechanical failure of a single uncemented modular femoral component, used in revision hip arthroplasty, as the end point and compared them to a control group treated with the same implant. Patient- and implant-specific risk factors for implant failure were analyzed. . Methods. All cases of a fractured uncemented modular revision femoral component from one manufacturer until April 2017 were identified and the total number of implants sold until April 2017 was used to calculate the fracture rate. The manufacturer provided data on patient demographics, time to failure, and implant details for all notified fractured devices. Patient- and implant-specific risk factors were evaluated using a logistic regression model with multiple imputations and compared to data from a previously published reference group, where no fractures had been observed. The results of a retrieval analysis of the fractured implants, performed by the manufacturer, were available for evaluation. Results. There were 113 recorded cases with fracture at the modular junction, resulting in a calculated fracture rate of 0.30% (113/37,600). The fracture rate of the implant without signs of improper use was 0.11% (41/37,600). In 79% (89/113) of cases with a failed implant, either a lateralized (high offset) neck segment, an extralong head, or the combination of both were used. Logistic regression analysis revealed male sex, high body mass index (BMI), straight component design, and small neck segments were significant risk factors for failure. Investigation of the implants (76/113) showed at least one sign of improper use in 72 cases. Conclusion. Implant failure at the modular junction is associated with patient- and implant-specific risk factors as well as technical errors during implantation. Whenever possible, the use of short and lateralized neck segments should be avoided with this revision system. Implantation instructions and contraindications need to be adhered to and respected. Cite this article: Bone Joint J 2020;102-B(5):573–579


The Bone & Joint Journal
Vol. 106-B, Issue 2 | Pages 151 - 157
1 Feb 2024
Dreyer L Bader C Flörkemeier T Wagner M

Aims. The risk of mechanical failure of modular revision hip stems is frequently mentioned in the literature, but little is currently known about the actual clinical failure rates of this type of prosthesis. The current retrospective long-term analysis examines the distal and modular failure patterns of the Prevision hip stem from 18 years of clinical use. A design improvement of the modular taper was introduced in 2008, and the data could also be used to compare the original and the current design of the modular connection. Methods. We performed an analysis of the Prevision modular hip stem using the manufacturer’s vigilance database and investigated different mechanical failure patterns of the hip stem from January 2004 to December 2022. Results. Two mechanical failure patterns were identified: fractures in the area of the distal fluted profile (distal stem fracture) and failure of the modular taper (modular fracture). A failure rate of 0.07% was observed for distal stem fracture, and modular fracture rates of 1.74% for the original and 0.013% for the current taper design. Conclusion. A low risk of mechanical failure for both fracture types was observed compared to other known complications in revision hip arthroplasty. In addition, the data show that a design change did significantly reduce the risk of a modular fracture. Cite this article: Bone Joint J 2024;106-B(2):151–157


The Bone & Joint Journal
Vol. 104-B, Issue 7 | Pages 801 - 810
1 Jul 2022
Krull P Steinbrück A Grimberg AW Melsheimer O Morlock M Perka C

Aims. Registry studies on modified acetabular polyethylene (PE) liner designs are limited. We investigated the influence of standard and modified PE acetabular liner designs on the revision rate for mechanical complications in primary cementless total hip arthroplasty (THA). Methods. We analyzed 151,096 primary cementless THAs from the German Arthroplasty Registry (EPRD) between November 2012 and November 2020. Cumulative incidence of revision for mechanical complications for standard and four modified PE liners (lipped, offset, angulated/offset, and angulated) was determined using competing risk analysis at one and seven years. Confounders were investigated with a Cox proportional-hazards model. Results. Median follow-up was 868 days (interquartile range 418 to 1,364). The offset liner design reduced the risk of revision (hazard ratio (HR) 0.68 (95% confidence interval (CI) 0.50 to 0.92)), while the angulated/offset liner increased the risk of revision for mechanical failure (HR 1.81 (95% CI 1.38 to 2.36)). The cumulative incidence of revision was lowest for the offset liner at one and seven years (1.0% (95% CI 0.7 to 1.3) and 1.8% (95% CI 1.0 to 3.0)). No difference was found between standard, lipped, and angulated liner designs. Higher age at index primary THA and an Elixhauser Comorbidity Index greater than 0 increased the revision risk in the first year after surgery. Implantation of a higher proportion of a single design of liner in a hospital reduced revision risk slightly but significantly (p = 0.001). Conclusion. The use of standard acetabular component liners remains a good choice in primary uncemented THA, as most modified liner designs were not associated with a reduced risk of revision for mechanical failure. Offset liner designs were found to be beneficial and angulated/offset liner designs were associated with higher risks of revision. Cite this article: Bone Joint J 2022;104-B(7):801–810


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_16 | Pages 1 - 1
19 Aug 2024
Terhune EB Carstens MF Fruth KM Hannon CP Bedard NA Berry DJ Abdel MP
Full Access

The relative advantages and disadvantages of two-stage versus one-stage management of infected total hip arthroplasties are the current subject of intense debate. To understand the merits of each approach detailed information on the short and, importantly, longer-term outcomes of each must be known. The purpose of this study was to assess the long-term results of two-stage exchange arthroplasty for THAs in one of the largest series to date.

We identified 331 infected THAs treated with a two-stage exchange arthroplasty between 1993 and 2021 at a single institution. Patients were excluded if they had prior treatment for infection. Mean age at reimplantation was 66 years, 38% were female, and mean BMI was 30 kg/m2. PJI diagnosis was based on the 2011 MSIS criteria. A competing risk model accounting for death was utilized. Mean follow up was 8 years.

The cumulative incidence of reinfection was 7% at 1 year and 11% at 5 and 10 years. Factors predictive of reinfection included BMI>30 kg/m2 (HR 2; p=0.049), and need for a spacer exchange (HR 3.2; p=0.006). The cumulative incidence of any revision was 13% at 5 and 10 years. The cumulative incidence of aseptic revision was 3% at 1 year, 7% at 5 years, and 8% at 10 years. Dislocation occurred in 33 hips (11% at 10 years; 45% required revision). Factors predictive of dislocation were female sex (HR 2; p=0.047) and BMI<30 kg/m2 (HR 3; p=0.02). The mean HHS improved from 54 to 75 at 10 years.

In this series of 331 two-stage exchange arthroplasties performed for infected hips, we found a low aseptic revision rate (8%) and a low rate of reinfection (11%) at 10 years. These long-term mechanical and infection data must be kept in mind when considering a paradigm shift to one-stage exchanges.

Level of Evidence: Level III


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_6 | Pages 55 - 55
2 May 2024
McCann C Ablett A Feng T Macaskill V Oliver W Keating J
Full Access

Subtrochanteric femoral fractures are a subset of hip fractures generally treated with cephalomedullary nail fixation\[1\]. Single lag screw devices are most commonly-used, but integrated dual screw constructs have become increasingly popular\[2,3\]. The aim of this study was to compare outcomes of fixation of subtrochanteric femoral fractures using a single lag screw (Gamma3 nail, GN) with a dual screw device (InterTAN nail, IN). The primary outcome was mechanical failure, defined as lag screw cut-out, back-out, nail breakage or peri-implant fracture. Consecutive adult patients (18yrs) with subtrochanteric femoral fracture treated in a single centre were retrospectively identified using electronic records. Patients that underwent surgical fixation using either a long GN (2010–2017) or IN (2017–2022) were included. Medical records and radiographs were reviewed to identify complications of fixation. Cox regression analysis was used to determine the risk of mechanical failure and secondary outcomes by implant design. Multivariable regression models were used to identify predictors of mechanical failure. The study included 622 patients, 354 in the GN group (median age 82yrs, 72% female) and 268 in the IN group (median age 82yrs, 69% female). The risk of any mechanical failure was increased two-fold in the GN group (HR 2.44 \[95%CI 1.13 to 5.26\]; _p=0.024_). Mechanical failure comprising screw cut-out (_p=0.032_), back-out (_p=0.032_) and nail breakage (_p=0.26_) was only observed in the GN group. Technical predictors of failure included varus >5° for cut-out (OR 19.98 \[2.06 to 193.88\]; _p=0.01_), TAD;25mm for back-out (8.96 \[1.36 to 58.86\]; p=0.022) and shortening 1cm for peri-implant fracture (7.81 \[2.92 to 20.91\]; _p=<0.001_). Our results demonstrate that an intercalated screw construct is associated with a lower risk of mechanical failure compared with the a single lag screw device. Intercalated screw designs may reduce the risk of mechanical complications for patients with subtrochanteric femoral fractures


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_16 | Pages 63 - 63
19 Aug 2024
Tsikandylakis G Mortensen KRL Gromov K Mohaddes M Malchau H Troelsen A
Full Access

Vitamin E-doped cross-linked polyethylene (VEPE) has encouraged the use of larger heads in thinner liners in total hip arthroplasty (THA). However, there are concerns about wear and mechanical failure of the thin liner, especially when metal heads are used. The aim of this randomized controlled trial was to investigate if the use of a large metal head in a thin VEPE liner would increase polyethylene wear compared with a standard 32-mm metal head and to compare periacetabular radiolucencies and patient-reported outcomes in THA. 96 candidates for uncemented THA were randomly allocated to either the largest possible metal head (36–44 mm) that could be fitted in the thinnest available VEPE liner (intervention group) or a standard 32-mm metal head (control group). The primary outcome was proximal head penetration (PHP) measured with model-based radiostereometric analysis (RSA). Secondary outcomes were periacetabular radiolucencies and patient-reported outcomes. The mid-term results of the trial at 5 years are presented. Median total PHP (interquartile range) was -0.04 mm (−0.12 to 0.02) in the intervention group and -0.03 mm (=0.14 to 0.05) in the control group (p=0.691). The rates of periacetabular radiolucencies were 1/44 and 4/42 (p=0.197), respectively. Patient-reported hip function and health-related quality of life did not differ between the groups, but participants in the intervention group reported a higher level of activity (median UCLA rank 7 vs 6, p=0.020). There were 5 revisions caused by dislocations (2), periprosthetic fracture (1), stem subsidence (1), or iliopsoas impingement (1). Large metal heads in thin VEPE liners did not increase liner wear and were not associated with liner failure 5 years after THA


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_16 | Pages 6 - 6
19 Aug 2024
Lausmann C Luck S Beil FT Citak M Gehrke T
Full Access

Mechanical failure due to dislocation, fracture and acetabular wear as well as persistence of infection are the main complications associated with the use of hip spacers in the treatment of periprosthetic joint infections (PJI). We have developed a novel, custom-made spacer as part of two-stage septic hip replacement and present the two- to five-year results after reimplantation. We prospectively examined a total of 73 patients over our study period in whom our new spacer technique was used. The technique includes a dual mobility inlay and a cemented straight stem in combination with antibiotic-loaded PMMA bone cement which allows full weight bearing meanwhile the interim period. The follow-up ranged between 24 and 60 months after reimplantation as second stage of a two-stage approach. The patients were contacted as part of the follow-up using a questionnaire concerning reoperation, reinfection as well as hip function by using the Harris Hip Score. 72 patients (98,6%) could be reimplanted, one patient is still using the spacer prothesis for 45 months because of excellent functional results with a Harris Hip Score of 95, nevertheless reimplantation is planned. The reinfection rate was less than 7% after reimplantation. The dislocation rate was 5%, and in total there was an overall complication rate of less than 10%. The Harris Hip Score was significantly improved. The ENDO spacer surgical technique is a promising option in the treatment of periprosthetic joint infections (PJI) for two-stage septic exchange with a low dislocation rate and good infection control after reimplantation. In addition, it enables early mobilization with the possibility of full weight-bearing in the interval between spacer implantation and reimplantation


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_12 | Pages 7 - 7
23 Jun 2023
van Hellemondt GG Faraj S de Windt T van Hooff M Spruit M
Full Access

Pelvic discontinuity (PD) is a detrimental complication following total hip arthroplasty (THA). The aim of this study was to assess the clinical and radiological results of patients with PD who were revised using a custom-made triflange acetabular component (CTAC). This is a single centre prospective study of patients with PD following THA who were treated with a CTAC. The Hip Disability and Osteoarthritis Outcome Score (HOOS), modified Oxford Hip Score (mOHS), EurQol five-dimension three-level (EQ-5D-3L) utility, and Numeric Rating Scale (NRS), including visual analogue score (VAS) for pain, were gathered at baseline, and at one- and two-year follow-up. Clinical and radiological complications, including reasons for revisions were registered. Trends over time are described and tested for significance and clinical relevance. 18 females with 22 CTACs were included with a mean age of 73.5 years (SD 7.7). There were significant improvements between baseline and final follow-up in HOOS (p<0.01), mOHS (p<0.01), EQ-5D-3L utility (p<0.01), EQ-5D-3L NRS (p<0.01), VAS pain rest (p<0.01), and VAS pain activity (p<0.01). A minimal clinically important improvement in mOHS and the HOOS was observed in 16 patients (73%) and 14 patients (64%), respectively. Definite healing of the PD was observed in 19 hips (86%). Complications included six cases with broken screws (27%), four cases (18%) with bony fractures, and one case (4.5%) with sciatic nerve paresthesia. One patient with concurrent bilateral PD had revision surgery due to recurrent dislocations. No revision surgery was performed for screw failure or implant breakage. This is the first prospective assessment in clinical outcome of patients with PD who were treated with a CTAC. We have demonstrated that CTAC in patients with THA acetabular loosening and PD can result in stable constructs with no mechanical failures. Moreover, clinically relevant improvements in health-related quality of life at two years’ follow-up was observed


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_14 | Pages 38 - 38
1 Nov 2021
Argenson J Jaubert M LeBaron M Ollivier M Flecher X
Full Access

Two-stage exchange revision total hip arthroplasty performed in case of infection has been considered during many years the gold standard for treatment of chronic infection. Nevertheless, during the last decade, concerns have risen regarding its safety and its efficiency. The purpose of our study was to, first investigate the spacer complications, then to analyze their risks factors. We retrospectively included 125 patients with chronic hip periprosthetic joint infection who underwent a two stage exchange revision arthroplasty performed between January 2013 and December 2019. All spacer complications were systematically collected and risk factors were analyzed. Statistical evaluation were performed using the Student and Mann-Whitney tests. Our study confirms the hypothesis of a high-risk strategy with 42% of patients who had a mechanical spacer failure and a 20% recurrence of infection during the average two years period of follow-up. We found a high rate of spacer migration (23%) and a low rate of spacer fracture (8%) compared to literature. The most important finding was that the majority of spacer complications and failures were found in a population with high medical comorbidities as highlighted by the ASA, Charlson and Lee score associations, as well as with the cardiac, pulmonary, kidney or hepatic chronic conditions. This study showed that a two-stage hip exchange revision is a high-risk procedure regarding complications and mechanical failures of PMMA spacers. In patients with high medical comorbidities, other strategies may be considered and interdisciplinary cooperation with other facilities are needed to identify and control each risk factor


The Bone & Joint Journal
Vol. 103-B, Issue 7 | Pages 1247 - 1253
1 Jul 2021
Slullitel PA Oñativia JI Zanotti G Comba F Piccaluga F Buttaro MA

Aims. There is a paucity of long-term studies analyzing risk factors for failure after single-stage revision for periprosthetic joint infection (PJI) following total hip arthroplasty (THA). We report the mid- to long-term septic and non-septic failure rate of single-stage revision for PJI after THA. Methods. We retrospectively reviewed 88 cases which met the Musculoskeletal Infection Society (MSIS) criteria for PJI. Mean follow-up was seven years (1 to 14). Septic failure was diagnosed with a Delphi-based consensus definition. Any reoperation for mechanical causes in the absence of evidence of infection was considered as non-septic failure. A competing risk regression model was used to evaluate factors associated with septic and non-septic failures. A Kaplan-Meier estimate was used to analyze mortality. Results. The cumulative incidence of septic failure was 8% (95% confidence interval (CI) 3.5 to 15) at one year, 13.8% (95% CI 7.6 to 22) at two years, and 19.7% (95% CI 12 to 28.6) at five and ten years of follow-up. A femoral bone defect worse than Paprosky IIIA (hazard ratio (HR) 13.58 (95% CI 4.86 to 37.93); p < 0.001) and obesity (BMI ≥ 30 kg/m. 2. ; HR 3.88 (95% CI 1.49 to 10.09); p = 0.005) were significantly associated with septic failure. Instability and periprosthetic fracture were the most common reasons for mechanical failure (5.7% and 4.5%, respectively). The cumulative incidence of aseptic failure was 2% (95% CI 0.4 to 7) at two years, 9% (95% CI 4 to 17) at five years, and 12% (95% CI 5 to 22) at ten years. A previous revision to treat PJI was significantly associated with non-septic failure (HR 9.93 (95% CI 1.77 to 55.46); p = 0.009). At the five-year timepoint, 93% of the patients were alive (95% CI 84% to 96%), which fell to 86% (95% CI 75% to 92%) at ten-year follow-up. Conclusion. Massive femoral bone loss was associated with greater chances of developing a further septic failure. All septic failures occurred within the first five years following the one-stage exchange. Surgeons should be aware of instability and periprosthetic fracture being potential causes of further aseptic revision surgery. Cite this article: Bone Joint J 2021;103-B(7):1247–1253


The Bone & Joint Journal
Vol. 103-B, Issue 7 Supple B | Pages 78 - 83
1 Jul 2021
Roedel GG Kildow BJ Sveom DS Garvin KL

Aims. Highly cross-linked polyethylene (HXLPE) has greatly improved the durability of total hip arthroplasty (THA) in young patients because of its improved wear characteristics. Few studies have followed this population into the second decade, and therefore the purpose of this investigation was to evaluate the clinical outcome for THA patients 50 years of age and younger at a minimum of 15 years postoperatively. The second purpose was to evaluate the radiological findings secondary to wear or mechanical failure of the implant. Methods. Between October 1999 and December 2005, 105 THAs were performed in 95 patients (53 female, 42 male) aged 50 years and younger (mean 42 years (20 to 50)). There were 87 patients (96 hips) that were followed for a minimum of 15 years (mean 17.3 years (15 to 21)) for analysis. Posterior approach was used with cementless fixation with a median head size of 28 mm. HXLPE was the acetabular bearing for all hips. Radiographs were evaluated for polyethylene wear, radiolucent lines, and osteolysis. Results. Clinical outcomes showed significant improvement of mean Harris Hip Scores from 52.8 (SD 13.5) preoperatively to 94.8 (SD 7.6) postoperatively. One hip was revised for recurrent instability, and there were no infections. No hips were revised for mechanical loosening or osteolysis. Mean polyethylene linear wear was 0.04 mm/year and volumetric wear was 6.22 mm. 3. /year, with no significant differences between head size or material. Osteolysis was not present in any of the hips. Conclusion. The use of HXLPE in THA for patients aged 50 years and younger has performed exceptionally well without evidence of significant wear causing mechanical loosening or necessitating revision. The radiolucent lines of the acetabular component must be followed to determine the prognostic significance. This investigation represents the longest clinical follow-up of a large, consecutive cohort of patients aged 50 years or younger with THA using HXLPE. This long-term analysis found negligible polyethylene wear, no incidence of aseptic loosening, and excellent clinical outcomes at and beyond 15 years of follow-up. Cite this article: Bone Joint J 2021;103-B(7 Supple B):78–83


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_13 | Pages 65 - 65
1 Oct 2018
Ayers DC Zheng H Lemay C Yang W Franklin PD
Full Access

Introduction. Historically, US arthroplasty revision rates are based on CMS data that cannot verify initial surgery date in patients under 65 years or laterality of revision. We calculated US one-year revision rates for primary total hip replacement (THR) using a representative cohort. Reasons for revision were documented. Methods. A multi-center cohort from US surgeons in 28 states collected sociodemographic data; medical, emotional, musculoskeletal comorbidities; BMI; and patient-reported pain and function (SF36, HOOS) for elective THR patients. Cases in 2011–2013 were matched with CMS data to ascertain 1 year revision through 2014. Predictors of revision were identified. Chart reviews to verify reasons for revision were performed. Results. Overall, 1.6% of 2926 primary THR surgeries were revised within 12 months. Mean age was 72 years, 59% female. No significant difference in pre-operative age, sex, BMI, pain, function, or emotional health was detected between revision and non-revision patients. Severe low back pain was twice as prevalent among revisions (23% vs. 11%; p<0.04) as was prior stroke (8.5% vs. 3%). Primary reasons for revision were infection (30%), mechanical failure including dislocation (26%), fracture (19%), metal ions (15%) and other (10%). In patients under 65 years of age, the distribution of reasons for revision are identical. Conclusion. The incidence of revision THR in the US within 12 months of the index procedure is 1.6%. Infection and mechanical failure, including dislocation, are the most common causes of early revision, followed by peri-prosthetic fracture. Patients with severe pre-op lumbar spine pain are at higher risk for early revision after THR. Abbreviations: total hip replacement (THR)


Bone & Joint Open
Vol. 3, Issue 6 | Pages 485 - 494
13 Jun 2022
Jaubert M Le Baron M Jacquet C Couvreur A Fabre-Aubrespy M Flecher X Ollivier M Argenson J

Aims. Two-stage exchange revision total hip arthroplasty (THA) performed in case of periprosthetic joint infection (PJI) has been considered for many years as being the gold standard for the treatment of chronic infection. However, over the past decade, there have been concerns about its safety and its effectiveness. The purposes of our study were to investigate our practice, collecting the overall spacer complications, and then to analyze their risk factors. Methods. We retrospectively included 125 patients with chronic hip PJI who underwent a staged THA revision performed between January 2013 and December 2019. All spacer complications were systematically collected, and risk factors were analyzed. Statistical evaluations were performed using the Student's t-test, Mann-Whitney U test, and Fisher's exact test. Results. Our staged exchange practice shows poor results, which means a 42% mechanical spacer failure rate, and a 20% recurrent infection rate over the two years average follow-up period. Moreover, we found a high rate of spacer dislocation (23%) and a low rate of spacer fracture (8%) compared to the previous literature. Our findings stress that the majority of spacer complications and failures is reflecting a population with high comorbid burden, highlighted by the American Society of Anesthesiology grade, Charlson Comorbidity Index, and Lee score associations, as well as the cardiac, pulmonary, kidney, or hepatic chronic conditions. Conclusion. Our experience of a two-stage hip exchange revision noted important complication rates associated with high failure rates of polymethylmethacrylate spacers. These findings must be interpreted in the light of the patient’s comorbidity profiles, as the elective population for staged exchange has an increasing comorbid burden leading to poor results. In order to provide better results for this specific population, our conclusion suggests that comparative strategy studies are required to improve our therapeutic indication. Cite this article: Bone Jt Open 2022;3(6):485–494


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_10 | Pages 19 - 19
1 Oct 2020
Roedel GG Sveom DS Garvin KL
Full Access

Introduction. THA for patient's 50 years and younger is a procedure at high risk for complications and failure because of the high level of activity of this population. Highly cross linked polyethylene (HXLPE) has greatly improved the durability of the implant because of the improved wear characteristics. Few studies have followed this population into the second decade and therefore the purpose of this investigation was to evaluate the clinical outcome for the patients 5o years of age and younger at a minimum of 15 years. The second purpose was to evaluate the radiographic findings secondary to wear or mechanical failure of the implant. Methods. Between October 1999 and December 2005, 105 THAs were performed in 95 patients (53 female, 42 male) age 50 years and younger (mean 42 years; range 20–50). Ten patients (10 hips) were lost to follow-up or deceased. The remaining 95 hips and 85 patients were followed for a minimum of 15 years (mean of 16.8, range 15–20.5) for analysis. HXLPE (Longevity, Zimmer Biomet) was the acetabular bearing for all hips. Radiographs were evaluated for radiolucent lines and osteolysis. Results. Clinical outcomes showed excellent results in 80 hips, good results in 10 hips, fair in 3 hips and poor in 2 hips. One hip was revised for recurrent instability, one patient suffered a DVT and there were no infections. None of the hips were revised for mechanical loosening or osteolysis. Radiolucent lines were present in Zone I of the acetabulum in 23 hips and Zone II in 6 and Zone III 7. Osteolysis was not present in any of the hips. Conclusions. The use of HXLPE in THA for patients 50 years of age and younger has performed exceptionally well without evidence of wear causing mechanical loosening or necessitating revision. Radiographic evidence of osteolysis was not identified, but progressive radiolucent lines of the acetabular component were noted


Bone & Joint Open
Vol. 4, Issue 2 | Pages 53 - 61
1 Feb 2023
Faraj S de Windt TS van Hooff ML van Hellemondt GG Spruit M

Aims

The aim of this study was to assess the clinical and radiological results of patients who were revised using a custom-made triflange acetabular component (CTAC) for component loosening and pelvic discontinuity (PD) after previous total hip arthroplasty (THA).

Methods

Data were extracted from a single centre prospective database of patients with PD who were treated with a CTAC. Patients were included if they had a follow-up of two years. The Hip Disability and Osteoarthritis Outcome Score (HOOS), modified Oxford Hip Score (mOHS), EurQol EuroQoL five-dimension three-level (EQ-5D-3L) utility, and Numeric Rating Scale (NRS), including visual analogue score (VAS) for pain, were gathered at baseline, and at one- and two-year follow-up. Reasons for revision, and radiological and clinical complications were registered. Trends over time are described and tested for significance and clinical relevance.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_12 | Pages 37 - 37
1 Oct 2019
Berend KR Crawford DA Adams JB Morris MJ Lombardi AV
Full Access

Background. Recurrent instability and dislocation after total hip arthroplasty are difficult complications. In certain cases, a constrained acetabular device can address these issues. The purpose of this study is to report the midterm outcomes and survivorship of a single constrained liner device. Methods. A retrospective review as performed on all procedures (except first stage exchange for infection) in which a Freedom® Constrained (Zimmer Biomet, Warsaw, IN) liner was used between December 2003 to November 2016. Patients with 2-year minimum follow-up or failure were included, yielding a cohort of 177 patients. Procedures were 130 revisions, 40 reimplantations following infection eradication, and 7 complex primaries. The constrained mechanism was implanted in 46 hips (26%) to treat active instability and 131 hips (74%) for increased risk of instability or intraoperative instability. Patients had on average 3.4 previous surgeries. Results. With an average 7.1-year follow-up, 11 hips dislocated (6.2%), and 13 hips (7.3%) were revised for aseptic loosening of the acetabular component, resulting in an overall constrained mechanism aseptic or mechanical failure rate of 14.1%. Nineteen hips (10.7%) failed from infection with 58% of these having had a previous infection. Patients with active instability had significantly higher failure for dislocation than patients who were at-risk (15.2% vs. 3%, p=0.01). All-cause survival rate at 7 years was 74.8%%, aseptic survival was 83.6%, and survival for instability was 91.8%. Conclusion. Revision for instability remains challenging as many patients have had numerous previous surgeries and at-risk anatomy. Constrained inserts are one option to manage instability, but a high rate of recurrence still occurs. For any tables or figures, please contact the authors directly


The Bone & Joint Journal
Vol. 105-B, Issue 1 | Pages 29 - 34
1 Jan 2023
Fransen BL Bengoa FJ Neufeld ME Sheridan GA Garbuz DS Howard LC

Aims

Several short- and mid-term studies have shown minimal liner wear of highly cross-linked polyethylene (HXLPE) in total hip arthroplasty (THA), but the safety of using thinner HXLPE liners to maximize femoral head size remains uncertain. The objective of this study was to analyze clinical survival and radiological wear rates of patients with HXLPE liners, a 36 mm femoral head, and a small acetabular component with a minimum of ten years’ follow-up.

Methods

We retrospectively identified 55 patients who underwent primary THA performed at a single centre, using HXLPE liners with 36 mm cobalt-chrome heads in acetabular components with an outer diameter of 52 mm or smaller. Patient demographic details, implant details, death, and all-cause revisions were recorded. Cox regression and Kaplan-Meier survival was used to determine all-cause and liner-specific revision. Of these 55 patients, 22 had a minimum radiological follow-up of seven years and were assessed radiologically for linear and volumetric wear.


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_6 | Pages 28 - 28
1 May 2019
Pryce G Al-Hajjar M Wilcox R Thompson J Board T Williams S
Full Access

Impingement of total hip replacements (THRs) can cause rim damage of polyethylene liners, and lead to dislocation and/or mechanical failure of liner locking mechanisms[1]. Previous work has focussed on the influence of femoral neck profile on impingement without consideration of neck-shaft angle. This study assessed the occurrence of impingement with two different stem designs (Corail standard [135°] and coxa vara [125°]) under different activities with varying acetabular cup orientation (30° to 70° inclination; 0° to 50° anteversion) using a geometric modelling tool. The tool was created in a computer aided design software programme, and incorporated an individual's hemi-pelvis and femur geometry[3] with a THR (DePuy Synthes Pinnacle. ®. shell and neutral liner; size 12 Corail. ®. standard or coxa vara and 32mm head). Kinematic data of activities associated with dislocation[2], such as stooping to pick an object from the floor was applied and incidences of impingement were recorded. Predicted implant impingement was influenced by stem design. The coxa vara stem was predicted to cause implant impingement less frequently across the range of activities and cup orientations investigated, compared to the standard stem [Fig. 1]. The cup orientations predicted to cause impingement the least frequently were at lower inclination and anteversion angles, relative to the standard stem [Fig. 1]. The coxa vara stem included a collar, while the standard stem was collarless; additional analysis indicated that differences were due to neck angle and not the presence of a collar. This study demonstrated that stem neck-shaft angle is an important variable in prosthetic impingement in THR and surgeons should be aware of this when choosing implants. Future work will consider further implant design and bone geometry variables. This tool has the potential for use in optimising stem design and position and could assist with patient specific stem selection based on an individual's activity profile. For any figures or tables, please contact the authors directly


Orthopaedic Proceedings
Vol. 101-B, Issue SUPP_12 | Pages 64 - 64
1 Oct 2019
McCalden RW Decker MM Klassen RJ Lanting B
Full Access

Introduction. HXLPE acetabular liners were introduced to reduce wear-related complications in THA. However, post-irradiation thermal free radical stabilization can compromise mechanical properties, leave oxidation-prone residual free radicals, or both. Reports of mechanical failure of HXLPE acetabular liner rims raise concerns about thermal free radical stabilization and in vivo oxidization on implant properties. The purpose of this study is to explore the differences in the mechanical, physical and chemical properties of HXLPE acetabular liner rims after extended time in vivo between liners manufactured with different thermal free radical stabilization techniques. Material and methods. Remelted, single annealed and sequentially annealed retrieved HXLPE acetabular liners with in vivo times greater than 4.5 years were obtained from our implant retrieval laboratory. All retrieved liners underwent an identical sanitation and storage protocol. For mechanical testing, a total of 55 explants and 13 control liners were tested. Explant in vivo time ranged from 4.6 – 14.0 years and ex vivo time ranged from 0 – 11.6 years. Rim mechanical properties were tested by microindentation hardness testing using a Micromet II Vickers microhardness tester following ASTM standards. A subset of 16 explants with ex vivo time under one year along with five control liners were assessed for oxidation by FTIR, crystallinity by Raman spectroscopy, and evidence of microcracking by SEM. Results. No significant difference in in vivo or ex vivo was found between thermal stabilization groups in either set of explants studied. In the mechanically tested explants, there was no significant correlation between in vivo time and Vickers hardness in any thermal stabilization group. A significant correlation was found between ex vivo time and hardness in remelted liners (Δ=.520, p=.011), but not in either annealed cohort. ANCOVA with ex vivo time as a covariate found a significant difference in hardness between the thermal free radical stabilization groups (p<.0005, η. 2. = 0.322). Post hoc analysis revealed hardness was significantly lower in the retrieved remelted group compared to both the single annealed (p=.001) and sequentially annealed (p<.0005) cohorts. Hardness was significantly higher in the retrieved remelted liners compared to controls (p=.007), with no different in either annealed cohort (figure 1). Detectable subsurface oxidation (OI > 0.1) was found in retrieved remelted (25%), single annealed (100%) and sequentially annealed (75%) liner rims (figure 2). Crystallinity was increased in the subsurface region relative to control liners for both annealed, but not remelted, liner rims. Hardness was increased in oxidized rims for both annealed cohorts but not in the remelted cohort. Microcracking was only found along the surface of one unoxidized remelted liner rim. Conclusion. Mechanical properties were reduced at baseline and worsened after in vivo time for remelted HXLPE liner rims. Rim oxidation was detected in all groups. Oxidation was associated with increased crystallinity and hardness in annealed cohorts, but not remelted liners. Increased crystallinity and oxidation do not appear to be directly causing the worsened mechanical behavior of remelted HXLPE liner rims after extended in vivo time. For any tables or figures, please contact the authors directly


Bone & Joint Open
Vol. 5, Issue 4 | Pages 286 - 293
9 Apr 2024
Upadhyay PK Kumar V Mirza SB Shah N

Aims

This study reports the results of 38 total hip arthroplasties (THAs) in 33 patients aged less than 50 years, using the JRI Furlong hydroxyapatite ceramic (HAC)-coated femoral component.

Methods

We describe the survival, radiological, and functional outcomes of 33 patients (38 THAs) at a mean follow-up of 27 years (25 to 32) between 1988 and 2018.