Aims. This study investigates head-neck taper corrosion with varying head size in a novel
One unpredictable clinical risk with THA may be impingement of a metal cup rim against a metal femoral-neck, with concomitant release of metal particles. Our objective was to determine if metal debris could be one trigger for catastrophic MOM wear. To simulate release of metal debris, we added metal particles (CoCr #230, Ti6Al4V #340) to six 38mm MOM bearings at beginning of each simulator test interval. The weekly 500,000 cycle intervals were replicated 10 times to acquire 5-million duty cycles. Flakes of polymerized bone cement (PMMA) were scraped from a retrieved TKR and used as control debris (N = 1,300 particles; 3 MOM).Introduction
Patients/Materials & Methods
Hip arthroscopy is a rapidly expanding technique that has a steep learning curve. Simulation may have a role in helping trainees overcome this. However there is as yet no validated
Introduction. Decreases in trainees' working hours, coupled with evidence of worse outcomes when hip arthroscopies are performed by inexperienced surgeons, mandate the development of additional means of arthroscopic training. Though virtual reality simulation training has been adopted by other surgical specialities, its slow uptake in arthroscopic training is due to a lack of evidence as to its benefits. These benefits can be demonstrated through learning curves associated with simulator training – with practice reflecting measurable increases in validated performance metrics. Materials & Methods. Twenty-five medical students completed seven simulated arthroscopies of a healthy virtual hip joint in the supine position on a simulator previously shown to have construct validity. Twelve targets had to be visualised within the central compartment; six via the anterior portal, three via the anterolateral portal and three via the posterolateral portal. Eight students proceeded to complete seven probe examinations of a healthy virtual hip joint. Eight targets were probed via the anterolateral portal. Task duration, number of collisions with soft tissue and bone, and distance travelled by arthroscope were measured by the simulator for every session. Results. A learning curve was demonstrated by the students, with significant improvements in time taken (P<0.01), number of collisions (P<0.01), collision severity (P<0.01), and efficiency of movement (P<0.01). The largest difference between consecutive sessions was seen between sessions 1 and 2, with sessions thereafter showing only minimal rates of improvement. Similar improvements were found in the probe examination with students showing significant improvements in time taken (P<0.01), number of collisions (P<0.01), collision severity (P<0.01) and distance travelled by arthroscope (P<0.01). Conclusion. The results of this study demonstrate a learning curve for a previously validated
Advances in total hip and knee replacement technologies have heretofore been largely driven by corporate marketing hype with each seeming advancement accompanied by a cost increase often out in front of peer-reviewed reports documenting their efficacy or not. As example, consider the growing use of ceramic femoral heads in primary total hip arthroplasty (THA). The question to consider is “Can an upcharge of $350 for a ceramic femoral head in primary THA be justified?” The answer to this question lies in an appreciation of whether the technology modifies the potential for costly revision arthroplasty procedures. Peer-Reviewed Laboratory & Clinical Review - According to the 2022 Australian National Joint Replacement Registry, the four leading causes of primary THA failure requiring revision are: 1.) infection, 2.) dislocation/instability, 3.) periprosthetic fracture and 4.) loosening, which constitute 87.5% of the reported reasons for revision. Focusing on these failure modes,
Introduction. Adverse local tissue reactions (ALTR) can result in devastating soft tissue and osseous destruction, while potentially increasing the risk of concomitant periprosthetic joint infection (PJI). The aims of this study were to evaluate cobalt (Co) and chromium (Cr) levels generated in simulators from metal-on-polyethylene (MoP) and ceramic-on-polyethylene (CoP) constructs, and determine their impact on native tissues and PJI risk through evaluation of human adipose-derived mesenchymal stem cells (AMSCs) and Staphylococcus epidermidis isolates. Methods. Ten
We report on an innovative surface grafting to highly crosslinked (HXLPE) bearing for THA using a biocompatible-phospholipid-polymer poly (2-methacryloyloxyethyl phosphorylcholine) (PMPC). Such hydrophilic surfaces mimic articular cartilage and are hypothesized to improve lubrication and thereby reduce friction and wear. We performed in vitro testing of wear and friction of ceramic-on-polyethylene THRs with the PMPC treatment, and compared them with untreated controls. Highly cross-linked UHMWPE bearings, gamma-ray-irradiated at different levels with and without vitamin E (HXL Vit. E: 125 kGy, HXL: 75 kGy, respectively) were divided so half were PMPC treated (n=3 for all four groups). All were paired with identical 40 mm diameter zirconia-toughened-alumina ceramic heads. Testing was carried-out on an AMTI
A significant reduction in wear at five and ten years was previously reported when comparing Durasul highly cross-linked polyethylene with nitrogen-sterilized Sulene polyethylene in total hip arthroplasty (THA). We investigated whether the improvement observed at the earlier follow-up continued, resulting in decreased osteolysis and revision surgery rates over the second decade. Between January 1999 and December 2001, 90 patients underwent surgery using the same acetabular and femoral components with a 28 mm metallic femoral head and either a Durasul or Sulene liner. A total of 66 hips of this prospective randomized study were available for a minimum follow-up of 20 years. The linear femoral head penetration rate was measured at six weeks, one year, and annually thereafter, using the Dorr method on digitized radiographs with a software package.Aims
Methods
Introduction. Because of concerns regarding excessive wear and short-term failures attributed to the metal-on-metal bearings, the use of metal-on-metal hip resurfacing arthroplasty (MOMHRA) has been greatly reduced since 2008, despite great mid-term results for well-designed implants and in certain patient populations. The true cause of excessive wear was then unknown. Therefore, identification of true risk factors for the procedure became paramount to refine indications and improve survivorship outcomes. Methods. Over the last 10 years, a systematic search of the US national library of Medicine and National institutes of health with the key words “metal-on-metal” and hip resurfacing” was conducted and returned 2186 items. Of these items, 862 were deemed relevant to our research purposes and entered in our center's reference database from which this review was performed. Results. Edge loading is the main culprit for high wear and high serum Co and Cr ion concentrations because it disrupts the fluid film lubrication of the device. Computation of the contact patch to rim distance (CPR), an estimate of the joint's functional coverage, is the best predictor of potential edge loading and excessive wear. Both in vivo and in vitro studies show that the wear of well-designed and well-positioned MOM bearings diminishes over time with continued use, an advantage only featured by MOM bearings. Systemic wear-related complications and hypersensitivity to metal once thought to be common are in fact rare occurrences. In addition, metal-related revisions only represent a small portion of the various modes of failure encountered with well-designed HRA. In our series of 1321 hips with only 0.5% lost to follow-up, 11 patients underwent revision surgery for excessive wear or adverse local tissue reaction. All but 2 had mal-positioned acetabular components (CPR distance <10mm). One of these 2 patients had serum cobalt and chromium levels of 13 and 9 µg/L respectively, despite a CPR distance of 18.3 mm, while the other showed a peri-prosthetic fluid collection estimated at 111cc on MRI. Component aseptic loosening (acetabular or femoral) remains, as is also the case for total hip arthroplasty (THA), the leading indication for revision surgery, even though substantial progress has been reported to reduce its incidence. Femoral neck fractures and loosening are associated with the surgeon's learning curve which can be avoided with proper training. The survivorship of the femoral component in our series after implementation of surgical technique changes is 99.2% at 10 years for an overall survivorship of 95.5%. A diagnosis of DDH is associated with a higher failure rate, particularly from socket loosening, while women without risk factors have a survivorship of 98.6% at 15 years. Conclusion. HRA has often been, but should not be associated with the results of large head metal-on-metal THA which often present with a different set of failure mechanisms. Now that most of the risk factors for HRA have been identified, and solutions found, a balanced perspective of its results is needed because the long-term data available validate the low wear of MOM bearings which was predicted in the early
This study reports on ceramic-on-metal (CoM) bearings in total hip replacement. Whole blood metal ion levels were measured. The median increase in chromium and cobalt at 12 months was 0.08 μg/1 and 0.22 μg/1, respectively, in CoM bearings. Comparable values for metal-on-metal (MoM) were 0.48 μg/1 and 0.32 μg/1. The chromium levels were significantly lower in CoM than in MoM bearings (p = 0.02). The cobalt levels were lower, but the difference was not significant. Examination of two explanted ceramic heads revealed areas of thin metal transfer. CoM bearings (one explanted head and acetabular component, one explanted head and new acetabular component, and three new heads and acetabular components) were tested in a
Head-taper corrosion is a cause of failure in total hip arthroplasty (THA). Recent reports have described an increasing number of V40 taper failures with adverse local tissue reaction (ALTR). However, the real incidence of V40 taper damage and its cause remain unknown. The aim of this study was to evaluate the long-term incidence of ALTR in a consecutive series of THAs using a V40 taper and identify potentially related factors. Between January 2006 and June 2007, a total of 121 patients underwent THA using either an uncemented (Accolade I, made of Ti12Mo6Zr2Fe; Stryker, USA) or a cemented (ABG II, made of cobalt-chrome-molybdenum (CoCrMo); Stryker) femoral component, both with a V40 taper (Stryker). Uncemented acetabular components (Trident; Stryker) with crosslinked polyethylene liners and CoCr femoral heads of 36 mm diameter were used in all patients. At a mean folllow-up of 10.8 years (SD 1.1), 94 patients (79%) were eligible for follow-up (six patients had already undergone a revision, 15 had died, and six were lost to follow-up). A total of 85 THAs in 80 patients (mean age 61 years (24 to 75); 47 (56%) were female) underwent clinical and radiological evaluation, including the measurement of whole blood levels of cobalt and chrome. Metal artifact reduction sequence MRI scans of the hip were performed in 71 patients.Aims
Methods
The aim of this study was to evaluate the performance of first-generation annealed highly cross-linked polyethylene (HXLPE) in cementless total hip arthroplasty (THA). We retrospectively evaluated 29 patients (35 hips) who underwent THA between December 2000 and February 2002. The survival rate was estimated using the Kaplan-Meier method. Hip joint function was evaluated using the Japanese Orthopaedic Association (JOA) score. Two-dimensional polyethylene wear was estimated using Martell’s Hip Analysis Suite. We calculated the wear rates between years 1 and 5, 5 and 10, 10 and 15, and 15 and final follow-up.Aims
Methods
This study reports the mid-term results of total hip arthroplasty (THA) performed using a monoblock acetabular component with a large-diameter head (LDH) ceramic-on-ceramic (CoC) bearing. Of the 276 hips (246 patients) included in this study, 264 (96%) were reviewed at a mean of 67 months (48 to 79) postoperatively. Procedures were performed with a mini posterior approach. Clinical and radiological outcomes were recorded at regular intervals. A noise assessment questionnaire was completed at last follow-up.Aims
Patients and Methods
The purpose of this study was to compare two different types
of metal-on-metal (MoM) bearing for total hip arthroplasty (THA):
one with a large femoral head (38 mm to 52 mm) and the other with
a conventional femoral head (28 mm or 32 mm). We compared clinical
outcome, blood metal ion levels, and the incidence of pseudotumour in
the two groups. Between December 2009 and December 2011, 62 patients underwent
MoM THA with a large femoral head (Magnum group) and 57 patients
an MoM THA with a conventional femoral head (conventional group).
Clinical outcome was assessed using the Harris Hip score, University
of California, Los Angeles (UCLA) activity score and EuroQol-5D
(EQ-5D). Blood metal ion levels were measured and MRI scans were
analyzed at a minimum of five years postoperatively.Aims
Patients and Methods
We reviewed the literature on the currently available
choices of bearing surface in total hip replacement (THR). We present
a detailed description of the properties of articulating surfaces
review the understanding of the advantages and disadvantages of
existing bearing couples. Recent technological developments in the
field of polyethylene and ceramics have altered the risk of fracture
and the rate of wear, although the use of metal-on-metal bearings has
largely fallen out of favour, owing to concerns about reactions
to metal debris. As expected, all bearing surface combinations have
advantages and disadvantages. A patient-based approach is recommended,
balancing the risks of different options against an individual’s
functional demands. Cite this article:
We have encountered patients who developed large joint fluid collections with massive elevations in chromium (Cr) and cobalt (Co) concentrations following metal-on-metal (MoM) hip arthroplasties. In some cases, retrieval analysis determined that these ion concentrations could not be explained simply by the wear rates of the components. We hypothesized that these effects may be associated with aseptic lymphocyte-dominated vasculitis-associated lesion (ALVAL). We examined the influence of the ALVAL grade on synovial fluid Co and Cr concentrations following adjustment for patient and device variables, including volumetric wear rates. Initially restricting the analysis to include only patients with one MoM hip resurfacing device, we performed multiple regression analyses of prospectively collected data. We then repeated the same statistical approach using results from a larger cohort with different MoM designs, including total hip arthroplasties.Objectives
Patients and Methods
Oxidised zirconium (OxZi) has been developed
as an alternative bearing surface for femoral heads in total hip arthroplasty
(THA). This study has investigated polyethylene wear, functional
outcomes and complications, comparing OxZi and cobalt–chrome (CoCr)
as part of a three-arm, multicentre randomised controlled trial.
Patients undergoing THA from four institutions were prospectively
randomised into three groups. Group A received a CoCr femoral head
and highly cross-linked polyethylene (XLPE) liner; Group B received
an OxZi femoral head and XLPE liner; Group C received an OxZi femoral
head and ultra-high molecular weight polyethylene (UHMWPE) liner.
At five years, 368 patients had no statistically significant differences
in short-form-36 (p = 0.176 mental, p = 0.756 physical), Western
Ontario and McMaster Universities Osteoarthritis Index (p = 0.847),
pain scores
(p = 0.458) or complications. The mean rate of linear wear was 0.028
mm/year (standard deviation ( Cite this article:
In a double-blinded randomised controlled trial,
83 patients with primary osteoarthritis of the hip received either
a ceramic-on-metal (CoM) or metal-on-metal (MoM) total hip replacement
(THR). The implants differed only in the bearing surfaces used.
The serum levels of cobalt and chromium and functional outcome scores
were compared pre-operatively and at six and 12 months post-operatively. Data were available for 41 CoM and 36 MoM THRs (four patients
were lost to follow-up, two received incorrect implants). The baseline
characteristics of both cohorts were similar. Femoral head size
measured 36 mm in all but two patients who had 28 mm heads. The
mean serum cobalt and chromium levels increased in both groups,
with no difference noted between groups at six months (cobalt p
= 0.67, chromium p = 0.87) and 12 months (cobalt p = 0.76, chromium
p = 0.76) post-operatively. Similarly, the mean Oxford hip scores,
Western Ontario and McMaster Universities Osteoarthritis index and
University of California, Los Angeles activity scores showed comparable improvement
at 12 months. Our findings indicate that CoM and MoM couplings are associated
with an equivalent increase in serum cobalt and chromium levels,
and comparable functional outcome scores at six and 12-months follow-up.
We report the five-year outcome of a randomised
controlled trial which used radiostereometric analysis (RSA) to assess
the influence of surface oxidised zirconium (OxZr, Oxinium) on polyethylene
wear A total of 120 patients, 85 women and 35 men with a mean age
of 70 years (59 to 80) who were scheduled for primary cemented total
hip arthroplasty were randomly allocated to four study groups. Patients
were blinded to their group assignment and received either a conventional
polyethylene (CPE) or a highly cross-linked (HXL) acetabular component
of identical design. On the femoral side patients received a 28
mm head made of either cobalt-chromium (CoCr) or OxZr. The proximal head penetration (wear) was measured with repeated
RSA examinations over five years. Clinical outcome was measured
using the Harris hip score. There was no difference in polyethylene wear between the two
head materials when used with either of the two types of acetabular
component (p = 0.3 to 0.6). When comparing the two types of polyethylene
there was a significant difference in favour of HXLPE, regardless
of the head material used (p <
0.001). In conclusion, we found no advantage of OxZr over CoCr in terms
of polyethylene wear after five years of follow-up. Our findings
do not support laboratory results which have shown a reduced rate
of wear with OxZr. They do however add to the evidence on the better
resistance to wear of HXLPE over CPE. Cite this article:
The Articular Surface Replacement (ASR) hip resurfacing arthroplasty has a failure rate of 12.0% at five years, compared with 4.3% for the Birmingham Hip Resurfacing (BHR). We analysed 66 ASR and 64 BHR explanted metal-on-metal hip replacements with the aim of understanding their mechanisms of failure. We measured the linear wear rates of the acetabular and femoral components and analysed the clinical cause of failure, pre-revision blood metal ion levels and orientation of the acetabular component. There was no significant difference in metal ion levels (chromium, p = 0.82; cobalt, p = 0.40) or head wear rate (p = 0.14) between the two groups. The ASR had a significantly increased rate of wear of the acetabular component (p = 0.03) and a significantly increased occurrence of edge loading (p <
0.005), which can be attributed to differences in design between the ASR and BHR. The effects of differences in design on the