Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_1 | Pages 23 - 23
2 Jan 2024
Ciatti C Quattrini F Asti C Maniscalco P
Full Access

Previous scientific studies have highlighted how coupling is an important element affecting total hip arthroplasty's survival.

This study aims to evaluate whether metal-on-metal (MOM) coupling could be a statistically significant risk factor.

The data from the regional joint registry (Registro dell'Impiantologia Protesica Ortopedica, RIPO) was used for analysis. The data collection accuracy of this registry was 97.2% in 2017.

We retrospective evaluate all MOM total hip arthroplasties (THAs) implanted in our department between January 01st 2000 and December 31st 2011. We used a control group composed by all other prosthesis implanted in our Department in the same time lapse.

We registered 660 MOM THAs. Mean age of patients was 66.9 years. 603 patients have a >36mm head, while 78 a <36 mm one. Neck modularity was present in half of patients. 676 implants were cementless. We registered 69 revisions, especially due to aseptic mobilization (16 THAs), implant breakage (9 THAs) and periprosthetic fracture (6 THAs).

The MOM THAs overall Kaplan-Meier survival rate was 87.2 at 15 years, and the difference between MOM THAs and other implants two curves is statistically significant (p<0.05). Male sex is a significant risk factors. Further evaluations are in progress to establish the presence of any additional risk factors. We think weight and/or BMI may be included in this category.

Our study confirms the data currently present in the literature regarding a lower survival of metal-on-metal hip prostheses.

The male sex is a statistically significant risk factor (p<0.05), while age, head size and modularity of the prosthetic neck are not statistically significant (p>0.05).

Any new finds will be presented at the congress venue.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_14 | Pages 3 - 3
1 Nov 2018
Ike H Inaba Y Kobayashi N Choe H Tezuka T Kobayashi D Watanabe S Higashihira S
Full Access

SL-PLUS MIA stem (Smith & Nephew Orthopaedics AG) is a modified implant of Zweymuller type SL-PLUS standard stem (Smith & Nephew Orthopaedics AG). We constructed finite element (FE) models and analysed equivalent stresses in the femur. In addition, we measured bone mineral density (BMD) in the femur by dual-energy X-ray absorptiometry (DEXA) after THA. The purpose of this study was to investigate the equivalent stress and to compare the results of the FE analyses with changes in BMD after THA. Twenty-one patients (18 women and 3 men) who underwent primary cementless THA with SL-PLUS MIA or SL-PLUS formed the basis of this study. Eleven patients received SL-PLUS MIA and ten patients received SL-PLUS. Zones were defined according to Gruen's system (zones 1∼7). Computed-tomography (CT) images of the femur were taken before and at 1 week after THA. FE models of the femur and prosthesis were obtained from CT data by Mechanical Finder (Research Center of Computational Mechanics Inc., Tokyo, Japan), software that creates FE models showing individual bone shape and density distribution. Equivalent stresses were analysed in zones 1 to 7 and compared to the DEXA data. FE studies revealed that there was no significant difference in equivalent stress between SL-PLUS MIA and SL-PLUS. BMD was maintained after THA in zones 3, 4, and 5, whereas BMD decreased in zones 2, 6, and 7. In zone 1, BMD decreased in SL-PLUS MIA stem group by 14%, while BMD was maintained in SL-PLUS standard stem.


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_4 | Pages 52 - 52
1 Apr 2018
Rieker C
Full Access

Total Hip Arthroplasty (THA) is a well-established, cost-effective treatment for improving function and alleviating pain in patients who have disabling hip disease with excellent long-term results. Based on the excellent results, there is an ongoing trend for THA to be performed in younger and more active patients, having higher physical demands on their new total joints.

Polyethylene (PE) wear and its biological consequences are one of the main causes of implant failure in THA. Macrophages phagocytise PE wear particles and this will result in osteolysis and loss of periprosthetic bone. The risk of these complications can be estimated in relation to the amount of volumetric wear based on two assumptions: that the number of PE particles dispersed in the peri-prosthetic tissues is controlled by the amount of PE wear; and that the development of osteolysis and the resulting aseptic loosening is triggered by these PE particles. Based on these assumptions, a model was developed to estimate the osteolysis-free life of a THA, depending on the Linear Wear Rate (LWR) and femoral head size of the PE bearing.

A review of the literature was conducted to provide an estimate of the radiologic osteolysis threshold based on the volumetric wear of the PE bearing. This review demonstrates that this radiologic osteolysis threshold is approximated 670 mm3 for conventional PE. The osteolysis-free life of the THA was estimated by simply dividing this threshold volume by the annual Volumetric Wear Rate (VWR) of the bearing. The annual VWR is basically controlled by two parameters: (1) annual LWR and (2) head size, and was calculated by using published formulae.

For 28 mm heads, following osteolysis-free life was determined in function of the annual LWR. LWR: 10 µm/y => 116.6 years / LWR: 25 µm/y => 46.6 years / LWR: 50 µm/y => 23.3 years / LWR: 100 µm/y => 11.6 years. For 40 mm heads, following osteolysis-free life was determined in function of the annual LWR. LWR: 10 µm/y => 57.1 years / LWR: 25 µm/y => 22.9 years / LWR: 50 µm/y => 11.4 years / LWR: 100 µm/y => 5.7 years.

The osteolysis-free life determined by this model is in good agreement with the clinical results of PE bearings having a 28 mm head size and demonstrates that extreme low LWRs are mandatory to assure a descent osteolysis-free life for THA (PE bearings) using large heads, such as 40 mm. For such head sizes, small variations of the LWR may have large impacts on the osteolysis-free life of the THA.


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_8 | Pages 107 - 107
1 Apr 2017
Kunutsor S Whitehouse M Blom A Beswick A
Full Access

Background

The two-stage revision strategy has been claimed as being the “gold standard” for treating prosthetic joint infection. The one-stage revision strategy remains an attractive alternative option, however, its effectiveness in comparison to the two-stage strategy remains uncertain. A systematic review and meta-analysis was conducted to compare the effectiveness of one- and two-stage revision strategies to prevent re-infection after prosthetic hip infection.

Methods

Cohort studies (prospective or retrospective) conducted in unselected patients with infection treated exclusively by one- or two-stage revision and reporting re-infection outcomes within two years of revision were retrieved from MEDLINE, EMBASE, Web of Science, Cochrane databases, manual search of bibliographies to March 2015, and email contact with investigators. Data were extracted by two independent investigators and a consensus was reached with involvement of a third. Rates of re-infection were aggregated using random-effect models after arcsine transformation, and were grouped by study and population level characteristics.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVI | Pages 92 - 92
1 Aug 2012
Joyce T Langton D Lord J Nargol A
Full Access

Background

The worldwide withdrawal of the DePuy Articular Surface Replacement (ASR) device in both its resurfacing and total hip replacement (THR) form on 26 August 2010, after 93,000 were implanted worldwide, has had major implications. The 2010 National Joint Registry for England and Wales quoted figures of 12-13% failure at five years; however these figures may be an underestimate.

Patients and methods

In 2004 a single surgeon prospective study of the ASR bearing surface was undertaken. Presented are the Adverse Reaction to Metal Debris (ARMD) failure rates of the ASR resurfacing and ASR THR systems. The diagnosis of ARMD was made by the senior author and was based on clinical history, examination, ultrasound findings, metal ion analysis of blood and joint fluid, operative findings and histopathological analysis of tissues retrieved at revision. Acetabular cup position in vivo was determined using EBRA software. Mean follow up was 52 months (24-81) and 70 patients were beyond 6 years of the procedure at the time of writing. Kaplan Meier survival analysis was carried out firstly with joints designated ‘failure’ if the patient had undergone revision surgery or if the patient had been listed for revision. A second survival analysis was carried out with a failure defined as a serum cobalt concentration > 7microgrammes/L (MHRA guideline from MDA-2010-069). Full explant analysis was carried out for retrieved prostheses.


Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_11 | Pages 45 - 45
1 Jul 2014
Vanhegan I Coathup M McCarthy I Haddad F Blunn G
Full Access

Summary Statement

Proximal femoral bony deficits present a surgical and biomechanical challenge to implant longevity in revision hip arthroplasty. This work finds comparable primary stability when a distally fixing tapered fluted stem was compared with a conical design in cadaveric tests.

Introduction

Proximal bony deficits complicate revision hip surgery and compromise implant survival. Longer distally fixing stems which bypass such defects are therefore required to achieve stability compatible with bony ingrowth and implant longevity.


The Journal of Bone & Joint Surgery British Volume
Vol. 83-B, Issue 4 | Pages 598 - 603
1 May 2001
Nevelos JE Prudhommeaux F Hamadouche M Doyle C Ingham E Meunier A Nevelos AB Sedel L Fisher J

We compared and quantified the modes of failure and patterns of wear of 11 Mittelmeier and 11 Ceraver-Ostal retrieved alumina-alumina hip prostheses with reference to the corresponding clinical and radiological histories.

Macroscopic wear was assessed using a three-dimensional co-ordinate measuring machine. Talysurf contacting profilometry was used to measure surface roughness on a microscopic scale and SEM to determine mechanisms of wear at the submicron level.

The components were classified into one of three categories of wear: low (no visible/measurable wear), stripe (elliptical wear stripe on the heads and larger worn areas on the cups) and severe (macroscopic wear, large volumes of material lost). Overall, the volumetric wear of the alumina-alumina prostheses was substantially less than the widely used metal and ceramic-on-polyethylene combinations. By identifying and eliminating the factors which accelerate wear, it is expected that the lifetime of these devices can be further increased.


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_8 | Pages 96 - 96
11 Apr 2023
Crippa Orlandi N De Sensi A Cacioppo M Saviori M Giacchè T Cazzola A Mondanelli N Giannotti S
Full Access

The computational modelling and 3D technology are finding more and more applications in the medical field. Orthopedic surgery is one of the specialties that can benefit the most from this solution. Three case reports drawn from the experience of the authors’ Orthopedic Clinic are illustraded to highlight the benefits of applying this technology. Drawing on the extensive experience gained within the authors’ Operating Unit, three cases regarding different body segments have been selected to prove the importance of 3D technology in preoperative planning and during the surgery. A sternal transplant by allograft from a cryopreserved cadaver, the realization of a custom made implant of the glenoid component in a two-stage revision of a reverse shoulder arthroplasty, and a case of revision on a hip prosthesis with acetabular bone loss (Paprosky 3B) treated with custom system. In all cases the surgery was planned using 3D processing software and models of the affected bone segments, printed by 3D printer, and based on CT scans of the patients. The surgical implant was managed with dedicated instruments. The use of 3D technology can improve the results of orthopedic surgery in many ways: by optimizing the outcomes of the operation as it allows a preliminary study of the bone loss and an evalutation of feasibility of the surgery, it improves the precision of the positioning of the implant, especially in the context of severe deformity and bone loss, and it reduces the operating time; by improving surgeon training; by increasing patient involvement in decision making and informed consent. 3D technology, by offering targeted and customized solutions, is a valid tool to obtain the tailored care that every patient needs and deserves, also providing the surgeon with an important help in cases of great complexity


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_7 | Pages 110 - 110
4 Apr 2023
Ding Y Li S Li C Chen Z Wu C
Full Access

Total joint replacement (TJR) was one of the most revolutionary breakthroughs in joint surgery. The majority studies had shown that most implants could last about 25 years, anyway, there is still variation in the longevity of implants. In US, for all the hip revisions from 2012 to 2017 in the United States, 12.0% of the patients were diagnosed as aseptic loosening. Variable studies have showed that any factor that could cause a systemic or partial bone loss, might be the risk of periprosthetic osteolysis and aseptic loosening. Breast cancer is the most frequent malignancy in women, more than 2.1 million women were newly diagnosed with breast cancer, 626,679 women with breast cancer died in 2018. It's been reported that the mean incidence of THA was 0.29% for medicare population with breast cancer in USA, of which the incidence was 3.46% in Norwegian. However, the effects of breast cancer chemotherapy and hormonotherapy, such as aromatase inhibitors (AI), significantly increased the risk of osteoporosis, and had been proved to become a great threat to hip implants survival. In this case, a 46-year-old female undertook chemotherapy and hormonotherapy of breast cancer 3 years after her primary THA, was diagnosed with aseptic loosening of the hip prosthesis. Her treatment was summarized and analyzed. Breast cancer chemotherapy and hormonotherapy might be a threat to the stability of THA prosthesis. More attention should be paid when a THA paitent occurred with breast cancer. More studies about the effect of breast cancer treatments on skeleton are required


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_4 | Pages 124 - 124
1 Mar 2021
Jelsma J Schotanus M Kleinveld H Grimm B Heyligers I
Full Access

An increase in metal ion levels is seen after implantation of all MoM hip prosthesis due to release from the surface directly, more so during articulation and corrosion of the bearing surfaces. The bearing surfaces in MoM prosthesis consist of cobalt, chromium and molybdenum. Several case-reports of cobalt toxicity due to a MoM prosthesis have been published in the last decade. Cobalt intoxication may lead to a variety of symptoms: neuro-ocular toxicity (tinnitus, vertigo, deafness, blindness, convulsions, headaches and peripheral neuropathy), cardiotoxicity and thyroid toxicity. Nausea, anorexia and unexplained weight loss have been described. Systemic effects from metal ions even with well functioning implants or with ion concentrations lower than those associated with known adverse effects may exist and warrant investigation. The aim of this study is to investigate self-reported systemic complaints in association with cobalt ion concentrations in patients with any type of MoM hip prosthesis. A cohort study was conducted. Patients with both unilateral and bilateral, resurfacing and large head metal on metal total hip arthroplasties were included for the current study. Blood metal ion concentrations (cobalt and chromium) were measured by inductively coupled plasma mass spectrometry (ICP-MS). Based on the known cobalt toxicity symptoms of case-reports and toxicology reports a new non-validated questionnaire was developed. questions were subdivided in general questions/symptoms, vestibular symptoms, neurological symptoms, emotional health and cardio- and thyroid toxicity symptoms. Independent samples T test, Fishers Exact Test and Pearsons (R) correlation were used. Analysis was performed on two groups; a low cobalt ion concentration group and a high cobalt ion concentration group A total of 62 patients, 36 (58%) men and 26 (42%) women, were included with a mean age at surgery of 60.8 ± 9.3 years (41.6 – 78.1) and a mean follow up of 6.3 ± 1.4years (3.7 – 9.6). In these patients a total of 71 prosthesis were implanted: 53 unilateral and 9 bilateral. Of these, 44 were resurfacing and 27 large head metal on metal (LHMoM) total hip arthroplasties. Mean cobalt and chromium ion concentrations were 104 ± 141 nmol/L (9 – 833) and 95 ± 130nmol/L (6 – 592), respectively. Based on the different thresholds (120 – 170 or 220 nmol/L) the low cobalt ion concentration group consisted of 44 (71%), 51 (82%) or 55 (89%) subjects respectively. No differences were found in general characteristics, independently of the threshold. The composite score of vestibular symptoms (vision, hearing, tinnitus, dizziness) was significantly higher (p < .050) in all high cobalt ion concentrations groups, independent of the threshold value This study aimed to detect a trend in self-reported systemic complaints in patients with metal-on-metal hip arthroplasty due to raised cobalt ion concentrations. Vestibular symptoms were more common in high cobalt ion concentration groups independent of the three threshold levels tested. The upper limit of acceptable cobalt ion concentrations remains uncertain. With regards to proactively inquired, self-reported symptoms the threshold where effects may be present could be lower than values currently applied in clinical follow-up. It is unknown what exposure to elevated metal ion concentrations for a longer period of time causes with aging subjects. Further research with a larger cohort and a more standardized questionnaire is necessary to detect previously undiscovered or under-reported effects


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_4 | Pages 10 - 10
1 Apr 2018
Wesseling M Vancleef S Meyer C Vander Sloten J Jonkers I
Full Access

Introduction. Modification in joint loading, and specifically shear stress, is found to be an important mechanical factor in the development of osteoarthritis (OA). Cartilage shear stresses can be investigated using finite element (FE) modelling, where typically in vivo joint loading as measured by an instrumented hip prosthesis is used as boundary condition. However, subject-specific gait characteristics substantially affect joint loading. The goal of this study is to investigate the effect of subject-specific joint loading as calculated using a subject-specific musculoskeletal model and integrated motion capture data on acetabular shear stress. Methods. Three healthy control subjects walked at self-selected speed while measuring marker trajectories (Vicon, Oxford Metrics, UK) and force data (two AMTI force platforms; Watertown, MA). A subject-specific MRI-based musculoskeletal model consisting of 14 segments, 19 degrees of freedom and 88 musculotendon actuators, and including wrapping surfaces around the hip joint, was used. All analyses were performed in OpenSim 3.1. The model was scaled to the dimensions of each subject using the marker positions of a static pose. A kalman smoother procedure was used to calculate joint angles. Muscle forces were calculated using static optimization, minimizing the sum of squared muscle activations, and hip contact forces (HCF) were calculated and normalized to body weight (BW). To calculate shear stress, HCFs and joint angles calculated during the stance phase of gait were imposed to a hip finite element model (hip_n10rb) using FFEbio 2.5. In the model, femoral and acetabular cartilage were represented using the Mooney-Rivlin formulation (c1=6.817, bulk modulus=1358.86) and the pelvis and femur bones as rigid bodies. Peak HCF as well as maximal acetabular shear stress, magnitude and location, and the HCF at the time of maximal shear stress were compared between subjects. Results. Maximal shear stress was lower for S3 compared to S1 and S2 (9.14, 9.48 and 7.14 MPa for S1, S2 and S3 respectively). Nevertheless, HCF at the time instance of peak stress as well as peak HCF were highest for S3 (S1: 2.40/4.54 BW, S2: 2.97/4.78 BW and S3: 3.13/6.46 BW respectively). Maximal shear stress also occurred earlier in the stance phase for S3 compared to S1 and S2 (31, 26 and 11% of the stance phase for S1, S2 and S3 respectively). In addition, the location of the peak maximal shear stress was found to be more superior for S3. Discussion. Subject-specific loading patterns clearly influence the calculated maximal shear stress in the acetabular cartilage, affecting both the magnitude and the location of the stress. In addition, higher shear stresses are not coinciding with higher HCFs. This finding highlights the need of subject-specific rather than generic loading patterns when assessing cartilage shear stresses and associated risk in OA development in individual patients


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_8 | Pages 47 - 47
1 Apr 2017
Jelsma J Senden R Schotanus M Kort N Grimm B Heyligers I
Full Access

Background. The second generation metal-on-metal (MoM) prosthesis of the hip became a worldwide success in the 90s. However, after the placement of a MoM prosthesis the cobalt ion concentrations raise significantly. This may lead to systemic complaints and even cobalt toxicity. Methods. Sixty-one patients (26F/35M) with both an unilateral and bilateral resurfacing or large-head MoM (LHMoM) hip prosthesis were included. At last follow-up (5.77 ± 1.57 yrs) cobalt concentrations in the blood were determined by ICP-MS. Based on the known cobalt toxicity symptoms we developed a non-validated questionnaire. Analysis was done on two groups; a low cobalt concentration group and a high cobalt concentration group. We used 170 nmol/L as the upper limit of well functioning prosthesis as defined by the Dutch Orthopaedic Society (NOV). Independent samples T test and Pearson correlation coefficient were done. Results. In 61 patients, 70 prosthesis were implanted: 44 resurfacings and 26 LHMoM. Mean cobalt and chromium ion concentrations were 110 ± 138 (9 – 833) and 104 ± 138 (6 – 592) nmol/L. The low cobalt concentration group consisted of 52 patients, while the high cobalt concentration group consisted of 9 hips. With respect to the systemic complaints, the subjects in the high or low group reported no significant differences. A trend towards more complaints with visual problems (75% vs 51.9%, P = .089) and dizziness (62.5% vs 38.9%, P = .098) was notable in the high cobalt concentration group. Conclusions. Metal on metal hip arthroplasty stays controversial for the possible side-effects. This study shows trends, but no significance, in systemic complaints from high cobalt ion concentrations. Complaints of neuro-ocular toxicity are most common. This is in correspondence with other literature, although the upper limit for cobalt ion concentrations remains uncertain. Level of Evidence. 2b. Disclosure. no financial or nonfinancial relationships to disclose


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_9 | Pages 56 - 56
1 May 2017
Jelsma J Senden R Schotanus M Kort N Heyligers I Grimm B
Full Access

Background. Metal-on-metal hip implants can produce adverse tissue reactions to wear debris. Increased metal ion concentrations in the blood are measured as a proxy to wear and the complications it can trigger. Many studies have examined various factors influencing the metal ion concentrations. This is the first study to investigate the effect of physical activity level, as objectively measured in daily life, on blood ion levels, expecting higher concentrations for higher patient activity. Methods. Thirty-three patients (13F/20M, 55.8 ± 6.2 years at surgery) with a unilateral resurfacing hip prosthesis were included. At last follow-up (6.8 ± 1.5 years) cobalt and chromium concentrations in the blood were determined by inductively coupled plasma mass spectrometry. Physical activity was measured during 4 successive days using a 3D-acceleration-based activity monitor. Data was analysed using validated algorithms, producing quantitative and qualitative parameters. Acetabular cup position was measured radiographically. Correlations were tested with Pearson's r'. Results. Wide ranges in cobalt (10-833nmol/l) and chromium (16-592nmol/l) concentrations were measured. No correlations were found between metal ion concentrations and patient characteristics. The mean time of walking per day of was 5475 ± 2730 seconds (≈ 91 minutes) per day, with 6953 ± 3104 steps made during the day. The cadence, the number of steps per minute, was 99.5 ± 7.4. The subjects performed 41 ± 13 sit-stand transfers per day. The number of peak intensity declines when peaks where more intense. No correlations were found between quantitative or qualitative parameters and metal ion concentrations. Conclusions. Higher metal ion concentrations were not correlated with higher activity levels contrary to original expectations. Based on the results the orthopaedic surgeons is able to reassure his or her patients that physical activity in daily living is presumably safe en does not influence metal ion concentrations. Level of Evidence. 2b. Approval. the ethics committee of the Atrium MC, Heerlen approved this study


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_8 | Pages 91 - 91
1 Apr 2017
Lerf R Reimelt I Dallmann F Delfosse D
Full Access

Background. When reversing the hard-soft articulation in inverse shoulder replacement, i.e. hard inlay and soft glenosphere, the tribological behaviour of such a pairing has to be tested thoroughly. Therefore, two hard materials for the inlay, CoCr alloy and alumina toughened zirconia ceramic (ATZ) articulating on two soft materials, conventional UHMWPE and vitamin E stabilised, highly cross-linked PE (E-XLPE) were tested. Methods. The simulator tests were performed analogue to standardised gravimetric wear tests for hip prosthesis (ISO 14242-1) with load and motion curves adapted to the shoulder. The test parameters differing from the standard were the maximum force (1.0 kN) and the range of motion. A servo-hydraulic six station joint simulator (EndoLab, Rosenheim) was used to run the tests up to 5 times 106 cycles with diluted calf serum at 37° C as lubricant. Results. The wear rates measured in the simulator when the CoCr alloy inlay articulated on UHMWPE and E-XLPE were respectively 32.50 +/− 3.48 mg/Mcycle and 10.65 +/− 2.36 mg/Mcycle. In comparison, when the ATZ inlay articulated on UHMWPE and E-XLPE the wear rates were 20.34 +/− 1.14 mg/Mcycle and 5.99 +/− 0.79 mg/Mcycle respectively. Conclusions. The simulator wear rate of the standard articulation CoCr – UHMWPE is similar to that found in the corresponding pairing for hip endoprosthesis. Replacing UHMWPE by E-XLPE, the wear rate is reduced to about 1/3 for both hard counterparts, CoCr and ZTA, respectively. Replacing the CoCr inlay by a part made from ZTA lowers wear by about 37 % in articulation against UHMWPE and about 44 % against E-XLPE. The lowest wear rate, with a reduction of about 80 % compared to the standard CoCr – UHMWPE, exhibits the pairing of both advanced materials, ZTA – E-XLPE. However, long-term clinical follow-up will confirm if this in-vitro wear reduction leads to longer in-vivo survival. Level of evidence. Laboratory test on sample implants. Study financed by Mathys Orthopaedie GmbH


Bone & Joint Research
Vol. 5, Issue 10 | Pages 461 - 469
1 Oct 2016
Liu YK Deng XX Yang H

Objectives. The cytotoxicity induced by cobalt ions (Co. 2+. ) and cobalt nanoparticles (Co-NPs) which released following the insertion of a total hip prosthesis, has been reported. However, little is known about the underlying mechanisms. In this study, we investigate the toxic effect of Co. 2+. and Co-NPs on liver cells, and explain further the potential mechanisms. Methods. Co-NPs were characterised for size, shape, elemental analysis, and hydrodynamic diameter, and were assessed by Transmission Electron Microscope, Scanning Electron Microscope, Energy Dispersive X-ray Spectroscopy and Dynamic Light Scattering. BRL-3A cells were used in this study. Cytotoxicity was evaluated by MTT and lactate dehydrogenase release assay. In order to clarify the potential mechanisms, reactive oxygen species, Bax/Bcl-2 mRNA expression, IL-8 mRNA expression and DNA damage were assessed on BRL-3A cells after Co. 2+. or Co-NPs treatment. Results. Results showed cytotoxic effects of Co. 2+. and Co-NPs were dependent upon time and dosage, and the cytotoxicity of Co-NPs was greater than that of Co. 2+. In addition, Co-NPs elicited a significant (p < 0.05) reduction in cell viability with a concomitant increase in lactic dehydrogenase release, reactive oxygen species generation, IL-8 mRNA expression, Bax/Bcl-2 mRNA expression and DNA damage after 24 hours of exposure. Conclusion. Co-NPs induced greater cytotoxicity and genotoxicity in BRL-3A cells than Co. 2+. Cell membrane damage, oxidative stress, immune inflammation and DNA damage may play an important role in the effects of Co-NPs on liver cells. Cite this article: Y. K. Liu, X. X. Deng, H.L. Yang. Cytotoxicity and genotoxicity in liver cells induced by cobalt nanoparticles and ions. Bone Joint Res 2016;5:461–469. DOI: 10.1302/2046-3758.510.BJR-2016-0016.R1


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_16 | Pages 101 - 101
1 Nov 2018
George A Ellis M Gill R
Full Access

It is well documented that implant loosening rate in sickle cell disease patients is higher than that seen in patients with hip arthroplasty from other indications. The Hypoxic inducible factor(HIF) - is activated in the microcellular hypoxic environment and this through a cascade of other enzymatic reactions promotes the activity of other factors and further help enhance angiogenesis and osteogenesis. The aim of this study was to investigate and propose a potential model for investigating osseointegration in a hypoxic microcellular environment using osteoblasts(MG63). Human MG63 osteoblastic cells were cultured under normoxia and hypoxic conditions (20%; and 1% oxygen saturation) for 72 hours under two different condition- with and without cobalt chloride. The samples cultured under normoxic condtions without cobalt chloride acted as control. Using qualitative polymerase chain reaction-(qPCR) - HIF expression was assessed under the above conditions in relation to the control. The results showed there was significant expression of the HIF 1 alpha protein under hypoxic condition with cobalt chloride in comparison with the control samples- all at 72hours incubation. Mann-Whitney U test was used to deduce level of significance of fold change.(p=0.002; <0.05). This was deemed as being a significant difference in the level of expression of HIF compared to the control. The results show that the hypoxic inducible factor can be expressed using the above tested. experimental invitro-model with significant results which can be a foundation for further research into improving hip implant prosthesis design to help enhance osseo-integration in sickle cell disease patient with AVN


Bone & Joint Research
Vol. 9, Issue 3 | Pages 146 - 151
1 Mar 2020
Waldstein W Koller U Springer B Kolbitsch P Brodner W Windhager R Lass R

Aims

Second-generation metal-on-metal (MoM) articulations in total hip arthroplasty (THA) were introduced in order to reduce wear-related complications. The current study reports on the serum cobalt levels and the clinical outcome at a minimum of 20 years following THA with a MoM (Metasul) or a ceramic-on-polyethylene (CoP) bearing.

Methods

The present study provides an update of a previously published prospective randomized controlled study, evaluating the serum cobalt levels of a consecutive cohort of 100 patients following THA with a MoM or a CoP articulation. A total of 31 patients were available for clinical and radiological follow-up examination. After exclusion of 11 patients because of other cobalt-containing implants, 20 patients (MoM (n = 11); CoP (n = 9)) with a mean age of 69 years (42 to 97) were analyzed. Serum cobalt levels were compared to serum cobalt levels five years out of surgery.


The Journal of Bone & Joint Surgery British Volume
Vol. 85-B, Issue 8 | Pages 1202 - 1206
1 Nov 2003
Fiorito S Magrini L Goalard C

We investigated the circulating levels of the main cytokines involved in bone resorption (IL-1β, IL-6, TNF-α), prostaglandins (PGE. 2. ) and metalloproteases (MMP-1), as possible early markers of osteolysis, in the serum of eight patients with periprosthetic osteolysis and ten patients without osteolysis. All had received a cementless hip prosthesis (ABG-1). We also assessed the serum levels of IL-11 and TGF-β anti-inflammatory cytokines exerting protective effect on bone resorption. The mean serum levels of IL-1β, IL-6, TNF-α, TGF-β, MMP-1, and PGE. 2. in patients with periprosthetic osteolysis did not differ significantly from those of patients without osteolysis or from those of normal controls. IL-11 serum levels were not detectable at all in any of the patients, while they were detected within normal reference values in the control subjects (significant inverse correlation). We believe that circulating cytokines cannot be regarded as markers of osteolysis, a condition characterised by a local inflammation without systemic signs of inflammation. On the contrary, the undetectable levels of IL-11 in implanted patients could provide evidence for a lack of balance between pro- and anti-inflammatory cytokines in these patients


The Journal of Bone & Joint Surgery British Volume
Vol. 81-B, Issue 4 | Pages 705 - 709
1 Jul 1999
Hara T Hayashi K Nakashima Y Kanemaru T Iwamoto Y

We have studied the effect of hydroxyapatite (HA) coating in 15 ovariectomised and 15 normal rats which had had a sham procedure. Twenty-four weeks after operation, HA-coated implants were inserted into the intramedullary canal of the right femur and uncoated implants into the left femur. The prostheses were removed four weeks after implantation. Twelve specimens in each group had mechanical push-out tests. Sagittal sections of the other three were evaluated by SEM. The bone mineral density (BMD) of the dissected left tibia was measured by dual-energy x-ray absorptiometry. The difference in BMD between the control and ovariectomised tibiae was 35.01 mg/cm. 2. (95% CI, 26.60 to 43.42). The push-out strength of the HA-coated implants was higher than that of the uncoated implants in both groups (p < 0.0001), but the HA-coated implants of the ovariectomised group had a reduction in push-out strength of 40.3% compared with the control group (p < 0.0001). Our findings suggest that HA-coated implants may improve the fixation of a cementless total hip prosthesis but that the presence of osteoporosis may limit the magnitude of this benefit


The Journal of Bone & Joint Surgery British Volume
Vol. 80-B, Issue 5 | Pages 912 - 917
1 Sep 1998
Granchi D Verri E Ciapetti G Stea S Savarino L Sudanese A Mieti M Rotini R Dallari D Zinghi G Montanaro L

Our aim was to determine if the serum levels of bone-resorbing cytokines (IL-1β, TNF-α, IL-6, GM-CSF) are altered in patients with aseptic loosening of a total hip prosthesis, and if such levels are influenced by the type of implant. We determined cytokine levels in sera from 35 patients before revision for failed total hip arthroplasty and compared them with those in 25 healthy donors. We also assessed the soluble receptor of interleukin-2 (sIL-2r) in serum as an indication of a specific immune reaction against the implant. Our findings showed that the sIL-2r and TNF-α serum level did not change. The IL-6 level was not significantly altered, but was higher in patients with TiAlV prostheses than in those with a CrCoMo implant and in patients with cemented prostheses. The IL-1β level was found to be higher in those with a TiAlV cemented prosthesis than in the control group (p = 0.0001) and other groups of patients (p = 0.003 v uncemented TiAlV, p = 0.01 v cemented CrCoMo, p = 0.001 v uncemented CrCoMo). The GM-CSF level significantly increased in patients compared with healthy subjects (p = 0.008), and it was higher in those with cemented than with uncemented implants (p = 0.01). Only patients with cementless CrCoMo prostheses had levels of GM-CSF similar to those of the control group. The highest GM-CSF concentrations were observed in patients treated with non-steroidal anti-inflammatory drugs (NSAIDs) in the last months before revision (p = 0.04). In addition, when massive osteolysis was observed, the level of GM-CSF tended to decrease to that of the control group