The eccentric glenosphere was principally introduced into reverse
shoulder arthroplasty to reduce the incidence of scapular notching.
There is only limited information about the influence of its design
on deltoid power and joint reaction forces. The aim of our study was to investigate how the diameter and
eccentricity of the glenosphere affect the biomechanics of the deltoid
and the resultant joint reaction forces. Different sizes of glenosphere and eccentricity were serially
tested in ten cadaveric shoulders using a custom shoulder movement
simulator.Aims
Methods
Scapular notching is a frequently observed radiographic phenomenon
in reverse shoulder arthroplasty (RSA), signifying impingement of
components. The purposes of this study were to evaluate the effect
of glenoid component size and glenosphere type on impingement-free
range of movement (ROM) for extension and internal and external
rotation in a virtual RSA model, and to determine the optimal configuration
to reduce the incidence of friction-type scapular notching. Preoperative CT scans obtained in 21 patients (three male, 18
female) with primary osteoarthritis were analyzed using modelling
software. Two concurrent factors were tested for impingement-free
ROM and translation of the centre of rotation: glenosphere diameter
(36 mm Aims
Materials and Methods
Introduction. Reverse Shoulder Arthroplasty (RSA) is recognized to be an effective solution for rotator cuff deficient arthritic shoulders, but there are still concerns about impingement and range of motion (ROM). Several RSA biomechanical studies have shown that humeral lateralization can increase ROM in planar motions (e.g. abduction). However, there is still a debate whether humeral lateralization should be achieved with a larger sphere diameter or by lateralizing the center of rotation (COR). The latter has shown to decrease the deltoid moment arm and increase shear forces, where the former may pose challenges in implanting the device in small patients. The aim of this study was to evaluate how humeral lateralization achieved by varying COR lateral offset and
There are a variety of sizes currently available for reverse total shoulder arthroplasty (RTSA) implant systems. Common sizing options include a smaller 36 to 38 mm or a larger 40 to 42 mm glenosphere, and are typically selected based on surgeon preference or patient size. Previous studies have only evaluated the abduction and adduction range of motion within a single plane of elevation, providing a limited view of the joint's possible range of motion. The purpose of this study was to use computer modeling to evaluate the abduction and adduction range of motion across multiple planes of elevation for a range of glenosphere sizes. Computed tomography images of four cadaveric specimens (age: 54 ± 24 years) were used to obtain the osseous anatomy to be utilised in the model. Solid-body motion studies of the RTSA models were constructed with varying
Scapular notching is a common problem following reverse shoulder arthroplasty (RSA). This is due to impingement between the humeral polyethylene cup and scapular neck in adduction and external rotation. Various glenoid component strategies have been described to combat scapular notching and enhance impingement-free range of motion (ROM). There is limited data available detailing optimal glenosphere position in RSA with an onlay configuration. The purpose of this study was to determine which glenosphere configurations would maximise impingement free ROM using an onlay RSA prosthesis. A three-dimensional (3D) computed tomography (CT) scan of a shoulder with Walch A1, Favard E0 glenoid morphology was segmented using validated software. An onlay RSA prosthesis was implanted and a computer model simulated external rotation and adduction motion of the virtual RSA prosthesis. Four glenosphere parameters were tested; diameter (36mm, 41mm), lateralization (0mm, 3mm, 6mm), inferior tilt (neutral, 5 degrees, 10 degrees), and inferior eccentric positioning (0.5mm, 1.5mm. 2.5mm, 3.5mm, 4.5mm). Eighty-four combinations were simulated. For each simulation, the humeral neck-shaft angle was 147 degrees and retroversion was 30 degrees. The largest increase in impingement-free range of motion resulted from increasing inferior eccentric positioning, gaining 15.0 degrees for external rotation and 18.8 degrees for adduction. Glenosphere lateralization increased external rotation motion by 13. 6 degrees and adduction by 4.3 degrees. Implanting larger
Is there an optimal way to place a glenoid in reverse total shoulder arthroplasty (RTSA)? Four of the six parameters that a surgeon can control in a RTSA involve the glenoid. The parameters are: inferior tilt, increased lateral or inferior offset and increased
Introduction. Acromial and scapular fractures are a rare but difficult complication with reverse total shoulder arthroplasty (rTSA), with an incidence rate reported from 1–10%. The risk factors associated with these fractures types is largely unknown. The goal of this study is to analyze the clinical outcomes, demographic and comorbidity data, and implant sizing and surgical technique information from 4125 patients who received a primary rTSA with one specific prosthesis (Equinoxe, Exactech, Inc) and were sorted based on the radiographic documentation of an acromial and/or scapula fracture (ASF) to identify factors associated with this complication. Methods. 4125 patients (2652F/1441M/32 unspecified; mean age: 72.5yrs) were treated with primary rTSA by 23 orthopaedic surgeons. Revision and fracture reverse arthroplasty cases were excluded. The radiographic presence of each fracture was documented and classified using the Levy classification method. 61 patients were identified as having ASF, 10 patients had fractures of the Type 1, 32 patients had Type 2, and 18 patients had Type 3 fractures according to Levy's classification. One fracture was not classifiable. Pre-op and post-op outcome scoring, ROM as well as demographic, comorbidity, implant, and surgical technique information were evaluated for these 61 patients and compared to the larger cohort of patients to identify any associations. A two-tailed, unpaired t-test identified differences (p<0.05). Results. The overall rate of ASF was 1.48% with the average time after surgery occurring at 12.9 ± 17.9 months (range 1 day to 78 months). Men had an ASF rate of 0.69% (10 of 1441); whereas women had a rate of 1.92% (51 of 2652). Patients with ASF were observed to be significantly shorter than patients without ASF (65.1 in vs 63.3 in, p=0.0004). ASF were more common in females (p=0.0019), have Rheumatoid Arthritis (p=0.0051), Cuff Tear Arthropathy (p=0.0093), or previous shoulder surgery (p=0.0189). Patient's weight did not correlate, nor did BMI. No difference was observed in humeral stem size,
The current evidence comparing the two most common approaches for reverse total shoulder arthroplasty (rTSA), the deltopectoral and anterosuperior approach, is limited. This study aims to compare the rate of loosening, instability, and implant survival between the two approaches for rTSA using data from the Dutch National Arthroplasty Registry with a minimum follow-up of five years. All patients in the registry who underwent a primary rTSA between January 2014 and December 2016 using an anterosuperior or deltopectoral approach were included, with a minimum follow-up of five years. Cox and logistic regression models were used to assess the association between the approach and the implant survival, instability, and glenoid loosening, independent of confounders.Aims
Methods
Introduction:. Given that factors like center of rotation (COR), neck shaft angle,
The liner design is a key determinant of the constraint of a reverse total shoulder arthroplasty (rTSA). The aim of this study was to compare the degree of constraint of rTSA liners between different implant systems. An implant company’s independent 3D shoulder arthroplasty planning software (mediCAD 3D shoulder v. 7.0, module v. 2.1.84.173.43) was used to determine the jump height of standard and constrained liners of different sizes (radius of curvature) of all available companies. The obtained parameters were used to calculate the stability ratio (degree of constraint) and angle of coverage (degree of glenosphere coverage by liner) of the different systems. Measurements were independently performed by two raters, and intraclass correlation coefficients were calculated to perform a reliability analysis. Additionally, measurements were compared with parameters provided by the companies themselves, when available, to ensure validity of the software-derived measurements.Aims
Methods
We report the use of the reverse shoulder prosthesis in the revision of a failed shoulder hemiarthroplasty in 19 shoulders in 18 patients (7 men, 11 women) with severe pain and loss of function. The primary procedure had been undertaken for glenohumeral arthritis associated with severe rotator cuff deficiency. Statistically significant improvements were seen in pain and functional outcome. After a mean follow-up of 44 months (24 to 89), mean forward flexion improved by 26.4° and mean abduction improved by 35°. There were six prosthesis-related complications in six shoulders (32%), five of which had severe bone loss of the glenoid, proximal humerus or both. Three shoulders (16%) had non-prosthesis related complications. The use of the reverse shoulder prosthesis provides improvement in pain and function for patients with failure of a hemiarthroplasty for glenohumeral arthritis and rotator cuff deficiency. However, high rates of complications were associated with glenoid and proximal humeral bone loss.