Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_18 | Pages 116 - 116
14 Nov 2024
Varga P Cameron P Hutchinson D Malkoch M Schwarzenberg P
Full Access

Introduction

When designing a new osteosynthesis device, the biomechanical competence must be evaluated with respect to the acting loads. In a previous study, the loads on the proximal phalanx during rehabilitation exercises were calculated. This study aimed to assess the safety of a novel customizable osteosynthesis device compared to those loads to determine when failure would occur.

Method

Forty proximal phalanges were dissected from skeletally mature female sheep and divided into four testing groups. A custom 3D printed cutting and drilling guide was used to create a reduced osteotomy and pilot holes to insert four 1.5 mm cortical screws. A novel light-curable polymer composite was used to fixate the bones with an in situ fixation patch. The constructs were tested in cyclic four-point bending in a bioreactor with ringer solution at 37°C with a valley load of 2 N. Four groups (N = 10) had increasing peak loads based on varying safety factors relative to the physiological loading (G1:100x, G2:150x, G3:175x, G4:250x). Each specimen was tested for 12,600 cycles (6 weeks of rehabilitation) or until failure occurred. After the test the thickness of the patch was measured with digital calipers and data analysis was performed in Python and R.


The Journal of Bone & Joint Surgery British Volume
Vol. 89-B, Issue 11 | Pages 1545 - 1550
1 Nov 2007
Koslowsky TC Mader K Dargel J Koebke J Hellmich M Pennig D

We have evaluated four different fixation techniques for the reconstruction of a standard Mason type-III fracture of the radial head in a sawbone model. The outcome measurements were the quality of the reduction, and stability. A total of 96 fractures was created. Six surgeons were involved in the study and each reconstructed 16 fractures with 1.6 mm fine-threaded wires (Fragment Fixation System (FFS)), T-miniplates, 2 mm miniscrews and 2 mm Kirschner (K-) wires; four fractures being allocated to each method using a standard reconstruction procedure. The quality of the reduction was measured after definitive fixation. Biomechanical testing was performed using a transverse plane shear load in two directions to the implants (parallel and perpendicular) with respect to ultimate failure load and displacement at 50 N. A significantly better quality of reduction was achieved using the FFS wires (Tukey’s post hoc tests, p < 0.001) than with the other devices with a mean step in the articular surface and the radial neck of 1.04 mm (. sd. 0.96) for the FFS, 4.25 mm (. sd. 1.29) for the miniplates, 2.21 mm (. sd. 1.06) for the miniscrews and 2.54 mm (. sd. 0.98) for the K-wires. The quality of reduction was similar for K-wires and miniscrews, but poor for miniplates. The ultimate failure load was similar for the FFS wires (parallel, 196.8 N (. sd. 46.8), perpendicular, 212.5 N (. sd. 25.6)), miniscrews (parallel, 211.8 N (. sd. 47.9), perpendicular, 208.0 N (. sd. 65.9)) and K-wires (parallel, 200.4 N (. sd. 54.5), perpendicular, 165.2 N (. sd. 37.9)), but significantly worse (Tukey’s post hoc tests, p < 0.001) for the miniplates (parallel, 101.6 N (. sd. 43.1), perpendicular, 122.7 N (. sd. 40.7)). There was a significant difference in the displacement at 50 N for the miniplate (parallel, 4.8 mm (. sd. 2.8), perpendicular, 4.8 mm (. sd. 1.7)) vs FFS (parallel, 2.1 mm (. sd. 0.8), perpendicular, 1.9 mm (. sd. 0.7)), miniscrews (parallel, 1.8 mm (. sd. 0.5), perpendicular, 2.3 mm (. sd. 0.8)) and K-wires (parallel, 2.2 mm (. sd. 1.8), perpendicular, 2.4 mm (. sd. 0.7; Tukey’s post hoc tests, p < 0.001)). The fixation of a standard Mason type-III fracture in a sawbone model using the FFS system provides a better quality of reduction than that when using conventional techniques. There was a significantly better stability using FFS implants, miniscrews and K-wires than when using miniplates


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_16 | Pages 1 - 1
1 Nov 2018
Warschawski Y Factor S Frenkel T Tudor A Steinberg E Snir N
Full Access

In Displaced Intracapsular Hip Fractures (ICHF) in young active patients, preservation of the femoral head and its blood supply are of high importance and urgent surgical treatment with anatomic reduction and internal fixation is the preferred intervention. Due to the strong varus displacement shear forces exerted across the hip, there are relatively high complication rates after fixation. There is no consensus regarding the optimal fixation device or technique. This retrospective study compared closed reduction internal fixation method using cannulated cancellous screw (CCS) with the Targon Femoral Neck (TFN) hip fixed angle screw. Data regarding, gender, operational data, duration of surgery, complications, NAS (Numerical Analogue Scale) pain score, Modified Harris Hip Score (MHHS) and SF-12 scores were retrieved for patients younger than 65 with displaced ICHF. Eighty-two patients were included in the study, 30 patients treated with CCS were compared to 52 patients treated with TFN. Fracture configuration (Garden and Pauwel classifications), mean time to surgery and complication rate did not differ significantly. Operative time did differ significantly between groups (CCS 56 minutes, TFN 92 minutes, p<0.001). At final follow-up the CCS group reported less pain (NAS 2.3 vs 3.5, p< 0.049) and better Mental Health Composite score of SF-12 (p=0.017) compared to the TFN group. Complication rates for the treatment of displaced ICHF with TFN and CCS showed no significant differences; however, the functional outcomes, as presented by the NAS and Mental Health Composite score of SF-12, showed superiority for CCS treatment. As this fixation method is related to reduce costs, we suggest CCS for the treatment of displaced ICHF in the young population.


Bone & Joint 360
Vol. 14, Issue 1 | Pages 48 - 49
1 Feb 2025
Marson BA

In this edition of Cochrane Corner, two updated reviews have been published in the Cochrane database. The first is a very broad review of rehabilitation strategies for patients with ankle fractures and includes early mobilization, the use of removable splints, and physical therapy techniques for patients treated with and without surgery. The second evaluates interventions for fixing tibial plateau fractures, including fixation techniques and bone graft substitutes


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_13 | Pages 69 - 69
1 Nov 2021
Pastor T Zderic I Richards G Gueorguiev B Knobe M
Full Access

Introduction and Objective. Distal femoral fractures are commonly treated with a straight plate fixed to the lateral aspects of both proximal and distal fragments. However, the lateral approach may not always be desirable due to persisting soft-tissue or additional vascular injury necessitating a medial approach. These problems may be overcome by pre-contouring the plate in helically shaped fashion, allowing its distal part to be fixed to the medial aspect of the femoral condyle. The objective of this study was to investigate the biomechanical competence of medial femoral helical plating versus conventional straight lateral plating in an artificial distal femoral fracture model. Materials and Methods. Twelve left artificial femora were instrumented with a 15-hole Locking Compression Plate – Distal Femur (LCP-DF) plate, using either conventional lateral plating technique with the plate left non-contoured, or the medial helical plating technique by pre-contouring the plate to a 180° helical shape and fixing its distal end to the medial femoral condyle (n=6). An unstable extraarticular distal femoral fracture was subsequently simulated by means of an osteotomy gap. All specimens were tested under quasi-static and progressively increasing cyclic axial und torsional loading until failure. Interfragmentary movements were monitored by means of optical motion tracking. Results. Initial axial stiffness was significantly higher for helical (185.6±50.1 N/mm) versus straight (56.0±14.4) plating, p<0.01. However, initial torsional stiffness in internal and external rotation remained not significantly different between the two fixation techniques (helical plating:1.59±0.17 Nm/° and 1.52±0.13 Nm/°; straight plating: 1.50±0.12 Nm/° and 1.43±0.13Nm/°), p≥0.21. Helical plating was associated with significantly higher initial interfragmentary movements under 500 N static compression compared to straight plating in terms of flexion (2.76±1.02° versus 0.87±0.77°) and shear displacement under 6 Nm static rotation in internal (1.23±0.28° versus 0.40±0.42°) and external (1.21±0.40° versus 0.57±0.33°) rotation, p≤0.01. In addition, helical plating demonstrated significantly lower initial varus/valgus deformation than straight plating (4.08±1.49° versus 6.60±0.47°), p<0.01. Within the first 10000 cycles of dynamic loading, helical plating revealed significantly bigger flexural movements and significantly lower varus/valgus deformation versus straight plating, p=0.02. No significant differences were observed between the two fixation techniques in terms of axial and shear displacement, p≥0.76. Cycles to failure was significantly higher for helical plating (13752±1518) compared to straight plating (9727±836), p<0.01. Conclusions. Although helical plating using a pre-contoured LCP-DF was associated with higher shear and flexion movements, it demonstrated improved initial axial stability and resistance against varus/valgus deformation compared to straight lateral plating. Moreover, helical plate constructs demonstrated significantly improved endurance to failure, which may be attributed to the less progressively increasing lever bending moment arm inherent to this novel fixation technique. From a biomechanical perspective, helical plating may be considered as a valid alternative fixation technique to standard straight lateral plating of unstable distal femoral fractures


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_1 | Pages 77 - 77
2 Jan 2024
Gueorguiev B Varga P
Full Access

Intramedullary nails (IMNs) are the current gold standard for treatment of long bone diaphyseal and selected metaphyseal fractures. Their design has undergone many revisions to improve fixation techniques, conform to the bone shape with appropriate anatomic fit, reduce operative time and radiation exposure, and extend the indication of the same implant for treatment of different fracture types with minimal soft tissue irritation. The IMNs are made or either titanium alloy or stainless steel and work as load-sharing internal splints along the long bone, usually accommodating locking elements – screws and blades, often featuring angular stability and offering different configurations for multiplanar fixation – to secure secondary fracture healing with callus formation in a relative-stability environment. Bone cement augmentation of the locking elements can modulate the construct stiffness, increase the surface area at the bone-implant interface, and prevent cut-through of the locking elements. The functional requirements of IMNs are related to maintaining fracture reduction in terms of length, alignment and rotation to enhance fracture healing. The load distribution during patient's activities is along the entire bone-nail interface, with nail length and anatomic fit being important factors to avoid stress risers


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_11 | Pages 124 - 124
1 Dec 2020
CETIN M SOYLEMEZ MS OZTURK BY MUTLU I KARAKUS O
Full Access

Bone-patellar tendon-bone autografts, hamstring autografts or allografts are widely used grafts for ACL revision surgeries. Also use of quadriceps autograft for both primary and revision ACL surgeries is in an increasing popularity due to its biomechanical superior properties and less donor site morbidity. However, although several fixation techniques and devices for quadriceps tendon graft fixation on femoral side have been reported, literature lacks about biomechanical studies comparing properties of these different fixation techniques and devices. We aimed to investigate whether there is a difference between the fixation techniques of quadriceps tendon graft by using different fixation techniques and devices on the femoral side in terms of stiffness and amount of slippage in the tunnel. Full-thickness central parts of quadriceps tendons from paired knees of twenty five calf knees were fixed through a 10mm x 25mm tunnel in twenty five paired sheep femurs. Quadriceps tendon patellar side with soft tissue ending fixed with four different fixation devices (adjustable suspensory system (group 1), absorbable interference screw (group 2), titanium interference screw (group 3) and adjustable suspensory system + absorbable interference screw (group 4)) and quadriceps tendon with a patellar bone plug fixed with a titanium interference screw (group 5) were tested in a servohydraulic materials testing machine. 10 samples were included in each group. After applying a preload of 10 N, a cyclic force was applied for 20 cycles from 10N to 110N at a 1 hertz frequency. Amount of slippage in the tunnel was calculated as the difference measured in millimeters between length at 10 N after 20 cycles and starting length at 10 N (Graph 1). To determine the stiffness, a single load-to-failure cycle was performed at a strain rate of 20 mm/min as the last step (Figure 1). Rupture of the graft was not seen in any of the samples. Median values of amount of slippage in the tunnel were 6,41mm, 5,99mm, 3,01mm, 4,83mm, and 3,94mm respectively. Median values of maximum load at failure were 464N, 160N, 350N, 350N and 389N respectively. Amount of slippage in the tunnel was highest in the group 1 and was lowest in the group 3 (p<0.001). Group 1 was found to be most resistant group against load-to-failure test and group 2 was the weakest (p<0.001). However inter-group analyses between group 3 and 5 revealed that, although group 3 had the least slippage in the tunnel, group 5 was better in terms of stiffness, but there was no statistically significant difference (p=0,124 and 0,119 respectively). There was a significant difference between group 2 and 3 in both amount of slippage in the tunnel and stiffness (p=0,001 and 0.028 respectively)(Table 1). Our study revealed that, although quadriceps graft with a bone plug fixed with metal interference screws is widely presumed to be a stable fixation technique, there was no significant difference in terms of stiffness when compared with quadriceps graft with soft tissue ending fixed with a metal interference screw. Although adjustable suspensory device group was the best in the terms of resistance against load-to-failure, it was the worst in terms of amount of slippage from the tunnel. Thus, if a suspensory device is to be used, it must be kept in mind that a strong 20 cycles of intra-operative tension force must be applied to prevent further slippage of the graft in the tunnel which can result in failure of reconstruction. For any figures or tables, please contact the authors directly


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_8 | Pages 25 - 25
11 Apr 2023
Richter J Ciric D Kalchschmidt K D'Aurelio C Pommer A Dauwe J Gueorguiev B
Full Access

Reorientating pelvic osteotomies are performed to improve femoral head coverage and secondary degenerative arthritis. A rectangular triple pelvic innominate osteotomy (3PIO) is performed in symptomatic cases. However, deciding optimal screw fixation type to avoid complications is questionable. Therefore, this study aimed to investigate the biomechanical behavior of two different acetabular screw configurations used for rectangular 3PIO osteosynthesis. It was hypothesized that bi-directional screw fixation would be biomechanically superior to mono-axial screw fixation technique. A rectangular 3PIO was performed in twelve right-side artificial Hemi-pelvises. Group 1 (G1) had two axial and one transversal screw in a bi-directional orientation. Group 2 (G2) had three screws in the axial direction through the iliac crest. Acetabular fragment was reoriented to 10.5° inclination in coronal plane, and 10.0° increased anteversion along axial plane. Specimens were biomechanically tested until failure under progressively increasing cyclic loading at 2Hz, starting at 50N peak compression, increasing 0.05N/cycle. Stiffness was calculated from machine data. Acetabular anteversion, inclination and medialization were evaluated from motion tracking data from 250-2500 at 250 cycle increments. Failure cycles and load were evaluated for 5° change in anteversion. Stiffness was higher in G1 (56.46±19.45N/mm) versus G2 (39.02±10.93N/mm) but not significantly, p=0.31. Acetabular fragment anteversion, inclination and medialization increased significantly each group (p≤0.02) and remained non-significantly different between the groups (p≥0.69). Cycles to failure and failure load were not significantly different between G1 (4406±882, 270.30±44.10N) and G2 (5059±682, 302.95±34.10N), p=0.78. From a biomechanical perspective, the present study demonstrates that a bi-directional screw orientation does not necessarily advantageous versus mono-axial alignment when the latter has all three screws evenly distributed over the osteotomy geometry. Moreover, the 3PIO fixation is susceptible to changes in anteversion, inclination and medialization of the acetabular fragment until the bone is healed. Therefore, cautious rehabilitation with partial weight-bearing is recommended


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_7 | Pages 37 - 37
4 Apr 2023
Pastor T Zderic I van Knegsel K Richards G Gueorguiev B Knobe M
Full Access

Proximal humeral shaft fractures are commonly treated with long straight plates or intramedullary nails. Helical plates might overcome the downsides of these techniques as they are able to avoid the radial nerve distally. The aim of this study was to investigate in an artificial bone model: (1) the biomechanical competence of different plate designs and (2) to compare them against the alternative treatment option of intramedullary nails. Twenty-four artificial humeri were assigned in 4 groups and instrumented as follows: group1 (straight 10-hole-PHILOS), group2 (MULTILOCK-nail), group3 (45°-helical-PHILOS) and group4 (90°-helical-PHILOS). An unstable proximal humeral shaft fracture was simulated. Specimens were tested under quasi-static loading in axial compression, internal/external rotation and bending in 4 directions monitored by optical motion tracking. Axial displacement (mm) was significantly lower in group2 (0.1±0.1) compared to all other groups (1: 3.7±0.6; 3: 3.8±0.8; 4: 3.5±0.4), p<0.001. Varus stiffness in group2 (0.8±0.1) was significantly higher compared to groups1+3, p≤0.013 (1: 0.7±0.1; 3: 0.7±0.1; 4: 0.8±0.1). Varus bending (°) was significantly lower in group2 compared to all other groups (p<0.001) and group4 to group1, p=0.022. Flexion stiffness in group1 was significantly higher compared to groups2+4 (p≤0,03) and group4 to group1, p≤0,029 (1: 0.8±0.1; 2: 0.7±0.1; 3: 0.7±0.1; 4: 0.6±0.1). Flexion bending (°) in group4 was higher compared to all other groups (p≤0.024) and lower in group2 compared to groups1+4, p≤0.024. Torsional stiffness remained non significantly different, p≥0.086. Torsional deformation in group2 was significantly higher compared to all other groups, p≤0.017. Shear displacement remained non significantly different, p≥0.112. From a biomechanical perspective, helical plating with 45° and 90° may be considered as a valid alternative fixation technique to standard straight plating of proximal third humeral fractures. Intramedullary nails demonstrated higher axial and bending stiffness as well as lower fracture gap movements during axial loading compared to all plate designs. However, despite similar torsional stiffness they were associated with higher torsional movements during internal/external rotation as compared to all investigated plate designs


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_4 | Pages 79 - 79
1 Mar 2021
Doodkorte R Roth A van Rietbergen B Arts J Lataster L van Rhijn L Willems P
Full Access

Complications after spinal fusion surgery are common, with implant loosening occurring in up to 50% of osteoporotic patients. Pedicle screw fixation strength reduces as a result of decreased trabecular bone density, whereas sublaminar wiring is less affected by these changes. Therefore, pedicle screw augmentation with radiopaque sublaminar wires (made with Dyneema Purity® Radiapque fibers, DSM Biomedical, Geleen, the Netherlands) may improve fixation strength. Furthermore, sublaminar tape could result in a gradual motion transition to distribute stress over multiple levels and thereby reduce implant loosening. The objective of this study is to test this hypothesis in a novel experimental setup in which a cantilever bending moment is applied to individual human vertebrae. Thirty-eight human cadaver vertebrae were stratified into four different groups: ultra-high molecular weight polyethylene sublaminar tape (ST), pedicle screw (PS), metal sublaminar wire (SW) and pedicle screw reinforced with sublaminar tape (PS+ST). The vertebrae were individually embedded in resin, and a cantilever bending moment was applied bilaterally through the spinal rods using a universal material testing machine. This cantilever bending setup closely resembles the loading of fixators at transitional levels of spinal instrumentation. The pull-out strength of the ST (3563 ± 476N) was not significantly different compared to PS, SW or PS+ST. The PS+ST group had a significantly higher pull-out strength (4522 ± 826N) compared to PS (2678 ± 292N) as well as SW (2931 ± 250N). The higher failure strength of PS + ST compared to PS indicates that PS augmentation with ST may be an effective measure to reduce the incidence of screw pullout, even in osteoporotic vertebrae. Moreover, the lower stiffness of sublaminar fixation techniques and the absence of damage to the cortices in the ST group suggest that ST as a stand-alone fixation technique in adult spinal deformity surgery may also be clinically feasible and offer clinical benefits


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_16 | Pages 26 - 26
1 Dec 2021
Edwards T Daly C Donovan R Whitehouse M
Full Access

Abstract. Objectives. There is debate regarding the optimal surgical technique for fixing femoral diaphyseal fractures in children aged 4 to 12 years. The aim of this study was to conduct a systematic review and meta-analysis to compare the complication rate following flexible intramedullary nailing (FIN), plate fixation and external fixation (EF) for traumatic femoral diaphyseal fractures in children aged 4 to 12. Methods. We searched MEDLINE, EMBASE and CENTRAL databases for interventional and observational studies. Two independent reviewers screened, assessed quality and extracted data from the identified studies. The primary outcome was the risk of any complication. Results. Nine randomised controlled trials (RCTs) and 19 observational studies fulfilled the eligibility criteria. Within the RCTs, five analysed FIN (n=161), two analysed plates (n=51) and five analysed EF (n=168). Within the observational studies, 13 analysed FIN (n=610), seven analysed plates (n=214) and six analysed EF (n=153). The overall risk of complications was lower following plate fixation when compared to FIN (RR 0.45, 95% CI 0.28 to 0.73, p=0.001) in the observational studies. The overall risk of complications was higher following EF when compared to FIN in both RCTs (RR 1.94, 95% CI 1.25 to 3.01, p=0.003) and observational studies (RR 1.97, 95% CI 1.50 to 2.58, p<0.001). The overall risk of complications was higher following EF when compared to plate fixation in both RCTs (RR 7.42, 95% CI 1.84 to 29.98, p=0.005) and observational studies (RR 4.39, 95% CI 2.64 to 7.30, p<0.001). Conclusions. This study reports a significantly decreased relative risk of complications when femoral diaphyseal fractures in children aged 4 to 12 are managed with plates. The overall quality of evidence is low, highlighting the need for a prospective multicentre randomised trial at low risk of bias due to randomisation and outcome measurement to identify if any fixation technique is superior


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_13 | Pages 132 - 132
1 Nov 2021
Chalak A Singh P Singh S Mehra S Samant PD Shetty S Kale S
Full Access

Introduction and Objective. Management of gap non-union of the tibia, the major weight bearing bone of the leg remains controversial. The different internal fixation techniques are often weighed down by relatively high complication rates that include fractures which fail to heal (non-union). Minimally invasive techniques with ring fixators and bone transport (distraction osteogenesis) have come into picture as an alternative allowing alignment and stabilization, avoiding a graduated approach. This study was focused on fractures that result in a gap non-union of > 6 cm. Ilizarov technique was employed for management of such non-unions in this case series. The Ilizarov apparatus consists of rings, rods and kirschner wires that encloses the limb as a cylinder and uses kirschner wires to create tension allowing early weight bearing and stimulating bone growth. Ilizarov technique works on the principle of distraction osteogenesis, that is, pulling apart of bone to stimulate new bone growth. Usually, 4–5 rings are used in the setup depending on fracture site and pattern for stable fixation. In this study, we demonstrate effective bone transport and formation of gap non-union more than 6 cm in 10 patients using only 3 rings construct Ilizarov apparatus. Materials and Methods. This case study was conducted at Dr. D. Y. Patil Medical Hospital, Navi Mumbai, Maharashtra, India. The study involved 10 patients with a non-union or gap > 6 cm after tibial fracture. 3 rings were used in the setup for the treatment of all the patients. Wires were passed percutaneously through the bone using a drill and the projecting ends of the wires were attached to the metal rings and tensioned to increase stability. The outcome of the study was measured using the Oxford Knee scoring system, Functional Mobility Scale, the American Foot and Ankle Score and Visual Analog Scale. Further, follow up of patients was done upto 2 years. Results. All the patients demonstrated good fixation as was assessed clinically and radiologically. 9 patients had a clinical score of > 65 which implied fair to excellent clinical rating. The patients showed good range of motion and were highly satisfied with the treatment as measured by different scoring parameters. Conclusions. In this case study, we demonstrate that the Ilizarov technique using 3 rings is equally effective in treating non-unions > 6 cm as when using 4–5 rings. Obtaining good clinical outcome and low complication rate in all 10 patients shows that this modified technique can be employed for patients with such difficulties in the future


Orthopaedic Proceedings
Vol. 103-B, Issue SUPP_13 | Pages 155 - 155
1 Nov 2021
Edwards T Daly C Donovan R Whitehouse M
Full Access

Introduction and Objective. The most common paediatric orthopaedic injury requiring hospital admission is a femoral fracture. There is debate regarding the optimal surgical technique for fixing femoral diaphyseal fractures in children aged 4 to 12 years. The National Institute for Health and Care Excellence (NICE) and the American Academy of Orthopaedic Surgeons (AAOS) have issued relevant guidelines, however, there is limited evidence to support these. The aim of this study was to conduct a systematic review and meta-analysis to compare the complication rate following flexible intramedullary nailing (FIN), plate fixation and external fixation (EF) for traumatic femoral diaphyseal fractures in children aged 4 to 12. Materials and Methods. We searched MEDLINE, EMBASE and CENTRAL databases for interventional and observational studies. Two independent reviewers screened, assessed quality and extracted data from the identified studies. The primary outcome was the risk of any complication. Secondary outcomes assessed the risk of pre-specified individual complications. Results. Nine randomised controlled trials (RCTs) and 19 observational studies (six prospective and 13 retrospective) fulfilled the eligibility criteria. Within the RCTs, five analysed FIN (n=161), two analysed plates (n=51) and five analysed EF (n=168). Within the observational studies, 13 analysed FIN (n=610), seven analysed plates (n=214) and six analysed EF (n=153). The overall risk of complications was lower following plate fixation when compared to FIN fixation (RR 0.45, 95% CI 0.28 to 0.73, p=0.001) in the observational studies. The overall risk of complications was higher following EF when compared to FIN fixation in both RCTs (RR 1.94, 95% CI 1.25 to 3.01, p=0.003) and observational studies (RR 1.97, 95% CI 1.50 to 2.58, p<0.001). The overall risk of complications was higher following EF when compared to plate fixation in both RCTs (RR 7.42, 95% CI 1.84 to 29.98, p=0.005) and observational studies (RR 4.39, 95% CI 2.64 to 7.30, p<0.001). Conclusions. Although NICE and the AAOS recommend FIN for femoral diaphyseal fractures in children aged 4 to 12, this study reports a significantly decreased relative risk of complications when these injuries are managed with plates. Our findings provide valuable information to healthcare professionals who are involved in discussing the risk and benefits of different management options with patients and their families. The overall quality of evidence is low, highlighting the need for a rigorous prospective multicentre randomised trial at low risk of bias due to randomisation and outcome measurement to identify if any fixation technique is superior


Orthopaedic Proceedings
Vol. 102-B, Issue SUPP_11 | Pages 77 - 77
1 Dec 2020
Ivanov S Stefanov A Zderic I Gehweiler D Richards G Raykov D Gueorguiev B
Full Access

Displaced intraarticular calcaneal fractures are debilitating injuries with significant socioeconomic and psychological effects primarily affecting patients in active age between 30 and 50 years. Recently, minimally and less invasive screw fixation techniques have become popular as alternative to locked plating. The aim of this study was to analyze biomechanically in direct comparison the primary stability of 3 different cannulated screw configurations for fixation of Sanders type II-B intraarticular calcaneal fractures. Fifteen fresh-frozen human cadaveric lower limbs were amputated mid-calf and through the Chopart joint. Following, soft tissues at the lateral foot side were removed, whereas the medial side and Achilles tendon were preserved. Reproducible Sanders type II-B intraarticular fracture patterns were created by means of osteotomies. The proximal tibia end and the anterior-inferior aspect of the calcaneus were then embedded in polymethylmethacrylate. Based on bone mineral density measurements, the specimens were randomized to 3 groups for fixation with 3 different screw configurations using two 6.5 mm and two 4.5 mm cannulated screws. In Group 1, two parallel longitudinal screws entered the tuber calcanei above the Achilles tendon insertion and proceeded to the anterior process, and two transverse screws fixed the posterior facet perpendicular to the fracture line. In Group 2, two parallel screws entered the tuber calcanei below the Achilles tendon insertion, aiming at the anterior process, and two transverse screws fixed the posterior facet. In Group 3, two screws were inserted along the bone axis, entering the tuber calcanei above the Achilles tendon insertion and proceeding to the central-inferior part of the anterior process. In addition, one transverse screw was inserted from lateral to medial for fixation of the posterior facet and one oblique screw – inserted from the posterior-plantar part of the tuber calcanei – supported the posterolateral part of the posterior facet. All specimens were tested in simulated midstance position under progressively increasing cyclic loading at 2 Hz. Starting from 200N, the peak load of each cycle increased at a rate of 0.1 N/cycle. Interfragmentary movements were captured by means of optical motion tracking and triggered mediolateral x-rays. Plantar movement, defined as displacement between the anterior process and the tuber calcanei at the most inferior side was biggest in Group 2 and increased significantly over test cycles in all groups (P = 0.001). Cycles to 2 mm plantar movement were significantly higher in both Group 1 (15847 ± 5250) and Group 3 (13323 ± 4363) compared to Group 2 (4875 ± 3480), P = 0.048. Medial gapping after 2500 cycles was significantly bigger in Group 2 versus Group 3, P = 0.024. No intraarticular displacement was observed in any group during testing. From biomechanical perspective, screw configuration implementing one oblique screw seems to provide sufficient hindfoot stability in Sanders Type II-B intraarticular calcaneal fractures under dynamic loading. Posterior facet support by means of buttress or superiorly inserted longitudinal screws results in less plantar movement between the tuber calcanei and anterior fragments. On the other hand, inferiorly inserted longitudinal screws seem to be associated with bigger interfragmentary movements


Orthopaedic Proceedings
Vol. 100-B, Issue SUPP_3 | Pages 8 - 8
1 Apr 2018
Marques E Fawsitt C Thom H Hunt LP Nemes S Lopez-Lopez J Beswick A Burston A Higgins JP Hollingworth W Welton NJ Rolfson O Garellick G Blom AW
Full Access

Background. Prosthetic implants used in primary total hip replacements have a range of bearing surface combinations (metal-on-polyethylene, ceramic-on-polyethylene, ceramic-on-ceramic, metal-on-metal); head sizes (small <36mm, large 36mm+); and fixation techniques (cemented, uncemented, hybrid, reverse hybrid), which influence prosthesis survival, patient quality of life, and healthcare costs. This study compared the lifetime cost-effectiveness of implants to determine the optimal choice for patients of different age and gender profiles. Methods. In an economic decision Markov model, the probability that patients required one or more revision surgeries was estimated from analyses of UK and Swedish hip joint registries, for males and females aged <55, 55–64, 65–74, 75–84, and 85+ years. Implant and healthcare costs were estimated from hospital procurement prices, national tariffs, and the literature. Quality-adjusted life years were calculated using utility estimates, taken from Patient-Reported Outcome Measures data for hip procedures in the UK. Results. Optimal choices varied between traditionally used cemented metal-on-polyethylene and cemented ceramic-on-polyethylene implants. Small head cemented ceramic-on-polyethylene implants were optimal for males and females aged under 65. The optimal choice for adults aged 65 and older was small head cemented metal-on-polyethylene implants. Conclusions. The older the patient, the higher the probability that small head cemented metal-on-polyethylene implants are optimal. Small head cemented ceramic-on-polyethelyne implants are optimal for adults aged under 65. Our findings can influence NICE guidance, clinical practice, and commissioning of services. Funding. NIHR Research for Patient Benefit programme PB-PG-0613-31032


Orthopaedic Proceedings
Vol. 96-B, Issue SUPP_11 | Pages 126 - 126
1 Jul 2014
Eguchi A Ochi M Adachi N Deie M Nakamae A Nakasa T
Full Access

Summary Statement. We evaluated the mechanical strength of two cortical suspension devices by reproducing clinical situation for ACL reconstruction. A most important factor affecting the displacement during cyclic load was the length of the tendon rather than the length of the device. Introduction. A definite consensus for the optimal graft fixation technique to the femur in an anterior cruciate ligament (ACL) reconstruction has not been reached, although there have been several fixation techniques such as cortical suspension devices, transfixation devices, and interference screws. The purpose of this study was to evaluate the mechanical strength of two cortical suspension devices by reproducing actual clinical situation for ACL reconstruction in order to compare the TightRope. TM. as a new adjustable-length loop device and the EndoButton. TM. as a well-known fixed-length loop device under the consistent conditions. Methods. Two cortical suspension devices were tested under cyclic and pull-to-failure loading conditions in both an isolated device setup and a specimen setup to make a complete bone-device-tendon construct in porcine femurs and bovine flexor tendons using a tensile testing machine. Especially to examine the influence of the length of the tendon and the device, a total length of the bone tunnel was fixed to 35 mm, and an effective length of tendon in the bone tunnel was adjusted to 15 mm for the EndoButton group (EB), the TightRope 15 group (TR15) and 21 mm for the TightRope 21 group (TR21). Results. In the isolated device testing, the ultimate tensile strength of the EB (1430 ± 148 N) had significantly higher than that of the TR (866 ± 53 N), and also had significant difference in the specimen testing. The displacement in the isolated device testing after preloading for the EB (1.09 ± 0.29 mm) showed statistically lower than that for the TR (2.57 ± 1.19 mm), and had a significant difference after the cyclic load. In the specimen testing, on the other hand, the displacement after preloading showed no statistical difference between the EB (1.06 ± 0.30 mm), the TR21 (1.76 ± 2.28 mm) and the TR15 (1.51 ± 1.78 mm). But limiting only to the displacement from 1000 to 2000 cycles, the TR21 (0.92 ± 0.44 mm) showed statistically higher than the TR15 (0.49 ± 0.22 mm). Discussion. Although current results indicated that the EB had greater mechanical strength than the TR, both devices are presumed to provide sufficient fixation strength under clinical conditions. An important new finding from the current study was the measurement of initial displacement from the state of initial fixation until loading began and 50 N of tension was applied. In isolated device testing, the TR had significantly more displacement than the EB during pre-loading. This may reflect the TR's loops stretch until a certain amount of tension is applied. In the comparison of the TR21 and the TR15, the TR21 had a significantly larger displacement with a cyclic load from 1000 to 2000 cycles. This result indicated that a most important factor affecting the displacement during cyclic load was the length of the tendon rather than the length of the device. Thus, we should decide the length of the tendon in the bone tunnel to avoid the displacement of the graft


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_9 | Pages 43 - 43
1 May 2017
Thakrar R Patel K Ghani Y Kotecha A Sikand M
Full Access

Background. The approach to Intramedullary (IM) fixation of long bone fractures remains a controversial issue. Early reports demonstrated less favourable results of retrograde nailing as compared with antegrade options due to higher non-union rates. The aim of this audit was to evaluate the outcomes of practice within the Trauma and Orthopaedic department with relation to IM nail fixation of diaphyseal femur fractures. Methodology. The Trauma database between February 2010 and September 2013 was used to identify all femur IM nailing procedures. Picture Archiving and Communication System (PACS) software was used to classify the fractures according to the Muller AO classification. All 3–2 (Diaphyseal femur fractures) were included in the audit. PACS imaging together with outpatient documentation was evaluated for radiological and clinical outcome. Results. A total sample size of 23 patients was identified (13 antegrade vs. 10 retrograde approach fixations). Mean patient age was 67 years and male to female ratios were similar (11M vs. 12F). Antegrade nailing was performed in a younger population as compared to retrograde nailing (mean age 60 vs. 73 respectively). Mean time to union was somewhat more protracted in the retrograde group (7 vs. 5 months), although all fractures united. The most common complication with relation to antegrade nailing was due to distal locking screws backing out. I case of infection was reported in the retrograde nail group, which was treated successfully with antibiotic therapy. There were 2 cases of nonunion observed in the antegrade group. Conclusions. The results of our practice were comparable to those published in recent literature. Overall, union rates for the two groups of fixation were similar. Each fixation technique is associated with its own specific set of complications. As a general rule antegrade nailing was reserved for a younger population so as to prevent trauma to the native knee joint


Orthopaedic Proceedings
Vol. 99-B, Issue SUPP_8 | Pages 64 - 64
1 Apr 2017
Jordan R Saithna A
Full Access

Background. Distal radius fractures are common injuries but no clear consensus regarding optimal management of unstable fractures exists. Open reduction and internal fixation with volar plates is an increasingly popular but the associated complication rate can be 10%. Intramedullary nails are an alternative offering the potential advantages of reduced risk of tendon injury and intra-articular screw penetration. This article systematically reviews the published literature evaluating the biomechanics, outcomes and complications of intramedullary nails in the management of distal radius fractures. Methods. A systematic review of Medline and EMBASE databases was performed for studies reporting the biomechanics, functional outcome or complications following intramedullary nailing of distal radius fractures. Critical appraisal was performed with respect to validated quality assessment scales. Results. 16 studies were included for review. The biomechanical studies concluded that intramedullary nails had at least comparable strength to locking plates. The clinical studies reported that IM nailing was associated with comparable ROM, functional outcome and grip strength to alternative fixation techniques. However, the mean complication rate was 17.6% (range 0 to 50%) with the most common complication being neurapraxia of the superficial radial nerve in 9.5%. Conclusion. This systematic review of pooled data from published series has shown that IM nailing can give comparable clinical results to current treatment modalities in extra-articular and simple intra-articular distal radius fractures. However the evidence is insufficient to determine whether IM nailing has any clinically important advantage over well-established alternatives. The complication rate reported is higher than that in contemporary studies for volar plating and this raises concerns about the role of this technique particularly when comparative studies have failed to show any major advantage to its use. Further adequately powered RCTs comparing the technique to both volar plates and percutaneous wire fixation are required. Level of evidence. IIa – systematic review of cohort studies. Conflict of Interests. The authors confirm that they have no relevant financial disclosures or conflicts of interest. Ethical approval was not sought as this was a systematic review


The Journal of Bone & Joint Surgery British Volume
Vol. 90-B, Issue 10 | Pages 1386 - 1391
1 Oct 2008
Ozbaydar M Elhassan B Esenyel C Atalar A Bozdag E Sunbuloglu E Kopuz N Demirhan M

We compared time-dependent changes in the biomechanical properties of single-and double-row repair of a simulated acute tear of the rotator cuff in rabbits to determine the effect of the fixation techniques on the healing process. A tear of the supraspinatus tendon was created in 80 rabbits which were separated into two equal groups. A single-row repair with two suture anchors was conducted in group 1 and a double-row repair with four suture anchors in group 2. A total of ten intact contralateral shoulder joints was used as a control group. Biomechanical testing was performed immediately post-operatively and at four and eight weeks, and histological analysis at four and eight weeks. The mean load to failure in group 2 animals was greater than in group 1, but both groups remained lower than the control group at all intervals. Histological analysis showed similar healing properties at four and eight weeks in both groups, but a significantly larger number of healed tendon-bone interfaces were identified in group 2 than in group 1 at eight weeks (p < 0.012). The ultimate load to failure increased with the number of suture anchors used immediately post-operatively, and at four and eight weeks. The increased load to failure at eight weeks seemed to be related to the increase in the surface area of healed tendon-to-bone in the double-row repair group


The Journal of Bone & Joint Surgery British Volume
Vol. 87-B, Issue 6 | Pages 868 - 872
1 Jun 2005
Metcalfe AJ Saleh M Yang L

Biomechanical studies involving all-wire and hybrid types of circular frame have shown that oblique tibial fractures remain unstable when they are loaded. We have assessed a range of techniques for enhancing the fixation of these fractures. Eight models were constructed using Sawbones tibiae and standard Sheffield ring fixators, to which six additional fixation techniques were applied sequentially. The major component of displacement was shear along the obliquity of the fracture. This was the most sensitive to any change in the method of fixation. All additional fixation systems were found to reduce shear movement significantly, the most effective being push-pull wires and arched wires with a three-hole bend. Less effective systems included an additional half pin and arched wires with a shallower arc. Angled pins were more effective at reducing shear than transverse pins. The choice of additional fixation should be made after consideration of both the amount of stability required and the practicalities of applying the method to a particular fracture