Advertisement for orthosearch.org.uk
Results 1 - 10 of 10
Results per page:
Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_10 | Pages 30 - 30
1 Jun 2023
Tissingh E Goodier D Wright J Timms A Campbell M Crook G Calder P
Full Access

Introduction. The FitBone lengthening nail (Orthofix UK) is an intramedullary device licensed for the lengthening of long bones in adults in the UK. It contains a motor powered by electricity transmitted via an induction coil placed underneath the skin. It was developed in Germany two decades ago but uptake in the UK has only started more recently. The aim of this study was to review the first cohort of FitBone lengthening nails in a unit with significant experience of other lengthening nails (including PRECICE and Stryde). Materials & Methods. Demographic, clinical and radiological data was prospectively collected on all FitBone cases starting in February 2022. Accuracy of lengthening rate, patient satisfaction and implant issues were all considered. Complications and learning points were recorded and discussed by the multidisciplinary team involved in the patients care. Results. Eleven lengthening nails were inserted between February and November 2022 (6 right femurs, 5 left femurs). The average patient age was 31 (16–57) with 4 females and 7 males. The average lengthening achieved was 44mm (13– 70) over an average of 59 days (35 to 104). Significant technical issues were encountered in this cohort of patients including slow opening up at osteotomy site (3 requiring speeding up of programme), early consolidation (one requiring re-do osteotomy) and backing out of locking screws (3 out of 11 nails). There were also patient use concerns with difficulty using the motor and the inability to reverse the lengthening without an additional component to the motor. Conclusions. We present the first UK cohort of patients with femoral lengthening using the FitBone implant and device. We highlight the technical and patient issues encountered during this learning curve and propose solutions to avoid these pitfalls


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_10 | Pages 28 - 28
1 Jun 2023
Musielak B Green N Giles S Madan S Fernandes J
Full Access

Introduction. Intramedullary lengthening devices have been in use in older children with closed /open growth plates with good success. This study aims to present the early experience of the FITBONE nail since withdrawal of the PRECICE nail. Materials & Methods. Retrospective analysis of both antegrade and retrograde techniques were utilized. Only patients where union was achieved and full weight bearing commenced were included. The complication rate, length gained, distraction index, weight bearing index (WBI) as well as mechanical axes were analysed. Results. 14 (7 males, 7 females) of a total of 16 (7 males, 7 females) patients with a mean age of 16.9 years with varied diagnosis of LLD were analysed. The mean length gained was 38 mm with an average distraction index of 0,74 mm/day. WBI in these patients on average was 59,6 days/cm lengthened. 6 complications were observed, including two nonunions (successfully treated) and a knee subluxation. Mechanical axis deviation improved from 13,3 mm to 6 mm on average. Overall there has been a nonsignificant tendency for WBI to decrease (Spearman's rank correlation coefficient −0.47, p=0.08) with increasing number of cases done, while no correlation between length gained and WBI (−0.01, p=0.96, respectively) was observed. Some nuances will be discussed. Conclusions. Limb lengthening with the FITBONE nail is relatively safe and efficient, however no significant change was seen in the outcome with previous motorized nails


The Bone & Joint Journal
Vol. 103-B, Issue 11 | Pages 1731 - 1735
1 Nov 2021
Iobst CA Frost MW Rölfing JD Rahbek O Bafor A Duncan M Kold S

Aims. Limb-lengthening nails have largely replaced external fixation in limb-lengthening and reconstructive surgery. However, the adverse events and high prevalence of radiological changes recently noted with the STRYDE lengthening nail have raised concerns about the use of internal lengthening nails. The aim of this study was to compare the prevalence of radiological bone abnormalities between STRYDE, PRECICE, and FITBONE nails prior to nail removal. Methods. This was a retrospective case series from three centres. Patients were included if they had either of the three limb-lengthening nails (STYDE, PRECICE, or FITBONE) removed. Standard orthogonal radiographs immediately prior to nail removal were examined for bone abnormalities at the junction of the telescoping nail parts. Results. In total, 306 patients (168 male, 138 female) had 366 limb-lengthening nails removed. The mean time from nail insertion to radiological evaluation was 434 days (36 to 3,015). Overall, 77% of STRYDE nails (20/26) had bone abnormalities at the interface compared with only 2% of FITBONE (4/242) and 1% of PRECICE nails (1/98; p < 0.001). Focal osteolysis in conjunction with periosteal reaction at the telescoping interface was only observed in STRYDE nails. Conclusion. Bone abnormalities at the interface of telescoping nail parts were seen in the majority of STRYDE nails, but only very rarely with FITBONE or PRECICE nails. We conclude that the low prevalence of radiological changes at the junctional interface of 242 FITBONE and 98 PRECICE nails at the time of nail removal does not warrant clinical concerns. Cite this article: Bone Joint J 2021;103-B(11):1731–1735


Orthopaedic Proceedings
Vol. 106-B, Issue SUPP_5 | Pages 24 - 24
23 Apr 2024
Thompson E James L Narayan B Peterson N
Full Access

Introduction. Management of deformity involving limb length discrepancy (LLD) using intramedullary devices offers significant benefits to both patients and clinicians over traditional external fixation. Following the withdrawal of the PRECICE nail, the Fitbone became the primary implant available for intramedullary lengthening and deformity correction within our service. This consecutive series illustrates the advantages and complications associated with the use of this device, and describes a novel technique modification for antegrade intramedullary lengthening nails. Materials & Methods. A retrospective cohort review was performed of patient outcomes after treatment with the Fitbone nail at two tertiary referral limb reconstruction services (one adult, one paediatric) between January 2021 to December 2023. Aetiology, indications, initial and final LLD, use of concomitant rail assisted deformity correction (ORDER), removal time and healing index were assessed. Complications of treatment were evaluated and described in detail, alongside technique modifications to reduce the rate of these complications. Results. 21 nails (18 femoral, 2 tibial, 1 humeral) were inserted in 6 adult and 13 paediatric patients. Post-traumatic and congenital/developmental LLD were the most common indications for surgery in the adult and paediatric cohorts respectively. ORDER was employed in 11 cases (9 femurs and 2 tibias). Treatment goals were achieved in all but one case. Complications included superficial infection, locking bolt migration, periprosthetic fracture and component failure. Seven patients required unplanned returns to theatre. Conclusions. The Fitbone nail is an established option for intramedullary limb lengthening, however its use in the UK has been relatively limited compared to the PRECICE until 2021. Our data helps to define its place for limb lengthening and complex deformity correction in both adult and paediatric patients, including in humeral lengthening and retrograde femoral insertion across an open physis. We have identified important potential risks and novel techniques to simplify surgery and avoid complications


Introduction

At Sheffield Children's Hospital, treatment of leg length discrepancy is a common procedure. Historically, this has been done with external fixators. With the development in intramedullary technology, internal nails have become the preferred modality for long bone lengthening in the adolescent population. However, it is important to review whether this technology practically reduces the known challenges seen and if it brings any new issues. Therefore, the aim of this review is to retrospectively evaluate the therapeutic challenges of 16 fit-bone intramedullary femoral lengthening's at Sheffield Children's Hospital between 2021–2022.

Materials & Methods

The international classification of function (ICF) framework was used to differentiate outcomes. The patient's therapy notes were retrospectively reviewed for themes around structural, activity and participation limitation. The findings were grouped for analysis and the main themes presented.


Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_5 | Pages 11 - 11
1 Apr 2022
Baumgart R
Full Access

Introduction. Fully implantable distraction nails are frequently used for lengthening of the lower leg. What are the indications for humerus lengthening and which results can be expected?. Materials and Methods. The humerus as the best healing bone in the human body is suitable for lengthening with a nail as well but the indications are rare in comparison with the lower legs. Especially when driving a car or when working on a tablet length discrepancy of the arms of more than 4–5cm may cause severe hardening of the cervical muscles and induce chronical pain in the upper spine. Results. A distraction nail (FITBONE) was used in 5 cases for humerus lengthening. The nail was inserted in all cases from proximal. The mean age of the patients was 34 years. The mean lengthening amount was 58mm (55–100). In one case the system was exchanged to reach the desired length. There was no infection, no radial nerve irritation and no chronical shoulder pain. The system was removed in 4 cases in an average of 15 months the other nail will be removed soon as well. Conclusions. The preliminary results of our 5 cases demonstrate, that the FITBONE device is advantageous for lengthening the humerus, if the initial length is sufficient for implantation. In comparison with the use of external fixators the functional outcome, the comfort of treatment and the cosmetic result is amazing


Bone & Joint Research
Vol. 10, Issue 7 | Pages 425 - 436
16 Jul 2021
Frommer A Roedl R Gosheger G Hasselmann J Fuest C Toporowski G Laufer A Tretow H Schulze M Vogt B

Aims

This study aims to enhance understanding of clinical and radiological consequences and involved mechanisms that led to corrosion of the Precice Stryde (Stryde) intramedullary lengthening nail in the post market surveillance era of the device. Between 2018 and 2021 more than 2,000 Stryde nails have been implanted worldwide. However, the outcome of treatment with the Stryde system is insufficiently reported.

Methods

This is a retrospective single-centre study analyzing outcome of 57 consecutive lengthening procedures performed with the Stryde nail at the authors’ institution from February 2019 until November 2020. Macro- and microscopic metallographic analysis of four retrieved nails was conducted. To investigate observed corrosion at telescoping junction, scanning electron microscopy (SEM) and energy dispersive x-ray spectroscopy (EDX) were performed.


The Journal of Bone & Joint Surgery British Volume
Vol. 88-B, Issue 7 | Pages 938 - 942
1 Jul 2006
Singh S Lahiri A Iqbal M

Limb lengthening by callus distraction and external fixation has a high rate of complications. We describe our experience using an intramedullary nail (Fitbone) which contains a motorised and programmable sliding mechanism for limb lengthening and bone transport. Between 2001 and 2004 we lengthened 13 femora and 11 tibiae in ten patients (seven men and three women) with a mean age of 32 years (21 to 47) using this nail. The indications for operation were short stature in six patients and developmental or acquired disorders in the rest.

The mean lengthening achieved was 40 mm (27 to 60). The mean length of stay in hospital was seven days (5 to 9). The mean healing index was 35 days/cm (18.8 to 70.9). There were no cases of implant-related infection or malunion.


The Bone & Joint Journal
Vol. 101-B, Issue 2 | Pages 178 - 188
1 Feb 2019
Chaudhary MM Lakhani PH

Aims

Double-level lengthening, bone transport, and bifocal compression-distraction are commonly undertaken using Ilizarov or other fixators. We performed double-level fixator-assisted nailing, mainly for the correction of deformity and lengthening in the same segment, using a straight intramedullary nail to reduce the time in a fixator.

Patients and Methods

A total of 23 patients underwent this surgery, involving 27 segments (23 femora and four tibiae), over a period of ten years. The most common indication was polio in ten segments and rickets in eight; 20 nails were inserted retrograde and seven antegrade. A total of 15 lengthenings were performed in 11 femora and four tibiae, and 12 double-level corrections of deformity without lengthening were performed in the femur. The mean follow-up was 4.9 years (1.1 to 11.4). Four patients with polio had tibial lengthening with arthrodesis of the ankle. We compared the length of time in a fixator and the external fixation index (EFI) with a control group of 27 patients (27 segments) who had double-level procedures with external fixation. The groups were matched for the gain in length, age, and level of difficulty score.


The Journal of Bone & Joint Surgery British Volume
Vol. 91-B, Issue 7 | Pages 955 - 961
1 Jul 2009
Simpson AHWR Shalaby H Keenan G

Femoral lengthening using the Intramedullary Skeletal Kinetic Distractor is a new technique. However, with intramedullary distraction the surgeon has less control over the lengthening process. Therefore, 33 femora lengthened with this device were assessed to evaluate the effect of operative variables under the surgeon’s control on the course of lengthening. The desired lengthening was achieved in 32 of 33 limbs. Problems encountered included difficulty in achieving length in eight femora (24%) and uncontrolled lengthening in seven (21%). Uncontrolled lengthening was more likely if the osteotomy was placed with less than 80 mm of the thick portion of the nail in the distal fragment (p = 0.052), and a failure to lengthen was more likely if there was over 125 mm in the distal fragment (p = 0.008). The latter problem was reduced with over-reaming by 2.5 mm to 3 mm. Previous intramedullary nailing also predisposed to uncontrolled lengthening (p = 0.042), and these patients required less reaming.

Using the Intramedullary Skeletal Kinetic Distractor, good outcomes were obtained; problems were minimised by optimising the position of the osteotomy and the amount of over-reaming performed.