Advertisement for orthosearch.org.uk
Results 1 - 19 of 19
Results per page:
Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 72 - 72
1 Sep 2012
Fujita H Katayama N Iwase T Otsuka H
Full Access

Purpose. Exeter stem was introduced to Japanese market at 1996. Since then, owing to its excellent clinical results, the number of the stem used has been increased year by year and more than 2000 stems have been implanted during the year 2009. The present study aims to prove its efficacy for Japanese patients by evaluating short term results of four major dedicated hip centers. Method. We present the short-term multi-center results of primary THA with Exeter stem combined with modern cementing technique in 881 patients (1000 hips). The average age of the patients at operation was 62.3years (ranging 23 to 89 years). Mean postoperative follow up period was 4.0 (ranging 2 to 9) years. Results. Intraoperative complications such as femoral fracture were observed in 6 hips. Postoperative cementing grade using Barrack's classification was judged as A for 735 hips, as B for 246 hips, and as C for 4 hips, respectively. Twenty-four patients died with unrelated disease and 52 patients were lost during follow-up period. Follow-up ratio was 94.8%. Thirty postoperative complications which include 9 infections, 14 dislocations, 5 deep vein thrombosis, and 2 sciatic nerve palsies were observed. Symptomatic pulmonary embolization or femoral fracture was not observed. Re-operations for infection and dislocation were performed for 8 and 6 hips, respectively. At the final follow-up, no radiolucent line at bone-cement interface was observed. Cortical hypertrophy was observed in 9.6% at Gruen zone 2 to 6. Kaplan-Meier survivorship analysis predicted a rate of survival at 5 years of 100% with radiolucent line at bone-cement interface of the femur as the endpoint, and of 98.8% when re-operation for any reason was used. Conclusion. The present study revealed excellent short-term result was obtained using Exeter stem combined with modern cementing technique for primary THAs in Japan


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 1 - 1
1 Sep 2012
Moideen AN Wolf C Donaldson O O'Doherty D
Full Access

Background. Exeter Trauma Stem (ETS) is a polished tapered collarless monopolar prosthesis used for cemented hemiarthroplasty for fracture neck of femur. Two modular rasps are available on the instrumentation set. The larger rasp, in practice, rarely fits into the femur making trial reduction impossible. Our hypothesis was that the absence of a trial reduction could affect leg length and stability of the prosthesis. Aim. To evaluate the leg length discrepancy following use of Exeter Trauma Stem for intracapsular fracture neck of femur. Methods. All patients who underwent ETS hemiarthroplasty between June 2009 to March 2010 (n = 100) were included in the study. Radiographic assessment of leg length was performed by two independent assessors based on measurements by Woolson et al. Results. Upto half of the patients had equal leg length measurements. Forty four patients had lengthening of the operated leg with a mean of 10mm (5–25). Further complications included four dislocations. There was no statistical significance between leg lengths and dislocations. Conclusion. Care is needed with regards to correct positioning of the ETS prosthesis to achieve equal leg lengths. Following this study a new intermediate sized modular rasp has been introduced to the instrumentation set


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_5 | Pages 8 - 8
13 Mar 2023
Powell-Bowns M Oag E Martin D Moran M Scott C
Full Access

The aim of the study was to report the survival of open reduction and internal fixation (ORIF) of Vancouver B fractures associated with the Exeter Stem (ES) at a minimum of 5 years. This retrospective cohort study assessed 129 consecutive patients with Vancouver B type fractures treated with ORIF from 2008-2016 at a minimum of 5 years. Patient records were examined, and the following recorded: details of primary prosthesis, details of injury, Vancouver classification, details of operative management, complications, and requirement for reoperation. Data was analysed using SPSS. Survival analysis was undertaken using the endpoint ‘reoperation for any reason’. Mean age at fracture was 78.2 (SD10.6, 46-96) and 54 (43%) were female. Vancouver subclassifications were: 24% B1, 70.5% B2 and 5.5% B3. For all Vancouver B fractures, Kaplan Meier analysis demonstrated a 5 year survival free from reoperation of 88.8% (82.0-94.7 95%CI). Fourteen patients required reoperation, most commonly within the first year for non-union and plate fracture (5.4%). Five-year survival for any reoperation differed significantly according to fracture type (p=0.016) and was worst in B1s: B1 76.6% (61.3-91.9); B2 92.6% 986.9-98.3); and 100% of B3. Univariate analysis identified B1 type (p=0.008) and a transverse fracture pattern (p=0.003) to be significantly associated with the need for reoperation. Adopting a strategy of fixation of all Vancouver B fractures involving the ES where the fracture was anatomically reducible and the bone cement interface was well-fixed was associated with a 5 year survival, free from reoperation of 88.8%


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 236 - 236
1 Sep 2012
Queiroz M Barros F Daniachi D Polesello G Guimarães R Ricioli W Ono N Honda E
Full Access

Introduction. One of the most common complications of ceramic on ceramic hip replacement is squeaking. The association of Accolade stem and Trident acetabular system has been reported to have squeaking incidence of up to 35,6%. There is doubt if this phenomenon occurs due to: the stem titanium alloy, the V40 femoral neck, the recessed liner of the trident cup or even the mal-seating of the trident insert on the cup. Objectives. Primary: The purpose of the present study was to determine the incidence of squeaking in association with the use of Exeter stem and Trident ceramic acetabular system. Secondary: Analysis of the correlation of the cup abduction angle and squeaking. Methods. During the period from March 2004 to December 2008, two surgeons performed 87 total hip arthroplasties in 77 patients with use of a ceramic-on-ceramic bearing (Exeter stem, alumina head, Trident ceramic acetabular system). Seventy six patients (86 THA) were available for review after at least 18 months follow-up. The incidence of squeaking and other noises was analyzed. Cup abduction angle was measured. The Pearson correlation coefficient was used to determine if a correlation existed between the cup abduction angle and squeaking. Results. The incidence of squeaking was 2,63% (2 patients). Both patients reported a “click” noise in hyperextension of the hip. The mean abduction angle was 44 degress (35–60), and 48 degrees (46 and 50) on the squeaking group. There was no statistically significant difference in the in the mean cup inclination between squeaky and quiet hips. Conclusion. The incidence of squeaking in association with the use of Exeter stem and Trident ceramic acetabular system was 2,63%. There was no correlation of the cup abduction angle and squeaking


The Bone & Joint Journal
Vol. 97-B, Issue 1 | Pages 94 - 99
1 Jan 2015
Grammatopoulos G Wilson HA Kendrick BJL Pulford EC Lippett J Deakin M Andrade AJ Kambouroglou G

National Institute of Clinical Excellence guidelines state that cemented stems with an Orthopaedic Data Evaluation Panel (ODEP) rating of > 3B should be used for hemiarthroplasty when treating an intracapsular fracture of the femoral neck. These recommendations are based on studies in which most, if not all stems, did not hold such a rating. . This case-control study compared the outcome of hemiarthroplasty using a cemented (Exeter) or uncemented (Corail) femoral stem. These are the two prostheses most commonly used in hip arthroplasty in the UK. Data were obtained from two centres; most patients had undergone hemiarthroplasty using a cemented Exeter stem (n = 292/412). Patients were matched for all factors that have been shown to influence mortality after an intracapsular fracture of the neck of the femur. Outcome measures included: complications, re-operations and mortality rates at two, seven, 30 and 365 days post-operatively. Comparable outcomes for the two stems were seen. . There were more intra-operative complications in the uncemented group (13% vs 0%), but the cemented group had a greater mortality in the early post-operative period (n = 6). There was no overall difference in the rate of re-operation (5%) or death (365 days: 26%) between the two groups at any time post-operatively. This study therefore supports the use of both cemented and uncemented stems of proven design, with an ODEP rating of 10A, in patients with an intracapsular fracture of the neck of the femur. Cite this article: Bone Joint J 2015;97-B:94–9


Bone & Joint Open
Vol. 5, Issue 6 | Pages 452 - 456
1 Jun 2024
Kennedy JW Rooney EJ Ryan PJ Siva S Kennedy MJ Wheelwright B Young D Meek RMD

Aims

Femoral periprosthetic fractures are rising in incidence. Their management is complex and carries a high associated mortality. Unlike native hip fractures, there are no guidelines advising on time to theatre in this group. We aim to determine whether delaying surgical intervention influences morbidity or mortality in femoral periprosthetic fractures.

Methods

We identified all periprosthetic fractures around a hip or knee arthroplasty from our prospectively collated database between 2012 and 2021. Patients were categorized into early or delayed intervention based on time from admission to surgery (early = ≤ 36 hours, delayed > 36 hours). Patient demographics, existing implants, Unified Classification System fracture subtype, acute medical issues on admission, preoperative haemoglobin, blood transfusion requirement, and length of hospital stay were identified for all patients. Complication and mortality rates were compared between groups.


Bone & Joint Open
Vol. 4, Issue 9 | Pages 659 - 667
1 Sep 2023
Nasser AAHH Osman K Chauhan GS Prakash R Handford C Nandra RS Mahmood A

Aims

Periprosthetic fractures (PPFs) following hip arthroplasty are complex injuries. This study evaluates patient demographic characteristics, management, outcomes, and risk factors associated with PPF subtypes over a decade.

Methods

Using a multicentre collaborative study design, independent of registry data, we identified adults from 29 centres with PPFs around the hip between January 2010 and December 2019. Radiographs were assessed for the Unified Classification System (UCS) grade. Patient and injury characteristics, management, and outcomes were compared between UCS grades. A multinomial logistic regression was performed to estimate relative risk ratios (RRR) of variables on UCS grade.


Bone & Joint Open
Vol. 3, Issue 9 | Pages 710 - 715
5 Sep 2022
Khan SK Tyas B Shenfine A Jameson SS Inman DS Muller SD Reed MR

Aims

Despite multiple trials and case series on hip hemiarthroplasty designs, guidance is still lacking on which implant to use. One particularly deficient area is long-term outcomes. We present over 1,000 consecutive cemented Thompson’s hemiarthroplasties over a ten-year period, recording all accessible patient and implant outcomes.

Methods

Patient identifiers for a consecutive cohort treated between 1 January 2003 and 31 December 2011 were linked to radiographs, surgical notes, clinic letters, and mortality data from a national dataset. This allowed charting of their postoperative course, complications, readmissions, returns to theatre, revisions, and deaths. We also identified all postoperative attendances at the Emergency and Outpatient Departments, and recorded any subsequent skeletal injuries.


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 241 - 241
1 Sep 2012
Field R Rajakulendran K Rushton N
Full Access

Introduction. The MITCH PCR is an anatomic, flexible, horse-shoe shaped acetabular component, with 2 polar fins. The rationale of the PCR cup design is to reproduce a near-physiological stress distribution in the bone adjacent to the prosthesis. The thin composite cup is designed to fuse and flex in harmony with the surrounding bony structure. Only the pathological acetabular cartilage and underlying subchondral bone of the horseshoe-shaped, load-bearing portion of the acetabular socket is replaced, thus preserving viable bone stock. The PCR is manufactured from injection moulded carbon fibre reinforced polyetheretherketone (PEEK), with a two layer outer surface comprising hydroxyapatite and plasma sprayed commercially pure titanium. It is implanted in conjunction with a large diameter low wear femoral head, producing a bearing that will generate minimal wear debris with relatively inert particles. Pre-clinical mechanical testing, finite element analysis and biocompatibility studies have been undertaken. FEA evaluation predicts preservation of host bone density in the load bearing segments. A pilot clinical study was completed on a proto-type version of the PCR cup (the “Cambridge” cup), achieving excellent 5 and 10 year results. Subjects and Methods. We report the three-year results from a two-centre, prospective clinical evaluation study of the MITCH PCR cup. Patient outcome has been assessed using standardised clinical and radiological examinations and validated questionnaires. The change in physical level of activity and quality of life has been assessed using the Oxford Hip Score, Harris Hip score and the EuroQol-5D score, at scheduled time-points. Serial radiographs have been analysed to monitor the fixation and stability of the components. Results and Conclusions. In total 25 PCR cups were implanted by 3 surgeons. There were 12 men and 13 women. The mean patient age at time of surgery was 67 years (range 57–74). An Accolade TMZF stem was used as the femoral component in 19 patients and an Exeter stem in 6. The mean Oxford Hip score improved from 19.8 pre-operatively to 45 at the latest follow-up. The mean Euroqol-5D score improved from 62.6 to 83.6 and the Harris Hip score improved from 49.9 to 90.6. Three adverse events were noted in 2 patients (2 chest infections and 1 deep vein thrombosis). One revision of the acetabular component was performed at 21 months for squeaking. This has been investigated and modification of the articular geometry has resolved the problem on in-vitro testing


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXVII | Pages 509 - 509
1 Sep 2012
Thomas G Hossain M Monk A Gill H Glyn-Jones S Andrew J Murray D Beard D Epos Group N
Full Access

Introduction. Malalignment of some designs of stem is associated with an increased risk of aseptic loosening and revision. We investigated whether the alignment of the cemented polished, double-taper design adversely affected outcome, in a multicentre prospective study. Methods. A multicentre prospective study of 1189 total hip replacements was undertaken to investigate whether there is an association between surgical outcome and femoral stem alignment. All patients underwent a primary THR with the Exeter femoral stem (Stryker Howmedica Osteonics, Mahwah, NJ) and a variety of acetabular components. The primary outcome measure was the Oxford hip score (OHS) and change in OHS at five years. Secondary outcomes included rate of dislocation and revision. Radiographic evaluation of the femoral component was also undertaken. The long axis of the Exeter femoral component and the long axis of the femoral canal were located, and the angle at the point of intersection measured. The cementing quality was determined as defined by Barrack et al. Radiolucent lines at the cement-stem and cement-bone interface in the five year radiographs were defined using the zones described by Gruen et al. Subsidence was measured as the vertical dimension of the radiolucency craniolateral to the shoulder of the stem in Gruen zone 1 as described by Fowler et al. Cement fractures were recorded. Results. The incidence of varus (>5 ° to the femoral axis), and valgus (>5 °) malignment were 3.7% and 0.8% respectively. Pre-operative demographics and OHS were similar in all groups (p > 0.4). There was no significant difference in OHS or change in OHS between neutral and malaligned groups at 5 years (neutral, mean=40.1, change=23.1; varus, mean=40.1, change=23.7; valgus, mean=42.0, change=26.6; p=0.46 and p=0.45 respectively). There was no significant difference in dislocation rate between the groups (p=0.66). There was also no significant difference in revision rate (p=0.34). There were no statistically significant differences in the incidence of femoral radiolucency, stem subsidence or cement fracture (p > 0.1). Conclusion. This study provides evidence that both varus and valgus implantation does not compromise the short to medium term clinical results of the cemented, polished, double-taper stems. Longer follow-up is required to establish the influence of stem alignment on the incidence of aseptic loosening and revision


The Bone & Joint Journal
Vol. 99-B, Issue 1_Supple_A | Pages 50 - 59
1 Jan 2017
Carli AV Negus JJ Haddad FS

Aims

Periprosthetic femoral fractures (PFF) following total hip arthroplasty (THA) are devastating complications that are associated with functional limitations and increased overall mortality. Although cementless implants have been associated with an increased risk of PFF, the precise contribution of implant geometry and design on the risk of both intra-operative and post-operative PFF remains poorly investigated. A systematic review was performed to aggregate all of the PFF literature with specific attention to the femoral implant used.

Patients and Methods

A systematic search strategy of several journal databases and recent proceedings from the American Academy of Orthopaedic Surgeons was performed. Clinical articles were included for analysis if sufficient implant description was provided. All articles were reviewed by two reviewers. A review of fundamental investigations of implant load-to-failure was performed, with the intent of identifying similar conclusions from the clinical and fundamental literature.


The Bone & Joint Journal
Vol. 100-B, Issue 12 | Pages 1618 - 1625
1 Dec 2018
Gill JR Kiliyanpilakkill B Parker MJ

Aims

This study describes and compares the operative management and outcomes in a consecutive case series of patients with dislocated hemiarthroplasties of the hip, and compares outcomes with those of patients not sustaining a dislocation.

Patients and Methods

Of 3326 consecutive patients treated with hemiarthroplasty for fractured neck of femur, 46 (1.4%) sustained dislocations. Of the 46 dislocations, there were 37 female patients (80.4%) and nine male patients (19.6%) with a mean age of 83.8 years (66 to 100). Operative intervention for each, and subsequent dislocations, were recorded. The following outcome measures were recorded: dislocation; mortality up to one-year post-injury; additional surgery; residential status; mobility; and pain score at one year.


Aims

This study aimed to compare the change in health-related quality of life of patients receiving a traditional cemented monoblock Thompson hemiarthroplasty compared with a modern cemented modular polished-taper stemmed hemiarthroplasty for displaced intracapsular hip fractures.

Patients and Methods

This was a pragmatic, multicentre, multisurgeon, two-arm, parallel group, randomized standard-of-care controlled trial. It was embedded within the WHiTE Comprehensive Cohort Study. The sample size was 964 patients. The setting was five National Health Service Trauma Hospitals in England. A total of 964 patients over 60 years of age who required hemiarthroplasty of the hip between February 2015 and March 2016 were included. A standardized measure of health outcome, the EuroQol (EQ-5D-5L) questionnaire, was carried out on admission and at four months following the operation.


The Bone & Joint Journal
Vol. 100-B, Issue 8 | Pages 1087 - 1093
1 Aug 2018
Barenius B Inngul C Alagic Z Enocson A

Aims

The aim of this study was to compare the functional and radiological outcomes in patients with a displaced fracture of the hip who were treated with a cemented or a cementless femoral stem.

Patients and Methods

A four-year follow-up of a randomized controlled study included 141 patients who underwent surgery for a displaced femoral neck fracture. Patients were randomized to receive either a cemented (n = 67) or a cementless (n = 74) stem at hemiarthroplasty (HA; n = 83) or total hip arthroplasty (THA; n = 58).


Bone & Joint Research
Vol. 5, Issue 1 | Pages 18 - 25
1 Jan 2016
Sims AL Parsons N Achten J Griffin XL Costa ML Reed MR

Background

Approximately half of all hip fractures are displaced intracapsular fractures. The standard treatment for these fractures is either hemiarthroplasty or total hip arthroplasty. The recent National Institute for Health and Care Excellence (NICE) guidance on hip fracture management recommends the use of ‘proven’ cemented stem arthroplasty with an Orthopaedic Device Evaluation Panel (ODEP) rating of at least 3B (97% survival at three years). The Thompsons prosthesis is currently lacking an ODEP rating despite over 50 years of clinical use, likely due to the paucity of implant survival data. Nationally, adherence to these guidelines is varied as there is debate as to which prosthesis optimises patient outcomes.

Design

This study design is a multi-centre, multi-surgeon, parallel, two arm, standard-of-care pragmatic randomised controlled trial. It will be embedded within the WHiTE Comprehensive Cohort Study (ISRCTN63982700). The main analysis is a two-way equivalence comparison between Hemi-Thompson and Hemi-Exeter polished taper with Unitrax head. Secondary outcomes will include radiological leg length discrepancy measured as per Bidwai and Willett, mortality, re-operation rate and indication for re-operation, length of index hospital stay and revision at four months. This study will be supplemented by the NHFD (National Hip Fracture Database) dataset.


The Bone & Joint Journal
Vol. 99-B, Issue 10 | Pages 1399 - 1408
1 Oct 2017
Scott CEH MacDonald D Moran M White TO Patton JT Keating JF

Aims

To evaluate the outcomes of cemented total hip arthroplasty (THA) following a fracture of the acetabulum, with evaluation of risk factors and comparison with a patient group with no history of fracture.

Patients and Methods

Between 1992 and 2016, 49 patients (33 male) with mean age of 57 years (25 to 87) underwent cemented THA at a mean of 6.5 years (0.1 to 25) following acetabular fracture. A total of 38 had undergone surgical fixation and 11 had been treated non-operatively; 13 patients died at a mean of 10.2 years after THA (0.6 to 19). Patients were assessed pre-operatively, at one year and at final follow-up (mean 9.1 years, 0.5 to 23) using the Oxford Hip Score (OHS). Implant survivorship was assessed. An age and gender-matched cohort of THAs performed for non-traumatic osteoarthritis (OA) or avascular necrosis (AVN) (n = 98) were used to compare complications and patient-reported outcome measures (PROMs).


The Bone & Joint Journal
Vol. 99-B, Issue 7 | Pages 958 - 963
1 Jul 2017
Mamarelis G Key S Snook J Aldam C

Aims

Hip hemiarthroplasty is a standard treatment for intracapsular proximal femoral fractures in the frail elderly. In this study we have explored the implications of early return to theatre, within 30 days, on patient outcome following hip hemiarthroplasty.

Patients and Methods

We retrospectively reviewed the hospital records of all hip hemiarthroplasties performed in our unit between January 2010 and January 2015. Demographic details, medical backround, details of the primary procedure, complications, subsequent procedures requiring return to theatre, re-admissions, discharge destination and death were collected.


The Bone & Joint Journal
Vol. 97-B, Issue 7 | Pages 992 - 996
1 Jul 2015
Parker MJ

A total of 56 male patients with a displaced intracapsular fracture of the hip and a mean age of 81 years (62 to 94), were randomised to be treated with either a cemented hemiarthroplasty (the Exeter Trauma Stem) or reduction and internal fixation using the Targon Femoral Plate. All surviving patients were reviewed one year after the injury, at which time restoration of function and pain in the hip was assessed. There was no statistically significant difference in mortality between the two groups (7/26; 26.9% for hemiarthroplasty vs 10/30; 33.3% for internal fixation). No patient treated with a hemiarthroplasty required further surgery, but eight patients treated by internal fixation did (p = 0.005), five requiring hemiarthroplasty and three requiring total hip arthroplasty. Those treated by internal fixation had significantly more pain (p = 0.02). The restoration of mobility and independence were similar in the two groups.

These results indicate that cemented hemiarthroplasty gives better results than internal fixation in elderly men with a displaced intracapsular fracture of the hip.

Cite this article: Bone Joint J 2015;97-B:992–6.


The Journal of Bone & Joint Surgery British Volume
Vol. 92-B, Issue 12 | Pages 1669 - 1677
1 Dec 2010
Thakar C Alsousou J Hamilton TW Willett K

We evaluated the cost and consequences of proximal femoral fractures requiring further surgery because of complications. The data were collected prospectively in a standard manner from all patients with a proximal femoral fracture presenting to the trauma unit at the John Radcliffe Hospital over a five-year period. The total cost of treatment for each patient was calculated by separating it into its various components. The risk factors for the complications that arose, the location of their discharge and the mortality rates for these patients were compared to those of a matched control group. There were 2360 proximal femoral fractures in 2257 patients, of which 144 (6.1%) required further surgery. The mean cost of treatment in patients with complications was £18 709 (£2606.30 to £60.827.10), compared with £8610 (£918.54 to £45 601.30) for uncomplicated cases (p < 0.01), with a mean length of stay of 62.8 (44.5 to 79.3) and 32.7 (23.8 to 35.0) days, respectively. The probability of mortality after one month in these cases was significantly higher than in the control group, with a mean survival of 209 days, compared with 496 days for the controls. Patients with complications were statistically less likely to return to their own home (p < 0.01).

Greater awareness and understanding are required to minimise the complications of proximal femoral fractures and consequently their cost.