Advertisement for orthosearch.org.uk
Results 1 - 2 of 2
Results per page:
The Journal of Bone & Joint Surgery British Volume
Vol. 93-B, Issue 5 | Pages 616 - 621
1 May 2011
O’Neill F Condon F McGloughlin T Lenehan B Coffey JC Walsh M

We biomechanically investigated whether the standard dynamic hip screw (DHS) or the DHS blade achieves better fixation in bone with regard to resistance to pushout, pullout and torsional stability. The experiments were undertaken in an artificial bone substrate in the form of polyurethane foam blocks with predefined mechanical properties. Pushout tests were also repeated in cadaveric femoral heads. The results showed that the DHS blade outperformed the DHS with regard to the two most important characteristics of implant fixation, namely resistance to pushout and rotational stability. We concluded that the DHS blade was the superior implant in this study


Bone & Joint Research
Vol. 6, Issue 5 | Pages 270 - 276
1 May 2017
Gosiewski JD Holsgrove TP Gill HS

Objectives

Fractures of the proximal femur are a common clinical problem, and a number of orthopaedic devices are available for the treatment of such fractures. The objective of this study was to assess the rotational stability, a common failure predictor, of three different rotational control design philosophies: a screw, a helical blade and a deployable crucifix.

Methods

Devices were compared in terms of the mechanical work (W) required to rotate the implant by 6° in a bone substitute material. The substitute material used was Sawbones polyurethane foam of three different densities (0.08 g/cm3, 0.16 g/cm3 and 0.24 g/cm3). Each torsion test comprised a steady ramp of 1°/minute up to an angular displacement of 10°.