Orthopaedic Proceedings
Vol. 104-B, Issue SUPP_10 | Pages 12 - 12
1 Oct 2022
Fes AF Leal AC Alier A Pardos SL Redó MLS Verdié LP Diaz SM Pérez-Prieto D
Full Access

Aim. The most frequent mechanical failure in the osteosynthesis of intertrochanteric fractures is the cut-out. Fracture pattern, reduction quality, tip-apex distance or the position of the cervico-cephalic screw are some of the factors that have been associated with higher cut-out rates. To date, it has not been established whether underlying bacterial colonization or concomitant infection may be the cause of osteosynthesis failure in proximal femur fractures (PFF). The primary objective of this study is to assess the incidence of infection in patients with cut-out after PFF osteosynthesis. Method. Retrospective cohort study on patients with cut-out after PFF osteosynthesis with endomedullary nail, from January 2007 to December 2020. Demographic data of patients (such as sex, age, ASA), fracture characteristics (pattern, laterality, causal mechanism) and initial surgery parameters were collected (time from fall to intervention, duration of surgery, intraoperative complications). Radiographic parameters were also analyzed (tip-apex distance and Chang criteria). In all cut-out cases, 5 microbiological cultures and 1 anatomopathological sample were taken and the osteosynthesis material was sent for sonication. Fracture-related infection (FRI) was diagnosed based on Metsemakers et al (2018) and McNally et al (2020) diagnostic criteria. Results. Of the 67 cut-out cases, 16 (23.9%) presented clinical, analytical or microbiological criteria of infection. Of these sixteen patients, only in 3 of them the presence of an underlying infection was suspected preoperatively. A new osteosynthesis was performed in 24 cases (35.8%) and a conversion to arthroplasty in the remaining 43 (64.2%). A comparative analysis was performed between cases with and without infection. The groups were comparable in terms of demographic data and postoperative radiological data (using Chang criteria and tip-apex distance). Patients with underlying infection had a higher rate of surgical wound complication (56.3% vs 22%, p = 0.014), higher rates of leukocytes counts (11.560 vs 7.890, p = 0.023) and time to surgery (5.88 vs 3.88 days, p = 0.072). Conclusions. One out of four osteosynthesis failure in PFF is due to underlying FRI and in almost 20% were not unsuspected before surgery. In PFF osteosynthesis failures, underlying infection should be taken into account as a possible etiological factor and thus a preoperative and intraoperative infection study should be always performed


Orthopaedic Proceedings
Vol. 105-B, Issue SUPP_2 | Pages 113 - 113
10 Feb 2023
Burrows K Lock A Smith Z McChesney S
Full Access

Failure of cephalomedullary fixation for proximal femur fractures is an uncommon event associated with significant morbidity to the patient and cost to the healthcare system. This institution changed nailing system from the PFNA (DePuy Synthes) to InterTan (Smith and Nephew) in February 2020. To assess for non-inferiority, a retrospective review was performed on 247 patients treated for unstable proximal femur fractures (AO 31 A2; A3). Patients were identified through manual review of fluoroscopic images. Stable fracture patterns were excluded (AO 31 A1). Pre/post operative imaging, demographic data, operative time and ASA scores assessed. Internal/external imaging and national joint registry data were reviewed for follow up. No significant difference was found in overall failure rate of PFNA vs InterTan (4.84% vs 3.23%; p = 0.748). Overall, short nails were more likely to fail by cut-out than long nails (7.5% vs 1.2%; p = 0.015). Nails which failed by varus cut-out had a higher tip/apex distance (TAD) (26.2mm vs 17.0mm; p < 0.001). Of concern, varus cut-out occurred in two InterTan nails with TAD of <25mm. The PFNA enjoyed a shorter operative time for both the short (59.1 vs 71.8 mins; p = 0.022) and long nails (98.8 vs 114.3 mins; p = 0.016) with no difference in 120-day survival rate. Overall failure rates of the PFNA and InterTan nailing systems were comparable. However, the failure rate of short nails in this study is concerning. Using long nails with a lag screw design for unstable intertrochanteric femoral fractures may reduce failure rates. Cumulative frequency analysis suggests stringent tip-apex distances of less than 21mm may reduce failure rates in lag-screw design cephulomedullary nails. This dataset suggests that unstable intertrochanteric fractures may be more reliably managed with a long cephalomedullary device


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXIX | Pages 164 - 164
1 Sep 2012
Gibson D Keogh C Morris S
Full Access

Introduction. Lag screw cut-out following fixation of unstable intertrochanteric fractures in osteoporotic bone remains an unsolved challenge. A novel new device is the X-Bolt which is an expanding type bolt that may offer superior fixation in osteoporotic bone compared to the standard DHS screw type device. Aims. The aim of this study was to test if there was a difference in cut-out using the X-Bolt implant compared with the standard DHS system. Methods. Specimens of low density surrogate bone (5pcf) were inserted into a simplified biomechanical fracture model and had either an X-Bolt or DHS implant inserted. There were eight samples in each group. The fracture model was tested with an incremental cyclical loading programme in a Material Test System. Displacement, cycle count and force exerted were continuously recorded until cut-out of the implant. Results. All of the specimens failed by varus collapse with superior cut-out and resulted in an automatic stop of the MTS. Specimens with the X-Bolt implant inserted lasted longer on cyclical count and withstood a greater force at cut-out compared with DHS specimens. The mean number of cycles to cutout in the DHS specimens was 4345 and in specimens with the X-Bolt inserted was 6898. The mean force at which cutout occurred in the DHS group was 1.025kN and in specimens with the X-Bolt inserted was 1.275kN. A statistically significant difference was observed with a P-value of 0.005 and a power of 87.2% with respect to cycle count and a P-value of 0.008 and power 84.8% with respect to force exerted at failure when comparing between the two groups. Conclusion. This study shows that the X-Bolt device demonstrated superior cut-out resistance and withstood greater loads compared to the DHS in low density surrogate bone in an unstable fracture model under cyclical axial loading


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXXII | Pages 7 - 7
1 Jul 2012
Agni N Sellers E Johnson R Gray A
Full Access

The aim of this study was to establish any association between implant cut-out and a Tip Apex Distance (TAD), ≥25mm, in proximal femoral fractures, following closed reduction and stabilisation, with either a Dynamic Hip Screw (DHS) or Intramedullary Hip Screw (IMHS) device. Furthermore, we investigated whether any difference in cut-out rate was related to fracture configuration or implant type. WE conducted a retrospective review of the full clinical records and radiographs of 65 consecutive patients, who underwent either DHS or IMHS fixation of proximal femoral fractures. The TAD was measured in the standard fashion using the combined measured AP and lateral radiograph distances. Fractures were classified according to the Muller AO classification. 35 patients underwent DHS fixation and 30 patients had IMHS fixation. 5 in each group had a TAD≥25mm. There were no cut-outs in the DHS group and 3 in the IMHS group. 2 of the cut-outs had a TAD≥25mm. The 3 cut-outs in the IMHS group had a fracture classification of 31-A2, 31-A3 and 32-A3.1 respectively. In addition, the fractures were inadequately reduced and fixed into a varus position. A TAD<25mm would appear to be associated with a lower rate of cut-out. The cut-out rate in the IMHS group was higher than the DHS group. Contributing factors may have included an unstable fracture configuration and inadequate closed fracture reduction at the time of surgery


Orthopaedic Proceedings
Vol. 98-B, Issue SUPP_9 | Pages 4 - 4
1 May 2016
Lo H
Full Access

Introduction. Osteoporotic intertrochanteric fracture (ITF) is frequent injuries affecting elderly, osteoporotic patients leading to significant morbidity and mortality. Successful prognosis including union and alignment is challenging even though initial successful reduction with internal fixation. Although many factors are related to the patient's final prognosis. Well reduction with stable fracture fixation is still the goal of treatment to improve the quality of life and decrease morbidity in patients with hip fractures, but this in turn depends on the type of fracture and bone quality. Poor bone quality is responsible for common complications, such as failure of fixation, varus collapse and lag screw cut-out, in elderly patients. Kim et al. found that the complication rate when using conventional DHS in unstable ITFs can be as high as 50% because of screw cut-out. We used the dynamic hip screws (DHS) strengthened by calcium phosphate cement (CPC) for treatment femoral intertrochanteric fracture and review the prognosis of our patients. Materials and Methods. From January of 2011 to January of 2014, 42 patients with femoral intertrochanteric fracture underwent surgery with DHS strengthened by CPC. Comparisons were made between the DHS plus CPC group with the other patients with only DHS used in our department. All patients were followed up for an average time of 14.8(6 to 24) months. X-ray was reviewed for the conditions of union and implant failure. Results. In DHS group, fixation failure happened in 3 case, delayed union and coax varus deformity in 2 cases. IN DHS plus CPC group, all fractures healed uneventfully, there is no non-union or malunion in this group. There is only 1 fixation cut-out and 1 secondary lag screw sliding was noted, however, union was still well over fracture site in this case, the patient had no clinical symptoms. Discussion. Residual bony defects present after DHS fixation in intertrochanteric fracture may lead to postoperative complications, including nonunion or implant failure. DHS strengthened by CPC is reliable fixation for old patients with intertrochanteric fracture, We demonstrated that augmentation of the bony defect with dynamic hip screw by reinforced calcium phosphate cement significantly improved the strength of osteoporotic bone, prevent screw loosening, and promote early healing of fracture. The patients can be decreased the risk of refracture and allow early weight bearing, especially in elderly patients with osteoporotic bone


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_18 | Pages 3 - 3
1 Apr 2013
Bradford OJ Niematallah I Berstock JR Trezies A
Full Access

Intra-operative Tip-Apex Distance (TAD) estimation optimises dynamic hip screw (DHS) placement during hip fracture fixation, reducing risk of cut-out. Thread-width of a standard DHS screw measures approximately 12.5 millimetres. We assessed the effect of introducing screw thread-width as an intra-operative distance reference to surgeons. The null hypothesis was that there were no differences between hip fracture fixation before and after this intervention. Primary outcome measure was TAD. Secondary outcome measures included position of the screw in the femoral head, quality of reduction, cut-out and surgeon accuracy of estimating TAD. 150 intra-operative DHS radiographs were assessed before and after introducing screw thread-width distance reference to surgeons. Mean TAD reduced from 19.37mm in the control group to 16.49mm in the prospective group (p=<0.001). The number of DHS with a TAD > 25mm reduced from 14% to 6%. Screw position on lateral radiographs was significantly improved (p=0.004). There were no significant differences in screw position on antero-posterior radiographs, quality of reduction, or rate of cut-out. Significant improvement in accuracy (p=0.05) and precision (p=0.005) of TAD estimation was demonstrated. Awareness and use of screw-thread width improves estimation and positioning of a DHS screw in the femoral head during fixation of hip fractures


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XLI | Pages 116 - 116
1 Sep 2012
Barton T Chesser T Harries W Gleeson R Topliss C Greenwood R
Full Access

Controversy exists whether to treat unstable pertrochanteric hip fractures with either intra-medullary or extra-medullary devices. A prospective randomised control trial was performed to compare the outcome of unstable pertrochanteric hip fractures stabilised with either a sliding hip screw or long Gamma Nail. The hypothesis was that there is no difference in outcome between the two modes of treatment. Over a four year period, 210 patients presenting with an unstable pertrochanteric hip fracture (AO/OTA 31 A2) were recruited into the study. Eligible patients were randomised on admission to either long Gamma Nail or sliding hip screw. Follow-up was arranged for three, six, and twelve months. Primary outcome measures were implant failure or ‘cut-out’. Secondary measures included mortality, length of hospital stay, transfusion rate, change in mobility and residence, and EuroQol outcome score. Five patients required revision surgery for implant cut-out (2.5%), of which three were long Gamma Nails and two were sliding hip screws (no significant difference). There were no incidences of implant failure or deep infection. Tip apex distance was found to correlate with implant cut-out. There was no statistically significant difference in either the EuroQol outcome scores or mortality rates between the two groups when corrected for mini mental score. There was no difference in transfusion rates, length of hospital stay, and change in mobility or residence. There was a clear cost difference between the implants. The sliding hip screw remains the gold standard in the treatment of unstable pertrochanteric fractures of the proximal femur


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XVII | Pages 35 - 35
1 May 2012
Baliga S Johnstone A McKenna S
Full Access

Background. Angular stable volar locking plates have become increasingly popular for more comminuted fractures of the distal radius. Newer designs of plates have been thicker in profile and incorporate more options for distal fragment fixation. Although they have been shown to be successful at maintaining reduction to allow early mobilisation the main drawback is from screw cut-out. In our practice we have noticed that the newer style of plates that offer more rigid fixation has lead to more instances of screw cut-out. We aimed to quantify the minimum number of locking pegs and or screws need to maintain the operative reduction. Method. We retrospectively looked at a series of 46 patients that had undergone volar plating. We assessed the fracture severity on pre-operative films (according to AO classification) and compared radiographic parameters (volar tilt VT, radial inclination RI and radial height RH) on post-operative films. We calculated the amount of reduction lost from initial post operative x-rays to radiographs taken when union was confirmed. We compared this to the number of locking units used to fix the distal radius and also the configuration they were inserted, i.e. the number in the radial and middle columns. Results. The mean loss of reduction in all plates was 0.9mm of RH, 2.2degs of RI and 2.8degs of VT. There was no difference in mean ‘reduction lost’ between plates that had a total of 2 or 3 locking units (RH 1mm and 1mm, RI 2.0deg and 2.7deg and VT 2.9deg and 3.2 deg respectively). This was also case when sub-analysing more severe OTA Type C fractures. Conclusion. Only one locking peg is needed under each column for adequate stability (i.e. a total of two distal locking units). Any more than this confers no additional benefit in maintaining reduction, ever in more severe fractures. The extra rigidity provided by more locking screws/pegs and also thicker plates: as provided by newer designs of plates, may make them more prone to cut out


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XXI | Pages 64 - 64
1 May 2012
M. P
Full Access

Fractures of the proximal femur at the level of the lesser trochanter (reversed and transverse fracture lines, Evans classification type II, AO classification A3 fractures) are known to have an increased risk of fixation failure. 58 patients with such a fracture were randomised to have the fracture fixed with either an intramedullary nail (220 mm Targon PF nail) or a sliding hip screw (SHS). The mean age of the patients was 82 and 11% were male. All patients were followed up for one year by a research nurse blinded to the treatment groups. Mean length of surgery was 50 minutes for the nail versus 52 minutes for the SHS. There were no differences between groups in the need for blood transfusion. Operative complications tended to be less for the nail group (1/27 versus 5/26). Mean hospital stay was 18 days for the nail group versus 29 days for those treated with the SHS. The only fracture healing complications were one case of cut-out in the SHS group and two cut-outs in the nail group, two of which required revision surgery. During follow-up those patients treated with the nail reported a tendency to lower pain scores than those treated with the SHS (p=0.04 at two months). This difference persisted even at one year from injury. Mortality and regain of mobility was similar between groups. These results indicate that for these difficult fractures types both types of fixation produce comparable outcomes


Orthopaedic Proceedings
Vol. 94-B, Issue SUPP_XVI | Pages 19 - 19
1 Apr 2012
Kaka R Kiwanis R
Full Access

Various operative treatments have been proposed for proximal humeral fractures. The purpose of our study was to compare complications of plate versus nail for these proximal humeral fractures and to determine whether it is the implant or fracture and surgeon related factors which result in complications. We had 74 patients operated from March 2006 till June 2008 for displaced 3 (49pts) or 4 (25pts) part proximal humeral fractures. 43 had plating (PHILOS) and 31 had a humeral nail inserted. 57 patients were over 60 years at presentation while 17 were younger than 60. All patients were followed regularly radiologically. The functional outcomes were assessed by Quick DASH score and were comparable in both groups at 1 year postoperatively.18 of the 43 patients in the Plating group had a radiological complication with 9 cases of screw cut-out, 5 fractures maluniting and 1 nonunion. There was no case of osteonecrosis. In the nailing group, 13 patients had radiological complications, with 8 patients having varus malunion, 3 having proximal screw loosening and 1 having osteonecrosis apart from the clinical complications of impingement and rotator cuff problems. Given the similar complication rate and similar functional outcome achieved by both these techniques it is hard to determine if any one in particular is better than the other. The key determining factors for a proximal humeral fracture might actually be patient and surgeon related. It is important to achieve medial continuity and good initial reduction in these fractures to prevent them from collapsing into varus, especially with the nails. Also, patient age, osteoporosis and functional demand are factors determining success of surgery in these patients. We did not find the severity of the fracture to be a determining factor, maybe this was because the relative proportion of type 4 fractures in our series was less


Orthopaedic Proceedings
Vol. 95-B, Issue SUPP_14 | Pages 19 - 19
1 Mar 2013
Naude P Maqungo S Roche S Nortje M
Full Access

Purpose of study. Unstable pelvic ring injuries usually occur in polytrauma patients and are associated with high mortality and morbidity. Percutaneous screw fixation of sacro-iliac joint dislocations, fracture-dislocations and sacral fractures is a well-recognised technique first described by Routt and is increasingly gaining popularity. This method is biomechanically comparable to open reduction and internal fixation with plates and screws but offers the advantages of minimally invasive surgical techniques. It is however a technically demanding procedure with reported complications including hardware failure, misplacement of screws, nerve injury and poor posterior reduction. The purpose of the study is to report clinical results of patients treated with closed reduction and percutaneous ilio-sacral screw fixation for unstable pelvic ring fractures by a single surgeon. Methods. A review of prospectively collected data was performed on all patients who had percutaneous sacro-iliac fixation between 2009 and 2012. Thirty five percutaneus sacro-iliac screws were inserted in 30 patients with a mean age of 25.6 years (range 17–62). Fracture types included 10 AO type B, and 20 AO type C. The mean follow-up period was 11.6 months (range 3–38). The complications assessed were screw misplacement, neurovascular complications, hardware breakage and loss of reduction. Results. All patients had a satisfactory initial reduction. One patient (2.8%) had misplacement of a screw with resultant temporary neurological fallout. One patient (2.8%) had screw misplacement without neurological fallout. Both of these patients initially had two screws inserted and the misplaced screws were removed and not reinserted. One patient (2.8%) had screw cut-out with loss of reduction. This screw was removed, open reduction peformed and the screw was re-inserted. Conclusion. The use of percutaneous sacro-iliac screws provides a safe and effective technique for the management of unstable posterior pelvic ring injuries. Our combined complication rate is comparable to published literature. NO DISCLOSURES


Bone & Joint Open
Vol. 1, Issue 5 | Pages 98 - 102
6 May 2020
Das De S Puhaindran ME Sechachalam S Wong KJH Chong CW Chin AYH

The COVID-19 pandemic has disrupted all segments of daily life, with the healthcare sector being at the forefront of this upheaval. Unprecedented efforts have been taken worldwide to curb this ongoing global catastrophe that has already resulted in many fatalities. One of the areas that has received little attention amid this turmoil is the disruption to trainee education, particularly in specialties that involve acquisition of procedural skills. Hand surgery in Singapore is a standalone combined programme that relies heavily on dedicated cross-hospital rotations, an extensive didactic curriculum and supervised hands-on training of increasing complexity. All aspects of this training programme have been affected because of the cancellation of elective surgical procedures, suspension of cross-hospital rotations, redeployment of residents, and an unsustainable duty roster. There is a real concern that trainees will not be able to meet their training requirements and suffer serious issues like burnout and depression. The long-term impact of suspending training indefinitely is a severe disruption of essential medical services. This article examines the impact of a global pandemic on trainee education in a demanding surgical speciality. We have outlined strategies to maintain trainee competencies based on the following considerations: 1) the safety and wellbeing of trainees is paramount; 2) resource utilization must be thoroughly rationalized; 3) technology and innovative learning methods must supplant traditional teaching methods; and 4) the changes implemented must be sustainable. We hope that these lessons will be valuable to other training programs struggling to deliver quality education to their trainees, even as we work together to battle this global catastrophe.